FDA Enforcement Class II Terminated

Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/ urologic closed and minimally invasive procedures. It is inserted via a trocar sleeve and is used to prepare, coagulate and cut tissue.

Recall: Z-2724-2014 · Reported October 8, 2014

Enforcement

Recall Number
Z-2724-2014
Event ID
68725
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Instrumed International, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 8, 2014
Initiation Date
May 30, 2014
Classification Date
September 30, 2014
Termination Date
November 29, 2017
Address
626 Cooper Ct, N/A, Schaumburg, IL, 60173-4537, United States

Description

Reusable laparoscopic instruments are packaged as non-sterile. Cleaning and sterilization of devices must occur prior to use. The instruments are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/ urologic closed and minimally invasive procedures. It is inserted via a trocar sleeve and is used to prepare, coagulate and cut tissue.

Reason

Additional caution statements have been added to the Instructions For Use for monopolar electrodes. 1) To avoid tissue carbonation, the operation voltage of the HF generator must not exceed 650 peak voltage (Vp) for all monopolar electrodes. and 2) For all monopolar electrodes, the useful life of this product is less than or equal to 50 cycles and less than or equal to 2 years.

Code Info

1) Instrumed Item No.: 250-08570; CareFusion Catalog Code: VM46-9112-1; Lot Number: 010111; 2) Instrumed Item No.: 250-08600; CareFusion Catalog Code: LA8523 ; Lot Number: 070211, 070213, 090412; 3) Instrumed Item No.: 250-08610; CareFusion Catalog Code: LA8523-45; Lot Number: 042011, 062209; 4) Instrumed Item No.: 250-08620; CareFusion Catalog Code: LA8515; Lot Number: 030511, 040310, 080111; 5) Instrumed Item No.: 250-08630; CareFusion Catalog Code: LA8515-45; Lot Number: 050212; 6) Instrumed Item No.: 250-08660; CareFusion Catalog Code: LA8514; Lot Number: 040110; 7) Instrumed Item No.: 250-08670; CareFusion Catalog Code: LA8514-45; Lot Number: 120410; 8) Instrumed Item No.: 250-08680; CareFusion Catalog Code: LA8526; Lot Number: 010313, 060311, 060312; 9) Instrumed Item No.: 250-08690; CareFusion Catalog Code: LA8526-45; Lot Number: 030314, 100410, 120110, 120312; 10) Instrumed Item No.: 250-08700; CareFusion Catalog Code: LA8518; Lot Number: 060111, 070110, 080211; 11) Instrumed Item No.: 250-08710; CareFusion Catalog Code: LA8518-45; Lot Number: 010114, 040113, 040313, 060410; 12) Instrumed Item No.: 250-08730; CareFusion Catalog Code: LA8520-45; Lot Number: 010213; 13) Instrumed Item No.: 250-08740; CareFusion Catalog Code: LA8519; Lot Number: 060108, 080310; 14) Instrumed Item No.: 250-08760; CareFusion Catalog Code: LA8521; Lot Number: 020212, 060113, 060213, 100513, 121410; 15) Instrumed Item No.: 250-08770; CareFusion Catalog Code: LA8521-45; Lot Number: 100210; 16) Instrumed Item No.: 250-08780; CareFusion Catalog Code: LA8513; Lot Number: 010110, 050313, 070112, 122110; 17) Instrumed Item No.: 250-08790; CareFusion Catalog Code: LA8513-45; Lot Number: 090411, 100309, 100312; 18) Instrumed Item No.: 250-08800; CareFusion Catalog Code: LA8525; Lot Number: 040114, 090410, 110310, 120111; 19) Instrumed Item No.: 250-08810; CareFusion Catalog Code: LA8525-45; Lot Number: 050308, 060310; 20) Instrumed Item No.: 250-08820; CareFusion Catalog Code: LA8517; Lot Number: 110410, 120212; 21) Instrumed Item No.: 250-08860; CareFusion Catalog Code: LA8507; Lot Number: 011111, 020111, 030210, 041411, 060112, 060313, 070312, 070712, 100413, 100512, 110112, 110313, 110413; 22) Instrumed Item No.: 250-08880; CareFusion Catalog Code: LA8507-45; Lot Number: 061011, 070313, 070513; 23) Instrumed Item No.: 250-08890; CareFusion Catalog Code: LA8506; Lot Number: 020211, 020410; 24) Instrumed Item No.: 250-08900; CareFusion Catalog Code: LA8506-45; Lot Number: 050407; 25) Instrumed Item No.: 250-08930; CareFusion Catalog Code: LA8505; Lot Number: 010314, 010513, 010611, 030113, 032211, 040112, 090313, 100111, 120710; 26) Instrumed Item No.: 250-08960; CareFusion Catalog Code: LA8505-45; Lot Number: 30313, 090113, 110110, 120211; 27) Instrumed Item No.: 250-08970; CareFusion Catalog Code: LA8504; Lot Number: 050110, 080510; 28) Instrumed Item No.: 250-08980; CareFusion Catalog Code: LA8504-45; Lot Number: 050311; 29) Instrumed Item No.: 250-08990; CareFusion Catalog Code: LA8503; Lot Number: 020414, 042310, 110411; 30) Instrumed Item No.: 250-09000; CareFusion Catalog Code: LA8503-45; Lot Number: 020512; 31) Instrumed Item No.: 250-09010; CareFusion Catalog Code: LA8509; Lot Number: 020413, 021511, 030211, 030212, 030G12, 03101816-; 32) Instrumed Item No.: 250-09030; CareFusion Catalog Code: LA8509-45; Lot Number: 070212; 33) Instrumed Item No.: 250-09040; CareFusion Catalog Code: LA8510; Lot Number: 040411; 34) Instrumed Item No.: 250-09070; CareFusion Catalog Code: LA8508-45; Lot Number: 050410; 35) Instrumed Item No.: 250-09100; CareFusion Catalog Code: VM46-9183; Lot Number: 060411; 36) Instrumed Item No.: 250-09120; CareFusion Catalog Code: VM46-9185; Lot Number: 090308

Distribution

Distribution US Nationwide in the states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.

Quantity

3,093 instruments