FDA Enforcement Class II Ongoing

Bag Decanter II, REF 2002S. Packaged on a form, fill and seal machine with a poly/Tyvek material then placed 50 products to a case (cs).

Recall: Z-2640-2024 · Reported August 28, 2024

Enforcement

Recall Number
Z-2640-2024
Event ID
94780
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Microtek Medical Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
August 28, 2024
Initiation Date
June 3, 2024
Classification Date
August 16, 2024
Address
1 Ecolab Pl, Saint Paul, MN, 55102-2739, United States

Description

Bag Decanter II, REF 2002S. Packaged on a form, fill and seal machine with a poly/Tyvek material then placed 50 products to a case (cs).

Reason

During sterile barrier testing performed on the decanter product line, the samples were identified to have pin holes and tears in the packaging film.

Code Info

SKU/UDI-Device Identifier: 2002S/00748426002399; Lot Numbers: 200403,211013,210626,200409,190913,200420,200422,200422,190510,190319,200108,190415,190702,190109,190827,191022,200313,200306,200312,191212,190226,190802,181113,181221,191213,191019,200617,181130,190114,200313,190620,191228,211109,211109,211107,211109,211021,211022,211012,211012,211026,211018,211105,211105,211105,211012,211007,210913,210913,210908,210802,210806,210810,210903,210902,210902,210812,210809,210718,210709,210729,210810,210328,210823,210729,210802,210802,210719,210728,210703,210625,210622,210704,210618,210618,210710,210610,210520,210519,210408,210518,210518,210518,210517,210512,210520,210318,210502,210502,210318,210317,210422,210409,210405,210403,210404,210329,210422,210327,210404,210325,210323,210402,210304,210220,210127,210121,210117,201208,201204,201204,201013,201127,201126,201102,201028,201022,201026,200906,200912,200904,200616,201007,190218,190729,190916,200513,200329,190426,181218,181221,200417,200514,200513,190823,190823,191021,190306,181124,191112,191213,191112,190108,190327,191127,190121,191004,191004,190725,200512,190419,200615,200611,190306,190506,191018,190227,190528,181107,190102,200102,181129,190716,190716,190116,181220,181225,181225,181214,190725,210618,181107,231129,200107,210618,181011,181214,181225,200421,190724,191212,200728,190802,190805,200422,191218,190912,191017,191112,200422,200403,190624,200221,190109,191019,190307,190802,181113,190409,190402,200227,190912,190402,181221,191212,190218,191111,190321,200811,190306,190114,191212,200103,200403,200511,190628,191228,200320,191212,211204,211021,211105,211011,211012,211108,211026,211026,211017,211017,211014,211013,211007,210913,210926,211007,210907,210902,210812,210826,210718,210806,210709,210726,210823,210729,210811,210802,210721,210728,210721,210703,210802,210704,210710,210704,210619,210618,210618,210610,210613,210408,210408,210613,210518,210517,210507,210509,210509,210317,210317,210409,210327,210326,210323,210323,210303,200421,210220,210121,210111,201126,201203,201204,201127,201126,201113,201113,201027,201028,201014,201014,201026,200929,200910,200928,200928,200928,200907,190322,191114,191112,191018,191018,190415,191111,191012,200728,200305,181128,181217,181213,200326,200417,190425,200513,191127,191111,190823,190215,190108,200912,190619,200417,191112,190725,181124,181109,191023,190916,190723,200320,181219,190318,190422,190422,191125,200302,190118,190813,190225,190320,200421,190520,181129,190915,190322,190704,200415,181106,190704,190416,190415,200108,200304,200320,200518,190226,191226,230711,230510,230524,230422,230419,230418,230323,230308,230303,230216,221016,221206,220510,220524,221202,221013,221013,221022,221010,221024,221008,221202,221026,221007,221205,221114,220720,221128,221127,221127,221009,221009,221127,220904,220904,220825,220825,220507,220912,0002626702,220508,220522,220522,220508,220517,220808,220517,211218,211218,220602,220305,220303,220303,220303,211220,220104,220325,211210,220304,220105,220223,220214,211203,211214