FDA Enforcement
Class II
Terminated
Justright Surgical, JustRight 5mm Stapler, REF JR-ST25-2.0, Sterile, EO, Rx Only, JR-ST25-2.0 product is packaged in cases of 6. The JustRight 5mm Stapler is intended for use in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomosis.
Recall: Z-2690-2014
·
Reported October 1, 2014
Enforcement
- Recall Number
- Z-2690-2014
- Event ID
- 69140
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Justright Surgical, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- October 1, 2014
- Initiation Date
- August 27, 2014
- Classification Date
- September 23, 2014
- Termination Date
- October 6, 2014
- Address
- 6325 Gunpark Dr, Suite G, Boulder, CO, 80301-3592, United States
Description
Justright Surgical, JustRight 5mm Stapler, REF JR-ST25-2.0, Sterile, EO, Rx Only, JR-ST25-2.0 product is packaged in cases of 6. The JustRight 5mm Stapler is intended for use in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomosis.
Reason
JustRight Surgical is recalling specific lots of JustRight 5mm Stapler and Reload products because they may deliver poor distal staple formation, past the end of the knife cut.
Code Info
Lot: 75DD0406
Distribution
Distributed in the states of Colorado, California and New York.
Quantity
3 cases containing 18 units