FDA Enforcement
Class II
Terminated
Safe-T PLUS Thora-Para 5 Fr Non-Valved Catheter Drainage Tray
Recall: Z-2647-2020
·
Reported July 29, 2020
Enforcement
- Recall Number
- Z-2647-2020
- Event ID
- 85751
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Carefusion 2200 Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 29, 2020
- Initiation Date
- May 28, 2020
- Classification Date
- July 18, 2020
- Termination Date
- September 20, 2024
- Address
- 75 N Fairway Dr, N/A, Vernon Hills, IL, 60061-1845, United States
Description
Safe-T PLUS Thora-Para 5 Fr Non-Valved Catheter Drainage Tray
Reason
The Universal Drainage Set (UDS) may contain valves that leak internally which may result in the introduction of up to 30cc of air into the patient.
Code Info
Catalog: OTP5000SP; UDI: 50885403284746; Batch 0001337784
Distribution
370 US customers and 13 distributors. Product distributed in: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MD, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY Customer listed, including Federal distribution, uploaded.
Quantity
N/A