,211214,211214,231227,231220,231121,231125,231106,231024,230915,230914,230906,230814,230619,230809,230721,230518,230422,230422,230411,230323,230307,220510,230224,221014,220502,220421,220325,190507,200108,220423,181225,190510,200227,190826,191021,190826,211206,210913,210826,210704,210704,210620,210618,210519,210502,210406,210304,201205,201205,201022,200928,200519,200329,190426,190723,190107,200108,200108,200108,191202,230304,221025,220902,221022,221114,221114,220502,220404,220315,211214,210402,210303,210303,210103,210704,201007,200907,200422,200728,191212,190802,200313,191022,210903,210325,210117,201022,191220,190725,200312,191223,200518,201029,201029,200422,200422,190415,200511,191223,190219,200704,201007,181113,181228,200305,190507,190109,200413,191213,200108,191111,200615,200107,190422,190422,200331,190623,200617,200615,200302,200416,190325,200519,191228,190325,200317,190218,200603,190716,181218,200617,200413,190219,200103,200320,190306,211206,211011,211109,211012,211108,211122,211108,211017,211026,211021,211017,211008,211007,211115,211013,211013,210914,210718,210908,210903,210826,210820,210809,210812,210718,210718,210809,210810,210809,210719,210823,210802,210726,210720,210723,210704,210626,210621,210625,210619,210718,210720,210620,210710,210612,210420,210611,210519,210517,210517,210511,210512,210515,210509,210502,210502,210317,210317,210407,210407,210409,210405,210408,210329,210406,210408,210403,210327,210327,210325,210323,210403,210402,210303,210220,210121,210102,201205,201207,201202,210117,201127,201103,201022,201027,201027,200929,201102,201026,201007,201028,200906,200928,200803,200907,201023,200611,200417,200422,190422,190826,190225,190219,200417,191112,190327,190320,200513,200306,190725,180629,180830,180814,180726,181124,181113,190419,190121,190409,200410,200320,181219,181219,181221,200611,200203,190306,190227,190219,200514,190528,181214,190103,190102,190318,190506,191226,191226,200421,190716,181107,200319,200102,200305,200221,190103,190322,190111,190117,190117,200604,200811,180912,180912,180912,181004,181011,181016,181025,181025,181102,181102,181102,190418,190219,200416,190515,181214,190702,200413,200102,191111,191112,231231,231225,231221,231212,231207,231205,231121,231124,231108,231102,231025,231008,231007,230915,230925,230814,230622,230622,230815,230725,230725,230724,230721,230711,230706,230613,230511,230512,230523,230421,230424,230415,230411,230209,230306,230307,230216,221015,221019,221206,220521,221023,221010,221128,221022,221025,221026,221204,221026,220905,221009,220915,220825,220912,220508,220522,220808,220824,220517,220818,220718,220423,220422,220317,0002621107,220429,220419,220415,220305,220303,220326,220413,220104,220325,220207,220121,220304,220224,220223,190827,210611,210820,210704,210826,210802,201013,201208,201113,191220,190716,191226,191226,190318,190322,231220,231020,230721,230606,230224,221115,220507,220422,220422,220304,211221,211213,231229,231221,231115,231015,230914,230912,230810,230706,230825,230720,230710,230512,230522,230522,230518,230421,230424,230308,230308,230308,230309,230216,221015,220904,220302,220207,220105,220304,190422,190325,190409,181218,200305,181213,190116,181113,191228,190402,200511,190729,181130,200418,190218,200302,191212,191212,191202,211107,211008,210913,210903,210718,210723,210728,210704,210618,210618,210507,210502,210507,210409,210329,210328,210310,210111,201113,201126,201113,200910,201023,190426,190426,190912,191017,191112,181204,190819,190619,191023,190318,190227,200421,181225,200312,181106,200317,231231,231216,231124,231103,231015,231012,230823,230725,230706,230522,230208,230321,220521,221206,221116,221205,220825,220502,211218,220315,220416,211217,220323,211210,211215,230529,190510,181213,200413,200421,200108,200108,190319,181113,181214,181214,191019,190118,191213,190306,200418,190702,190620,200616,181128,181128,190111,190114,200728,200514,181218,211107,211115