FDA Enforcement Class II Terminated

Transonic Flow Probe, 2 mm, Catalog Number HQC2FMC Product Usage: to measure flow intra-operatively.

Recall: Z-2720-2015 · Reported September 9, 2015

Enforcement

Recall Number
Z-2720-2015
Event ID
71907
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Transonic Systems Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 9, 2015
Initiation Date
July 27, 2015
Classification Date
September 2, 2015
Termination Date
July 18, 2016
Address
34 Dutch Mill Rd, N/A, Ithaca, NY, 14850-9785, United States

Description

Transonic Flow Probe, 2 mm, Catalog Number HQC2FMC Product Usage: to measure flow intra-operatively.

Reason

Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.

Code Info

Serial Numbers: HQC2FMC1801, HQC2FMC 1802, HQC2FMC1805, HQC2FMC1806

Distribution

Worldwide Distribution - US Nationwide and the countries of: Canada, Japan, Russia and The Netherlands.

Quantity

Domestic: 4 units