FDA Enforcement Class II Terminated

The synthetic non-absorbable surgical sutures CARDIOFLON Evolution are composed of polyester [poly(ethylene terephtalate)] braids, coated, with or without PLEDGETS of different sizes. CARDIOFLON Evolution sutures are green dyed (D&C green # 6, CI 61565) and white (undyed). Some packs are composed of green and white braids to facilitate location during use. CARDIOFLON Evolution sutures comply with the requirements of the United States Pharmacopoeia (USP) and the European Pharmacopoeia (EP) for non-absorbable surgical sutures, except for some diameters. Type of packaging: - Unit: Sutures are packed in a support card Onepak or in a pack under an external polyethylene/ paper peel pouch. - Box: 36 units for single or double armed (36 sutures), 24 units of ligatures (24 reels), 12 packs for double armed (10 Green or 5 Green/5 White) (120 sutures).

Recall: Z-2676-2014 · Reported October 1, 2014

Enforcement

Recall Number
Z-2676-2014
Event ID
68721
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Peters Surgical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 1, 2014
Initiation Date
June 25, 2014
Classification Date
September 22, 2014
Termination Date
December 13, 2017
Address
42 Rue Benoit Frachon, N/A, Bobigny, N/A, N/A, France

Description

The synthetic non-absorbable surgical sutures CARDIOFLON Evolution are composed of polyester [poly(ethylene terephtalate)] braids, coated, with or without PLEDGETS of different sizes. CARDIOFLON Evolution sutures are green dyed (D&C green # 6, CI 61565) and white (undyed). Some packs are composed of green and white braids to facilitate location during use. CARDIOFLON Evolution sutures comply with the requirements of the United States Pharmacopoeia (USP) and the European Pharmacopoeia (EP) for non-absorbable surgical sutures, except for some diameters. Type of packaging: - Unit: Sutures are packed in a support card Onepak or in a pack under an external polyethylene/ paper peel pouch. - Box: 36 units for single or double armed (36 sutures), 24 units of ligatures (24 reels), 12 packs for double armed (10 Green or 5 Green/5 White) (120 sutures).

Reason

A potential splitting of the surface treatment of the thread ends on certain sutures has recently been identified.

