FDA Enforcement
Class II
Terminated
Invivo Sentinelle Venguard Breast MRI Auxiliary Table/Tabletop with 8/16 Channel Coil Array for GE 1.5T/3T MRI Systems.
Recall: Z-2688-2016
·
Reported September 7, 2016
Enforcement
- Recall Number
- Z-2688-2016
- Event ID
- 74064
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Invivo Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- September 7, 2016
- Initiation Date
- May 6, 2016
- Classification Date
- August 30, 2016
- Termination Date
- August 21, 2019
- Address
- 3545 SW 47th Ave, N/A, Gainesville, FL, 32608-7691, United States
Description
Invivo Sentinelle Venguard Breast MRI Auxiliary Table/Tabletop with 8/16 Channel Coil Array for GE 1.5T/3T MRI Systems.
Reason
A diode was inadvertently not assembled into the system, which may result in coil overheating
Code Info
Model #: 9896-032-14382; Serial #: 13 (production order 301128446), 14 (production order 301153586), and 15 (production order 301153587).
Distribution
WI
Quantity
3