Medline medical procedure kits labeled as follows: a) ARTHROSCOPY, REF DYNJ909699; b) BACK CDS, REF CDS984000C; c) EXTREMITY, REF DYNJ900847A; d) HAND-LF, REF DYNJ902619G; e) HIP PINNING CDS, REF CDS984009C; f) HIP PINNING/TFN/ANTERIOR HIP, REF DYNJ902380K; g) KNEE, REF DYNJ904771B; h) KNEE ARTHROSCOPY, REF DYNJ905519D; i) KNEE ARTHROSCOPY, REF DYNJ906273F; j) KNEE ARTHROSCOPY, REF DYNJ909747; k) KNEE ARTHROSCOPY V, REF DYNJS2029D; l) LAMINECTOMY, REF DYNJ907159A; m) LOWER EXTREMITY, REF DYNJ904767A; n) LOWER EXTREMITY, REF DYNJ905518F; o) LOWER EXTREMITY, REF DYNJ909938; p) MINOR EXTREMITY, REF DYNJ909939; q) PODIATRY, REF DYNJ906275D; r) SHOULDER, REF DYNJ904769C; s) SHOULDER, REF DYNJ905511D; t) SHOULDER, REF DYNJ909749; u) SHOULDER ARTHROSCOPY, REF DYNJ906274F; v) SHOULDER ARTHROSCOPY CDS, REF CDS984008C; w) SOUTH EXTREMITY, REF DYNJ909212A; x) SPINE, REF DYNJ905512F; y) TOTAL, REF DYNJ909704; z) TOTAL HIP, REF DYNJ907157A; aa) TOTAL HIP BASIC, REF DYNJ902368K; bb) TOTAL JOINT, REF DYNJ905513G; cc) TOTAL JOINT, REF DYNJ908375C; dd) TOTAL JOINT II, REF DYNJS2027F; ee) TOTAL KNEE, REF DYNJ907158A; ff) TOTAL KNEE CDS, REF CDS983991C; gg) TOTAL SHOULDER, REF DYNJ902367J; hh) TURK-HAND, REF DYNJ905852C; ii) UPPER EXTREMITY, REF DYNJ905514D
Enforcement
- Recall Number
- Z-2714-2024
- Event ID
- 94356
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- MEDLINE INDUSTRIES, LP - Northfield
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- August 28, 2024
- Initiation Date
- March 22, 2024
- Classification Date
- August 22, 2024
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Medline medical procedure kits labeled as follows: a) ARTHROSCOPY, REF DYNJ909699; b) BACK CDS, REF CDS984000C; c) EXTREMITY, REF DYNJ900847A; d) HAND-LF, REF DYNJ902619G; e) HIP PINNING CDS, REF CDS984009C; f) HIP PINNING/TFN/ANTERIOR HIP, REF DYNJ902380K; g) KNEE, REF DYNJ904771B; h) KNEE ARTHROSCOPY, REF DYNJ905519D; i) KNEE ARTHROSCOPY, REF DYNJ906273F; j) KNEE ARTHROSCOPY, REF DYNJ909747; k) KNEE ARTHROSCOPY V, REF DYNJS2029D; l) LAMINECTOMY, REF DYNJ907159A; m) LOWER EXTREMITY, REF DYNJ904767A; n) LOWER EXTREMITY, REF DYNJ905518F; o) LOWER EXTREMITY, REF DYNJ909938; p) MINOR EXTREMITY, REF DYNJ909939; q) PODIATRY, REF DYNJ906275D; r) SHOULDER, REF DYNJ904769C; s) SHOULDER, REF DYNJ905511D; t) SHOULDER, REF DYNJ909749; u) SHOULDER ARTHROSCOPY, REF DYNJ906274F; v) SHOULDER ARTHROSCOPY CDS, REF CDS984008C; w) SOUTH EXTREMITY, REF DYNJ909212A; x) SPINE, REF DYNJ905512F; y) TOTAL, REF DYNJ909704; z) TOTAL HIP, REF DYNJ907157A; aa) TOTAL HIP BASIC, REF DYNJ902368K; bb) TOTAL JOINT, REF DYNJ905513G; cc) TOTAL JOINT, REF DYNJ908375C; dd) TOTAL JOINT II, REF DYNJS2027F; ee) TOTAL KNEE, REF DYNJ907158A; ff) TOTAL KNEE CDS, REF CDS983991C; gg) TOTAL SHOULDER, REF DYNJ902367J; hh) TURK-HAND, REF DYNJ905852C; ii) UPPER EXTREMITY, REF DYNJ905514D
A slight dimensional variation which has the potential for increased difficulty fitting into outer canister.
