StealthStation" S7 System with Polaris Spectra Camera part number PSU 9733437 - Product Usage: are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures.
Enforcement
- Recall Number
- Z-2707-2020
- Event ID
- 85901
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Medtronic Navigation, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Telephone
- Report Date
- August 5, 2020
- Initiation Date
- June 5, 2020
- Classification Date
- July 27, 2020
- Termination Date
- May 28, 2024
- Address
- 826 Coal Creek Cir, N/A, Louisville, CO, 80027-9710, United States
Description
StealthStation" S7 System with Polaris Spectra Camera part number PSU 9733437 - Product Usage: are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures.
There is a potential for specific camera units to be affected by a characterization issue during manufacturing that results in inaccurate navigation while utilizing both the active and passive trackers simultaneously. While no injuries or illnesses have been reported, If impacted cameras were to be used with both active and passive trackers simultaneously, it would result in system navigation inaccuracy of 6mm.
StealthStation S7 Serial Number N06164844 with Polaris Spectra Camera Serial Number P7-01896; and StealthStation S7 Serial Number N07308697 with Polaris Spectra Camera Serial Number P7-18852.
US Nationwide distribution including in the states of Alabama and Massachusetts.
2