FDA Enforcement Class II Terminated

Diamond Coated Tips and Burs are accessories to the air powered dental handpiece (S-MAX pico) The intended use of the tips is based on the parent device. For burs used with the dental handpieces, the intended use is general dental actions such as removing carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations and restorations and polishing teeth.

Recall: Z-2674-2016 · Reported September 7, 2016

Enforcement

Recall Number
Z-2674-2016
Event ID
74666
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Nakanishi Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 7, 2016
Initiation Date
July 8, 2016
Classification Date
August 26, 2016
Termination Date
June 25, 2018
Address
700, Shimohinata, Kanuma, N/A, N/A, Japan

Description

Diamond Coated Tips and Burs are accessories to the air powered dental handpiece (S-MAX pico) The intended use of the tips is based on the parent device. For burs used with the dental handpieces, the intended use is general dental actions such as removing carious material, reducing hard tooth structure, cavity preparation, finishing tooth preparations and restorations and polishing teeth.

Reason

Through an error, the labeling failed to include the symbol or other text indicating that the diamond coated tips were single use only. A contributing factor to this error is that in other markets such as Japan, the diamond coated tips are multiple use and are not limited to single use only. The tips are stainless steel with a coating of fine diamond powder. The tips are sold non-sterile and are to be sterilized prior to first use. The tips were cleared as single use only. This problem was discovered during an ongoing, proactive, quality system improvement plan that was initiated by Nakanishi in October 2014. Because the correct instructions for sterilization are not available, it is possible that the devices might not be sterile prior to use. The absence of the single use statement introduces a risk of cross-contamination if the tips are reprocessed and reused.

Code Info

Waiting for clarification from the firm.

Distribution

Nationwide Distribution to Illinois, Florida and Georgia.

Quantity

4284 units