FDA Enforcement Class II Terminated

rHead Stem Implant Plasma Coated, Size 4. Orthopedic implant for replacement of the proximal end of the radius in the Stryker rHead Radial Head and Uni-Elbow system.

Recall: Z-2680-2018 · Reported August 15, 2018

Enforcement

Recall Number
Z-2680-2018
Event ID
79315
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Howmedica Osteonics Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 15, 2018
Initiation Date
October 24, 2017
Classification Date
August 8, 2018
Termination Date
May 18, 2023
Address
325 Corporate Dr, N/A, Mahwah, NJ, 07430-2006, United States

Description

rHead Stem Implant Plasma Coated, Size 4. Orthopedic implant for replacement of the proximal end of the radius in the Stryker rHead Radial Head and Uni-Elbow system.

Reason

Revision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.

Code Info

310-2013

Distribution

Distribution US Nationwide.

Quantity

301