a) REF DYNJ909699, UDI/DI 40195327446759 (case), 10195327446758 (unit), Lot Numbers: DYNJ909699, DYNJ909699, DYNJ909699, DYNJ909699, DYNJ909699, DYNJ909699; b) REF CDS984000C, UDI/DI 40889942011552 (case), 10889942011551 (unit), Lot Numbers: 22GBT604, 22IBI040, 23ABA955, 23DBT170, 23EBU747, 23HBB409; c) REF DYNJ900847A, UDI/DI 40888277771209 (case), 10888277771208 (unit), Lot Numbers: d) REF DYNJ902619G, UDI/DI 40195327214488 (case), 10195327214487 (unit), Lot Numbers: 22FBG650, 22FBK152, 22FBW287, 22HBA416, 22HBG372, 22HBS529, 22JBA184, 23EBC466, 23HBL953; e) REF CDS984009C, UDI/DI 40888277768971 (case), 10888277768970 (unit), Lot Numbers: 22FBL426, 22GBU282, 23CBD236, 23HBY917; f) REF DYNJ902380K, UDI/DI 40195327507009 (case), 10195327507008 (unit), Lot Numbers: 23KBX257; g) REF DYNJ904771B, UDI/DI 40195327137190 (case), 10195327137199 (unit), Lot Numbers: 22EMB001, 22IMF542, 22KMC063, 23AME421, 23CMC716, 23DME033, 23FMI125, 23HMC273, 23HMI602, 23LME735; h) REF DYNJ905519D, UDI/DI 40193489967396 (case), 10193489967395 (unit), Lot Numbers: 22GMG221, 22HMB217, 22LMJ233, 23AME962, 23AMI318, 23AMJ640, 23DMI673, 23IMG465, 23LMD163; i) REF DYNJ906273F, UDI/DI 40195327511013 (case), 10195327511012 (unit), Lot Numbers: 23JBH985; j) REF DYNJ909747, UDI/DI 40195327468201 (case), 10195327468200 (unit), Lot Numbers: 23HBZ535, 23KBF559; k) REF DYNJS2029D, UDI/DI 40193489214414 (case), 10193489214413 (unit), Lot Numbers: 22IDA884, 22LDA526, 23CDA885, 23FDA459, 23JDB741; l) REF DYNJ907159A, UDI/DI 40193489970419 (case), 10193489970418 (unit), Lot Numbers: 22EME489, 22FMH850, 22IMA282, 22JME894, 22LME340, 23AMB899, 23AMB937, 23AMF082, 23AMG074, 23AMJ063, 23BMF089, 23CMA038, 23FMF222, 23GMB990, 23GMF747, 23HMH495, 23JMC384, 23LMD528; m) REF DYNJ904767A, UDI/DI 40889942850953 (case), 10889942850952 (unit), Lot Numbers: 22EMI503, 22IMC879, 22KMD093, 22LME746, 23AMC120, 23DMA417, 23DMD168, 23FME590, 23GMH321, 23HMG168; n) REF DYNJ905518F, UDI/DI 40195327040407 (case), 10195327040406 (unit), Lot Numbers: 22FMI412, 22HMD694, 22IMA909, 22JMC316, 22KMC860, 22LMD325, 23AMF318, 23BMG871, 23EMF948, 23FMA357, 23GMD311, 23HMG163, 23JMA417, 23KMB755, 23LME256; o) REF DYNJ909938, UDI/DI 40195327511228 (case), 10195327511227 (unit), Lot Numbers: 23JBI008; p) REF DYNJ909939, UDI/DI 40195327511235 (case), 10195327511234 (unit), Lot Numbers: 23JBH778, 23KBU569; q) REF DYNJ906275D, UDI/DI 40195327510986 (case), 10195327510985 (unit), Lot Numbers: 23JBI291, 23LBO829; r) REF DYNJ904769C, UDI/DI 40195327137039 (case), 10195327137038 (unit), Lot Numbers: 22FMB469, 22IMA132, 22KMD488, 22LME616, 23GMF950, 23HME593, 23JMG901, 23KMG320; s) REF DYNJ905511D, UDI/DI 40193489967549 (case), 10193489967548 (unit), Lot Numbers: DYNJ905511D, DYNJ905511D, DYNJ905511D, DYNJ905511D, DYNJ905511D, DYNJ905511D, DYNJ905511D; t) REF DYNJ909749, UDI/DI 40195327468225 (case), 10195327468224 (unit), Lot Numbers: 23IBM342; u) REF DYNJ906274F, UDI/DI 40195327511020 (case), 10195327511029 (unit), Lot Numbers: 23JBQ500, 23LBI456; v) REF CDS984008C, UDI/DI 40888277768988 (case), 10888277768987 (unit), Lot Numbers: 22GBL020, 22JBE903, 23ABI396, 23ABM540, 23CBO804; w) REF DYNJ909212A, UDI/DI 40195327493241 (case), 10195327493240 (unit), Lot Numbers: 23JBA160, 23KBA970; x) REF DYNJ905512F, UDI/DI 40195327225378 (case), 10195327225377 (unit), Lot Numbers: 22JME529, 22KMD611, 22LMI893, 23BMI268, 23CMF328, 23FMG242, 23HMH500, 23IMC073, 23IMG464, 23JMI252, 23LMC546; y) REF DYNJ909704, UDI/DI 40195327446919 (case), 10195327446918 (unit), Lot Numbers: 23GBN059, 23KBA374; z) REF DYNJ907157A, UDI/DI 40193489970358 (case), 10193489970357 (unit), Lot Numbers: 22FMD804, 22JMG489, 22LMC517, 23BMA251, 23CMC137, 23CME146, 23DMA129, 23EMF952, 23FMG076, 23GMI555, 23IME182, 23JMI348, 23KMA447; aa) REF DYNJ902368K, UDI/DI 40195327505890 (case), 10195327505899 (unit), Lot Numbers: 23JBI661, 23KBT583; bb) REF DYNJ905513G, UDI/DI 40193489968027 (case), 10193489968026 (unit), Lot Numbers: 22FMG260, 22JMB517, 22JMF062, 23BME809, 23BMF391, 23BMG366, 23BMG367, 23BMI412, 23EMB428, 23FMA677, 23GMD648, 23JMC096, 23KMB846, 23LMC515; cc) REF DYNJ908375C, UDI/DI 40195327484263 (case), 10195327484262 (unit), Lot Numbers: 23IBL834, 23KBD095, 23LBB392; dd) REF DYNJS2027F, UDI/DI 40195327360314 (case), 10195327360313 (unit), Lot Numbers: 23EDA949, 23GDA203, 23HDB800, 23JDB140, 23KDA685, 23KDC088; ee) REF DYNJ907158A, UDI/DI 40183489970402 (case), 10195327137038 (unit), Lot Numbers: 22EME052, 22FMG234, 22IMC193, 22IMH124, 22LME781, 23CMD417, 23CME132, 23CMF812, 23DMF106, 23GMB888, 23HMH494, 23IME266, 23JMA733, 23KMH135; ff) REF CDS983991C, UDI/DI 40889942131410 (case), 10889942131419 (unit), Lot Numbers: 22FBB136, 22FBU282, 22GBN805, 22HBG300, 22HBS717, 22IBI043, 23CBT616, 23EBA215, 23EBU789, 23HBI490; gg) REF DYNJ902367J, UDI/DI 40195327505838 (case), 10195327505837 (unit), Lot Numbers: 23JBG975, 23KBN775, 23KBP709; hh) REF DYNJ905852C, UDI/DI 40195327214204 (case), 10195327214203 (unit), Lot Numbers: 22IBK464, 22IBM480, 23ABO113, 23ABO114, 23ABQ791, 23DBP733, 23EBR661, 23GBO389, 23HBX998, 23KBG164, 23LBF311; ii) REF DYNJ905514D, UDI/DI 40193489968034 (case), 10193489968033 (unit), Lot Numbers: 22EMA529, 22EMH888, 22FMB880, 22GME946, 22GMH574, 22GMI092, 22HMH887, 22LMD324, 22LMI453, 23AMG866, 23IMG780
Worldwide - US Nationwide and the countries of Panama, Mexico, Canada, Qatar, El Salvador, Oman, Colombia, Singapore.
14977 units