FDA Enforcement Class II Ongoing

Alinity hq Analyzer, REF: 09P68-01

Recall: Z-2681-2024 · Reported August 28, 2024

Enforcement

Recall Number
Z-2681-2024
Event ID
94958
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Abbott Laboratories
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 28, 2024
Initiation Date
June 14, 2024
Classification Date
August 16, 2024
Address
4551 Great America Pkwy, Santa Clara, CA, 95054-1208, United States

Description

Alinity hq Analyzer, REF: 09P68-01

Reason

1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating falsely high Mean Cell Hemoglobin(MCH)/MCHC and falsely low hematocrit results. 2)Cell events may be incorrectly counted as basophil(BASO), resulting in increased BASO and %BASO counts

Code Info

UDI-DI: 380740138851. Software Versions: 5.7 (US) and 5.6 and below (OUS), Serial Numbers: HQ00105, HQ00108, HQ00117, HQ00118, HQ00119, HQ00120, HQ00126, HQ00127, HQ00128, HQ00132, HQ00136, HQ00139, HQ00140, HQ00141, HQ00142, HQ00144, HQ00145, HQ00147, HQ00148, HQ00149, HQ00152, HQ00153, HQ00154, HQ00156, HQ00157, HQ00161, HQ00162, HQ00166, HQ00169, HQ00171, HQ00172, HQ00173, HQ00174, HQ00175, HQ00183, HQ00184, HQ00185, HQ00186, HQ00187, HQ00188, HQ00266, HQ00268, HQ00269, HQ00270, HQ00271, HQ00273, HQ00274, HQ00276, HQ00277, HQ00278, HQ00279, HQ00281, HQ00282, HQ00283, HQ00286, HQ00287, HQ00289, HQ00291, HQ00295, HQ00296, HQ00299, HQ00300, HQ00302, HQ00303, HQ00312, HQ00313, HQ00314, HQ00316, HQ00318, HQ00319, HQ00321, HQ00324, HQ00327, HQ00328, HQ00329, HQ00331, HQ00332, HQ00334, HQ00335, HQ00336, HQ00337, HQ00338, HQ00417, HQ00418, HQ00419, HQ00420, HQ00422, HQ00424, HQ00425, HQ00430, HQ00432, HQ00433, HQ00434, HQ00435, HQ00436, HQ00438, HQ00439, HQ00440, HQ00441, HQ00442, HQ00443, HQ00444, HQ00445, HQ00446, HQ00447, HQ00448, HQ00451, HQ00453, HQ00455, HQ00456, HQ00457, HQ00458, HQ00460, HQ00461, HQ00463, HQ00464, HQ00466, HQ00468, HQ00469, HQ00471, HQ00472, HQ00474, HQ00475, HQ00476, HQ00536, HQ00539, HQ00540, HQ00541, HQ00542, HQ00544, HQ00545, HQ00546, HQ00547, HQ00548, HQ00550, HQ00551, HQ00552, HQ00553, HQ00554, HQ00555, HQ00558, HQ00561, HQ00562, HQ00563, HQ00564, HQ00565, HQ00567, HQ00568, HQ00569, HQ00570, HQ00571, HQ00573, HQ00574, HQ00575, HQ00576, HQ00577, HQ00578, HQ00579, HQ00580, HQ00581, HQ00582, HQ00583, HQ00584, HQ00585, HQ00586, HQ00587, HQ00638, HQ00639, HQ00640, HQ00641, HQ00642, HQ00643, HQ00644, HQ00645, HQ00646, HQ00647, HQ00648, HQ00649, HQ00650, HQ00651, HQ00654, HQ00655, HQ00657, HQ00660, HQ00662, HQ00663, HQ00665, HQ00666, HQ00667, HQ00668, HQ00669, HQ00670, HQ00671, HQ00672, HQ00673, HQ00674, HQ00675, HQ00676, HQ00677, HQ00678, HQ00679, HQ00680, HQ00681, HQ00682, HQ00683, HQ00684, HQ00685, HQ00686, HQ00731, HQ00732, HQ00733, HQ00734, HQ00735, HQ00736, HQ00737, HQ00738, HQ00740, HQ00741, HQ00742, HQ00743, HQ00744, HQ00745, HQ00746, HQ00747, HQ00748, HQ00749, HQ00750, HQ00751, HQ00753, HQ00754, HQ00756, HQ00757, HQ00758, HQ00759, HQ00760, HQ00761, HQ00762, HQ00763, HQ00764, HQ00765, HQ00766, HQ00767, HQ00768, HQ00769, HQ00770, HQ00771, HQ00772, HQ00773, HQ00774, HQ00775, HQ00819, HQ00820, HQ00821, HQ00822, HQ00823, HQ00824, HQ00825, HQ00826, HQ00827, HQ00828, HQ00829, HQ00830, HQ00831, HQ00832, HQ00833, HQ00834, HQ00835, HQ00836, HQ00837, HQ00838, HQ00839, HQ00840, HQ00841, HQ00842, HQ00844, HQ00845, HQ00846, HQ00847, HQ00848, HQ00849, HQ00852, HQ00853, HQ00856, HQ00857, HQ00858, HQ00859, HQ00860, HQ00861, HQ00864, HQ00865, HQ00866, HQ00867, HQ00923, HQ00924, HQ00925, HQ00926, HQ00927, HQ00929, HQ00932, HQ00933, HQ00934, HQ00935, HQ00936, HQ00937, HQ00938, HQ00939, HQ00940, HQ00941, HQ00942, HQ00943, HQ00944, HQ00945, HQ00946, HQ00947, HQ00949, HQ00950, HQ00951, HQ00952, HQ00953, HQ00954, HQ00955, HQ00956, HQ00957, HQ00958, HQ00959, HQ00960, HQ00961, HQ00962, HQ00963, HQ00964, HQ00965, HQ00966, HQ00967, HQ00968, HQ01018, HQ01019, HQ01020, HQ01021, HQ01022, HQ01023, HQ01024, HQ01025, HQ01026, HQ01027, HQ01028, HQ01029, HQ01030, HQ01031, HQ01032, HQ01033, HQ01034, HQ01035, HQ01037, HQ01038, HQ01039, HQ01040, HQ01043, HQ01044, HQ01046, HQ01047, HQ01049, HQ01050, HQ01051, HQ01052, HQ01054, HQ01055, HQ01056, HQ01057, HQ01058, HQ01060, HQ01061, HQ01062, HQ01063, HQ01064, HQ01065, HQ01067, HQ01122, HQ01123, HQ01124, HQ01125, HQ01126, HQ01127, HQ01128, HQ01129, HQ01130, HQ01131, HQ01132, HQ01133, HQ01134, HQ01135, HQ01136, HQ01137, HQ01138, HQ01139, HQ01140, HQ01141, HQ01142, HQ01143, HQ01144, HQ01145, HQ01146, HQ01147, HQ01148, HQ01149, HQ01150, HQ01151, HQ01152, HQ01153, HQ00191, HQ00192, HQ00193, HQ00194, HQ00196, HQ00197, HQ00198, HQ00201, HQ00202, HQ00203, HQ00204, HQ00208, HQ00209, HQ00210, HQ00214, HQ00215, HQ00217, HQ00218, HQ00222, HQ00223, HQ00230, HQ00231, HQ00233, HQ00242, HQ00243, HQ00244, HQ00245, HQ00246, HQ00248, HQ00250, HQ00252, HQ00254, HQ00255, HQ00257, HQ00258, HQ00259, HQ00260, HQ00261, HQ00264, HQ00265, HQ00343, HQ00344, HQ00346, HQ00347, HQ00348, HQ00349, HQ00350, HQ00352, HQ00353, HQ00356, HQ00357, HQ00358, HQ00359, HQ00361, HQ00362, HQ00364, HQ00365, HQ00368, HQ00371, HQ00372, HQ00376, HQ00378, HQ00379, HQ00380, HQ00382, HQ00383, HQ00384, HQ00385, HQ00386, HQ00387, HQ00389, HQ00390, HQ00391, HQ00393, HQ00404, HQ00406, HQ00410, HQ00411, HQ00413, HQ00414, HQ00415, HQ00416, HQ00477, HQ00478, HQ00479, HQ00481, HQ00483, HQ00484, HQ00485, HQ00487, HQ00489, HQ00490, HQ00491, HQ00492, HQ00493, HQ00494, HQ00495, HQ00496, HQ00497, HQ00498, HQ00499, HQ00500, HQ00501, HQ00503, HQ00504, HQ00505, HQ00506, HQ00507, HQ00508, HQ00509, HQ00510, HQ00511, HQ00512, HQ00513, HQ00514, HQ00516, HQ00517, HQ00519, HQ00524, HQ00527, HQ00528, HQ00530, HQ00534, HQ00535, HQ00588, HQ00589, HQ00590, HQ00591, HQ00592, HQ00593, HQ00594, HQ00595, HQ00596, HQ00598, HQ00599, HQ00600, HQ00601, HQ00602, HQ00603, HQ00604, HQ00605, HQ00606, HQ00607, HQ00608, HQ00609, HQ00