Vial Decanter, REF 2006S. Packaged on a form, fill and seal machine with a poly/Tyvek material then placed 50 products to a case (cs).
Enforcement
- Recall Number
- Z-2642-2024
- Event ID
- 94780
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Microtek Medical Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Report Date
- August 28, 2024
- Initiation Date
- June 3, 2024
- Classification Date
- August 16, 2024
- Address
- 1 Ecolab Pl, Saint Paul, MN, 55102-2739, United States
Description
Vial Decanter, REF 2006S. Packaged on a form, fill and seal machine with a poly/Tyvek material then placed 50 products to a case (cs).
During sterile barrier testing performed on the decanter product line, the samples were identified to have pin holes and tears in the packaging film.
SKU/UDI-Device Identifier: 2006S/00748426002511; Lot Numbers: 191216,191122,211019,210916,210604,210313,201017,181211,181212,191211,190627,190812,190617,200506,190531,190618,190808,210503,190609,190808,191216,190609,190722,190307,191015,190513,190524,190703,190808,190807,190531,200402,190326,221029,220920,220914,220427,220404,220420,230612,230607,231028,230320,230320,221207,221027,221027,221008,220926,220920,220914,220909,220515,211111,220322,220408,211117,210708,211023,210523,210313,210523,210308,210114,201017,191216,191026,190925,190513,190808,181212,190808,191216,201008,190530,191107,190910,200505,191119,200320,181108,191216,181108,181212,190423,200618,190531,190927,191218,190627,190423,190531,200804,200218,190627,190424,200507,191011,211113,211103,211023,210910,210909,210708,210605,210413,210412,210412,210412,210319,210126,210114,210218,210107,201010,201006,201006,200605,181126,190722,191024,191024,191205,191011,200819,190609,190609,181212,191024,190910,200106,191024,190905,191120,181121,190926,191227,190513,200324,190531,200504,190812,190424,181023,231204,231204,231107,231107,231014,230707,230707,230515,230413,230328,221221,221216,221029,221028,221020,221018,221008,220928,221122,220923,220924,220922,220914,220505,220515,220504,200820,200602,200414,190609,200804,191216,190311,200905,181121,181112,181108,181203,200318,190104,191216,200605,190424,200226,190812,181212,191122,191107,200824,191205,190530,200605,191122,191122,191120,190910,200320,200210,190311,200605,181211,181203,200824,200403,200226,191205,190326,190531,211009,211110,210708,211113,211113,211110,211110,211103,210311,210604,210708,211113,211020,210915,210803,210915,211020,210911,210803,210712,210503,210712,211010,210916,210803,211010,210916,210910,210909,210722,210722,210713,210707,210708,210605,210605,210604,210523,210604,210603,210523,210523,210523,210510,210524,210524,210503,210524,210503,210413,210412,210319,210410,210410,210319,210319,210410,210412,210410,2103067,210313,210312,210311,210308,210306,210129,210126,210115,210114,210219,210218,210218,210106,201021,201017,201021,210107,210107,210105,201017,201017,210112,201010,201006,201006,201008,200618,200505,200414,190722,181212,181212,200602,200318,190530,190513,200320,190910,181108,181108,200226,190423,200403,181121,200820,200228,190417,200504,200703,190722,190722,181203,181203,181206,181206,181206,190530,190513,191028,191205,200403,200318,190807,181108,200324,190530,200113,200228,200226,191218,190305,181112,190326,200504,190627,181203,200819,200403,191216,191026,200106,221215,221207,221122,221018,221020,221017,220928,220926,220404,220514,220928,220426,220404,220322,211209,220122,220409,211209,211208,220322,220322,210708,211023,210712,210712,220411,220411,231107,231028,230607,221222,221222,221216,221216,221028,221018,221018,221008,220928,220923,220408,211111,181108,210112,210112,231204,231005,230707,200228,200602,181121,200414,190104,190925,200218,191119,191205,191014,200824,190423,200507,190808,181212,191107,200605,210115,210112,190530,190311,200403,191105,200228,191205,191120,190513,190609,190927,190703,200310,200425,200402,200804,190326,231116,231114,230707,220914,220412,220428,191014,191014,200320,200403,191216,200605,200403,200210,200820,181112,210308,200218,200804,200506,200414,191107,210114,210105,210105,201017,201017,201008,191119,191205,200324,200403,191227,191011,191011,190910,200228,191028,200226,200504,200324,200113,200113,210413,210107,201012,201006,201006,200425,200820,200819,231028,190423,190424,190326,190513,190424,190417,190311,190417,181206,181206,190513,190401,181206,190401,190812,190722,190609,190722,190807,220909,211209,190618,190424,190311,190305,191122,191205,191026,191026,191026,190925,200320,200228,181112,200414,200226,200425,200218,190307,190307,181203,190104,181206,190530,211019,211020,211020,211009,210803,210911,210712,210712,210712,210707,210712,211019,211010,201012,210114,210523,210603,210312,210410,210410,210315,210312,210126,210129,201021,210108,210107,201017,201006,200824,200425,191227,200226,191024,190618,200819,200505,181121,181108,190524,190530,191028,191105,200228,200113,181023,181108,181108,181108,190926,230516,230602,230607,230603,230602,230603,230603,230516,230602,230602,230603,230515,230607,230515,230516,230515,230515,230515,230515,230516,230515,230515,230516,230515,230516,230516,230516,230515,230516,230516,230516,230328,221018,220928,220519,220922,220920,220920,220427,220420,220504,220408,200602,181112,200804,181212,190305,190311,191107,190812,200506,200403,181203,181203,190423,190423,210915,210712,210712,210803,210803,210722,210604,210523,210524,210412,210312,210306,210306,210115,210112,201010,190513,190609,200210,190812,190910,181108,181211,191011,190609,200318,200403,190722,190722,191227,200425,200504,190926,200425,200318,191211,200403,231204,231013,221017,211104,220409,210916,210916,210523,210410,201218,181212,190807,181203,200320,190424,181121,200824,190925,191205,200507,191014,200820,190808,211009,211020,211113,211023,210915,210712,210803,210605,210603,210523,210523,210524,210503,210410,210315,210311,210126,210126,210219,210219,200605,210107,201017,200210,190618,200210,190417,191211,191011,190305,181121,190905,200402,190513,190531,191028,190424,191120,181203,200804,191218,191227,231116,231228,231114,231028,231013,230707,230612,230612,230412,230320,221222,221207,221029,221018,221018,221017,220929,220926,221028,221122,220920,220920,220420,220420,220411,211209,211113,211111,200602,200218,191015,191026,200824,191211,200310,200320,200605,211112,211103,211023,210915,210803,210712,210916,211010,210722,210722,210603,210603,210503,210503,210313,210129,210129,210219,210219,201010,201008,200819,181211,181121,191011,200820,200820,190808,200504,200106,190905,200318,190905,190905,180927,181108,190531,200402,200228,190305,181203,200113,200414,190530,190627,200824,200506,200618,200106,190627,190531,200318,191216,181203,190617,181203,200507,191024,200310,211110,211019,210708,211113,211023,211020,211020,211020,211020,210911,211019,211019,210916,211010,211010,210909,210916,210910,210713,210713,210707,210707,210605,210605,210604,210603,210523,210523,210510,210524,210524,210503,210503,210413,210412,210412,210410,210316,210316,210313,210311,210311,210308,210306,210126,210126,210114,210115,210219,210219,210218,210218,210108,201021,201021,210105,201017,201012,201010,201006,201006,201006,200824,200425,190910,200505,181212,181212,200602,200602,200210,200403,191105,190423,200819,200505,190618,181108,190609,200113,200228,190530,200228,191011,191024,191211,200703,200703,190812,191119,191105,190307,200318,190513,190609,191105,191120,190305,200318,181108,190513,200820,200504,181121,190926,190927,191105,191028,190807,190326,181108,200425,190627,200819,200403,181112,200113,190812,181023,190326,190627,181112,231204,231230,231116,231114,231114,231028,231014,231005,230612,230612,230612,230516,230413,230412,230413,230328,230320,221216,221221,221207,221029,221029,221018,221028,221018,221018,221018,221017,221017,220928,221028,221122,221222,221122,220924,220923,220923,220505,220514,220514,220909,220327,220427,220515,211104,211222,220411,220411,220411,220322,220122,211209,220409,211111,211117,211114,211117,231107,230707,230320,230620,230612,230504,230504,230412,230328,221207,221207,221018,221020,221018,220928,220519,220923,220920,220427,220327,220327,220420,220412,211222,211104,211114,211117,211009,211103,211020,211020,210916,210916,210910,210707,210605,210603,210524,210503,210316,210115,210115,200605,210105,201006,231230,231107,231228,231114,231028,230612,230621,230328,221215,221029,220928,220928,220923,220914,220412,220408,190617,200824,211020,210915,210909,210410,190910,200320,190910,231014,221018,221018,220514,220914,220420,200602,200905,190925,191119,200218,200414,200507,200605,200425,211114,211113,211113,211103,211020,210712,211019,211010,210909,210722,210708,210605,210603,210510,210105,210105,210112,190618,190618,200819,200228,191024,200505,200318,200402,200228,200820,190703,181206,191218,191218,191028,200113,200402,190530,190531,181203,181203,190812,231114,231107,231005,230621,230515,230413,230413,221215,221221,220516,221018,220519,220504,220920,220909,220505,220515,211104,220412,211111,211118,211117,190812,190808,181112,231114,220122,220924,191014,190925,190617,211110,210712,210604,210308,201017,190722,181212,200819,190530,191024,190417,190307,200228,200402,181112,211110,211112,211113,211103,211103,211023,211113,211020,210713,210707,210803,210707,210910,210713,210605,210603,210523,210523,210523,210510,210603,210503,210413,210410,210316,210410,210312,210311,210308,210126,210115,210115,210114,210114,210219,210106,201021,210105,201017,191122,190925,190925,200506,200602,190627,200507,200507,190812,190104,191026,191026,191120,200414,200425,200414,190104,190311,190617,190424,190617,200820,200605,190807,191211,201017,201017,210112,201008,190808,191107,191107,200310,181212,200819,191105,190808,200403,191227,200820,190423,190417,191211,190417,190307,181108,190513,200504,191120,200228,190722,191015,200819,181212,181112,190305,191218,200804,211019,211112,211023,211009,210916,210909,210722,210707,200318,190910,190627,190326,181212,191014,200618,190424,191211,190812,190530,200804,191122,191122,190925,190925,200218,190311,181112,190618,200403,181203,191205,191011,190423,211009,211113,210909,211023,211019,210909,210603,210707,210708,210605,210523,210524,210503,210410,210319,210316,210316,210312,210308,210306,210126,210115,210114,210218,210106,210107,210105,201017,201012,201006,201008,201008,190808,191107,191216,200318,190530,190530,200602,191119,190910,200210,190311,200403,200820,181121,191011,200228,190417,200505,200703,190722,181206,191015,191015,190513,191105,190905,190401,191120,200819,190722,200226,200504,190703,181203,191026,191015,190401,190925,200414,200320,200320,200106,231014,230607,230320,230320,220516,221017,220928,220928,220929,220404,220922,220514,220404,220428,211222,211114,211117,231230,231013,231010,231010,230707,230620,230707,230620,230612,230607,230607,230607,230413,230413,230413,230412,230320,221221,221207,220516,221122,221028,221027,221028,221221,221017,220404,221122,220924,220514,220514,220914,220327,220504,220428,211117,211117,211117,231228,231228,231028,231028,231010,230621,230612,230515,230412,230412,230328,230320,230318,230318,230318,221216,221215,221216,221207,220516,221018,221017,221008,220928,220922,220920,220514,220505,220909,220515,211104,220504,220412,220411,220411,211118,211117,191105,200605,191026,231230,220428,190627,191014,190423,210916,211112,210915,210722,210713,210722,210708,210503,190926,191015,231204,231230,231204,231116,231116,231228,231114,231107,231028,231014,231005,231005,230707,230707,230621,230612,230516,230504,230412,230412,230328,230320,221221,220516,221018,221027,221018,220914,220909,220505,220327,220426,220420,220412,220428,220411,220408,191014,190808,191119,211023,210605,210523,210129,210129,210115,210115,190424,200505,191105,231013,230413,221216,221122,220909,220420,220322,191014,210306,190609,200310,191216,211113,211112,211112,210712,210712,210916,211010,210803,210604,210604,210605,210510,210523,200605,200820,200506,200504,200602,200506,200602,200618,200425,200324,200324,200402,200402,200602,200618,231204,231204,231014,231014,231014,230621,230620,230621,230620,230707,230515,230504,221207,220516,221207,221028,221028,220519,220929,220929,220929,220922,220922,220505,220428,220515,220515,220515,220420,211111,211110,220409,220409,231013,210712,190617,210712,231204,231230,231114,231228,231107,231114,231107,231010,231013,231005,230620,230707,230620,230612,230515,230412,230412,230328,230320,230516,221221,221018,220411,181212,181203,190311,191026,190530,190925,181126,200824,211103,210722,210524,200210,200218,181212,190530,200228,190401,190417,191218,191120,230504,220428,220411,220909,220909,220505,220426,220122,220409,211117,220322,220408,211019,211019,211110,210708,210503,211020,211020,210707,211019,210707,210605,210605,210604,210316,210523,210523,210510,210114,210114,210219,210218,210218,201006,210107,210105,201017,190925,190627,190424,181112,200506,200905,200507,181203,190424,190401,190925,191107,190617,190311,200403,181203,191205,200505,190423,201017,210112,201006,201008,190530,190808,191024,190910,200310,200602,190618,190423,200403,200820,200324,191015,190417,200505,200425,190722,190513,190513,190905,200318,191105,200226,200819,190926,200504,190326,190703,191218,200505,200507,191026,190305,181023,190531,200218,210105,231014,230612,230612,230412,221216,230328,221018,220928,220519,221122,220920,220327,220411,220122,211209,211118,211009,210915,210915,211019,210707,210909,210604,210311,210306,211009,210712,210712,191011,181203,181203,181108,181212,211209,181203,191014,191014,200414,190617,181121,190807,190609,191119,211010,210803,210410,210319,210315,190910,181126,181211,200505,190905,200820,190609,180927,190926,200504,190703,200113,200318,191120,231005,230707,230707,230612,230515,230504,230412,230328,230328,230328,230318,221222,221221,221216,221215,220516,221122,221029,221018,221027,220504,220427,220923,220909,200602,190812,191026,181112,191211,190925,190617,191205,211019,211009,211113,211023,210712,210713,210306,210604,210412,210412,210312,210219,210105,201008,200425,190530,200505,190609,181121,181206,190307,200318,191011,190417,191105,200226,200226,231114,231014,230620,230504,230328,221221,221029,221028,221020,221008,220926,220404,220404,220408,211117,191014,231116,231114,231107,231013,230620,230612,230515,230320,221216,221028,220504,220504,221008,220926,220922,220909,220914,220505,220505,220515,220420,220428,220122,211117,211117,200318,190627,181203,191026,190104,200320,200320,200804,200824,190925,181206,181126,210916,211103,211020,210911,211019,210910,210713,210605,210524,210413,210315,210114,210129,201021,201017,210112,190531,190910,191105,191120,190513,200506,211110,220411,211209,210106,210916,210105,191218,191216,190925,200507,190311,190424,181112,190627,190617,200218,191026,190617,190104,181212,191216,190311,200824,200804,190311,200310,190808,191107,200310,200310,211019,211019,211113,211110,211114,211009,210712,210707,210803,210722,210916,210713,210713,210604,210604,210523,210524,210503,210503,210413,210412,210313,210312,210311,210308,210306,210114,210114,200605,201021,201021,201017,201017,210112,201012,200824,191122,200504,190722,191119,200602,191011,191015,200505,200402,181108,181121,200504,190307,191015,200228,190609,190905,190417,190812,191028,200324,190531,200605,190926,190926,200804,190812,200226,230707,230318,221216,221017,220504,220924,220920,220909,220426,220412,211222,220322,220408,221207,221017,221008,220920,220914,211113,220408,220408,211117,210306,231028,210911,190627,190609,191026,200310,181126,191211,200210,190910,200403,191011,200505,181206,190926,190104,200602,181112,200804,200804,200506,200602,201010,200504,220914,191015,190326,181112,190905,181008,200602,210308,181211,181121,221018,220519,211211,191014,190609,191211,200106,200210,181212,190417,191105,200228,191024,181121,200320,190722,190808,191026,181203,201017,201012,190722,190609,181108,190424,220428,220428,220411,190627,190627,190925,190925,190305,191216,191216,190424,191216,200804,200804,200425,200425,191107