,210913,210812,210806,210824,210709,210626,210723,210511,210318,210318,210407,210409,210405,210403,210304,210117,210102,201013,201104,201028,201028,200929,200803,190325,191127,200203,190119,190729,200326,190426,191023,191025,200615,200615,191111,190623,190725,191202,181205,181228,190108,190819,191025,190225,191223,191017,200221,200222,190704,181106,191112,190415,181120,181120,190813,200604,190219,231231,231225,231221,231216,231212,231220,231121,231020,231123,231108,231103,231029,231023,231015,231012,231006,230920,230914,230828,230823,230706,230821,230725,230721,230720,230711,230616,230616,230606,230606,230510,230518,230522,230522,230524,230522,230424,230418,230419,230411,230323,230208,230308,230306,230304,220510,230216,221205,221015,221019,221021,220720,220720,220825,220502,220824,211218,220509,220314,220422,211221,220325,220324,220324,220325,220325,220304,220304,220223,211214,190528,181225,191025,190802,190507,200204,200214,200615,190827,190307,190624,190117,200227,200617,190729,191111,200519,181130,190426,190218,190121,200329,190823,211206,211122,211107,211027,211018,211026,211021,211013,210926,210914,210908,210904,211115,210810,210719,210718,210718,210718,210716,210823,210811,210823,210723,210703,210622,210625,210620,210618,210420,210611,200910,210512,210507,210505,210317,210407,210409,210405,210403,210326,210405,210305,210303,210121,210127,210102,201202,201202,201106,201113,201104,201102,201029,201022,201026,201007,200910,200908,200704,200803,200907,201023,191104,190621,190702,190912,191220,200304,191112,190327,191127,190725,181124,181219,200320,190419,200302,190819,200912,190227,200410,190108,190724,190716,191111,200615,200319,190119,190528,181128,181130,191125,200320,190318,190418,190809,200610,190702,190913,191223,191021,190114,200511,200108,230624,230821,230821,230821,230821,230821,230825,230724,230711,230616,230606,230522,230508,230422,230419,230323,230410,230209,230304,230220,230220,230131,221014,221208,221203,220524,221021,211216,221024,221021,221114,221009,220906,220906,221009,220825,220823,220506,220507,220522,220720,220808,211218,220421,220315,220416,220421,220302,220326,211221,220325,220323,220111,220304,200728,190805,200403,181213,190418,191021,181218,190307,200317,200227,190402,190506,191228,190623,200222,190218,190306,200519,190515,211021,211026,211021,211015,210926,210914,210913,210902,210812,210812,210809,210806,210802,210802,210718,210618,210619,210408,210408,210518,210517,210505,210505,210407,210502,210405,210324,210303,210103,210103,201013,201102,201022,201022,200907,200904,200417,190306,181205,200514,191205,190515,181210,190409,190621,190827,191217,191217,191023,181109,200319,190119,190813,181228,191017,190915,200420,200304,181106,181120,181102,231224,231205,231207,231025,231008,230906,230524,220808,211218,220423,220305,220302,220303,220207,220121,220111,220223,211213,211214,230925,230823,230726,230721,230510,230518,230421,230411,230306,221014,220524,220524,221022,221114,190813,210403,201026,200511,200331,200331,191220,200313,200728,200519,200519,191205,191228,191228,191016,191017,211107,211106,211015,210926,210908,210809,210729,210620,210620,210520,210515,210515,210509,210302,210302,210103,210111,210111,201208,201102,201023,200928,200928,200928,201029,200106,191220,191220,190823,190720,190823,200103,200103,200319,200220,190819,200320,190725,200417,191226,200512,200319,191017,200906,200929,200617,200418,200418,211204,211108,211108,210913,210904,210826,210820,210508,210806,210709,210710,210612,210404,210405,210323,210323,201127,201029,201029,201027,201102,200929,191114,191012,200911,200615,200611,200615,200320,200409,231220,231220,231121,231129,231029,231029,231029,231029,220823,230912,230828,230825,230823,230906,230811,230822,230725,230726,230721,230720,230706,230512,230524,230529,230421,230415,230415,230415,230415,230209,230321,230321,230303,230303,230303,221014,221014,221208,221023,220