FDA Enforcement Class II Terminated

DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures

Recall: Z-2651-2017 · Reported July 5, 2017

Enforcement

Recall Number
Z-2651-2017
Event ID
77250
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Carestream Health Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 5, 2017
Initiation Date
May 25, 2017
Classification Date
June 29, 2017
Termination Date
June 13, 2018
Address
150 Verona St, N/A, Rochester, NY, 14608-1733, United States

Description

DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures

Reason

Carestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly.

Code Info

Serial Numbers: Worldwide except China: 101 through 2623; China only: 800101 through 800482

Distribution

Worldwide Distribution - US (nationwide) Internationally to Australia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, India, Japan, Israel, Korea, Malaysia, Maldives, Mexico, New Zealand, Peru, Puerto Rico, Singapore, Switzerland, Taipei, Thailand, Turkey, and UAE.

Quantity

Domestic: 1449 units; Foreign: 1416 units