FDA Enforcement Class II Terminated

The Verigene Gram Negative Blood Culture Nucleic Acid Test (BC-GN), performed using the sample-to-result Verigene System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-negative bacteria and resistance markers. BC-GN is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram negative bacteria as determined by gram stain.

Recall: Z-2681-2014 · Reported October 1, 2014

Enforcement

Recall Number
Z-2681-2014
Event ID
69034
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Nanosphere, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 1, 2014
Initiation Date
October 10, 2013
Classification Date
September 23, 2014
Termination Date
June 16, 2015
Address
4088 Commercial Ave, N/A, Northbrook, IL, 60062-1829, United States

Description

The Verigene Gram Negative Blood Culture Nucleic Acid Test (BC-GN), performed using the sample-to-result Verigene System, is a qualitative multiplexed in vitro diagnostic test for the simultaneous detection and identification of selected gram-negative bacteria and resistance markers. BC-GN is performed directly on blood culture media using blood culture bottles identified as positive by a continuous monitoring blood culture system and which contain gram negative bacteria as determined by gram stain.

Reason

Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies for the following products and lots of Extraction Trays.

Code Info

Product Number: 20-009-021; Lot Number and Expiration Date: Lot: 100113021C, Expiration Date: 03/30/14

Distribution

Worldwide Distribution - USA including AZ, CA, FL, GA, IA, IL, IN, KY, MA, MD, MN, MO, MT, NC, NE, NJ, NY, OH, OR, PA, TX, UT, VA, WA, WI, OH and Internationally to Bulgaria, Germany, Italy, Japan, South Korea, and Spain.

Quantity

400 units