Diamond Coated Tips and Burs are accessories to the Ultrasonic Scaler (VARIOS 350 LUX, VARIOS 350, VARIOS 150 LUX, VARIOS 150). The intended use of the tips is based on the parent device. For the ultrasonic scalers and the sonic air scaler, the tips are used for various actions such as scaling (removal of calculus), periodontics (root planing and maintenance), endodontics (root canal enlarging, irrigation, and filling), restorative (caries removal and tooth surface detailing).
Enforcement
- Recall Number
- Z-2672-2016
- Event ID
- 74666
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Nakanishi Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- September 7, 2016
- Initiation Date
- July 8, 2016
- Classification Date
- August 26, 2016
- Termination Date
- June 25, 2018
- Address
- 700, Shimohinata, Kanuma, N/A, N/A, Japan
Description
Diamond Coated Tips and Burs are accessories to the Ultrasonic Scaler (VARIOS 350 LUX, VARIOS 350, VARIOS 150 LUX, VARIOS 150). The intended use of the tips is based on the parent device. For the ultrasonic scalers and the sonic air scaler, the tips are used for various actions such as scaling (removal of calculus), periodontics (root planing and maintenance), endodontics (root canal enlarging, irrigation, and filling), restorative (caries removal and tooth surface detailing).
Through an error, the labeling failed to include the symbol or other text indicating that the diamond coated tips were single use only. A contributing factor to this error is that in other markets such as Japan, the diamond coated tips are multiple use and are not limited to single use only. The tips are stainless steel with a coating of fine diamond powder. The tips are sold non-sterile and are to be sterilized prior to first use. The tips were cleared as single use only. This problem was discovered during an ongoing, proactive, quality system improvement plan that was initiated by Nakanishi in October 2014. Because the correct instructions for sterilization are not available, it is possible that the devices might not be sterile prior to use. The absence of the single use statement introduces a risk of cross-contamination if the tips are reprocessed and reused.
Waiting for clarification from the firm.
Nationwide Distribution to Illinois, Florida and Georgia.
30719 units