FDA Enforcement Class II Terminated

SafeStep Huber Needle Set AllPoints Port Access System; Catalog Numbers AN152075 (UDI 00801741119484), AN152075G (UDI 00801741119583), AN162010 (UDI 00801741119682), AN162010G (UDI 00801741119446), AN162075 (UDI 00801741119675), AN162075G (UDI 00801741119439)

Recall: Z-2664-2020 · Reported July 29, 2020

Enforcement

Recall Number
Z-2664-2020
Event ID
85853
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Becton Dickinson & Company
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 29, 2020
Initiation Date
June 19, 2020
Classification Date
July 18, 2020
Termination Date
July 29, 2022
Address
1 Becton Dr, N/A, Franklin Lakes, NJ, 07417-1815, United States

Description

SafeStep Huber Needle Set AllPoints Port Access System; Catalog Numbers AN152075 (UDI 00801741119484), AN152075G (UDI 00801741119583), AN162010 (UDI 00801741119682), AN162010G (UDI 00801741119446), AN162075 (UDI 00801741119675), AN162075G (UDI 00801741119439)

Reason

Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

Code Info

Catalog Numbers AN152075 (UDI 00801741119484), AN152075G (UDI 00801741119583), AN162010 (UDI 00801741119682), AN162010G (UDI 00801741119446), AN162075 (UDI 00801741119675), AN162075G (UDI 00801741119439). Lot Numbers: REDT0554 REDW1158 REDW2451 REDX1491 REDY1817 REDY2798 REEN3076 REEP3245 REEP3545 REDR2659 REDS4631 REDT0709 REDT4233 REDU2308 REDV1300 REDV3619 REDX2864 REDY3766 REEP3542 REEQ0644 REER1047 REDX4729 REDZ0700 REDZ1905 REEN1900 REEN2797 REEP3240 REEP4079 REEQ0719 REEQ1251 REER1105 REEQ1250 REDR2870 REDS0199 REDS3479 REDT1592 REDU0734 REDU1349 REDU2234 REDU3360 REDU4418 REDV1066 REDV3684 REDW0388 REDW1159 REDY1595 REDY2936 REDY3545 REDZ1827 REEN2784 REEP0777 REER0610 REER2163 REDS4692 REDT0555 REDU1332 REDU2236 REDV1956 REDW3670 REDX1492 REDX3881 REDX4827 REDY1818 REDY2791 REDY3767 REDZ0388 REDZ1906 REDZ3147 REEN3311 REEQ1172 REER1106 REER2125

Distribution

Worldwide Distribution

Quantity

5,296,132 total