,191107,191211,211019,211019,210604,210603,210603,210316,210312,210306,210306,210114,210114,210114,210219,210219,210219,210218,210105,210105,200210,200218,200218,190531,181108,181108,181108,181108,191024,191024,200318,200425,200425,200703,190905,190905,200403,200318,181211,190513,190513,190305,190305,191227,190401,190401,181108,181203,181203,190531,190618,190531,191120,200113,211222,211118,200602,200824,210803,210916,210910,210605,210412,210410,210115,210115,210219,201017,201012,190910,200210,190530,181108,191014,181206,200318,200820,190513,190926,200226,200504,231028,230320,221017,221222,220515,220122,220322,191211,190722,211103,210803,210722,210707,210605,210604,210412,210315,210115,210218,210107,210107,210105,201012,200505,190311,181112,181108,181212,190910,190910,200318,200506,181211,191105,191119,200210,191205,200819,190530,200618,190524,190524,190722,190926,190531,191028,200905,191015,200804,181203,190326,181112,200506,200425,191216,200824,190424,191014,191026,190104,191122,200310,200804,190424,200506,200403,181203,181212,191011,191205,211103,211110,211019,211009,211009,210708,211113,211020,210712,210712,210712,210803,210909,210909,210713,210713,210707,210708,210708,210603,210603,210503,210306,210129,210126,210114,210106,201021,210107,190609,181212,200403,190417,200506,181121,181121,200402,191011,200820,200820,181206,200403,190305,181108,181108,200402,190513,181121,190927,200819,190326,190927,181203,181121,191218,190305,181023,181023,231005,230620,230707,230621,230607,230621,230515,230504,230413,230412,230412,230328,230328,230320,230320,230320,230318,221222,220516,221122,221029,221028,221028,221028,221027,221018,221017,221018,221017,221017,220928,221122,221222,221122,220404,220924,220924,220427,220923,220914,0002626701,220505,220327,220515,220428,220404,220420,220420,220428,211114,211111,220411,220322,220122,211209,211111,220408,211117,211117,221018,190104,200218,190423,211019,210605,210313,201012,190722,190808,191024,200605,181023,231114,230620,230320,221029,221008,220514,220922,220327,220428,220408,231230,231230,231230,231116,231116,231107,231228,231107,231028,231014,231014,231013,231013,231013,231010,231010,230707,230707,230621,230612,230504,230504,230413,230412,230412,230412,230328,230328,230320,221222,221222,221222,221216,221216,221216,221216,221216,221207,221027,221027,221020,221017,221008,221008,221008,221008,220928,221028,221122,221029,220923,220404,220923,220926,220914,220909,220505,220505,220909,220327,220427,220515,220426,211104,220504,211111,220409,220409,220409,220409,220409,211209,211118,211111,220322,220408,220322,220322,220408,211114,211019,211009,211009,211113,211112,211112,211110,211103,210708,211020,211023,211023,211112,211020,210915,210915,211020,210911,210911,210712,210712,210712,210712,210916,210707,211010,210604,210316,210523,210316,210524,210412,210412,210410,210410,210410,210410,210319,210312,210316,210315,210315,210315,210313,210313,210312,210312,210311,210308,210308,210306,210306,210114,210115,210126,210126,210115,210115,210114,210114,210219,210218,210218,210218,210218,210106,201218,201021,210107,201017,190925,190925,190627,200804,190424,190401,200905,181112,181112,200506,200506,181203,181212,191014,200602,200106,181203,191014,191211,191216,190812,190812,191120,190925,190925,200804,191122,200218,181112,181112,181108,181112,200403,200414,200414,190311,200824,181203,191011,191011,191011,190618,200505,200106,201010,201006,201006,200824,201008,201008,190808,190808,191024,181108,191024,190910,200602,200602,200210,191015,190910,181108,200506,200403,190609,190617,200210,190311,181203,181121,181211,200310,191011,200820,200820,200505,200425,200703,200425,190722,190722,181206,191105,190905,200113,190905,181108,181112,190530,200113,200226,200226,200819,181112,190926,200504,190627,191227,191227,181206,191218,190531,190807,191218,200113,190305,181108,200605,200113,190531,200218,191218,200218,200804,191122,200824,231228,230707,230612,230612,230602,221215,221029,221028,221027,220924,220427,220514,220514,211104,220322,191122,190531,211110,211118,211020,211010,210114,200210,200618,181121,190722,200425,220928,181203,191014,191211,190812,191014,200414,200320,200320,181112,200506,200414,190104,190104,200210,190513,181126,200507,200210,191205,200106,200824,191122,190423,211112,211103,210713,210707,210605,210603,210524,210410,210410,210311,210115,210107,201012,201006,191227,200318,181211,190722,190417,191120,190910,190905,190513,191211,200226,191028,181206,190927,190618,181108,200402,181203,191015,200106,221008,220504,220922,211114,220909,220909,211104,220409,211211,231116,231107,231028,231014,231013,230612,221221,221221,221221,221216,220516,221027,221018,221018,221017,221017,221008,221008,220404,221122,220924,220923,220922,220914,220505,220327,220515,220404,211104,220420,220412,220428,211111,211211,211211,220322,211009,211019,211110,210708,210915,210915,210911,210712,211019,210707,210510,210503,210604,210413,210413,210413,210410,210410,201021,190609,200507,191122,210112,191119,190326,231114,230328,221018,221020,220928,220926,220327,210605,210605,210308,191014,200310,210308,191120,200403,181108,191120,180927,191122,190326,190531,190424,190513,181112,190627,190627,200106,181203,181203,181203,181203,191014,190627,200226,181203,190104,190104,190104,191014,191026,190812,191216,190104,191011,190424,191211,190524,190311,181121,190925,181108,190424,180821,190513,190617,190617,190609,191122,190311,190311,190311,191026,190617,190617,181203,190104,190311,191211,181112,191014,190925,190925,190423,190617,181203,181203,190311,190424,181126,181126,181126,200106,181203,181203,181212,181212,181212,191014,191211,190807,190808,191011,191011,191119,191119,190910,190910,191107,190722,190722,190531,190531,190722,190722,191119,181212,191024,181126,190609,191119,191119,190311,191211,190530,190618,190530,181211,190910,191211,181108,190618,190808,181121,190417,190609,191216,191011,191026,191024,190531,190722,181121,181206,181206,190812,190910,190513,190722,191216,191105,190513,190609,190812,190812,200106,191105,191105,181203,191015,191024,190905,190513,191105,181206,191015,190524,190617,190524,190305,191011,190807,190807,190807,191227,190926,190926,190401,190530,190926,191015,190513,190812,190703,190927,190326,190703,191015,190807,190305,191227,190808,190807,181015,191028,190925,190925,190807,190807,190807,190807,191120,200113,190627,181112,191120,190305,190326,190326,200320,200602,181203,200310,190424,191014,190627,200824,200905,191211,191014,191107,191026,200320,190424,191216,200218,190104,190424,200804,200804,200320,200414,190617,200804,190311,191014,181112,181112,181203,191205,200507,190808,181212,191107,190531,190531,190524,200226,191107,201006,201008,190531,190326,190531,190618,191119,191119,181108,200403,200602,181211,181212,190618,181211,191205,180927,200819,190531,181108,200310,191011,190722,191011,191011,191024,190417,200703,190307,190905,191026,200318,200820,200402,200402,190609,191120,181206,190305,181121,191105,190417,200402,200402,181112,191227,190722,191227,200226,200504,200504,190927,191028,191015,191218,190703,191227,191026,191028,190401,190807,190807,200605,191107,180927,190925,181023,190618,181112,190326,200106,190305,191120,230412,221221,220519,220909,220426,220411,220411,211114,211117,200804,200602,200605,211110,211110,211009,211112,211023,210915,210803,211010,210524,210524,210524,210413,210412,210126,201017,201017,210112,201012,200824,200824,190417,190530,190513,190618,190417,181108,190513,200228,190530,190722,181108,181108,211020,201017,221027,220504,220327,211222,211110,211020,190104,190104,190104,211103,211103,210803,210803,210910,210707,210523,210524,210503,210313,210308,210308,210114,190326,200403,191205,191120,200820,181108,181121,191011,190417,190531,230318,221029,220929,221017,220920,220514,220420,220122,220122,220322,190609,190627,190104,191122,200605,190423,190423,211110,210915,210915,211010,210915,210712,211010,210713,210713,210605,210316,210524,210311,210306,210115,210219,201021,201017,201012,201008,191107,200505,190417,181108,181108,200318,191211,181211,191105,190910,190609,190910,191011,181203,200228,191015,191105,190926,200226,200226,200504,200605,200113,190812,181023,231228,231114,231028,231010,231005,230612,230413,230320,221215,221018,220926,220909,220515,220322,211117,190423,191024,211019,211009,210916,210708,211113,211103,211023,211023,211020,210915,210915,210909,210722,210605,210603,210503,210316,190513,190609,191028,190307,231204,231114,231107,231028,231028,231014,231013,230707,230612,230413,230412,221216,221216,221029,221028,221027,221020,221017,221017,221008,220519,221122,221122,220926,220923,220427,220922,220920,220920,220514,211104,220412,220505,220909,220427,220515,220420,220420,220412,211110,211111,220409,211111,220408,210603,190812,210803,181211,230320,221221,221122,221018,221018,221008,180821,210112,181112,181126,181108,181108,201008,181112,181112,200506,181203,210916,210916,210910,210605,210510,210316,210316,210218,210112,201008,200605,190618,191211,181108,191120,181112,230607,230621,230516,230320,221207,221018,221215,220327,220327,220428,211222,191205,190807,200106,190618,190311,211019,210708,210708,210412,210106,200507,190910,200318,190905,181206,231010,221222,220404,211118,200824,200605,200403,190424,181212,211103,210722,210605,210503,210412,210315,210308,210114,210105,210112,190910,200310,190423,200218,181211,190618,191011,191011,190513,190807,190722,190926,180822,181108,230621,221122,221029,220427,220411,200320,231228,231114,231013,231005,230320,220505,220515,220420,220322,220408,231013,230504,230413,221017,220928,220914,220909,220327,220515,220515,220420,220420,220412,220412,220409,220409,220327,211111,220408,211009,210129,210605,210713,211112,211113,211112,211112,211113,211113,211113,211103,210708,210708,211023,211020,211023,211023,210915,210915,210915,210803,210911,210911,210712,211019,210708,210708,210916,210707,211010,211010,210916,210910,210909,210909,210909,210708,210708,210708,210605,210605,210605,210605,210605,210604,210311,210523,210316,210316,210315,210315,210316,210412,210410,210410,210319,210319,210316,210316,210316,210315,210315,210410,210313,210313,210315,210312,210311,210311,210308,210306,210306,210126,210126,210126,210114,210114,210114,210219,210219,210218,210218,210218,210218,210106,210107,201017,190326,190423,200905,200506,200506,190627,191216,191014,181203,181203,200605,190424,190812,191026,191122,190513,190311,200310,200310,181108,190617,200218,200824,200824,200403,200403,191205,200505,190423,190423,201017,210112,210112,210112,201010,201010,201006,201006,201008,201008,190808,190808,191024,200310,191119,191119,191024,190910,200602,200602,200403,200403,181121,200210,181121,200820,200324,191015,200820,200505,200505,190417,200425,190722,191105,190905,190905,200318,190513,200113,200226,200819,200819,181112,200226,190326,200504,190703,191026,191105,191218,200605,200505,181023,200414,200804,191205,200403,190618,190618,191011,191011,200703,200820,181206,181206,200113,220420,211020,210105,200804,221029,220411,211209,200106,200228,211113,211113,210911,210712,210603,210523,210510,210604,210503,210316,210315,210115,210218,210108,210106,201017,210112,191218,190531,191211,200318,200504,200504,200113,200819,200402,200804,191014,181203,200618,181212,190423,190531,200318,200218,190104,190104,190925,190531,181112,200218,200210,191216,200310,190617,190808,190925,190423,200507,190808,200218,191205,190423,191216,191211,211019,210708,211112,211112,211112,211103,211020,210915,210712,211010,211019,210916,210916,210910,210707,210707,210707,210707,210316,210524,210503,210413,210410,210315,210311,210126,210115,210115,210219,210106,210106,210107,210107,201017,201010,201006,201006,181126,200403,191119,181108,191216,190531,200504,200403,191011,200505,190722,191024,181206,190423,200403,191011,191015,200402,181112,190927,191218,200318,191028,190627,191107,191028,191015,190910,181112,200425,190531,181112,181112,200106,200218,200218,190305,181023,181023,231204,231116,231107,231028,231028,231010,231013,230707,230620,230612,230607,230612,230621,230504,230515,230504,230413,230328,230328,230328,230328,221222,221221,221215,221207,221122,221122,221029,221027,221027,221017,221008,220924,220427,220926,220923,220923,220922,220514,220514,220914,220505,220909,220428,220420,220412,220322,220122,211111,220122,211211,211117,211117,211117,210712,210712,220404,220420,211020,200414,210604,220404,181112,181121,200226,181108,200703,181108,191011,190104,211010,210604,210523,210315,210412,210412,210316,210115,201010,200605,200602,200403,190627,191216,200506,200602,191026,191122,200824,200310,210712,210707,210523,210523,210410,210306,210114,201021,200804,190812,200226,200403,181112,181121,190305,190808,191028,190326,181203,191216,200602,200605,200106,190424,181112,200402,200824,200320,191122,191122,190104,190104,190530,200414,190617,200605,191119,181212,191014,200507,200824,190910,200310,200425,211019,211010,210708,211113,211113,211110,211103,211103,211023,211020,210910,210911,210712,210803,211010,210916,210803,210909,210713,210605,210523,210524,210524,210524,210412,210410,191122,191211,181112,190617,190808,181203,190104,190531,191014,190925,190424,190423,191107,191211,181108,190530,190910,190808,190417,190618,181212,181211,191120,191122,190423,190417,181212,191024,190618,191211,190808,190722,190307,181108,190627,181121,191105,191105,190905,190417,190417,190513,181206,191107,191028,190807,191211,181108,181112,190927,190618,190513,190326,190524,190531,190417,191026,181203,190305,191014,191105,190812,190627,200106,210604,210316,210524,210410,210315,210308,210115,210115,210219,210218,201006,210105,201017,190925,200804,200506,191014,200905,191216,191107,200310,191011,201017,191024,200310,200210,191011,200820,200507,200113,200819,200602,191218,200605,200320,221122,220519,220923,220122,220409,220409,211209,211009,210916,210708,210605,210316,210523,210523,210510,210316,210524,210503,210412,210412,210410,210410,210312,231204,231204,231230,231230,231114,231116,231228,231107,231107,231114,231107,231028,231014,231010,231010,231005,230612,230621,230621,230621,230612,230607,230621,230504,230515,230413,230413,230412,230320,221215,221222,221221,221221,221221,221215,221216,221216,221221,221221,221207,221029,221207,221029,221029,221018,221029,221020,221008,221008,221008,221017,221008,221008,221008,220928,220928,220928,221122,220928,220926,220519,220926,220926,220404,220404,220926,220504,221027,220924,220924,220924,220924,220923,220914,220920,220514,220504,220514,220914,220914,220909,220909,220505,220505,220505,220909,220909,220909,220327,211104,211104,220420,220420,220428,220408,211209,220122,220409,211209,211209,211209,211209,211118,211118,211117,211211,220408,211009,211009,211110,211222,211113,211112,211112,211112,211113,211113,211103,210708,211103,211114,211023,211020,210915,210911,210712,210712,211019,211019,210916,210916,210707,210707,210722,210722,210916,211010,211010,210910,210909,210713,210713,210722,210707,210605,210605,210605,210604,210413,210313,210524,210524,210524,210524,210503,210503,210413,210412,210319,210316,210315,210412,210313,210313,210313,210312,210311,210308,210306,210306,210115,210108,210108,201021,210107,210107,210105,210107,201017,190311,190423,181112,200506,190627,191216,191014,191014,190424,191211,190812,191122,191122,191026,190925,190617,190104,200310,200324,200218,181112,181112,190618,190617,190617,191205,191011,191014,191014,191014,200507,191211,190423,210112,201010,201010,201006,200905,190808,191119,181108,190530,191024,200210,191024,190910,200403,190618,181211,181211,200228,200402,191011,191015,190417,200505,190722,191105,190513,190905,190905,200113,181108,190531,190531,200226,181121,190926,190326,190513,200504,190627,190703,190703,190703,191227,181203,181206,191015,190812,190807,191028,181023,190925,200320,200324,190530,200218,191122,210316,230612,220516,221122,220914,211104,211209,191216,190530,190104,181112,200414,190618,211010,210915,210712,210909,210605,210523,210412,210308,210114,210106,201010,190910,191216,190307,200605,191026,231010,231005,181112,181203,181112,181211,181212,181108,181212,180723,181212,181121,181108,181206,181212,211111,231230,231230,231116,231114,231228,231010,231013,230607,230504,221221,221215,221029,221027,221018,221017,221017,220924,220924,220924,220922,220920,211104,220420,211222,220122,211209,220409,211019,211103,211020,210915,210915,210803,210911,210712,210916,210722,210916,210910,210713,210713,210605,210605,210605,210603,210603,210315,210313,210312,210313,210311,210311,210308,210306,210306,210115,210126,210115,210114,210114,210114,210107,201017,191120,190627,200804,200804,190424,200905,200506,191216,181203,191014,191211,190812,190812,190925,200804,200605,200218,200218,200403,190618,190627,181203,191205,191011,191211,200505,201017,210112,201012,201010,201006,201006,201006,201008,190808,190910,200602,191015,200403,190609,200210,191011,200820,191015,200505,190722,190905,200113,200226,200819,190926,191120,190627,191227,191218,190807,200113,190925,200320,200402,200218,200602,201008,190627,190424,190812,190104,191107,181126,191205,211103,211019,211113,211020,211020,210911,210712,211019,210916,210916,210713,210722,210707,210713,210708,210603,210523,210315,210524,210413,210315,210312,210129,210114,210106,201017,191119,190311,190307,200820,221018,220926,220923,220926,220420,220412,220428,220122,211209,211111,220408,200505,210315,210311,210308,210219,210218,210106,201021,201021,210107,210112,201012,201010,200605,200210,190311,191205,191024,200819,200820,181211,190417,1904