907,221114,221116,221115,221114,220907,220906,220523,220502,220506,220508,220829,220829,220517,220829,220314,220315,220429,220429,220415,211221,220324,211213,220214,211214,231121,231102,231102,231102,231102,180910,181004,181113,190108,181213,190318,190319,190218,190219,181220,190114,190419,190322,190402,190121,190107,190108,181128,181219,190219,190219,190109,190507,190426,190528,190528,190515,221022,221010,221024,190510,181213,190402,190306,190325,190121,190409,190723,190426,190725,190621,190109,190219,181219,190107,181105,181109,190318,190318,190826,181225,181214,190416,181120,190805,190108,190510,200326,200313,190826,190325,211115,211115,211022,210907,201202,201113,200929,200910,200907,200417,200418,190215,191220,181124,190724,200102,190320,190702,200616,190219,211210,191021,190802,190121,210509,200907,200422,181221,181221,191017,181214,190117,190118,190118,190322,190802,190805,181113,181113,200204,200204,191114,191016,200106,200912,190325,200304,211206,211027,211018,211026,211026,211126,211007,211006,211015,210926,210914,210913,210913,210914,210904,210904,210826,210903,210824,210812,210809,210806,210717,210718,210806,210907,210709,210626,210806,210709,210902,210811,210802,210907,210904,210721,210723,210719,210723,210720,210704,210704,210626,210618,210519,210612,210518,210511,210511,210502,210422,210407,210405,210405,210328,210328,210323,210322,210405,210303,210303,210117,210117,201208,201208,210111,201104,201103,201113,201113,201027,201014,200910,200929,201109,201028,200911,200904,200906,191025,191025,191025,200326,200326,190425,190425,190723,200203,200203,200410,200410,181219,190215,190215,190107,190107,191025,190320,190506,190325,191125,191125,200305,181228,200912,200928,200416,200416,200416,191017,191017,181106,181106,181107,190416,190416,190418,200317,191223,200413,191025,200102,200102,230811,230811,230811,230705,230705,230705,230705,230705,230705,230705,230705,230705,230705,230705,230705,230705,230630,230625,230625,230630,230630,230625,230625,230630,230630,230630,230625,230622,230622,230619,230619,230622,230619,230619,230619,230619,230622,230619,230622,230622,230619,230622,230619,230619,230619,230619,230622,230622,230815,230725,230711,230720,230706,230706,230613,230511,230209,230321,230323,230321,220906,220908,220905,221008,221010,221127,221021,230306,221128,221022,221022,221114,221114,220905,220508,220508,220808,220422,220419,220303,220305,220223,220104,220104,220104,220325,220212,220212,220214,220214,211204,200417,200512,200611,200512,200320,200421,211204,211204,211107,211107,211105,211105,211107,211105,211011,211011,211018,211102,211026,211021,211021,211017,211017,211014,230920,231006,230821,230511,230511,230510,230411,230323,220510,230131,221206,221021,221010,221204,221127,220825,220522,220517,220602,220305,220302,211221,220207,220105,220223,211213,190515,191104,191112,191112,200305,200420,200108,200704,200422,191021,191019,190624,191217,181113,200403,190409,190118,200331,190506,190218,190827,191212,191021,190219,181130,181130,190114,200313,200519,200519,191202,211204,211204,211107,211122,211115,211011,211108,211026,211017,211013,211007,210926,210926,210913,210926,210718,210718,210810,210811,210823,210719,210723,210719,210621,210621,210619,210610,210420,210515,210402,200611,210517,210507,210507,210507,210508,210407,210422,210407,210404,210328,210422,210324,210323,210402,210304,210220,210111,210102,201202,201102,201113,201102,201029,201029,200929,201023,201023,200928,200928,200106,190322,190218,190306,200512,190809,190720,200329,200514,191127,191127,190823,200803,200326,200417,200615,200103,200410,200304,190425,200103,191127,190723,190416,191023,190419,181219,191202,190107,190107,190819,190422,190528,191109,200615,190510,200320,181228,181228,190510,190724,190813,190704,191202,191017,200420,191223,200413,191021,190219,231020,231008,231006,230524,230914,230823,230705,230622,230809,230616,230512,230524