The Verigene Respiratory Virus Plus Nucleic Acid Test (RV+) on the Verigene System is a qualitative nucleic acid multiplex test intended to simultaneously detect and identify multiple respiratory virus nucleic acids in nasopharyngeal (NP) swab specimens from individuals with signs and symptoms of respiratory tract infection. The following virus types and subtypes are identified using the RV+: Influenza A, Influenza A subtype H1, Influenza A subtype H3, 2009 H1N1, Influenza B, Respiratory Syncytial Virus (RSV) subtype A, and RSV subtype B. The test is not intended to detect Influenza C virus. Detecting and identifying specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection, if used in conjunction with other clinical and laboratory findings. Negative results for Influenza A, Influenza B, or RSV do not preclude influenza virus or RSV infection and should not be used as the sole basis for diagnosis, treatment, or patient management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered in order to obtain the final diagnosis of respiratory viral infection. Performance characteristics for Influenza A Virus were established when Influenza A/H3, A/H1, and 2009 H1N1 were the predominant Influenza A viruses circulating. These characteristics may vary when other Influenza A viruses are emerging. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions used specifically for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Enforcement
- Recall Number
- Z-2684-2014
- Event ID
- 69034
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Nanosphere, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 1, 2014
- Initiation Date
- October 10, 2013
- Classification Date
- September 23, 2014
- Termination Date
- June 16, 2015
- Address
- 4088 Commercial Ave, N/A, Northbrook, IL, 60062-1829, United States
Description
The Verigene Respiratory Virus Plus Nucleic Acid Test (RV+) on the Verigene System is a qualitative nucleic acid multiplex test intended to simultaneously detect and identify multiple respiratory virus nucleic acids in nasopharyngeal (NP) swab specimens from individuals with signs and symptoms of respiratory tract infection. The following virus types and subtypes are identified using the RV+: Influenza A, Influenza A subtype H1, Influenza A subtype H3, 2009 H1N1, Influenza B, Respiratory Syncytial Virus (RSV) subtype A, and RSV subtype B. The test is not intended to detect Influenza C virus. Detecting and identifying specific viral nucleic acids from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection, if used in conjunction with other clinical and laboratory findings. Negative results for Influenza A, Influenza B, or RSV do not preclude influenza virus or RSV infection and should not be used as the sole basis for diagnosis, treatment, or patient management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered in order to obtain the final diagnosis of respiratory viral infection. Performance characteristics for Influenza A Virus were established when Influenza A/H3, A/H1, and 2009 H1N1 were the predominant Influenza A viruses circulating. These characteristics may vary when other Influenza A viruses are emerging. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions used specifically for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.
Product Number: 20-009-020; Lot Number and Expiration Date: Lot: 091113020B, Expiration Date: 03/10/14
Worldwide Distribution - USA including AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, South Korea, and Spain.
780 units