23,200505,231230,231230,231204,231116,231116,231228,231228,231028,231028,220505,220427,220327,220412,220412,220504,211111,211110,220411,220122,220122,181203,190925,191026,190311,190311,210803,210910,210603,210503,210412,210410,210313,210108,201021,201021,201017,190524,191024,181211,190417,200425,190609,200228,191028,200226,190927,190807,200218,200226,230707,230412,230412,230328,230328,221216,221122,221029,221018,221221,221008,220928,220920,220514,220426,220426,220404,220411,211117,231107,231114,231107,231028,231010,231010,231010,230707,230707,230620,230620,230412,221215,221207,221027,221018,221008,221008,220926,220926,220924,211211,211209,220409,220327,211211,211111,220408,220322,211117,211110,211019,211113,211114,210708,211020,210915,210712,210916,210708,210523,210524,210503,210503,210413,210315,210312,210311,210114,210219,200505,190925,190627,181112,200506,191216,200507,190513,191120,191107,181112,200403,200505,190618,191205,191011,190808,190423,210112,201012,191119,200602,190910,181108,190609,181121,181121,200210,200505,200425,200318,181112,181112,181121,181108,190530,200226,191026,191015,190627,200505,181023,200320,200905,190609,181203,181206,190104,200824,191205,190311,200819,191205,211103,211019,211019,211009,211009,210708,211110,211103,211010,211023,211020,210915,210915,210915,211020,210712,210916,210916,210707,210708,210604,210524,210503,210412,210410,210316,210319,210319,210311,210129,210114,210218,201218,201021,190424,190423,190808,190722,191024,191105,191119,181121,190423,190401,200226,200310,190926,190326,190627,191026,190812,190905,220404,220924,220924,220926,220922,220922,220514,220920,211104,220914,220909,220427,220515,220515,220420,220504,220428,211209,211211,220322,220408,210713,210722,210708,200602,200905,190925,200507,200605,181112,190925,181211,191011,200824,211020,210910,210722,210707,210708,210605,210604,210218,210106,201021,210107,201017,201017,210112,201012,201010,201006,200505,190531,200602,200820,190531,200504,200703,181121,181108,181112,190926,200504,200819,191218,230328,220929,221122,221122,220504,220404,220926,220924,220428,220428,211222,220409,220322,200403,190627,181008,181112,200506,200905,200602,191216,190424,191211,190812,200210,190925,191107,190925,190617,200804,191122,190104,200310,200218,200218,190311,200414,190618,200824,191205,190925,191014,200505,191227,190808,191227,191024,211110,211019,211019,211009,211112,211113,211103,211114,211103,211103,210713,210713,210915,211009,211020,211020,211023,210911,210712,210712,210916,210523,210803,211010,210910,210909,210910,210722,210713,210708,210708,210605,210605,210604,210603,210603,210523,210503,210524,210503,210412,210412,210412,210410,210410,210410,210410,210316,210315,210312,210311,210311,210308,210306,210129,210126,210126,210115,210114,210219,210218,210218,210106,210106,201021,210107,210105,201017,201017,210112,201012,201012,201010,201006,201006,201008,201008,190910,200310,200310,200602,190311,181108,190618,181121,191011,191014,200820,190417,200505,190722,190722,200425,191218,200507,200106,191015,200403,190513,191105,200402,200618,200618,190417,190513,190530,200819,200504,181206,190326,190703,191227,190927,191026,191218,200113,190305,200320,190617,231204,231230,231230,231116,231228,231107,231228,231107,231028,231028,231010,231014,231010,231013,231005,231005,230707,230707,230620,230621,230612,230621,230612,230504,230504,230412,230413,230320,230602,230320,221215,230318,230515,221221,221221,221215,221207,221207,230320,221029,221018,221028,221027,221027,221020,220519,221017,221017,221008,221222,220928,220928,220926,220926,220404,220504,220924,220926,220514,220920,211104,220914,220505,220505,220909,220327,220327,220327,220504,220426,220404,211110,220411,220411,220411,220411,211118,220409,220404,220404,220404,220404,211118,211118,211222,220322,220322,211222,211222,211114,211104,211104,211104,231230,231228,231114,231114,230621,230612,230515,230412,230328,221215,221216,221029,221008,220928,220926,220924,220923,220920,220914,220909,220515,220420,220412,211111,220408,190326,200605,190531,200804,200804,181112,190104,190812,200218,190609,191107,200507,191024,190925,191211,191205,211110,211009,211023,210911,210803,210315,210604,210503,210604,210412,210306,210114,210129,210107,201012,201006,190311,181211,181212,191024,191024,190807,190722,200318,190513,190513,200113,181212,191120,181023,181023,200402,200506,200505,190627,200602,200320,211009,211009,210915,210915,210712,210708,210503,210316,210106,201012,191119,191120,190530,200228,200310,181212,191107,200504,190926,190326,231116,231114,231028,231005,230707,230607,230412,230320,230320,221216,221029,221018,221122,220924,220923,220920,220920,220505,220420,220428,211209,211211,231204,231230,231204,231116,231116,231107,231107,231114,231114,231107,231010,231013,231005,230612,230607,230621,230607,230707,230504,230607,230412,230621,230412,230412,221215,230328,230320,230320,221215,221221,230318,221216,221216,221221,221216,221207,220516,221216,221029,221029,221018,221222,221028,220519,221017,221017,220504,221008,220929,220928,221215,220928,220928,220928,220404,220404,220924,220924,220923,220427,220920,220920,220920,220914,220514,220514,220909,211111,220505,220505,220909,220505,220505,220909,220909,220909,220515,220515,220327,220327,220426,220411,211117,220512,211111,211111,211222,220411,211211,220411,220122,220122,211111,211211,220322,220408,211114,211117,211114,211110,211009,211113,211103,211103,211112,211112,211113,211103,211103,211103,210708,211023,211023,211023,211020,210915,211020,210915,210915,210915,211020,210911,210712,210712,210712,211019,211019,210707,210722,211010,210916,210910,210909,210909,210722,210722,210713,210713,210713,210708,210708,210605,210708,210605,210605,210605,210604,210604,210604,210603,210523,210604,210523,210510,210510,210524,210524,210524,210524,210503,210503,210503,210413,210412,210412,210412,210410,210410,210319,210316,210315,210313,210313,210312,210311,210311,210308,210308,210114,210306,210308,210306,210114,210219,200605,210108,210108,210107,210107,191122,200804,190326,190326,200905,181112,200506,200506,181203,181203,191216,200106,200602,191211,191122,190104,190311,200310,181112,200414,200824,190617,190617,181203,200507,200106,181212,191211,210112,210112,201012,201012,201010,201010,201006,201008,201008,201008,200310,191119,191024,190910,200425,190311,181108,200403,200210,190609,181211,181211,200820,200505,191211,200425,200318,191105,190905,191122,191120,191120,190305,181108,181112,200819,190530,191120,200504,191227,191026,181203,191218,200210,200228,200310,200113,200113,190305,190305,190305,200320,181023,200324,200324,190722,200824,210707,210605,210413,210315,210126,210218,201012,200310,191014,181203,210503,210503,210503,200318,190326,191014,180927,200218,200106,191216,200320,190424,190703,200218,190424,190925,190925,190617,200824,200507,181203,200106,191205,181212,191211,211110,211009,210803,211114,211103,210915,210915,210712,210916,210722,210910,210722,210708,210604,210604,210523,210523,210413,210413,210316,210316,210313,210308,210126,210129,210129,201218,210107,201017,210112,201012,201008,190326,200425,200210,200320,191024,190618,200425,200820,181108,181108,190530,191227,191216,191024,191216,190722,191015,200507,190905,191120,190305,200820,200318,200310,181121,190305,181206,191218,200414,200504,200320,200618,191218,190524,200226,190925,231204,231204,231116,231116,231228,231107,231028,230707,230707,230620,230504,230504,230320,230320,230516,230318,221207,221029,220516,221028,221017,221017,221029,220926,220929,220924,220922,220920,220505,220505,220920,220909,211104,211104,220411,220411,211211,220408,181112,210129,210129,181108,190417,200402,200402,191014,190618,200320,200106,191014,211019,211009,211009,210916,211113,210915,210915,210712,210916,210910,210713,210713,210605,210603,210523,210510,210413,210412,210312,210315,210313,210308,210126,210126,210219,210218,210106,210105,201010,201006,201006,200824,191119,191011,191211,200618,191024,200820,200504,200106,200106,200106,231028,230320,221216,221216,221029,221018,221017,220928,220928,220926,220924,220922,220920,220909,220505,220427,211104,220412,211111,211104,211211,220322,231230,231230,231230,231230,231114,231116,231116,231228,231107,231114,231107,231028,231014,231010,231010,231010,231013,231013,231005,230707,230620,230621,230620,230620,230612,230328,230328,230328,230328,230328,230320,221222,221222,230320,221221,230602,221216,221207,220516,220516,220516,221029,221018,221029,221020,221020,221017,221017,221008,221008,221008,221027,221008,221008,220928,220928,221122,220928,220926,220926,220404,220404,220519,220519,220924,220924,220923,220920,220920,220514,220920,220514,220914,220914,220909,220515,211104,220420,220420,220412,220504,211111,211111,211211,220411,220411,211209,220122,211209,220122,220122,211209,211209,220122,211114,211117,211118,211111,211222,211211,220408,211117,211117,211110,211114,211019,211009,211019,211009,211009,211113,211112,211112,211113,211110,211103,210708,210708,211023,211023,211020,210915,210915,210915,210915,210915,211020,210911,210911,211019,211009,210707,210722,210803,210916,211010,210910,210910,210909,210909,210713,210604,210713,210713,210713,210708,210605,210605,210604,210604,210603,210316,210523,210523,210523,210510,210524,210510,210316,210524,210503,210503,210503,210413,210412,210412,210313,210313,210313,210410,210410,210410,210412,210316,210413,210316,210313,210312,210312,210311,210308,210126,210115,210115,210115,210115,210114,210219,210219,210218,210218,210218,210106,210108,201021,201021,210107,210105,210105,190617,190808,190326,181112,200506,181203,191216,200226,200507,200602,200602,190812,190424,191026,190925,191122,190530,190617,200218,190311,200310,181112,200414,190618,191205,200505,200507,190423,201012,201012,201006,201008,190808,190808,191024,191119,190910,190530,191015,181108,200403,200506,190609,181211,200228,190417,200505,190722,190513,191105,191105,190513,190905,181108,200113,190530,181112,190926,190326,200504,190627,191218,191026,191015,190807,181023,200113,190925,200320,190617,200218,191122,200506,200804,191216,200402,190424,200414,190530,191122,190311,181112,200414,200506,200824,191014,200425,211019,211110,211019,211009,211009,211103,211112,211113,211103,211020,211103,210915,210712,210712,211019,210916,210707,210722,210803,211010,211010,210910,210909,210722,210722,210722,210523,210510,210510,210413,210413,210410,210312,210311,210308,210126,210115,210114,210218,210129,210108,201021,210108,210107,201017,210112,201010,201010,201006,200824,200425,181108,200506,190618,190618,200228,200819,200820,200505,181206,181206,200507,190905,191120,200226,181121,190926,200504,191015,200402,190305,231013,231005,230612,221221,221221,221216,221215,221207,220516,221122,221029,221018,221029,221028,221027,221018,221018,221017,221017,221008,220929,220928,221122,221215,220519,220404,220427,220920,220514,220514,220914,220914,220327,220427,211208,211208,220408,220122,220322,200414,200506,211019,211009,210410,210218,210108,201006,191024,181211,190423,190307,200210,190807,200218,200218,190925,200905,181212,190401,190424,200602,200602,190812,191026,190812,190925,200414,190617,200804,200310,191014,181203,200507,191205,190530,190808,211114,210708,211114,210708,210915,210911,210503,210708,210722,210722,210707,210605,210605,210523,210523,210510,210524,210503,210413,210319,210319,210312,210315,210312,210308,210306,210126,210115,210219,210218,210106,201021,210105,210107,210112,201012,201006,200824,200824,191119,190910,190910,190910,200602,200819,200505,200113,181212,191011,191024,200403,190722,200403,200210,190905,200228,190524,191105,190305,190305,200820,191015,190807,191218,181023,200402,200504,181112,190326,200106,221122,221122,220922,220914,220505,220327,220327,220412,211111,220122,220322,211117,181112,191024,191205,200602,191105,190808,190104,181008,181211,181211,181112,190627,191216,180806,191014,191211,190925,190424,190424,200804,191122,190104,190424,190311,181126,181203,190925,190531,200226,210112,210112,201006,201008,181212,191024,200602,200403,191011,200820,190417,191211,200504,200106,200425,181206,181108,191119,190305,190513,181112,181121,200320,200226,190326,190812,191227,191028,181212,190807,191218,190305,200218,200824,200602,191216,191015,200605,200228,190627,191107,191211,190812,200210,191205,190311,181203,200824,200310,210524,210524,210319,210410,210316,210315,210308,210306,210306,210126,210106,210107,201017,201017,201012,190423,200403,200605,200602,181212,190530,191105,190910,200505,190423,200403,200820,200318,191011,191120,191205,191120,191120,191105,190807,190807,200504,200113,190305,200113,191105,191105,190524,200226,220922,231116,231116,231005,230320,230320,220408,211118,210413,201008,200318,210112,210114,190326,200905,200602,190311,190423,181112,191014,190925,190530,190617,181211,200403,191122,211118,211019,211023,211020,210915,210712,210712,211019,210510,210524,210524,210503,210503,210319,210316,210311,210308,210114,210106,201021,210112,200605,191024,181108,200504,200703,190722,200820,200403,181206,191107,191028,190305,190807,181108,181112,190513,200819,190812,230412,220926,220922,220514,220505,220909,220515,220404,220428,211114,211118,220322,230612,181203,210112,181112,190609,190531,181126,210916,211103,210911,211010,201010,181206,191024,181108,181112,190326,190812,200226,231014,231010,231005,230707,230620,230612,230516,230504,230412,230516,230328,230328,221222,221216,221122,221018,221018,221017,220929,220926,220924,220514,220914,220122,211009,210915,211117,231230,231230,231114,231028,230707,230621,230620,230612,230412,230320,230320,221017,220404,220926,220427,211104,220914,220427,220428,211110,220409,220409,220408,211110,211009,210916,211110,211112,211020,210911,210712,211019,210713,210713,210708,210605,210503,210412,210412,210410,210115,210114,191122,190423,200403,190618,181112,200506,191014,200106,191211,190812,190530,190104,200218,181112,200403,191011,190423,210112,210112,201012,201010,200905,190808,181212,191024,190530,190307,190417,191015,181108,200403,191011,200820,200425,191216,190513,190905,190513,200618,190305,181108,190530,200226,190926,191120,190627,191227,181203,191015,190812,190925,190617,200218,220516,210312,211118,211118,211009,210916,211113,211112,211113,211103,211103,211023,211023,211023,211020,210915,210915,211009,211009,210909,210712,210803,211010,211019,210803,210916,210803,211010,210707,210909,210713,210707,210708,210605,210603,210603,210523,210604,210510,210510,210603,210524,210524,210503