FDA Enforcement Class II Ongoing

DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.

Recall: Z-2691-2024 · Reported August 28, 2024

Enforcement

Recall Number
Z-2691-2024
Event ID
95018
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter Mishima K.K.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 28, 2024
Initiation Date
July 10, 2024
Classification Date
August 21, 2024
Address
454-32, Higashino; Nagaizumi-Cho, Sunto-Gun, N/A, Japan

Description

DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.

Reason

Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.

Code Info

Catalog Number: C63520 UDI-DI code: 14987666545065 Serial Numbers: 2023070010 2023060005 2023090015 2023070009 2023080011 2023080012 2023100019 2023100018 2023100020 2024010037 2024010036 2023110022 2024010038 2023090016 2024020040 2024020041 2024020039 2024020043 2024020042 2023110023 2024020045 2024020046 2024010024 2024010025

Distribution

Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, South Africa.

Quantity

24 analyzers