610, HQ00611, HQ00612, HQ00613, HQ00614, HQ00615, HQ00616, HQ00620, HQ00623, HQ00625, HQ00626, HQ00627, HQ00628, HQ00629, HQ00630, HQ00631, HQ00632, HQ00633, HQ00634, HQ00635, HQ00637, HQ00687, HQ00688, HQ00689, HQ00690, HQ00691, HQ00692, HQ00693, HQ00694, HQ00695, HQ00696, HQ00697, HQ00698, HQ00699, HQ00700, HQ00701, HQ00703, HQ00704, HQ00705, HQ00706, HQ00707, HQ00708, HQ00709, HQ00710, HQ00711, HQ00712, HQ00713, HQ00715, HQ00716, HQ00717, HQ00718, HQ00719, HQ00720, HQ00721, HQ00722, HQ00723, HQ00724, HQ00725, HQ00726, HQ00727, HQ00728, HQ00729, HQ00730, HQ00776, HQ00777, HQ00778, HQ00779, HQ00780, HQ00781, HQ00782, HQ00783, HQ00785, HQ00786, HQ00787, HQ00788, HQ00789, HQ00790, HQ00791, HQ00792, HQ00793, HQ00794, HQ00795, HQ00796, HQ00797, HQ00798, HQ00799, HQ00800, HQ00801, HQ00802, HQ00803, HQ00804, HQ00805, HQ00806, HQ00807, HQ00808, HQ00809, HQ00810, HQ00811, HQ00812, HQ00813, HQ00814, HQ00815, HQ00816, HQ00817, HQ00818, HQ00868, HQ00870, HQ00871, HQ00873, HQ00874, HQ00875, HQ00876, HQ00879, HQ00880, HQ00881, HQ00882, HQ00883, HQ00885, HQ00887, HQ00888, HQ00890, HQ00891, HQ00892, HQ00893, HQ00894, HQ00895, HQ00896, HQ00899, HQ00900, HQ00901, HQ00902, HQ00903, HQ00904, HQ00905, HQ00906, HQ00907, HQ00911, HQ00913, HQ00914, HQ00915, HQ00916, HQ00917, HQ00918, HQ00919, HQ00920, HQ00921, HQ00922, HQ00969, HQ00970, HQ00971, HQ00972, HQ00973, HQ00974, HQ00975, HQ00976, HQ00977, HQ00978, HQ00979, HQ00980, HQ00981, HQ00982, HQ00983, HQ00984, HQ00985, HQ00986, HQ00987, HQ00989, HQ00990, HQ00991, HQ00994, HQ00995, HQ00996, HQ00997, HQ00998, HQ00999, HQ01000, HQ01001, HQ01002, HQ01003, HQ01006, HQ01007, HQ01008, HQ01009, HQ01011, HQ01013, HQ01014, HQ01015, HQ01016, HQ01017, HQ01068, HQ01069, HQ01070, HQ01071, HQ01072, HQ01073, HQ01074, HQ01075, HQ01076, HQ01077, HQ01078, HQ01079, HQ01080, HQ01081, HQ01082, HQ01083, HQ01084, HQ01085, HQ01086, HQ01087, HQ01088, HQ01089, HQ01091, HQ01092, HQ01093, HQ01094, HQ01095, HQ01096, HQ01097, HQ01098, HQ01102, HQ01103, HQ01104, HQ01105, HQ01108, HQ01113, HQ01114, HQ01116, HQ01118, HQ01119, HQ01120, HQ01121, HQ01155, HQ01156, HQ01157, HQ01158, HQ01159, HQ01160, HQ01161, HQ01162, HQ01165, HQ01167, HQ01168, HQ01169, HQ01170, HQ01171, HQ01172, HQ01173, HQ01174, HQ01175, HQ01177, HQ01180, HQ01181, HQ01183, HQ01187, HQ01188, HQ01190, HQ01196, HQ01231, HQ01233, HQ01236, HQ01237, HQ01244, HQ01246

Distribution

Worldwide - US Nationwide distribution in the states of NY, OK, NC and the countries of ALBANIA, ARGENTINA, AUSTRALIA, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, Botswana, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FRANCE, GERMANY, GREECE, HONDURAS, HONG KONG, HUNGARY, INDIA, INDONESIA, ITALY, JAMAICA, JAPAN, JORDAN, KUWAIT, LEBANON, LITHUANIA, MALAYSIA, MALDIVES, MEXICO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, SERBIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UK, URUGUAY, VIETNAM.

Quantity

816