,210312,210412,210412,210319,210319,210316,210115,210115,231204,231204,231230,231116,231228,231107,231114,231028,231013,231010,231005,230707,230620,230607,230620,230607,230413,230413,230328,221222,221221,221216,221215,220516,221029,221027,221018,221018,221221,221028,220504,221008,221008,220504,221122,220504,220929,220924,220924,220926,220920,220920,220909,220505,220427,220327,220404,220426,220412,220412,220411,220411,211209,220408,211209,211118,211118,211208,211211,211114,210712,210712,200106,190104,211114,211112,211023,211103,211103,211023,211023,211020,210915,210915,211009,210803,210712,211020,210707,210722,210713,210722,210708,210605,210315,210604,210603,210604,210523,210510,210510,210524,210524,210503,210503,210413,210410,210412,210410,210319,210316,210315,210312,210311,210308,210126,210306,210126,210114,210219,210218,210129,210112,190326,230620,230707,230621,230612,230515,230515,230413,230607,230412,230328,230328,230328,230328,230320,230318,221222,221216,221215,221207,221122,221029,221018,221207,221018,221020,221008,220923,220922,220920,220514,220914,220505,220909,220327,220515,220404,211104,220420,211222,220322,220411,220122,211208,220408,211111,220408,220322,231204,231230,230412,221017,220411,220122,220122,211209,211209,231230,231204,231116,231228,231114,231114,231028,231005,230620,230621,230612,230607,230504,230504,230412,230328,230320,230516,221216,220516,221029,221029,221216,221008,221008,220926,220926,220926,220923,220920,220920,220514,220914,220914,220327,211104,211222,220411,220408,220409,211209,211117,221027,190423,190609,190609,190627,190618,190627,190424,190609,200106,191216,200414,191211,200824,200210,200218,191014,181126,190609,191122,211118,211114,211009,211112,211112,211110,211103,211103,210708,211020,210915,210915,210915,210803,211020,210911,210712,210712,211010,210707,210803,211010,210910,210910,210909,210909,210713,210722,210707,210713,210707,210707,210708,210605,210605,210605,210604,210603,210523,210523,210523,210510,210524,210510,210313,210413,210410,210319,210316,210316,210316,210316,210308,210306,210306,210107,210107,210107,210107,210112,210112,191122,190722,181126,181211,191211,181203,200310,181108,181108,190530,200703,200703,191120,200310,190905,200403,190609,200618,191015,191015,191211,190812,190926,191015,181203,191026,181203,190305,200804,191218,190305,200804,190326,191216,200602,200106,200905,191026,191026,200320,190530,190530,191122,190104,200218,190311,181112,181112,190617,200605,200414,181203,191119,191205,191014,200507,200824,190423,200210,181212,191107,200824,200310,200605,191205,190910,211019,211019,211009,210916,210708,211103,211023,211023,211020,210523,210803,210915,211009,210909,210910,210712,210803,210916,210910,210910,210909,210722,210707,210707,210605,210605,210605,210605,210604,210523,210524,210503,210319,210413,210410,210319,210316,210315,210311,210308,210308,210306,210126,210115,210115,210114,210219,210114,210218,210219,210129,210129,210108,210105,201012,201017,201021,210107,210105,201017,210112,210112,201012,201010,200210,181108,200403,190618,181211,190530,200505,190423,190423,181108,190722,190722,200318,191105,190905,190513,200618,200703,200703,191120,200414,190609,190609,190307,190307,200820,200504,181121,181206,200226,191015,191218,190926,200113,190807,200402,200402,190617,181112,190812,200506,200804,190326,190627,231230,231228,231107,231114,231107,231028,231014,231005,230707,230707,230620,230620,230621,230612,230412,230412,230328,230320,230320,230621,221222,221216,221207,230318,221122,221018,221027,220519,221216,221020,221017,221027,221008,220928,221122,220928,220924,220504,220922,220505,220909,220427,220515,220426,220420,220412,220428,220428,211110,220411,211118,211208,201008,200218,200425,200804,200106,190424,181112,190617,181206,190104,200320,190513,190531,191205,200403,191014,200507,190531,200414,200218,181212,200824,190925,190617,200605,190910,190618,190808,200605,200403,200106,181211,200824,200320,200310,200324,191011,211110,211019,211009,211010,210708,211114,211110,211103,210708,211023,211023,211103,210915,210915,210915,211020,210712,211020,211019,210916,210916,211020,211010,211010,210910,211010,210910,210909,210713,210722,210722,210708,210708,210605,210604,210603,210523,210316,210523,210510,210524,210524,210503,210503,210503,210413,210313,210412,210412,210410,210410,210316,210315,210313,210311,210308,210308,210306,210306,210129,210126,210115,210114,210114,210114,210219,210219,210218,210218,210106,210108,210105,201017,210112,210112,201010,201006,201006,200824,201008,200505,200210,190424,190417,190531,181206,191024,200506,200318,190618,190722,200210,200310,200210,200820,200506,181121,200618,200228,190927,190905,200605,200504,200703,190417,190531,190609,190722,190513,191107,191028,190807,190905,200113,190305,190609,200226,200226,190927,190513,190524,190326,200402,181203,200819,181112,181112,181206,181203,191105,191216,200506,191026,190812,190812,231010,231010,231013,231010,231005,230620,230612,230621,230621,230612,230515,230504,230328,230328,230320,230318,221222,221221,221216,221215,221207,221029,221020,221027,221028,221027,230320,221018,221020,221017,221017,221017,221008,220929,220928,221122,220504,220404,220929,220924,220924,220427,220923,220922,220514,220514,220914,220914,220505,220909,220327,220515,220426,220404,220420,220420,220428,220122,211111,220322,220122,211209,220409,211209,211117,211118,211208,220322,220322,210523,210722,210707,190513,220924,220924,220909,220515,211104,220420,220412,220428,211111,211111,220411,200618,200703,200210,200402,190609,190307,190307,200228,200504,191024,190703,181206,191015,200602,200320,190609,181112,190703,200506,191216,191216,190424,190424,200824,200605,200210,190311,200824,200505,200324,191205,211019,211009,211222,210708,210708,211103,211103,211023,211020,210915,210915,211020,211009,210911,210712,211019,211009,210722,210916,210910,210909,210713,210722,210708,210708,210605,210604,210603,210523,210523,210523,210524,210503,210503,210413,210412,210315,210311,210311,210308,210306,210306,210129,210126,210114,210219,210106,201218,201021,210107,210105,201017,201012,201006,201008,200318,200602,190530,190910,180807,200403,190609,181121,191011,200504,200703,191105,181206,181206,191015,200318,200402,191227,200226,200819,191218,200106,230620,230620,230612,230515,230504,230412,230412,230328,230328,230328,230320,230607,221221,221216,221216,221207,221028,221027,221018,221020,221017,220929,220928,221122,220926,220404,220924,220920,220914,220514,220327,220122,220409,211209,220322,220322,220322,191218,191028,200226,190627,191216,200507,201006,191011,200505,181108,190417,190703,200905,190311,181203,200226,190104,181212,181126,181211,181212,181023,181203,190423,181203,181108,181108,200226,181023,200602,230612,230504,230413,221215,221207,221122,221028,221027,221027,221017,221008,220928,220519,220427,220914,220412,220122,190926,181121,190807,200402,211113,211110,231230,211211,200113,190812,190812,230328,210523,190424,210316,200618,200228,220914,211117,231204,231116,231116,231114,231114,231028,231010,231005,230707,230620,230621,230515,230412,230412,230328,230318,221222,221215,221207,221029,221028,221027,221020,221008,220929,221122,220427,220922,220920,220427,220914,220327,220420,220420,220412,220428,220428,211110,211111,220322,200804,200106,181203,191015,191014,191026,200618,191216,190424,181112,190617,200310,200320,190424,200824,190812,200403,190905,191205,211009,210708,211110,210915,210915,211009,211019,210803,210909,210713,210722,210708,210605,210605,210316,210412,210523,210510,210503,210503,210410,210410,210316,210312,210308,210126,210129,210114,210218,210106,201021,210107,201017,201010,201006,201008,190423,181212,191024,190910,190417,200320,200318,181108,200820,190530,200703,200210,181206,191107,191015,200504,190524,181112,190926,200504,190703,181206,190401,181211,181203,190807,190910,200425,191218,200226,210115,190722,191105,191120,191211,181203,181112,190311,181203,181108,181121,181108,190417,211113,221008,220412,200905,190424,190104,200320,190808,210605,200218,190905,231204,231230,231116,231114,231028,230621,230412,221027,230318,200106,190617,181108,200210,200506,200804,190326,190627,200506,190424,200602,181203,190104,191026,200320,190925,200605,190531,211010,210916,200226,231116,231116,230328,221215,221207,221207,221018,220924,220926,220428,220411,220122,210915,210708,210708,210524,210219,201012,201010,200824,201008,190530,181121,200402,191227,200402,191119,200507,190808,191107,200605,210316,210115,210218,201017,210112,201006,201008,200403,181126,181112,190424,190905,191026,181126,181126,181212,190104,190311,181203,190417,181108,181211,181206,181212,181108,190104,190401,190524,181121,181206,190401,181023,181015,190326,190423,190423,230621,221027,181203,191216,191107,190609,191122,191205,191205,191205,200218,190617,181203,191119,191014,200507,200824,190808,191105,200310,211110,210916,211010,210605,210503,210410,210412,210115,210218,210108,201021,201017,210112,201010,201010,201006,200824,200605,191024,190513,200703,200402,200228,190401,191227,191015,200113,200804,231114,230620,230612,230620,230413,221216,221207,221122,221122,220404,220923,220923,220514,220515,220515,220122,211209,211114,190311,221020,220505,191122,210510,210311,201008,190530,200402,181206,200320,190609,191122,181112,181112,190627,200218,191216,191107,191014,190424,191107,200824,191211,200210,190311,200403,191205,191011,210911,210503,210413,210412,210308,210306,201017,190722,191024,191105,191105,181212,181108,181121,190609,200210,190722,190513,190513,190807,191211,200402,190531,200113,190812,191218,200106,190305,190305,221215,221122,221018,221008,220914,220409,220322,200320,200414,181112,200506,200602,181203,191014,191211,200602,191107,190530,190617,200804,200218,200218,190424,200824,200106,191107,200226,200106,190424,190423,211110,211113,211103,211020,210911,211020,210712,210707,210722,210708,210603,210412,210316,210306,210115,210218,210106,210105,210112,201010,190808,190808,191119,190311,190910,190618,190618,190609,190618,181121,200402,200228,191015,191015,200318,200403,200318,181206,190513,190927,191227,190326,191120,190627,190812,190807,190305,231204,231116,231228,231028,231010,231013,231013,230504,230707,230707,230707,230504,230516,230413,230328,230320,221221,221215,221029,221027,221215,221122,221017,220924,220920,220909,220914,220327,220404,220411,220409,211222,211211,211019,210916,210722,210713,220504,220924,220427,220426,220420,220122,200905,181112,191026,200320,190104,191014,200507,200605,190423,191107,210916,211009,210713,210523,210503,210316,210126,210108,210112,181211,200505,200605,181108,190722,190513,191105,191120,200402,190609,190401,190307,190305,200820,191227,200226,181112,181108,190910,190812,231107,230707,230607,230504,230318,221029,221008,220926,220427,220505,220412,220411,220322,220322,181203,181203,191205,200605,200210,200824,181211,181203,210604,210313,210306,210129,210218,201017,201008,181121,181211,200403,200403,230620,230328,221215,221029,220924,181108,191120,190927,200504,191105,190926,190807,200402,191218,200113,200804,190424,190530,191119,191119,190311,190618,190722,190513,190305,190927,190703,191028,190905,190807,190531,190812,210524,200324,181212,181211,200226,230620,230515,220516,220519,181203,190531,190925,190311,181023,231114,231013,220122,181112,190618,180709,181211,190722,181121,191015,210915,210312,210115,200505,190513,230620,220412,221008,231228,231028,220909,220909,220515,220420,200403,211209,211209,211117,190627,191216,200425,200804,200602,190424,181203,191014,200210,190524,200218,200228,181112,190812,200824,190925,190311,181203,181212,191205,211019,210708,211113,211110,211020,210915,210712,210712,211019,210803,210713,210713,210707,210708,210315,210605,210523,210523,210524,210503,210413,210412,210410,210316,210315,210313,210308,210306,210126,210126,210218,210108,210108,210107,201017,210112,201012,201006,201008,200505,190618,191119,200210,181108,200403,200819,191011,190530,190722,191024,190417,200618,181206,190905,200403,190609,200403,200504,190513,191227,200113,191026,191120,181112,190326,190326,191216,190326,200226,191024,200905,190424,190524,191122,200414,190617,190104,190104,191218,190424,190925,200506,190703,190617,200824,190530,200507,190808,210915,211019,210916,210916,210605,210605,210524,210503,210413,210412,210316,210315,210311,210115,210115,210219,201021,201021,210108,201021,201008,200310,200310,200226,190311,190531,200425,190910,181121,181108,181108,191227,190910,190417,191120,200318,181108,181121,200226,191028,191028,191011,210604,210603,210115,201006,190530,181211,191227,200228,190305,200506,231230,231010,230412,220505,190627,191119,191122,191216,191216,200506,200506,200506,200602,200602,200602,200824,200905,200402,200320,200320,200320,190530,200218,200218,200414,200414,200414,190617,200605,200804,191205,191205,200507,200824,200824,200824,190808,181212,181212,191107,200820,200824,200310,200310,200425,200605,200324,191011,191011,190910,190910,211114,211019,210916,210916,211113,211112,211113,211103,211103,211023,211023,211020,210915,210915,210915,210803,211020,211009,211020,210712,210712,210803,210803,210803,211010,211010,210910,210909,210910,210909,210722,210713,210713,210713,210708,210604,210523,210523,210523,210523,210510,210510,210510,210524,210524,210312,210503,210312,210413,210412,210412,210410,210410,210410,210316,210316,210316,210315,210313,210308,210308,210308,210308,210306,210115,210126,210115,210219,210219,210114,210219,210219,210108,210129,210218,210218,210108,201021,201021,201021,201021,210107,210107,210107,210105,210105,210105,210105,201017,201017,210112,210112,201010,200824,201010,200819,201006,201006,201008,200210,200210,181126,181108,181126,181126,200403,190618,200505,200505,190722,190423,190423,190423,200318,200318,200618,200618,200703,191120,200402,200403,200402,190609,231204,231204,231114,231114,231114,231230,231230,231116,231107,231228,231114,231228,231107,231107,231204,231014,231028,231014,231013,231013,231013,231005,231005,231005,230707,230707,230620,230620,230620,230620,230707,230620,230612,230612,230612,230607,230607,230607,230515,230515,230515,230413,230413,230413,230413,230328,230328,230320,230320,230320,230320,221215,230318,230318,221215,221221,221207,221207,221207,221207,221029,220516,220516,221028,221027,221027,220519,220519,221020,221020,220519,221017,220519,221008,221027,221017,221008,220504,220928,220928,220928,220928,220504,220504,221122,220929,220923,220922,220920,220920,220514,211104,220920,220504,220920,220914,220914,220909,220909,220505,220427,220427,220515,211104,220420,220412,220412,220412,220504,211110,211111,220411,220411,220411,220122,220122,211209,211118,211211,211211,211208,211208,211208,211117,211110,220923,190627,191216,200602,200602,190311,200106,191024,190925,181126,200505,190812,200602,190530,190618,181121,190307,190812,181203,200402,191028,190627,190627,190925,190311,190423,181112,181203,190104,190424,190812,190905,190530,190618,191107,191107,191107,200320,190617,190617,200218,181126,200414,200414,191205,190808,181203,190424,191015,200310,211110,211009,211112,211103,211019,210803,211010,211010,210916,210916,210909,210909,210722,210707,210604,210603,210510,191024,190423,181212,190722,190722,190423,191211,181108,181121,190417,190927,191011,190513,181203,181206,191015,190305,190305,200504,190703,190808,191218,200402,190531,191028,191015,231228,231228,231114,231107,231010,230707,230612,230607,230621,230612,230504,230328,221216,221029,221029,221028,220514,220514,220514,220914,220909,220404,220404,220426,220420,220420,220420,220412,220412,211222,220411,220411,211114,211117,211117,231114,231013,221018,220427,211209,220122,220122,190812,210915,210106,190513,210915,210106,191211,181126,190627,200228,200228,200228,190927,200820,200820,200504,200504,181121,181121,190703,190305,190305,180911,200226,200226,191218,191218,190926,190927,190807,190807,200113,200402,200402,190531,200819,200804,181203,191216,200506,200602,200905,200320,200602,190812,200218,200310,181126,190617,211113,211113,210605,210503,210503,210503,210503,210311,210306,200605,201010,190618,200820,181206,190927,190531,200113,190513,221027,220505,220122,190531,190609,190104,191026,200218,190617,190617,200605,200414,191119,191205,200507,200824,190423,200106,190808,191107,200820,200310,200605,190910,210319,210312,210316,210315,210312,210308,210306,210306,210126,210126,210114,210219,210218,210129,210108,201021,210107,210105,210112,210112,210112,201010,190609,200819,200403,181211,191024,200505,181206,200318,200618,200703,200210,200210,190307,200402,200820,200820,200228,190927,200820,200504,200226,191015,191028,190807,200402,200804,190326,190326,190424,181121,190104,190104,190925,190808,210603,220920,211110,210909,211010,210413,201021,210107,231107,231114,230621,230607,230516,230413,221017,220926,220923,220428,211209,211114,211010,190326,200505,190627,200905,190424,191211,200602,190812,200106,191026,190530,190808,200804,190311,181112,200414,190617,181203,191014,200507,191205,190424,181212,201010,201006,201008,190808,190524,200425,190910,200310,200820,181108,181211,191024,190618,191011,200507,200228,191024,190417,200703,181206,191105,190905,200318,190305,181108,190524,191120,200402,181121,190417,190703,191028,181206,191105,190807,191218,190305,200504,190618,181023,200402,181112,200318,211009,211103,210503,210312,210410,210412,210315,210114,191119,191218,231230,231005,230612,221215,220922,220922,220914,220428,211222,200218,210712,210112,201008,200320,200210,220914,190305,191014,200414,211222,210713,210308,201012,220412,210722,201017,190307,191015,191218,231204,231230,231230,231116,231228,231107,231107,231028,230707,230621,230612,230612,230504,230504,230412,230412,230412,230328,230516,230318,221221,221221,220514,220515,180620,190513,210315,191024,191024,181108,231114,220924,220924,190910,191216,200218,200226,210306,210129,210126,191026,191120,200819,191227,231204,231114,211117,190618,191105,190703,190807,190618,210909,200310,211020,210910,210909,210707,210524,210315,201010,221029,231028,210708,190609,181206,181023,181112,200318,190627,181203,190926,200824,190423,190424,190808,190513,190104,191122,190925,200605,190703,181206,191011,200505,211019,211019,211009,211009,211103,211019,210915,210911,210712,211010,210713,210713,210605,210523,210604,210413,21041 