FDA Enforcement Class II Terminated

Sensis Vibe Systems with Software Version VD10B, Model Numbers 10765502, 10910620, 11007641, 6648161 --- Programmable diagnostic computer,

Recall: Z-2688-2017 · Reported July 12, 2017

Enforcement

Recall Number
Z-2688-2017
Event ID
77324
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 12, 2017
Initiation Date
May 16, 2017
Classification Date
July 6, 2017
Termination Date
August 22, 2018
Address
40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States

Description

Sensis Vibe Systems with Software Version VD10B, Model Numbers 10765502, 10910620, 11007641, 6648161 --- Programmable diagnostic computer,

Reason

Software error. In Sensis Vibe systems with software version VD10B, a software error can result in: problems generating a report and/ or - information from different examinations of the same patient being combined into one report. --- The error causes information from two examinations to be combined into one report. This does not result in a hazardous situation; however, if the erroneous report has been used for treatment or diagnosis, this can potentially result in an incorrect treatment or diagnosis.

Code Info

Software Version VD10B

Distribution

US Distribution to the states of : IN, MI, IA

Quantity

US: 13 units