FDA Enforcement Class II Terminated

VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Recall: Z-2653-2014 · Reported September 24, 2014

Enforcement

Recall Number
Z-2653-2014
Event ID
68909
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
September 24, 2014
Initiation Date
July 5, 2014
Classification Date
September 16, 2014
Termination Date
April 1, 2015
Address
6-22, 4-CHOME, KAN-ON-SHIN-MACHI, NISHI-KU, HIROSHIMA, N/A, N/A, Japan

Description

VERO Linear Accelerator System, Model Number MHI-TM2000; Common Name: Medical Linear Accelerator. MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.

Reason

Software Anomaly: If a user changes the calendar setting from Workday to Holiday or vice versa, that would erroneously change the status of treatment/fractions, which are completed and to become Treated (completed) status on the exact day when such change is made, to Untreated status. Furthermore, such treatment would be cloned and mistakenly added to the schedule as Untreated treatment

Code Info

Software Version 3.0.0 and after. Serial Numbers 201902, 203901, and 203919

Distribution

Distributed in the states of NY, FL, and TX.

Quantity

3 units in use in the US