3,210112,210319,210312,210312,210311,210306,210306,210126,210115,210114,210218,210218,201021,210105,201017,201010,201008,181212,191105,190910,181108,190609,190307,200324,190808,181206,200318,190513,191105,190305,200403,181112,190926,190531,181203,191015,191119,190812,190305,181023,231116,231028,231014,231010,230620,230707,230504,230412,230412,230328,230320,230516,221216,221029,221018,220924,220427,220924,220505,220427,220420,220428,220412,220122,220409,211117,211117,210410,231114,231010,230620,230320,221222,221207,221028,220923,220504,220514,211104,220412,220428,220122,211111,211117,190104,191014,191205,211023,210803,210708,210523,201006,200602,190530,191211,200320,200505,181121,200506,220516,221221,221027,220920,220514,220505,211104,220420,211111,190618,210803,210803,210712,210413,200324,200505,190722,200402,190926,190627,181203,191216,191216,190531,191211,191211,200113,231114,231114,231005,230413,231116,231010,231013,231013,231005,230612,230504,230504,230320,230320,221221,221221,221018,221018,221017,221017,220926,220928,190424,191205,210911,210803,210105,201006,200703,190305,200602,200905,190104,231114,230515,220516,220923,211117,181108,190617,210605,200507,210916,210707,190326,181212,181211,181203,200226,190531,200602,190812,191216,191026,200320,200218,200218,190104,190617,181121,181203,181203,181112,200414,200804,191014,190617,200824,200414,191014,191122,191011,190910,190925,210510,210524,210510,210503,210413,210413,210412,210412,210112,210410,210319,210315,210315,210313,210311,210308,210306,210306,210126,210115,210115,210114,210219,210218,210218,210218,210108,201021,201021,201021,210105,210105,210112,210112,210112,201012,201012,201006,201006,200824,201008,201008,200505,200310,190524,200425,190722,190808,191227,191107,190311,191024,200210,181108,191014,191105,191211,191205,200819,190531,200506,200403,190808,190722,181121,181206,200703,190905,191011,191120,190305,200504,181121,190531,181108,190305,190423,190513,190417,200226,190927,191015,200318,200402,191227,190424,200819,190524,181203,181023,190627,181112,190305,200320,191119,191119,190311,190424,200414,200605,191014,191014,200310,200425,210916,210910,210707,210708,210708,210604,210524,210524,210413,210412,210319,210410,210311,210308,210126,210129,210129,201021,201017,201012,190530,190609,200210,200218,200210,190311,200819,191205,190423,200505,200318,200618,200703,191120,200402,200402,190609,200820,191028,191105,200226,190926,190807,190807,200402,200804,200506,190627,191216,190531,191026,190925,191119,190104,191014,190423,201008,181203,191122,190311,190627,200218,200506,190424,190812,190530,200824,190925,200310,191218,191205,181126,191011,200824,211019,211009,211009,211009,210708,211020,210915,210915,210915,210911,210712,210712,211020,210803,210916,210713,210713,210722,210707,210707,210603,210503,210413,210410,210315,210312,210312,210306,210129,210114,210114,210114,210218,201218,210107,201017,201006,201006,190531,190722,181212,191026,200318,200602,190910,181121,190311,190307,200403,200228,191011,200820,190531,200504,200703,190524,191105,200318,191211,181108,181108,190307,200402,200403,190513,190513,181112,190926,190627,190926,200819,191026,200226,191227,200106,190305,230413,230412,230328,230320,221221,221215,221122,221028,221017,221008,220504,220926,220514,220914,220515,220426,211104,220428,220428,220411,220411,220411,220409,211118,211009,231005,191216,191014,200905,190424,191014,200218,190104,190617,190925,200414,200218,200824,200403,190808,191205,210722,210510,210603,210503,210413,210316,210316,210315,210308,210126,210126,210115,210219,201008,210108,201021,210105,210112,210112,201010,201006,190609,200310,190703,200602,190910,190910,191211,190530,191216,200210,190417,181206,191105,190926,200310,191218,190423,200113,200605,191026,191026,191105,190305,220926,220420,211110,211111,211111,200905,190424,210510,210308,210311,210311,210311,210308,210306,200905,201006,221008,220420,220408,200320,210915,211019,210803,210315,210605,210524,210319,210315,210306,210114,210219,210107,201006,200505,200507,200320,191028,200310,191218,190926,191120,200605,200106,220408,220408,190311,210915,210713,210413,210306,201008,181203,191216,181203,181203,190812,230607,220914,231204,231116,231114,231107,231114,231107,231107,231028,221222,230607,230607,220519,221018,220504,221122,220926,221028,220404,220926,220924,220924,220514,220514,220514,220914,220914,220928,220909,220404,220428,220428,211222,220322,220122,211209,220409,220327,211111,220322,211222,211112,211113,210708,211023,210915,211020,210712,210712,211019,210707,210722,210916,210722,210909,210713,210722,210713,210713,210605,210605,210605,210605,210604,210603,210523,210523,210510,210413,210412,210410,210412,210316,210313,210313,210311,210308,210129,210126,210114,210114,210219,210218,210108,201021,201017,200905,181112,200506,190627,181203,200618,200226,190424,200804,181112,190104,190513,200403,190617,190311,181203,181203,200505,201017,210112,201006,201006,201008,201008,190808,200226,190530,190910,200403,200403,190401,200228,200820,200505,181121,191015,191015,190513,200106,200318,190513,190905,181108,200113,190305,181112,200226,200819,191015,191218,190807,181023,190305,190925,200804,211114,211114,211009,211019,210803,210910,210803,210909,210524,231204,231230,231114,231116,231228,231107,231028,231028,230621,230620,230621,230612,230612,230515,230607,230413,230412,230412,230320,230320,221221,221216,221216,221221,221207,220516,230318,221029,221027,221018,220519,221018,220519,221008,220928,220928,220926,220926,220514,220514,220914,220505,220412,220412,220504,211211,211111,220411,211209,211118,211208,211117,190926,231116,231116,230621,230612,230413,191216,200106,190311,190618,190618,181211,191024,181108,190722,200318,181206,190618,190926,190807,211009,211111,190423,201012,231013,220504,190326,190326,181203,191216,200506,200602,200905,191216,200320,200320,191122,191026,190925,190618,200218,200414,191119,191205,200507,191014,200824,190910,191211,190808,191216,191107,200824,200310,200425,191011,210312,210523,210524,210524,210312,210410,210412,210312,210312,210319,210316,210315,210311,210308,210115,210114,210219,210218,210218,201021,210107,210105,201017,210112,210112,201012,201006,201006,201006,200605,200210,190618,200228,200228,191024,200505,190722,200113,200318,190905,190910,200618,200703,200210,200403,200228,190927,200820,190417,190703,200226,191227,181212,191015,191218,191028,191028,200113,190807,200402,200402,190609,231208,231230,231230,231230,231204,231204,231116,231114,231114,231114,231014,231010,231005,230620,230621,230612,230620,230607,230621,230516,230504,230413,230328,230412,230328,221216,230320,230318,221215,230607,221215,221215,221215,221207,221122,220516,221027,220504,220504,220519,221017,221017,220519,221008,220504,220519,221122,220926,220923,220504,220514,220914,220914,220909,220505,220427,220427,211104,220412,220412,220428,220504,220428,220428,211111,220322,211209,211222,211208,211208,211117,231107,221020,220926,220926,220409,210916,210603,210603,210312,181108,190305,220515,220428,211209,191211,190617,181112,210712,210503,210312,210312,210115,201218,200505,190530,190910,200505,181108,200228,190524,181112,190807,200504,190305,231014,231005,221215,221028,221027,220923,200310,181108,191216,200320,190307,190311,181126,200507,181203,181212,191107,210112,191119,190910,200113,190417,200703,200402,190609,200504,190722,190417,190513,181212,190925,190305,181023,230320,221018,220929,220922,220505,220404,220420,220411,220411,211019,211103,211019,210604,210523,210510,210503,210413,210319,210315,210311,210115,210219,210108,211209,190925,190531,181108,181108,190531,190627,190530,190104,181112,181126,190311,190423,181112,190417,190423,181206,181203,190807,191014,200320,190925,200218,191120,200602,200414,191216,191216,191011,211019,210915,211020,210712,210916,210713,210510,210410,210313,210306,210115,210219,210106,201021,201017,201008,201008,200820,191105,191024,200318,200504,200113,231230,231116,231114,231028,231014,230612,230413,230320,221122,221018,221029,221020,221008,220516,220923,220505,220909,220515,220404,220412,220122,220408,211009,211113,210915,210604,200504,211222,190812,221017,191120,190326,190808,181112,200402,190513,200602,181203,190401,191216,191014,191026,191122,191026,190807,200320,190424,191014,200414,200414,191122,190531,181126,200507,200106,191015,190609,181212,201006,201006,200605,200210,190910,191119,190311,191024,190618,200602,200602,181211,200820,190530,190910,190531,191211,200506,191011,200505,190722,200820,200318,191216,200703,190417,190417,181108,190905,190513,191120,190401,200318,190609,200218,200310,181108,190417,190305,191227,190513,200504,180905,181206,190927,191026,191015,200318,191227,190807,200605,191028,200113,190531,200804,190326,200605,190617,191122,221222,221027,220929,220411,220411,211110,211019,211023,211010,211010,210510,210316,210311,200819,200320,190808,190617,190925,200403,200403,211103,210911,210713,210713,210510,210316,210108,210106,210106,201017,201017,210112,201012,200113,200113,200228,200618,200618,191024,191024,190513,181203,231114,230620,230328,221216,221018,221122,220920,220515,211117,190925,190812,191011,191015,190910,191011,190905,231228,231228,221221,200310,210114,200602,181112,191026,191119,200824,191211,181212,200605,190910,211009,210910,210910,210909,210510,210410,210311,210126,210114,201021,201021,190609,190311,181121,190104,200228,200324,181108,190722,200113,200402,200820,200228,190417,190531,200506,231107,231107,231014,230612,230413,221215,221018,221029,221122,220922,220514,220420,220408,220322,200602,191216,191216,191216,191216,191211,191026,200506,191216,200310,200506,200507,200218,200507,191205,191205,200403,191205,200310,211110,211019,211009,211019,211019,211009,211112,211110,211103,211023,211020,211020,210712,210803,211010,210916,210713,210707,210707,210605,210604,210523,210523,210524,210503,210413,210413,210413,210311,210410,210319,210313,210313,210313,210312,210126,210306,210129,210115,210115,210115,210114,210114,210114,210219,210218,210108,210106,210107,210108,201021,210112,201012,201010,201010,201008,201008,200106,200320,191211,200403,200210,191205,191227,200228,191205,200318,200618,200703,191120,200318,191120,200228,200228,200113,200820,191227,200318,200425,200504,191227,200226,200504,200819,200605,200804,200425,200402,200113,200226,200425,231028,230328,230328,230320,230320,230320,221222,221221,221221,221216,221215,221215,221221,221207,221207,221207,221207,221122,221029,221029,220519,221008,221122,221122,220924,220926,220920,220920,211104,220327,220427,220427,220427,220327,220428,220404,220412,220504,220428,220428,220411,220122,220122,211111,220408,211117,211117,211117,211117,190424,190617,200218,190104,191011,190926,190807,201021,210107,191122,190925,190627,200506,200226,200106,200507,191211,190925,191122,200310,190530,200403,190808,190808,201010,201006,200310,191024,181121,200602,191015,191105,191015,200210,200403,191211,200505,200318,200425,190722,190513,191105,200113,190910,181108,190513,191120,200504,190627,191227,191015,190807,190807,200605,191026,200113,200320,200320,191026,210106,211114,231116,231116,231114,231005,230620,230612,230607,230412,230328,221221,221221,221215,221207,221207,221027,220519,221020,221018,221017,221008,221008,221122,220404,220404,220924,220519,220924,220923,220922,220920,220909,220909,220427,220515,211104,220420,220122,220409,220409,220409,211118,211111,220408,211009,211114,211112,211110,211103,210708,210915,210803,210916,210707,211010,210722,210916,210909,210909,210722,210713,210722,210315,210605,210707,210707,210605,210604,210510,210316,210315,210311,210308,210306,210306,210115,210126,210115,210114,210114,210114,210219,210218,210105,191122,190925,200506,191216,200905,200507,200226,181203,191107,190925,191122,190311,200310,181112,200218,200824,200414,191205,200505,190808,201017,210112,201012,201010,201006,201006,201008,191024,200310,181121,190513,200602,190910,181108,200210,190307,190311,200403,181121,200820,200228,191015,191211,200505,190513,190722,190905,190910,200318,181108,190513,181121,200113,200226,200819,190326,190417,191120,200504,190627,191227,191015,190807,191218,200113,181112,181023,200320,200804,181023,181112,231230,231230,231230,231230,231230,231116,231116,231116,231107,231107,231114,231010,231010,231005,231005,231005,231005,231005,230707,230612,230621,230515,230504,230413,230412,230328,230328,221215,230516,230318,221221,221221,221221,221216,221216,221221,221221,221216,221207,221207,221207,221207,221029,221221,221018,221207,221018,221018,221017,221008,221008,221008,221008,221008,220404,220926,220924,220923,220923,220920,220920,220514,211114,220514,220914,220914,220504,220909,220505,220505,220428,211222,211209,211111,211111,220411,220411,220122,211209,220327,211117,211111,220322,220322,220408,211117,211019,210916,211113,211112,211113,211103,211103,211023,211020,210915,211103,210915,210915,210915,211020,210911,210712,210712,211019,210916,210916,210707,210707,210722,211010,211010,210909,210722,210713,210713,210713,210722,210603,210713,210605,210605,210605,210604,210603,210603,210523,210523,210523,210510,210316,210316,210524,210604,210524,210524,210503,210503,210413,210604,210412,210410,210315,210313,210313,210312,210311,210308,210306,210126,210306,210306,210126,210115,210115,210114,210219,210218,210218,210218,200905,181112,200506,190627,191216,191014,191211,200507,190812,191026,190925,190617,191122,190104,200218,181112,200403,190311,190618,191205,191011,200507,201010,201006,200804,201008,201008,190808,200310,181108,190609,191119,181212,191024,191024,200602,200425,191015,191024,190910,190618,181211,181211,200228,191011,191011,200505,190812,181206,200318,181211,181108,181108,191105,190905,190905,191122,200819,181121,181121,190926