Code Info

1) Item reference: 19P20Q , Batch number: 1135 , Expiration date: 10/2018 , GTIN on box label: 3661522034989 ; 2) Item reference: 19P30AC , Batch number: 942720 , Expiration date: 03/2018 , GTIN on box label: 3661522034446 ; 3) Item reference: 19P30AE , Batch number: 7627 , Expiration date: 04/2019 , GTIN on box label: 3661522034477 ; 4) Item reference: 19P30AE , Batch number: 942910 , Expiration date: 02/2018 , GTIN on box label: 3661522034477 ; 5) Item reference: 19P30AE , Batch number: 971670 , Expiration date: 05/2018 , GTIN on box label: 3661522034477 ; 6) Item reference: 19P30AK , Batch number: 942530 , Expiration date: 03/2018 , GTIN on box label: 3661522034552 ; 7) Item reference: 19P30AK , Batch number: 978580 , Expiration date: 06/2018 , GTIN on box label: 3661522034552 ; 8) Item reference: 19P30AN , Batch number: 942500 , Expiration date: 02/2018 , GTIN on box label: 3661522034576 ; 9) Item reference: 19P30AQ , Batch number: 2480 , Expiration date: 10/2018 , GTIN on box label: 3661522034606 ; 10) Item reference: 19P30AQ , Batch number: 5508 , Expiration date: 02/2019 , GTIN on box label: 3661522034606 ; 11) Item reference: 19P30AQ , Batch number: 994990 , Expiration date: 07/2018 , GTIN on box label: 3661522034606 ; 12) Item reference: 19P30AQ , Batch number: 1008710 , Expiration date: 08/2018 , GTIN on box label: 3661522034606 ; 13) Item reference: 19P30N , Batch number: 942420 , Expiration date: 02/2018 , GTIN on box label: 3661522034262 ; 14) Item reference: 19P30R , Batch number: 4370 , Expiration date: 01/2019 , GTIN on box label: 3661522034279 ; 15) Item reference: 19P30R , Batch number: 5021 , Expiration date: 01/2019 , GTIN on box label: 3661522034279 ; 16) Item reference: 19P30R , Batch number: 5557 , Expiration date: 02/2019 , GTIN on box label: 3661522034279 ; 17) Item reference: 19P30R , Batch number: 948700 , Expiration date: 03/2018 , GTIN on box label: 3661522034279 ; 18) Item reference: 19P30R , Batch number: 978410 , Expiration date: 06/2018 , GTIN on box label: 3661522034279 ; 19) Item reference: 19P30S , Batch number: 994580 , Expiration date: 07/2018 , GTIN on box label: 3661522034323 ; 20) Item reference: 19P30T , Batch number: 4371 , Expiration date: 01/2019 , GTIN on box label: 3661522034347 ; 21) Item reference: 19P30T , Batch number: 4974 , Expiration date: 02/2019 , GTIN on box label: 3661522034347 ; 22) Item reference: 19P30T , Batch number: 948740 , Expiration date: 03/2018 , GTIN on box label: 3661522034347 ; 23) Item reference: 19P30U , Batch number: 948770 , Expiration date: 03/2018 , GTIN on box label: 3661522034354 ; 24) Item reference: 19P30V , Batch number: 4372 , Expiration date: 01/2019 , GTIN on box label: 3661522034286 ; 25) Item reference: 19P30V , Batch number: 948790 , Expiration date: 03/2018 , GTIN on box label: 3661522034286 ; 26) Item reference: 19P30V , Batch number: 992210 , Expiration date: 07/2018 , GTIN on box label: 3661522034286 ; 27) Item reference: 19P30X , Batch number: 967150 , Expiration date: 05/2018 , GTIN on box label: 3661522034392 ; 28) Item reference: 19P30X , Batch number: 999390 , Expiration date: 09/2018 , GTIN on box label: 3661522034392 ; 29) Item reference: 19S15A , Batch number: 2192 , Expiration date: 10/2018 , GTIN on box label: 3661522035009 ; 30) Item reference: 19S15G , Batch number: 2195 , Expiration date: 10/2018 , GTIN on box label: 3661522035061 ; 31) Item reference: 19S15G , Batch number: 2440 , Expiration date: 10/2018 , GTIN on box label: 3661522035061 ; 32) Item reference: 19S15G , Batch number: 942960 , Expiration date: 02/2018 , GTIN on box label: 3661522035061 ; 33) Item reference: 19S15G , Batch number: 970060 , Expiration date: 05/2018 , GTIN on box label: 3661522035061 ; 34) Item reference: 19S15G , Batch number: 982470 , Expiration date: 06/2018 , GTIN on box label: 3661522035061 ; 35) Item reference: 19S15G , Batch number: 