,230508,230411,230209,230307,230302,220510,230220,230131,220521,221016,221203,221013,220506,221113,221127,220825,220720,220720,220522,220517,220422,220603,220315,220323,220104,211214,211213,211214,190416,191212,191228,211022,211014,210812,210519,200306,200320,230906,190729,230906,230814,230810,230906,230814,191223,190510,190826,211206,210718,210625,210709,210329,200514,181106,210914,201203,191218,190215,190415,200410,190306,190426,190809,191012,181219,190102,181213,191017,191212,201022,201007,200803,191228,200312,190823,190823,190725,200410,190121,190725,190318,190915,200304,200312,181213,200617,190306,190306,191228,211007,210806,210802,210618,210323,201203,191127,200409,190619,190108,181106,231102,230307,221024,220303,190319,190111,210907,210907,210720,210509,210325,201027,200417,181218,190520,190702,190514,190510,190319,190506,200103,211011,211102,211022,211017,211014,211006,210925,210812,210806,210802,210723,210622,210622,210618,210512,210508,210422,210121,201203,201013,201126,201027,191111,190826,190723,190416,190915,231231,231216,231124,231121,231121,231106,231102,231020,231008,230912,230825,230725,230721,230616,230512,230518,230421,230424,230419,230323,221019,221208,221204,221128,221010,221010,221010,221114,220907,221127,220912,220829,220523,220502,220507,220522,220829,220517,220818,220423,220602,220317,220305,220302,220413,220325,211220,220121,220105,220212,211213,200403,181214,200617,200422,191206,191206,181113,190117,190826,190827,190827,200418,200616,190620,191016,190325,190325,200514,211126,211126,211011,211122,211126,211018,211014,211006,210926,210914,210907,210824,210715,210716,210721,210721,210721,210626,210621,210720,210422,210611,210420,210515,210407,210409,210323,210304,210103,201204,201126,201106,201109,201014,200908,200912,200912,200911,201022,200912,200907,191205,191205,200326,190426,200615,191111,190215,200302,200319,190419,200227,190520,191212,200604,200615,191226,190716,200317,181120,181213,190510,190514,240104,231223,231216,231115,231115,231106,231029,231012,231012,230920,230920,230920,230920,230920,230920,230915,230828,230823,230823,230810,230624,230821,230724,230710,230710,230616,230606,230510,230510,230522,230420,230424,230323,230323,230308,230309,230304,230224,230216,230220,221205,221019,221206,220524,221021,221203,221007,221007,221115,221113,220510,220904,221009,220824,211218,220429,220603,220418,220423,220422,220416,220424,220424,220305,220323,220223,220413,220111,220212,220212,211203,211008,211008,211008,210925,210926,210925,210925,210908,210908,210904,210904,211015,210902,210820,210820,210810,210810,210719,210718,190116,200420,190402,190402,210327,201127,200906,200904,191114,190107,190915,211203,210323,191104,200422,191228,190506,210926,210903,210812,210723,210626,210613,210611,210509,210327,210404,210323,210303,201023,200811,201007,201022,190628,190720,190111,200617,190327,190915,190227,190121,200302,190723,191202,200227,200319,181129,200417,191112,181106,210709,210717,210716,210709,210719,210729,210728,210723,210721,210728,210703,210703,210703,210625,210618,210625,210710,210710,210620,210610,210610,210618,210609,210609,210520,210520,210612,210420,210408,210408,231015,181214,191218,191019,191217,190107,190723,200222,190215,210403,210318,210408,210318,210420,210507,210505,210505,210508,210508,210317,210502,210502,210422,210406,210406,210404,210404,210403,210329,231221,231124,231129,231029,231102,230724,230725,230912,230828,230725,230821,230809,230711,230711,230518,230524,230424,210728,230411,221208,221208,221015,221015,221206,220524,221202,221205,221024,220517,221023,231223,231221,231227,231123,231123,231029,231015,231008,230915,230829,230811,230811,230809,230720,230711,230710,230710,230606,230512,230518,230518,230524,230422,230411,230410,230308,230309,230303,230216,221015,221208,221206,221022,221116,221007,221007,221116,221115,220907,221128,220912,220904,220825,220720,220503,220