,200504,200504,190703,191227,191227,181023,191015,191015,191218,191218,191026,190812,181023,190305,200320,190925,190617,200218,200804,191122,231005,200210,200602,200804,191122,200905,200320,191205,191216,191216,191014,200507,200218,200824,200605,200210,200824,200310,191122,200324,191011,211118,211019,211019,211009,211019,210708,211113,211110,211103,210708,211023,211020,210915,210915,211020,210911,210712,210712,211019,210916,211010,211010,210910,210909,210713,210722,210707,210708,210708,210605,210605,210604,210603,210523,210523,210603,210503,210510,210524,210604,210524,210503,210503,210413,210313,210410,210410,210316,210308,210312,210315,210313,210311,210308,210126,210308,210306,210306,210306,210129,210126,210115,210114,210114,210219,210218,210218,210106,210108,201021,210107,210105,210105,201017,210112,210112,210112,201010,201006,201006,200505,191024,200318,191105,191227,200210,200310,200403,200820,200618,200324,200228,190927,200504,200703,191107,191218,191028,200402,200226,200402,190927,190926,200113,200113,200819,191026,191014,191216,200506,200106,231010,231010,231010,231005,231005,230707,230612,230621,230620,230612,230707,230621,230515,230504,230516,230412,230328,230328,230320,230320,230318,221222,221221,221216,221215,221207,220516,221122,221029,221018,221028,221221,221027,221018,221020,221018,221017,221017,221008,220929,220928,220924,220920,220514,220514,220914,220914,220909,220327,220427,211209,220409,211209,211118,211211,211211,220322,220408,220408,220408,200320,190609,191119,190617,200605,191205,190423,190808,181112,191105,191107,200310,211103,211009,190617,190617,210308,190609,200228,200228,220420,211118,190609,220914,220914,220427,220515,220426,220426,220420,211211,211222,220411,220408,200324,210126,211019,211019,211010,210523,210315,221017,211117,191205,191026,210604,201021,210105,191105,200504,200403,210707,210605,200605,191122,190925,190627,200905,200106,200320,190424,190424,191026,191211,190104,200320,190424,190617,190617,200824,181203,200218,200403,191014,191011,200507,200605,190808,190808,190531,210410,210412,210319,210319,210316,210316,210315,210313,210311,210308,210306,210306,210129,210126,210115,210114,210219,210218,210218,210106,210107,201021,210107,210107,201017,201017,210112,201010,201006,191107,191105,190808,191227,190417,181108,200602,191119,190311,181211,191011,200819,200210,200310,200402,181108,200505,190618,200820,191211,191105,181206,190910,200703,200318,190531,200403,190307,191015,191120,200820,200318,190513,181121,200228,181108,190703,190927,190807,181206,191015,181203,181203,181206,191026,191218,200605,200113,200320,190326,200506,181112,191122,210524,230413,220408,181203,191107,200605,200318,200804,191216,200605,200605,200905,190104,200228,190531,200218,181121,181203,181206,181212,191205,191024,191014,200618,191024,190530,211114,211019,211009,210916,211112,211113,211103,211103,211023,211023,211020,210915,210915,211009,210712,210803,210712,211019,211010,211010,210909,210722,210713,210708,210605,210605,210604,210603,210603,210503,210510,210510,210524,210524,210503,210503,210413,210603,210413,210410,210319,210316,210315,210311,210308,210306,210306,210126,210115,210114,210219,210218,210108,210108,210107,210108,201021,210107,210105,201017,201010,200824,200425,191119,191120,190910,200320,191211,190311,190417,181108,200506,190618,181211,181212,200324,191011,200819,190417,200505,200505,181108,200318,191024,190513,200618,200703,191105,200504,190524,190609,191120,191120,190401,190305,181121,190926,200504,200228,190401,190417,190423,190609,190530,191014,191028,181206,200226,190927,200402,191218,200113,200819,181023,200425,190812,190424,200318,190627,230621,230612,221215,221216,221207,220516,221122,221029,221029,221222,220516,221020,221027,221020,221018,221017,221017,220909,220327,220515,220404,220426,220122,220122,211209,220322,231230,231114,231028,230621,221215,221215,221029,221028,220504,220519,220922,220514,220920,220515,220412,220411,211117,200506,190424,191216,200618,200320,190104,191211,191216,190910,211113,210909,210713,210713,210713,210605,210523,210319,210312,210219,210108,210112,190609,190722,191211,181121,191024,200820,190722,180927,190807,190305,190531,190531,190617,190808,191107,191107,211114,211010,211113,211110,211023,211023,211020,210915,210803,211009,210909,210803,210916,210803,210803,210909,210910,210909,210713,210713,210708,210708,210605,210603,210604,210523,210510,210510,210524,210413,210413,210410,210410,210410,210315,210315,210312,210313,210308,210126,210115,210218,210107,210112,210112,190618,181206,190905,181206,190417,190703,190926,190807,190531,231204,231230,231204,231204,231107,231114,231114,231028,231028,231013,231013,231005,230707,230621,230612,230612,230707,230707,230504,230412,230328,230318,221215,221027,221027,221027,221221,221018,220519,220519,220928,220928,221017,220928,220928,220926,221028,220924,220924,220519,220504,220926,220922,220920,220920,220914,220914,220909,220909,220515,220412,220412,220428,211111,220411,220411,211118,211208,190905,191211,181203,181203,181203,200602,191211,181203,190104,200320,200320,190104,190104,190925,181112,190812,200414,191014,200824,190618,191014,190530,200605,211019,211019,210916,211103,211103,211103,210915,211020,211020,210712,210707,210803,210722,211010,210909,210315,210707,210605,210605,210523,210523,210510,210524,210503,210503,210316,210315,210308,210308,210126,210126,210115,210114,210219,210218,210108,210107,201021,210107,201012,201010,201010,201006,190808,191119,191024,190910,181211,191211,191024,200403,200403,190530,200505,200228,190307,190423,191024,190530,200618,190722,181108,181108,191105,200228,191011,200820,200318,190513,190531,181206,190513,190703,191218,200113,181023,191028,191218,200605,191205,200226,191122,231204,231116,231114,231028,231028,231014,231010,230707,230707,230320,230620,230504,230412,230412,230320,230318,230318,210605,221215,220516,221020,221020,221008,221008,220928,220928,221028,220519,220504,220427,220926,220922,220920,220514,211104,220914,220327,220420,220412,220412,220428,211111,220411,220409,211211,220408,210218,231116,231028,231010,230621,230320,230516,230320,230320,221207,220505,221029,221029,221018,221017,220928,221222,220923,220514,220909,220420,220428,211118,191216,200106,191205,200507,200507,190423,181126,191014,181211,211110,210916,211110,210713,210722,211112,211020,210911,211019,210916,210708,210604,210603,210510,210524,210412,210410,210410,210315,210312,210311,210115,210114,201017,201017,201008,201008,200602,191026,191024,191211,190530,190307,190307,200318,190722,181206,190905,190905,200403,200403,200820,191120,200218,190417,190417,190703,190703,191028,191028,190927,190305,191216,191218,200504,200504,200226,210909,210909,210605,210503,210503,210315,210115,210218,210108,201021,210105,190424,190104,190524,190618,181211,181108,190722,190722,181206,190531,190927,190807,181212,231107,230320,221029,220920,220408,200506,191122,191205,210708,211020,210909,210605,210510,210523,210315,210126,210108,210218,210218,210107,200819,191105,230320,221216,221029,221029,221020,220519,221008,220404,220929,220926,220924,221017,210112,201012,201006,190609,200605,200403,200819,191216,190905,200703,200228,181206,200113,200218,200804,190104,181112,181108,180821,210713,200602,190424,190513,200106,191026,200320,190530,191119,191205,190617,200414,191014,191015,200507,200824,190808,191107,200310,200605,200605,190910,210115,210115,210218,210108,210107,210105,210112,200210,181206,181108,181121,181112,231116,231107,230707,230707,230612,230612,230504,230413,230328,230328,200414,190531,190627,190424,191026,190531,190104,181112,190812,191014,200506,200824,200605,191015,210503,210315,210312,210126,210306,210306,210126,210219,201218,201021,210107,201017,210112,200505,191216,200218,200210,191024,190910,190910,181108,200820,190618,181211,200324,190424,190423,190417,181108,181211,191211,190618,191014,200703,181206,200403,190927,190807,191211,190722,200318,190423,190807,200113,190926,200504,200605,200226,190401,200905,191211,191211,191014,191014,190104,190423,190424,190513,190617,190424,190311,190722,190808,200310,210915,210916,210722,210605,210523,210413,210410,210306,210114,201218,201012,201012,181108,200402,190609,190424,190513,200425,200210,181121,190905,190531,181206,190927,190807,190627,191119,200506,231230,231005,230504,230318,220909,231014,231010,231005,231005,220515,191120,191120,191120,231230,231116,231116,231228,231204,231114,231107,231028,231028,231028,231014,231010,231010,231005,230707,230707,230621,230620,230612,230612,230504,230412,230413,230516,230412,230328,230320,230320,230320,230320,230318,221207,220516,221122,221018,221027,221018,220519,221020,221020,221017,220929,220519,191216,200106,200106,190401,200905,191026,200320,191122,200218,181212,190104,181203,191014,200106,200425,211009,211019,211009,211023,210916,211010,211113,211103,211023,211020,210803,211020,210916,210916,211010,210909,210722,210708,210605,210605,210603,210523,210524,210503,210410,210316,210311,210219,210126,210115,210219,210219,210129,210106,201010,201008,200804,201017,190311,200403,190618,190627,191014,200505,191105,200318,181206,200703,200402,200228,190927,181121,190417,190703,190722,190722,190401,191218,190807,200819,200804,231230,231204,231114,231107,231114,231107,231013,231005,231005,230620,230620,230620,230516,230612,230607,230607,230515,230413,230413,230328,230320,230412,230320,230318,221221,221207,220516,221029,221122,221029,220516,220516,221027,221020,221018,221028,221008,220928,221018,220928,221122,220926,220922,220920,220914,220926,220914,220909,220412,220412,220428,211111,220411,220122,220409,211117,220327,190424,191216,190104,190326,200218,191119,211009,211019,210803,211010,210910,210909,210713,210523,210503,210410,210308,210306,210126,210107,191211,181211,200819,191211,190905,191205,200318,200820,191218,200113,200804,190424,231005,230607,230516,230328,221018,221008,220928,220504,220926,220428,211009,220924,190311,211019,211009,211019,210708,211112,211113,211110,190618,190910,221222,221222,221215,221020,221018,220516,220924,220504,220504,220928,221215,221122,220404,220924,220504,220515,220412,220428,220428,200602,191026,190925,190530,181112,200507,211019,211009,210916,211112,211103,211103,211020,211020,210915,210712,210712,211019,210713,210722,210605,210603,210510,210524,210311,210129,210219,201006,201008,201008,191105,200605,200605,221221,220516,221029,221029,221027,221018,221018,221017,221017,220929,221028,221122,221008,221122,220924,220923,220922,220514,220514,220327,220428,220412,211009,211209,211209,211020,231005,221207,211111,211209,230620,221020,220928,220926,231114,231010,230504,230328,221215,221216,221018,221017,220928,221122,220920,220914,220428,220411,220411,220408,211117,191216,200320,190424,200320,200605,191105,190808,211019,211103,210915,210911,211010,210722,210604,210503,210410,210311,210306,210115,210218,201218,201012,201008,191024,191105,190618,190401,181121,200504,190926,191120,181206,200605,200824,191026,200605,211023,211020,210909,210605,210524,210524,210306,210219,210108,210107,200320,181121,190609,190703,190910,191120,230707,230621,230621,230504,230328,221221,221029,221020,220928,220926,220519,220929,220504,220909,220515,211111,220122,211209,220920,211010,191216,191028,190423,200414,200824,211019,210708,210712,210603,210603,210315,210312,190910,190513,211209,200506,200218,200507,210708,211103,211103,210915,210910,210803,210909,210910,210707,210707,201021,201006,200324,200113,200804,231230,231230,231204,231107,231107,231013,230707,230707,230612,230504,230412,221215,221216,221122,220516,221027,221018,221008,221122,220922,220515,211104,220428,220428,211114,220411,211209,220409,211117,190423,200804,181211,181203,200506,190424,190104,190925,181126,181112,190424,181203,211103,210803,210916,210604,210316,210218,190530,181212,190311,181211,181121,190417,181206,190401,181108,190609,200228,181112,190531,181203,181203,190305,181023,220504,221122,220412,191218,200310,191014,200218,190808,181108,200820,190609,190423,190531,230707,221008,220327,220122,190910,181203,181203,191218,190627,200804,191014,191014,190618,190627,190627,200905,200506,200425,190627,200106,200402,191014,191026,191211,191122,190104,190925,190925,181112,200804,190812,191026,191107,190104,181212,200824,181203,191205,191028,200218,190424,200824,181126,190808,190808,181203,181203,190618,181203,191011,200824,200824,191211,190609,200425,200310,190808,200226,191216,191122,191205,190910,190423,201008,200605,190808,200226,200218,191024,190910,200310,190618,190722,181212,200320,200310,190703,200425,191205,190910,181108,181108,181112,200403,190530,190530,200310,200505,200425,191227,200310,200210,200403,200228,200820,200318,190530,190513,200505,200228,191024,190417,190417,190417,191211,200504,200703,200703,190722,190417,191015,190531,190609,190609,181206,181206,190905,200318,200505,190513,190609,191120,190401,181108,181121,200228,191011,190417,190417,200402,200504,200504,181112,200228,200819,200318,191120,191216,190703,190703,190812,181206,190807,190401,190807,190807,191218,190326,181108,190807,181023,181023,181112,200425,191026,190326,190401,200402,200106,200106,190104,211110,211114,211019,210915,211009,210713,211103,211103,211103,211020,210915,210915,211020,211020,210911,210712,210712,210712,210916,210707,210803,211010,211010,210909,210722,210713,210713,210722,210708,210708,210708,210605,210605,210605,210604,210604,210523,210523,210523,210523,210510,210524,210524,210524,210524,210503,210503,210413,210412,210410,210410,210315,210313,210311,210311,210308,210308,210306,210306,210306,210126,210126,210115,210115,210114,210114,210114,210108,210106,210106,210106,210107,210107,201017,201017,210112,201012,201010,201006,231204,231204,231230,231230,231230,231116,231116,231228,231107,231114,231114,231014,231014,231014,231010,231010,231005,230707,230707,230620,230620,230607,230612,230515,230412,230413,230412,230412,230412,230412,230328,230328,230328,230320,230320,230318,221216,221215,221215,221207,221029,230320,221027,221027,221027,221027,220519,221018,221020,221020,221020,221017,221017,221017,221017,221028,220929,220929,220928,220928,220504,220928,220504,220924,220924,220924,220923,220427,220427,220926,220923,220922,220920,220514,220514,211114,211104,220914,211104,220914,220914,220914,220909,220327,220427,220327,220327,220515,220515,220404,220404,220404,211104,220420,220420,220504,220428,220504,220428,211009,211211,220411,220411,220411,220411,220411,211209,211209,211209,211111,211111,211211,211211,211211,211114,220322,220408,220408,211114,211114,211117,211117,211114,231010,190627,200506,190627,190627,200824,200226,191216,191216,191014,191026,200824,191026,200320,191211,200602,200218,200506,190627,191026,200618,200506,191107,200414,200218,200824,190812,190808,200210,191205,191205,191211,200507,200425,200605,190808,191122,191011,191107,200310,211110,211118,211019,211019,211009,211009,211009,210916,211103,210708,211113,211112,211110,211103,211103,210708,211023,211023,211023,211023,211020,211020,210915,210915,210915,210915,210915,210713,210915,211020,210911,211009,210911,210911,210712,210712,210312,210909,211019,211019,210916,210916,210707,210803,210523,210803,211010,210916,210916,210910,210910,210909,210722,210722,210713,210713,210722,210722,210707,210707,210707,210708,210708,210605,210605,210605,210605,210605,210605,210605,210604,210604,210604,210604,210603,210603,210523,210523,210523,210523,210523,210523,210604,210510,210524,210510,210524,210524,210524,210524,210503,210503,210503,210503,210503,210413,210603,210413,210410,210412,210412,210412,210412,210412,210410,210311,210410,210410,210319,210312,210319,210316,210316,210315,210315,210315,210313,210313,210312,210312,210311,210311,210308,210308,210308,210308,210308,210308,210306,210306,210306,210306,210306,210126,210129,210115,210115,210114,210114,210114,210114,210114,210114,210219,210218,210218,210108,210108,210106,201006,200605,201008,201021,201021,201021,210107,210107,210107,210105,201017,201017,210112,210112,210112,201006,201006,200824