FDA Enforcement Class II Ongoing

Computed Tomography X-ray System; Model uCT 530; Rx; UDI: (01) GTIN: 06971576831012 - Product Usage: intended to produce cross-sectional images of the body using computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac, and vascular).

Recall: Z-2704-2020 · Reported August 5, 2020

Enforcement

Recall Number
Z-2704-2020
Event ID
85868
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Shanghai United Imaging Healthcare Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
August 5, 2020
Initiation Date
June 20, 2020
Classification Date
July 27, 2020
Address
No. 2258 Chengbei Rd, Jiading Ind., N/A, Shanghai, N/A, N/A, China

Description

Computed Tomography X-ray System; Model uCT 530; Rx; UDI: (01) GTIN: 06971576831012 - Product Usage: intended to produce cross-sectional images of the body using computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, cardiac, and vascular).

Reason

The metal edge overlap of the mylar strip in the gantry was uneven, which may create a sharp edge.

Code Info

SN 300162, 300163, 300174, 300176

Distribution

US Nationwide distribution including in the states of CA, IN and TX. See uploaded distribution list for more details.

Quantity

4