502,220506,220507,220517,211218,220314,220314,220315,220416,211018,220302,220421,220421,220303,211221,220324,220207,220105,220212,220214,211213,230823,220214,181214,190528,200704,200603,190912,190307,181113,190507,190416,200313,200519,190729,191213,191212,190218,200107,190823,200103,191016,211206,211107,211011,211122,211021,211013,210812,210810,210809,210806,210806,210811,210625,210621,210619,210609,210519,210511,210507,210318,210420,210404,210329,210326,210305,210220,210220,201023,201013,201127,201113,201029,200616,200929,200928,200803,200904,190402,190325,190419,200415,200513,191220,190819,200302,190619,190108,190107,190321,190724,200611,181205,190716,200409,200409,200222,190520,190418,190624,200319,190813,191223,181217,181106,200107,190109,231225,231205,231015,230925,230825,230809,230710,230518,230411,230323,230220,220524,221007,221009,220825,220915,220523,220423,220414,220121,220105,211214,200728,190624,200418,211012,211021,211008,210926,210907,210806,210718,210811,210811,210723,210622,210520,210508,210508,210502,210404,210328,210310,210304,210302,201205,201202,201102,200910,190325,190306,200329,191127,200320,190307,190507,200415,181106,190119,190823,210327,191019,191012,211204,211126,211122,211105,210908,211102,211022,211018,211014,210925,210913,210907,210723,210802,210723,210703,210622,210622,210619,210408,210511,210408,210508,210111,210103,201208,201208,201202,201013,201127,201113,190121,190119,190107,191017,190915,191018,190103,240104,231229,231223,231223,231216,231121,231207,231205,231121,231123,231123,231106,231102,231102,231025,231007,231006,230920,230918,230823,230823,230809,230815,230809,230721,230721,230720,230710,230613,230511,230512,230518,230519,230529,230508,230508,230422,230517,230415,230410,230209,230209,230309,230224,221202,221202,221023,221013,221024,221022,220907,221026,221116,220905,221009,220825,220502,220507,220508,220829,220802,220509,220423,220422,220317,220603,220718,220718,220419,220302,220415,220303,220305,220326,220326,220324,220121,211210,220214,220212,211202,190415,190507,191019,190121,190409,190306,190111,190121,190704,191025,200214,200317,190805,190319,200306,181221,190827,200615,190307,190624,200302,181113,190402,190402,200511,200519,191228,200416,200313,181130,190507,190425,190426,191212,200617,181210,181220,200417,200417,211206,211204,211108,211108,211115,211011,211021,211008,211006,210904,210902,210716,210709,210622,210704,210719,210720,210610,210612,210420,210519,210518,210518,210618,210512,210509,210318,210422,210422,210404,210328,210326,210404,210310,210402,210220,210111,210127,210102,201203,201203,201127,201106,201102,201029,201014,200929,200908,200906,200907,200908,200617,190823,190628,190111,191012,200513,191220,190725,190322,191127,190121,181124,200403,200320,200615,181204,190819,190102,190724,200302,191226,191223,190716,200403,200420,200421,200312,190409,190409,191108,190913,190514,200312,190915,190109,181102,200422,191226,230914,231006,230912,230823,230811,230811,230622,230624,230724,230724,230512,230512,230523,230523,230422,230422,230420,230419,230410,230321,230307,230306,230306,230302,230224,230131,230131,221015,221015,221208,220521,221023,221008,221203,221114,221128,220902,220503,220506,220522,211219,220509,220422,220423,220315,220416,220429,220302,211213,211216,220207,220121,220111,211214,220212,220212,190510,190510,211206,211122,210826,210420,210317,210305,210103,201026,181204,190409,200513,191023,181219,191018,191018,191223,191223,190725,240104,231012,230725,230220,221023,220523,220423,211214,231025,231025,231008,231006,230925,230915,230823,230823,230705,230809,230724,230720,230721,211122,211109,211126,211126,211109,211202,211126,210826,210907,210908,210809,210710,210610,210619,210519,210408,210408,200904,200904,200811,200904,200604,200928,200416,200415,200416,200513,200416,200416,200415,200415,200513,200416,200416,200323,200323,200306,200306,200306,200409,