,201006,201006,201008,201008,190808,191119,200320,200403,200226,191216,190910,191216,200310,200506,200602,191211,191205,200505,190618,191105,191119,200819,200819,190910,190910,200210,200310,191211,200403,191024,200228,191015,200618,200703,200703,200106,191119,200210,190910,200403,200403,191120,191011,200820,190905,200318,200228,190807,200504,200113,200210,200504,191028,190927,200318,200504,200310,191218,191119,191028,200113,191218,191218,200324,191026,200113,191015,190926,190807,191216,200226,200320,200506,200106,200226,211117,211117,211117,211117,211117,231116,231028,231028,231013,230320,230320,221221,221216,211118,211009,210916,211112,210708,211023,211023,211112,211019,211103,210803,211020,211009,210911,210916,210803,210916,210803,211010,211010,211010,210910,210910,210909,210713,210713,210713,210707,210605,210605,210605,210605,210604,210523,210523,210523,210311,210523,210315,210316,210603,210503,210510,210524,210510,210524,210312,210413,210413,210412,210315,210313,231230,231228,231028,231028,231013,231010,231005,231005,230707,230707,230620,230612,230621,230412,230412,230412,230328,230328,230320,230320,221215,221222,221216,221207,221018,221028,221028,221028,221018,221017,221020,220519,221008,221017,221028,221122,220926,221027,220924,220923,220505,220920,220514,220909,220505,220515,220515,220420,220420,220412,220428,211114,220409,211117,211117,210916,200824,200506,190627,181203,191014,200824,200905,181112,191216,190424,191211,200618,190812,190905,200210,200106,191026,190530,191122,200320,200310,200403,200414,200506,200804,200320,181112,181112,190617,190104,190104,190424,190311,190925,181203,200824,191205,191205,191014,191211,200414,200507,200414,200507,200210,190423,190808,200310,211114,211019,211009,211009,210916,211112,211113,211103,211103,211103,211023,211103,211103,211023,211023,211020,210915,210915,211020,210911,210712,211019,210312,211019,210707,210803,210803,210803,210803,210910,210910,210910,210722,210722,210713,210722,210722,210722,210707,210306,210707,210708,210605,210315,210604,210604,210604,210603,210603,210523,210523,210510,210510,210524,210503,210524,210503,210412,210413,210413,210413,210412,210410,210410,210319,210316,210313,210311,210308,210126,210306,210129,201021,210108,210126,210115,210115,210114,210219,210218,210129,210108,200605,210106,201012,201021,201021,210107,210108,210105,210105,201017,210112,210112,210112,201012,201012,201010,201006,201008,201008,190609,191119,191024,190910,200425,191211,190417,190311,191024,181108,181108,181211,200403,181211,200228,190618,191011,191011,200819,200505,200506,190417,200113,191211,190808,190722,181206,200318,191205,200618,200507,200703,181108,190513,200507,191105,190401,190905,200402,191120,200819,190513,190524,181121,181121,200226,190926,200820,190417,190531,200504,190703,191028,190927,191218,181206,190807,200113,181023,200402,190305,190531,200226,200804,191122,190326,231010,231010,231013,231005,231005,231005,230621,230620,230612,230612,230515,230504,230412,230413,230412,230328,230328 ,230320,230320,230328,230328,230320,230318,221222,221221,221221,221221,221215,221222,230318,221029,221029,221029,221029,221027,221027,221027,221018,221020,221018,221017,221017,221008,220929,220928,220928,221122,220926,220504,220404,220924,220923,220427,220427,220923,220922,220920,220514,220514,220914,220914,220505,220505,220909,220327,220427,220515,220515,220428,220426,220404,220420,220504,211208,211111,211208,211208,211211,211104,211208,211104,211208,211019,210909,210803,210909,210524,210603,210603,210412,210412,231230,231107,231010,231005,230607,230504,230328,221207,230320,221017,221122,220920,220412,211111,220504,220428,220122,220411,220408,220322,220322,211114,231230,221122,220514,220420,231204,231010,230320,230320,221018,191014,191026,191122,200218,181203,210605,210524,210308,210219,190311,181211,190910,200504,200703,200318,190513,190530,190703,190424,181112,191024,181126,181108,181023,200113,200425,200819,190618,190531,181203,190925,210129,201012,200505,200218,190423,190722,200504,191105,181023,200402,200804,191120,191216,200506,190424,200824,200320,191026,190925,200605,191119,191205,200824,200507,200210,190808,191107,200605,191015,211114,210916,211110,211110,211020,210910,210909,210722,210605,210523,210503,210412,210412,210316,210311,210311,210126,210219,201021,210107,210112,210112,201006,190417,181023,190618,191107,190311,200403,181211,200324,200819,191105,191119,200505,200318,190905,200618,200703,191120,191205,200402,200820,200228,190531,190401,190703,181206,190926,190807,191218,200113,200113,200402,190531,200226,200804,190627,221017,221017,220909,220922,220427,220412,211209,220122,220322,231228,230620,230515,230412,221221,220504,221020,221020,220914,220428,210523,210126,231013,221207,221207,221017,221017,220914,211114,211103,211112,211113,211114,211114,210803,211113,211023,211023,211113,210915,210911,210911,211023,211009,210911,210911,211010,211020,210910,210910,210707,210604,210604,210603,210605,210523,210510,210510,210503,210503,210503,210524,210319,210413,210312,210319,210319,210413,210311,210315,210129,210129,210114,210106,210106,210105,210105,201012,201010,201010,201008,200905,200905,200425,200506,200618,181108,190513,200602,180620,200703,200106,200106,200320,190524,190524,190531,190617,200703,200618,181121,181121,190104,181206,181206,200324,200226,190927,190927,190926,190926,190401,190401,190326,190326,190401,190401,190401,190401,190531,191120,191120,191218,191105,191107,191105,191026,191028,181126,181126,200618,190703,190703,190812,190307,190307,190307,190401,190926,191015,200819,200310,200324,200324,200324,200402,200820,200804,200819,190807,190905,190905,191218,191227,191211,200113,231116,231114,231116,231028,231028,231005,231013,231013,231013,230612,230602,230504,230412,230413,230504,230621,230328,230318,230318,221216,221216,221215,221216,230320,221216,220519,221020,221027,221020,221018,221008,220929,220922,220923,220516,220926,220926,220519,220923,220923,220920,211222,220327,220515,220327,220426,220420,220426,220412,220412,211222,211211,211222,211111,220322,220322,220322,211117,211117,190722,181203,200506,200605,190925,190424,181112,190627,190812,190104,190424,191120,191107,190925,190925,190812,200403,181203,181212,191211,190423,211103,210712,210803,210707,210713,210605,210503,210413,210308,210114,210108,201017,201006,190531,200605,190808,190812,200226,190910,181108,190618,191119,191122,181211,190910,190609,191011,191011,200402,191120,200318,191120,200504,200504,190531,190722,181112,200320,200228,190531,190703,190927,190807,190423,191028,190401,190326,190524,181112,200226,190305,190627,200226,190305,221029,221017,220519,220920,220505,211104,220122,211209,220909,220427,220412,220412,211110,211209,211209,211009,181126,190910,181108,191227,231230,231028,231005,221017,220914,190424,191105,191105,200703,191120,191211,200218,190424,191107,211114,210712,210712,210712,211010,210916,210603,210524,210412,210319,210316,210129,210126,190531,200425,200425,200602,181211,200318,200403,200703,191015,190305,181121,190905,191105,181206,191028,200506,231013,220428,220428,220428,220411,220122,220122,211118,211114,220327,211118,181203,190326,180821,190617,190104,191205,190311,181203,190424,190311,190618,190618,210708,181212,190609,181108,181212,181212,181211,190307,181212,190311,190307,181206,181108,181112,190305,190423,190417,190513,190531,181023,181023,190305,190305,190305,181023,220428,181108,181108,181108,231013,231230,230707,230328,221222,221027,220929,210915,200602,190807,190807,181015,180911,181211,181203,191107,191014,200824,190808,210503,210412,210319,210410,210312,210308,210126,210306,210126,210114,210114,210218,210107,210105,201017,201012,201006,190423,181108,200403,181211,200505,190910,200228,200618,191120,200210,181206,200318,190531,190417,191028,190926,191015,190812,191227,191227,190305,200804,190326,231028,230412,221008,220909,220428,220122,220409,220420,211117,181212,200414,181203,200605,190808,191122,210915,211110,211023,210712,211010,210722,210708,210603,210510,210410,210316,210308,210129,210114,201008,210105,201012,181126,181211,200820,190401,200210,200318,191227,190703,191028,211111,210916,211112,211023,210911,210707,210604,210604,210604,210412,210312,210315,210306,231204,231107,231116,231028,231005,230504,230413,230412,230413,230413,230320,221221,221216,221018,220928,221122,220928,220923,220922,220920,220426,211209,211118,211111,211117,220122,200320,210114,181211,211019,211023,211020,210910,210605,210524,220922,220515,220122,200402,190531,190609,190618,201012,200320,200210,200226,200506,200505,190722,190722,181121,190417,191026,200228,200228,190927,200226,191218,190926,190807,231230,231204,231116,231116,231114,231114,231107,231028,231014,231010,231010,230707,230620,230607,230515,230607,230328,230328,230607,230318,221207,221029,221029,221029,221027,221020,220920,220924,220923,211104,211104,220914,220909,220909,220505,220515,190926,200905,200506,190627,190722,190807,190617,191026,190722,211112,210915,211020,210722,210316,210413,210316,210312,210308,210126,210114,210218,210105,201012,200318,191011,181121,191028,190417,190417,190417,200618,200106,200318,190513,200402,200402,190104,231114,231228,231228,231107,231028,231014,231010,230707,230621,230607,230621,230413,230328,220519,220909,220924,220920,220909,220515,211104,220420,220408,211222,211209,211209,211117,190627,181112,200506,200506,190311,190311,200310,190530,190925,191211,190617,190812,200824,211009,211103,210915,210712,210707,210707,210524,210115,200605,210108,210107,201008,200605,190531,200210,191024,200403,190609,200228,190926,200113,181212,181203,181023,190305,191216,191122,231230,231230,231107,231228,231107,231028,231028,231013,230707,230707,230620,230620,230612,230515,230413,230412,230328,230320,221215,221221,221221,221207,221122,221222,221028,221027,221020,221017,221221,221008,221017,220519,220928,220404,221122,220924,220514,220920,220427,220514,220909,220928,220427,220411,211211,211117,200602,190530,200804,181112,190326,200506,181212,200507,190104,211113,210911,210916,210713,210713,210909,210312,210114,201010,201006,200505,181121,191119,190423,191122,200228,191011,200504,200703,181206,190513,200819,190812,191218,230621,230412,230320,221216,221122,221029,221028,221018,220928,220504,220404,220420,220409,211209,220322,210708,190618,181112,231013,231013,230612,230504,230328,230328,230516,221215,221221,221216,221216,221215,221008,220928,221122,220926,220924,220924,220923,220920,220514,220327,220428,220411,211209,220409,211019,210915,210911,210712,210712,210707,211010,210316,210603,210503,210503,210312,210311,210308,210114,210126,210115,210114,210114,210114,190925,200905,200602,191014,200507,190925,200403,191011,200505,181203,201010,201010,201006,201006,201008,190812,181212,191024,191015,190910,181108,200210,190311,190417,200703,200113,190905,181108,190513,200819,190926,190326,190627,191218,200113,200804,200804,200804,181112,200318,200218,200905,190627,200106,190104,181203,191014,190722,191011,200310,210107,201017,210112,201010,201008,200505,190722,200602,191119,191120,190531,191211,191211,200210,200210,200506,181203,181121,181206,190905,190424,200703,191119,190307,191105,200318,200402,190513,190927,181108,190311,190305,200226,200226,200504,190812,231010,231010,200804,190326,190627,190627,211009,211010,210909,210722,210410,210311,210306,220922,230516,221028,221017,221017,220920,221008,220926,221122,220924,220923,220926,220922,220514,211111,211111,220411,220411,211208,221215,231014,181121,220923,221216,230504,221029,221222,221207,220922,230707,211009,211113,211113,210911,210712,210722,210523,210412,210412,210316,201006,231107,231028,231028,230328,230328,230328,230320,230607,230320,221222,221215,221215,221222,221221,221221,230318,221221,221221,221216,221216,221215,221221,221215,221215,221216,221207,221207,220516,220516,221207,221122,221029,221029,221018,221018,221018,221029,221027,230320,230320,221027,221027,221027,220519,221020,221020,221020,221020,221020,221018,221018,221017,221017,221017,220926,220926,221017,221017,221008,221008,221008,221018,221008,221008,221008,220929,221008,220928,220928,220928,220928,221122,221122,220926,220926,220404,221122,220404,220924,220924,220926,220929,220924,220924,220924,220924,220427,220427,220923,220922,220922,220920,220920,220514,220514,220514,211104,220914,220914,220909,220909,220909,220427,220427,220515,220515,220428,211104,220426,220420,220420,220428,220420,220412,220412,220504,220428,220428,220428,220428,211111,211111,211111,220411,211209,220122,211209,220409,220408,220408,211117,211117,211117,231228,231014,231005,231005,231005,231005,231005,210605,211112,211112,190104,190401,190513,220920,191122,190627,191014,200602,191014,200905,200506,200106,200218,191026,200506,200310,191211,200403,200507,191011,200226,190910,211114,211009,210916,210708,211112,211113,210708,211023,210915,210915,211009,210712,211019,210722,210722,210707,210708,210605,210604,210523,210523,210523,210315,210312,210319,210315,210313,210308,210308,210308,210126,210115,210115,210219,210219,210129,210106,210106,210108,201021,210107,201021,210107,210105,201017,201010,190910,200602,190910,200210,200403,200505,191011,191024,191024,200820,200820,191216,200228,200507,200618,191105,200318,191205,200403,200228,200820,200318,191105,191120,191216,200504,191105,200324,191218,200804,181108,231204,231204,231114,231116,231116,231114,231028,231028,231013,231005,230707,230707,230612,230612,230412,230413,230328,230328,230328,230515,230318,221221,221216,221207,220516,221029,220519,221018,221018,221018,221017,221017,220929,221008,221122,220920,220920,220914,220914,220505,220427,220515,211104,220504,211209,220504,220428,220122,211209,211114,211117,211117,211110,190627,181211,200226,181206,190609,190925,190925,190311,190424,190617,181112,181112,190808,190311,210605,210510,210524,210129,190609,190424,190808,190311,190530,190618,190618,181121,181121,190417,181206,190307,190531,190513,190722,190812,190305,190401,190531,190423,190513,190807,181108,190807,181112,181108,190305,231228,230620,230612,221221,221215,221215,220519,220928,220924,220409,211110,210708,210712,210707,210713,210722,210604,210510,210523,210916,210313,190627,190104,200320,191014,190808,200106,211118,211010,210708,210523,210523,210510,210524,210524,210503,210413,210313,210413,210412,210412,210410,210316,210313,210308,210129,210114,210105,210112,201006,190531,190423,191218,181206,191107,200402,190910,231010,231005,231005,230707,230707,230515,230412,230412,220516,221027,221018,221020,221017,221008,221122,220926,220514,220914,211118,211117,220322,220408,220408,220408,200318,181112,181203,200824,200106,200210,210803,210604,210510,210413,210306,210114,210112,190927,191011,191107,191028,190807,190305,190812,191120,200228,211118,220322,220322,220322,220322,200804,190627,181126,211009,210524,210503,210410,210114,201008,191227,191227,230612,230515,220920,220920,220909,220122,211117,190311,190531,190423,211019,210722,210106,200605,210112,210112,190812,190910,190812,191028,221008,220920,220504,231107,220914,220928,220505,220515,220412,211114,211110,211111,220411,220411,220122,220122,220122,211118,211110,211110,211019,211009,211009,211112,211114,211113,211103,211103,211023,211020,210915,210915,210712,211019,210916,210916,210909,210713,210605,210605,210604,210605,210316,210523,210523,210510,210316,210603,210503,210503,210413,210412,210410,210410,210316,210315,210313,210312,210311,210308,210306,210306,210306,210126,210126,210115,210114,210114,210219,210218,210218,210106,201006,210108,201021,210107,210105,201017,200905,200506,200618,181203,190424,190812,191026,190925,190530,200218,190311,181112,200414,200824,200618,190618,200507,190423,201017,210112,190808,191024,200310,190530,190609,190910,190910,190307,181108,200403,181121,200402,200820,191015,190417,190423,200505,200505,190722,190513,190513,190905,190513,181108,190305,190530,181112,200226,190926,190311,190326,200504,190927,190326,190627,181203,191218,191015,190807,191028,200618,200804,181023,181112,200804,200320,190617,190531,200