994930 , Expiration date: 07/2018 , GTIN on box label: 3661522035061 ; 36) Item reference: 19S20J , Batch number: 942940 , Expiration date: 05/2018 , GTIN on box label: 3661522034910 ; 37) Item reference: 19S20J , Batch number: 966100 , Expiration date: 08/2018 , GTIN on box label: 3661522034910 ; 38) Item reference: 19S20R , Batch number: 1872 , Expiration date: 09/2018 , GTIN on box label: 3661522037539 ; 39) Item reference: 19S30AB , Batch number: 7057 , Expiration date: 04/2019 , GTIN on box label: 3661522034439 ; 40) Item reference: 19S30AB , Batch number: 991250 , Expiration date: 07/2018 , GTIN on box label: 3661522034439 ; 41) Item reference: 19S30AB , Batch number: 999910 , Expiration date: 08/2018 , GTIN on box label: 3661522034439 ; 42) Item reference: 19S30AE , Batch number: 942400 , Expiration date: 03/2018 , GTIN on box label: 3661522034460 ; 43) Item reference: 19S30AE , Batch number: 984600 , Expiration date: 06/2018 , GTIN on box label: 3661522034460 ; 44) Item reference: 19S30AK , Batch number: 942510 , Expiration date: 02/2018 , GTIN on box label: 3661522034538 ; 45) Item reference: 19S30AK , Batch number: 992600 , Expiration date: 07/2018 , GTIN on box label: 3661522034538 ; 46) Item reference: 19S30AK , Batch number: 1011370 , Expiration date: 09/2018 , GTIN on box label: 3661522034538 ; 47) Item reference: 19S30AQ , Batch number: 942580 , Expiration date: 03/2018 , GTIN on box label: 3661522034590 ; 48) Item reference: 19S30AW , Batch number: 1485 , Expiration date: 09/2018 , GTIN on box label: 3661522037492 ; 49) Item reference: 19S30AX , Batch number: 1486 , Expiration date: 09/2018 , GTIN on box label: 3661522037508 ; 50) Item reference: 19S30AY , Batch number: 1487 , Expiration date: 09/2018 , GTIN on box label: 3661522037515 ; 51) Item reference: 19S30B , Batch number: 3403 , Expiration date: 01/2019 , GTIN on box label: 3661522034118 ; 52) Item reference: 19S30B , Batch number: 5136 , Expiration date: 02/2019 , GTIN on box label: 3661522034118 ; 53) Item reference: 19S30B , Batch number: 6222 , Expiration date: 02/2019 , GTIN on box label: 3661522034118 ; 54) Item reference: 19S30B , Batch number: 942350 , Expiration date: 03/2018 , GTIN on box label: 3661522034118 ; 55) Item reference: 19S30B , Batch number: 1009510 , Expiration date: 08/2018 , GTIN on box label: 3661522034118 ; 56) Item reference: 19S30C , Batch number: 3995 , Expiration date: 01/2019 , GTIN on box label: 3661522034125 ; 57) Item reference: 19S30C , Batch number: 4920 , Expiration date: 01/2019 , GTIN on box label: 3661522034125 ; 58) Item reference: 19S30C , Batch number: 5158 , Expiration date: 02/2019 , GTIN on box label: 3661522034125 ; 59) Item reference: 19S30C , Batch number: 6554 , Expiration date: 03/2019 , GTIN on box label: 3661522034125 ; 60) Item reference: 19S30C , Batch number: 977760 , Expiration date: 06/2018 , GTIN on box label: 3661522034125 ; 61) Item reference: 19S30C , Batch number: 988910 , Expiration date: 07/2018 , GTIN on box label: 3661522034125 ; 62) Item reference: 19S30Z , Batch number: 942670 , Expiration date: 03/2018 , GTIN on box label: 3661522034415 ; 63) Item reference: 19S35A , Batch number: 942760 , Expiration date: 04/2018 , GTIN on box label: 3661522034675

Distribution

Worldwide Distribution- US: Nationwide (CA, FL, GA, IL, IN, LA, MD, MI, MO, MS, NC, NY, OH, OR, PA, SC, TX, VA, WA); and countries: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, CAMBODIA, CHINA, CYPRUS, CZECH REPUBLIC, DOMINICAN REPUBLIC, EGYPT, FRANCE, GEORGIA, GERMANY, HUNGARY, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, LEBANON, LIBYA, MALAYSIA, MAURITIUS, MOLDOVA, MOROCCO, MYANMAR, NETHERLANDS, NEW ZEALAND, NORTH KOREA, PAKISTAN, PERU, POLAND, ROMANIA, RUSSIA, SAUDI ARABIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWITZERLAND, TAIWAN, THAILAND, UNITED KINGDOM, VENEZUELA, VIETNAM, YEMEN

Quantity

22,190 units