FDA Enforcement Class II Terminated

Nova Biomedical Stat Profile Prime Analyzer as follows: PN 52855 Stat Profile Prime ABG Analyzer PN 53655 Stat Profile Prime ABG Analyzer with Scanner PN 52856 Stat Profile Prime CCS Analyzer PN 53656 Stat Profile Prime CCS Analyzer with Scanner PN 53657 Stat Profile Prime CCS Comp Analyzer with Scanner PN 52857 Stat Profile Prime CCS Comp Analyzer Blood Gas/Electrolyte/Metabolite/CO-Oximetry Analyzer

Recall: Z-2732-2017 · Reported July 19, 2017

Enforcement

Recall Number
Z-2732-2017
Event ID
77532
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Nova Biomedical Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 19, 2017
Initiation Date
June 19, 2017
Classification Date
July 13, 2017
Termination Date
August 14, 2018
Address
200 Prospect St, Waltham, MA, 02453-3457, United States

Description

Nova Biomedical Stat Profile Prime Analyzer as follows: PN 52855 Stat Profile Prime ABG Analyzer PN 53655 Stat Profile Prime ABG Analyzer with Scanner PN 52856 Stat Profile Prime CCS Analyzer PN 53656 Stat Profile Prime CCS Analyzer with Scanner PN 53657 Stat Profile Prime CCS Comp Analyzer with Scanner PN 52857 Stat Profile Prime CCS Comp Analyzer Blood Gas/Electrolyte/Metabolite/CO-Oximetry Analyzer

Reason

Software- scheduled A-QC analysis could initiate prior to a patient result being released by the operator and result in the pending patient result being overwritten with the A-QC test results.

Code Info

All current software versions

Distribution

Worldwide Distribution - US (Nationwide) Foreign: Albania Algeria Argentina Australia Austria Bangladesh Benin Bolivia Brazil Canada Chile China Colombia Costa Rica Czech Republic ECUADOR Egypt El Salvador Finland France Germany Ghana Greece Guatemala Guyana Honduras Hong Kong Hungary ICELAND India Iran Israel Italy Ivory Coast Japan Jordan Kenya Korea Latvia Macedonia Malaysia Mexico Nepal Netherlands Nigeria Pakistan Pakistan Palestine Peru Philippines Poland Puerto Rico Republic of Panama Republica Dominicana Russia Saudi Arabia Slovak Republic Slovenia South Africa Sri Lanka Switzerland Taiwan Thailand The Netherlands Trinidad,W.I. Tunisia Turkey UK Uruguay Vietnam

Quantity

179 units (US) and 904 (OUS) Total 1083