FDA Enforcement
Class II
Terminated
Transonic Flow Probe, 2 mm, Catalog Number HQD2FMC -- Product Usage: to measure flow intra-operatively.
Recall: Z-2726-2015
·
Reported September 9, 2015
Enforcement
- Recall Number
- Z-2726-2015
- Event ID
- 71907
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Transonic Systems Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 9, 2015
- Initiation Date
- July 27, 2015
- Classification Date
- September 2, 2015
- Termination Date
- July 18, 2016
- Address
- 34 Dutch Mill Rd, N/A, Ithaca, NY, 14850-9785, United States
Description
Transonic Flow Probe, 2 mm, Catalog Number HQD2FMC -- Product Usage: to measure flow intra-operatively.
Reason
Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.
Code Info
Serial Numbers: HQD2FMC2797, HQD2FMC2873, HQD2FMC2874, HQD2FMC2875, HQD2FMC2876, HQD2FMC2877, HQD2FMC2878, HQD2FMC2879, HQD2FMC2880, HQD2FMC2881, HQD2FMC2882, HQD2FMC2883, HQD2FMC2884, HQD2FMC2885, HQD2FMC2886, HQD2FMC2888, HQD2FMC2889, HQD2FMC2890, HQD2FMC2908
Distribution
Worldwide Distribution - US Nationwide and the countries of: Canada, Japan, Russia and The Netherlands.
Quantity
Domestic: 8 units, foreign: 11 units