Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/ urologic closed and minimally invasive procedures. It is inserted via a trocar sleeve and is used to prepare, coagulate and cut tissue.
Enforcement
- Recall Number
- Z-2727-2014
- Event ID
- 68725
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Instrumed International, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 8, 2014
- Initiation Date
- May 30, 2014
- Classification Date
- September 30, 2014
- Termination Date
- November 29, 2017
- Address
- 626 Cooper Ct, N/A, Schaumburg, IL, 60173-4537, United States
Description
Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/ urologic closed and minimally invasive procedures. It is inserted via a trocar sleeve and is used to prepare, coagulate and cut tissue.
Additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must not exceed 650 peak voltage (Vp) for all monopolar electrodes. and 2) For all monopolar electrodes, the useful life of this product is less than or equal to 50 cycles and less than or equal to 2 years.
1) Instrumed Item No.: 250-08860; Symmetry Surgical Inc. Catalog Code: 90-9525; Lot Number: 030214; 2) Instrumed Item No.: 250-08930; Symmetry Surgical Inc. Catalog Code: 90-9522; Lot Number: 070511; 3) Instrumed Item No.: 250-08990; Symmetry Surgical Inc. Catalog Code: 90-9526; Lot Number: 031511; 4) Instrumed Item No.: 250-09010; Symmetry Surgical Inc. Catalog Code: 90-9523; Lot Number: 080112
Distribution US Nationwide in the states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.
895 instruments