FDA Enforcement
Class II
Terminated
Sensis Vibe Systems with Software Version VD10B, Model Numbers 10765502, 10910620, 11007641, 6648161 --- Programmable diagnostic computer,
Recall: Z-2688-2017
·
Reported July 12, 2017
Enforcement
- Recall Number
- Z-2688-2017
- Event ID
- 77324
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 12, 2017
- Initiation Date
- May 16, 2017
- Classification Date
- July 6, 2017
- Termination Date
- August 22, 2018
- Address
- 40 Liberty Blvd, N/A, Malvern, PA, 19355-1418, United States
Description
Sensis Vibe Systems with Software Version VD10B, Model Numbers 10765502, 10910620, 11007641, 6648161 --- Programmable diagnostic computer,
Reason
Software error. In Sensis Vibe systems with software version VD10B, a software error can result in: problems generating a report and/ or - information from different examinations of the same patient being combined into one report. --- The error causes information from two examinations to be combined into one report. This does not result in a hazardous situation; however, if the erroneous report has been used for treatment or diagnosis, this can potentially result in an incorrect treatment or diagnosis.
Code Info
Software Version VD10B
Distribution
US Distribution to the states of : IN, MI, IA
Quantity
US: 13 units