FDA Enforcement Class II Completed

Stryker iBur 4.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-012-040D

Recall: Z-2726-2024 · Reported September 4, 2024

Enforcement

Recall Number
Z-2726-2024
Event ID
95040
Classification
Class II
Status
Completed
Product Type
Devices
Firm
Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 4, 2024
Initiation Date
July 24, 2024
Classification Date
August 26, 2024
Address
1941 Stryker Way, Portage, MI, 49002-9711, United States

Description

Stryker iBur 4.0mm Diamond Round, Distal Bend- intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-012-040D

Reason

Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.

Code Info

GTIN: 07613327501278 Lot Numbers: 22334017 22339017 23004017 23016017 23085017 23115017 23195017 23299017

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.

Quantity

1706 units