218,191216,221017,220909,220420,220504,220408,211114,191122,191122,191119,231116,231114,230707,230621,230612,230612,230504,230413,230607,230412,220404,190627,191014,190910,211114,211103,210523,210603,210308,201017,201006,200425,181212,181211,181108,190423,190417,200324,231230,231116,231107,231114,231028,231013,231005,230707,230612,230328,230328,221222,221221,221207,221029,221029,221027,221018,220926,220920,220514,220420,190531,191014,191211,191014,200226,191014,191205,200218,180821,200506,200414,181203,191205,191105,181203,210708,210708,211103,210915,210911,210712,210712,210722,211010,210909,210713,210713,210707,210707,210707,210315,210603,210603,210523,210503,210412,210308,210308,210115,210114,210218,201021,210105,201017,200505,190808,200403,200226,190910,190609,200318,200210,191211,191024,200618,200703,181206,190905,191120,190401,191015,190905,191120,191015,181121,191011,191120,190530,200504,191028,181206,200226,190531,190326,191211,200218,190807,190627,181023,181023,190524,190524,200605,191028,181206,191218,191120,200402,200106,231230,231230,230412,230413,230621,221029,221207,220514,220514,220420,211104,220322,211104,231230,231116,231230,231116,231107,231114,231028,231010,231005,231005,230620,230620,230621,230621,230607,230602,221222,230318,221216,221216,221216,221207,221207,221029,221028,221029,221028,220504,221017,220928,221122,220926,220924,220519,220924,220909,220327,220427,220427,220420,220504,220412,220428,220409,211118,211111,220322,211117,211019,211009,211009,211113,211113,210708,211023,211020,210915,210915,211020,210712,210712,210708,210412,210412,210316,210524,210413,210412,210412,210410,210313,210311,210306,210126,210115,210219,210106,210108,201218,201021,190627,190326,200905,181112,181203,200507,191107,200804,191122,190104,200218,181112,190617,200824,181203,200505,201017,201012,201006,201008,181212,200425,200210,190618,200820,200505,191211,200425,181108,200618,200113,200226,200819,200819,181112,191216,200504,190703,191211,181023,181023,200218,200804,190423,190524,200226,200824,190812,190424,191205,200507,210911,210916,210604,210524,210503,210524,210316,210316,210306,210126,210115,210114,210219,201021,210105,200425,190609,200318,191205,191205,190910,191120,191120,190401,191105,181206,200425,191026,200226,190617,190722,210707,201006,200403,181108,181121,221008,191216,191014,200602,190424,190104,200320,190618,200218,190424,181112,190617,200605,191119,200414,181203,191205,191014,200507,200507,200824,190423,200210,200106,190808,181212,191107,211114,211103,211110,211113,211023,210911,210803,210803,210708,210605,210510,210312,210410,210313,210308,210126,210219,210108,210105,200605,200819,181108,191205,190722,200318,200310,200618,191105,190905,200618,191120,200210,200210,200402,190609,200820,200228,200504,181121,190417,181206,191015,191218,190926,190807,200402,190812,191120,200804,190326,190326,190326,231204,231114,231116,231228,231114,231028,231028,231013,230707,230707,230707,230612,230621,230515,230607,230413,230328,230318,221221,221216,221207,220519,221018,220928,220926,220924,220926,220924,220514,220914,220505,220428,220428,211111,211209,211211,190925,190104,190609,190618,191107,191028,190925,190910,190423,190417,181121,190927,190905,200318,190513,181108,190703,181108,190326,191105,191028,191120,231230,231228,231114,231010,231005,230504,230413,230412,230320,230320,221029,221027,220519,221008,220928,221122,220920,220514,220504,220428,220322,220122,200804,181121,191205,211112,211019,211112,211113,211020,210915,210803,210722,210722,210315,210523,210410,210115,210114,201006,201006,201006,200824,201008,200210,190417,181206,200402,190722,191120,190927,190927,191227,190807,200402,231204,231230,231230,231116,231114,231028,231014,231014,231010,231010,231005,231005,231005,230621,230612,230612,230612,230612,230515,230413,230412,230607,230412,230412,230412,230328,230328,230328,230320,230320,221221,221216,220516,221027,221018,221018,221027,220516,221020,221017,221008,220519,221122,220926,220404,220404,220924,220923,220514,220505,220515,220515,211118,220408,200425,200403,191211,181203,191026,190424,190423,200507,191216,191122,200228,190424,200506,190513,190617,190808,190617,200804,190307,190925,191122,191205,200824,190812,200403,181212,200226,191205,191011,190423,211103,211009,211019,211103,211020,210915,211020,210911,210712,210312,211019,211009,210916,210916,210713,210707,210604,210603,210523,210524,210524,210510,210413,210412,210312,210313,210308,210129,210129,210126,210126,210219,190807,201021,201021,210107,210107,210107,210105,210112,201006,200605,200605,200403,190722,200226,191024,190311,181108,191105,200210,190910,181121,200402,190609,191011,190905,200318,190531,190531,191015,191015,200318,200228,200504,181112,200113,200819,190423,180627,190926,181206,191026,190807,200425,190305,200804,181023,200602,190812,191227,190311,190311,210713,210306,210306,210114,201010,190722,191211,200324,190722,191120,190531,200504,181212,221215,221029,220926,220923,211209,231028,231028,211209,210311,200318,200318,191216,200804,210312,210218,210108,201021,230328,220428,181112,190627,191216,190424,200106,190104,191026,190531,190609,191122,191205,191119,191119,200414,190104,190311,191014,200507,181203,190423,190808,181212,200824,200310,191216,200425,211114,211118,210916,211112,211112,211103,210915,211023,211023,210915,210915,210803,210911,210911,210712,210803,210916,211010,210909,210713,210605,210605,210603,210603,210523,210604,210412,210410,210316,210315,210311,210308,210306,210126,210115,210219,210219,210218,210129,210108,210108,201021,210107,210105,210112,210112,201006,200210,181108,190618,191024,200820,200318,200618,191105,190905,200703,200402,190609,181212,200228,200504,190417,190531,190703,181112,181206,191015,191218,190807,200402,200804,200506,231230,231116,231114,231028,231107,231005,230707,230607,230515,230413,230328,230328,221215,221222,221207,221122,221029,221020,220929,220504,220929,220504,220514,220909,220427,220504,220515,220428,211104,220420,220428,220428,220122,220408,220408,211118,211117,190424,190531,181212,181126,181212,190307,190311,190530,190618,190417,181112,181203,201008,191211,200507,190808,191011,181121,191015,191024,200403,200228,190530,190305,211113,210708,210911,210707,210713,210713,210524,210503,210313,210106,210108,231204,231116,231005,230320,230318,230318,221020,221008,221008,220514,220327,220322,200605,190424,190627,191211,181203,191026,200602,191211,190531,191216,190925,200320,190423,190617,190104,190617,200218,200310,190513,190513,181212,200824,200506,191216,191205,200403,181203,181212,191011,181211,201006,201006,201006,201006,201006,200905,190722,191216,191024,181126,200403,200602,200602,200602,181212,181212,191011,191211,181108,200506,181211,191105,191105,191205,200403,200505,200210,200402,191011,181108,200403,190722,191227,181121,191211,190618,200505,190722,190722,200228,191024,190417,200504,200618,200618,200703,200703,200507,200703,190812,190910,190905,190513,190812,191105,200324,191120,191011,200318,200318,181108,191105,190722,200113,190307,190531,190926,200318,190926,190926,191026,200113,190326,200402,190703,191026,200113,211019,211009,211009,211009,211009,211009,211222,210708,211023,211023,211023,211023,211023,211023,211020,211020,210915,210712,210712,211019,210803,211010,210722,210604,210713,210707,210707,210707,210708,210708,210605,210605,210604,210604,210603,210412,210523,210510,210510,210524,210524,210523,210413,210412,210312,210412,210412,210410,210410,210312,210319,210316,210316,210315,210312,210312,210311,210311,210311,210306,210306,210306,210306,210126,210115,210115,210114,210114,210219,210218,201017,210112,210112,201012,201010,201010,201010,201006,201006,231204,231116,231116,231228,231228,231114,231114,231114,231114,231114,231107,231028,231028,231014,231013,231010,231005,230620,230320,230612,230612,230612,230504,230504,230413,230412,230328,230328,230318,221215,221215,221222,221222,230318,221221,221215,221216,221216,221207,221221,221029,221029,221018,221020,221017,221017,221028,221008,221008,220929,221122,220926,220926,220923,220922,220923,220514,220514,220514,220514,220505,220505,220427,220504,220428,220428,211211,211208,220411,220411,220411,220122,220409,211209,211209,220409,220409,211209,211009,220322,220322,210722,231107,231013,230607,221122,200618,191107,211110,211019,211009,211009,211009,211020,211110,211103,210708,211023,211023,210915,210915,210911,210911,210712,211019,210707,210722,210722,210713,210722,210707,210708,210708,210605,210315,210604,210603,210523,210523,210413,210410,210311,201006,201017,200605,180709,190905,200403,190326,220412,220504,220428,211209,220322,220409,211208,211114,211117,211117,211117,231204,231116,231116,231114,231114,231028,231013,231010,230612,230607,230515,230413,230320,230318,230328,221028,221018,220928,220504,220926,220514,220914,220909,220515,220412,220322,211118,211208,211117,211110,211103,211009,211113,211110,211020,210712,210916,210523,210604,210315,210413,210413,210313,210412,210319,210114,210219,201021,190531,201017,210112,201012,191216,191011,200324,200228,190722,200228,181112,200228,191028,200804,181203,210911,210722,210722,211010,210605,210604,210523,210523,210129,210218,210108,210107,201008,190513,191024,190722,181121,191011,200318,200504,190305,190305,200820,181108,200228,200106,190305,231204,231014,230612,221215,221018,220909,220909,220923,211104,211117,221221,181203,220926,220922,220920,220920,211104,211104,191216,191122,200605,191205,200618,211023,211009,210722,210510,210315,210308,201017,210112,201008,200605,191028,200804,231230,231228,230707,230707,230413,230412,221215,221029,221020,220929,220924,220412,211209,211209,211114,220920,190104,181212,191014,210712,210115,220428,190326,181112,200218,200507,191211,190925,190104,200310,200310,191205,181203,181206,200403,200824,200505,191026,191107,211110,211114,211009,211009,210708,211112,211103,201010,201006,201008,190722,190807,190531,181212,200210,181108,190307,200228,191011,190417,200703,190513,190722,190905,191105,190524,190524,190530,190305,200819,200504,200504,190627,181203,191015,191015,191218,200106,200113,181023,200402,221020,221008,221018,221008,220924,220922,220922,220514,220909,220427,220426,211104,220420,220412,220504,220428,220428,211211,220411,211209,220322,220408,211114,231107,231228,231228,231114,231028,231028,230328,230328,230320,230320,221215,221222,230318,221221,221215,221216,221207,220516,221207,221221,221122,221029,221207,221029,221029,221027,221027,221027,221018,221020,221020,221020,221018,220926,221008,220519,221008,221008,221008,220929,220928,220928,220519,221122,220926,220404,220929,220924,220427,220922,220922,220920,220514,220514,220514,211104,220914,220505,220327,211209,220327,220515,211104,220420,220412,220428,211114,211111,211111,220411,211209,220409,220409,211209,220322,220408,220408,220408,211117,190627,191122,190609,200804,191014,200905,190925,191014,200320,190618,181112,190617,190627,200320,200425,181203,190104,191216,200507,190424,200824,190925,190424,181126,200403,181203,190808,190808,211110,211110,211019,210916,210916,211113,211103,211023,211023,211020,210915,211020,210909,210909,210712,210916,210916,210722,210605,210605,210605,210603,210523,210523,210510,210524,210413,210412,210412,210319,210319,210316,210312,210315,210308,210308,210308,210306,210115,210114,210219,210218,210218,201008,201021,201017,201012,201010,201006,201008,200605,200403,200425,191227,191119,181212,191024,190910,181211,200318,191205,191011,200819,200403,200507,190618,200210,181211,200425,190722,191227,190530,190531,190910,190417,200318,200403,191105,181108,191120,190417,181112,190305,190513,181206,181206,200226,191107,200819,200106,200403,200113,200402,200504,200113,191227,200318,190104,190401,230328,220411,210713,210713,210708,210707,210605,210605,210605,210603,210603,210523,210523,210523,210510,210524,210524,210503,210410,210319,210315,210313,210312,210311,210308,210306,210306,210306,210114,210219,210218,210129,210218,210106,210108,210108,201021,231230,231204,231230,231116,231107,231114,231107,231028,231028,230621,230612,230621,230607,230318,221221,220516,220428,220122,220122,220409,220409,200905,181112,200506,190627,191216,191216,190618,200602,200618,190812,190104,190104,191026,190925,190617,200605,200310,200218,181112,200414,200414,190311,200824,190618,191205,200507,200507,190423,201012,201012,201010,201010,201010,201006,201008,201008,201008,190808,200310,181108,191119,191119,181212,190618,200425,200602,190910,200403,200210,200210,190311,190311,190618,181211,200228,200820,200505,191024,191024,200402,200210,191026,200403,190417,190417,200425,190722,181206,200318,181108,191105,190905,200618,200113,191119,191122,190305,200113,200819,190531,190530,190627,181121,200226,191120,200218,190326,200504,190703,191227,191227,181206,181206,191105,191218,191026,191028,190812,181023,200324,200425,190305,190925,190617,200218,200804,200824,200402,200228,190531,200804,200506,200905,200602,200605,200824,210909,201008,191011,190627,200804,200210,200320,181203,190424,190531,190530,190104,190925,181112,181121,190609,181203,190423,190531,190808,181212,190609,190627,190925,200618,190618,191122,190618,211019,211112,210708,210712,210510,210510,210524,200605,200402,190305,190326,190326,190423,231228,230707,230504,230504,230328,221207,221029,221027,221020,220412,200602,190424,201012,191014,200824,181126,190812,181203,211113,210713,210316,210311,210114,201006,201017,201012,201010,201008,200505,191107,200228,200703,190812,191015,191015,190807,190807,200819,191218,200504,190305,190812,230412,221008,221008,220926,211104,211117,210523,190104,220909,190925,190910,181108,190311,190618,181211,190531,190417,181108,181206,190722,190618,190905,190401,190609,181121,181023,181203,190513,190531,190703,190808,181023,181206,181206,181206,190807,181211,181023,190609,181112,190531,190531,190618,190812,190424,190530,190531,200506,191216,200602,191216,200106,200310,200403,211113,210915,211020,210712,210916,210909,210605,210523,210510,210524,210312,210410,210308,210126,210115,210218,201021,210107,210112,201006,201006,200605,200403,200703,200403,200504,191227,200804,231005,230620,230620,230504,230328,220516,221018,220404,220924,220909,220412,211222,220322,220408,220408,211222,211114,190627,181112,190104,181203,181023,181121,181023,191026,190417,181211,231014,230320,221008,220922,220327,211110,191205,191107,181212,191205,211103,210315,210523,210312,210316,210114,210107,200820,181206,190808,190513,181112,191120,200425,191227,231013,231013,220914,190424,190617,211103,210722,210603,210524,210315,210115,210107,181211,181203,181108,190618,181206,190905,231107,230320,220929,220914,211111,191120,191026,200320,200824,200507,191011,191011,200228,190722,210112,190531,200602,190925,200605,200824,211009,211103,211112,210605,210523,210412,210306,210115,210107,190326,190703,191120,190703,181206,200506,190424,231114,230621,230620,221008,221122,220928,221122,220909,220515,211110,211118,200113,200113,210108,201017,231028,231005,230504,230328,221028,220928,220519,220926,220923,220322,190812,190104,200320,190617,200804,191211,191011,211009,211113,211113,210708,210915,211019,210916,210722,210605,210524,210413,210316,210311,210126,210219,200605,210107,210112,201012,200905,200602,200403,191024,191105,190417,200310,190926,200504,200226,200403,191218,200605,190627,231116,231228,231028,231014,231005,230620,230516,230413,230320,221216,221207,220519,221122,220920,220515,220515,220420,220322,220408,231116,221216,210524,210315,210107,200218,200602,210112,190423,191216,200320,200414,200507,211113,210708,210915,210712,211010,210722,210713,210605,210523,210315,210412,210315,210306,210126,210218,201021,210105,200602,181108,181211,190417,190530,190722,190722,190524,190305,191028,181206,190927,190807,200113,190925,231116,231028,230612,230412,221222,221222,221018,221008,221122,220514,220327,220420,220428,220411,211111,220408,220322,200824,191227,191205
Distributed nationwide and internationally to Canada, Mexico, New Zealand, Belize, and Hong Kong.
30,435 cases (1,521,750 units)