200317,200323,200323,200323,200323,200325,200409,231224,231224,231224,231224,231224,231224,231121,231023,231024,231023,231024,231024,231024,231023,231023,231024,231024,231023,230829,230829,230829,230829,230821,230815,230823,230815,230817,230817,230817,230817,230817,230817,230517,230517,230517,230518,230517,230517,230420,230420,230420,230420,230323,230209,230208,230208,230208,230208,230208,230303,230303,230303,230303,221016,221019,221019,221016,221016,221128,221021,220925,220925,220925,220925,211219,211219,211219,211219,220423,220314,220422,220422,220326,220413,211217,220413,220326,220326,220105,220105,220105,220105,220105,220105,210812,231015,230411,220912,190227,190227,181120,190510,181220,181220,211204,211122,211122,211006,210925,210907,210812,210904,210717,210520,210518,210518,210511,210318,210502,210422,210404,210326,210323,210402,201026,210303,210304,210220,210127,210103,201127,201127,201126,201113,201103,190322,190628,181205,181106,231231,231225,231221,231216,231207,231205,231125,231115,231125,231108,231103,231029,231015,231012,231006,230920,230915,230912,230925,230823,230906,230705,230821,230821,230721,230720,230720,230706,230726,230616,230613,230522,230523,230424,230419,230323,230323,230303,230216,221208,221019,221206,220524,221128,221022,221022,221208,221204,221010,221010,221007,221114,221114,220907,220905,221128,220912,220915,220902,220503,220823,220502,220912,220522,220523,220424,220517,220818,220423,220419,220416,220214,220212,220303,220304,220413,220413,220326,220325,220121,220121,211210,220105,220212,211213,190506,211122,191212,191112,191114,200203,191212,190119,190729,211027,210907,210621,210407,200928,191226,181129,181214,190416,200107,240104,230828,230517,230131,221113,220303,231231,231223,231223,231216,231129,231220,231124,231129,231108,230821,231102,231025,231015,231008,231007,230925,230918,230914,230912,230828,230823,230811,230705,230809,230724,230720,230616,230606,230510,230512,230522,230523,230508,230421,230422,230411,221014,221019,221204,221202,221202,221010,221021,221204,221115,221203,220908,220907,221127,221127,220502,220317,220317,220824,220507,220507,0002626702,220508,220802,220718,220603,220602,220422,220418,220416,211018,220303,220302,211214,211220,211217,211221,220223,220224,220223,211216,211202,211215,211214,211011,211107,211011,211105,211109,211011,211011,211115,211011,211018,211026,211021,211018,211013,211014,211017,210926,210925,210907,210904,210907,210902,210902,210820,210810,210718,210809,210709,210716,210709,210719,210728,210721,210703,210626,210710,210620,210610,210609,210520,210420,210517,210512,210508,210317,210422,210404,210329,210327,210326,210324,210305,210402,210402,210303,210303,210121,210111,210102,201205,201103,201013,201127,201103,201106,201102,201029,201127,190805,190507,181120,200413,190827,200617,200420,181221,200107,190307,191111,190624,190327,190402,191228,191019,200313,181130,190506,191205,181113,200331,190226,191021,190826,200616,190623,200103,190306,201013,200929,200908,200906,200912,201022,201026,201007,190322,191202,200513,191220,190111,190720,200304,191212,190425,190119,190415,181204,190819,200616,200227,191004,191109,190619,191112,190320,190320,190723,181102,200421,191018,190418,190628,181107,181129,191112,200222,190813,200417,200102,181228,181204,190913,190704,200220,191212,190915,190109,181106,200304,200518,200312,190415,200214,181214,181218,181228,181106,181210,181219,181113,220304,220223,211203,211203,211203,211213,211206,211122,211022,211013,211013,211006,210925,210925,210810,210826,210812,210812,210826,210709,210811,210719,210704,210622,210626,210618,210618,210420,210517,210612,210508,210502,210502,210317,210407,210406,210405,210329,210323,210325,210402,210121,210111,201203,201203,201202,201013,201103,201022,201029,201022,201022,201126,191104,200421,200312,200317,190409,190912,200422,190520,200420,191212,191019,190402,190307,190623,190826,200617,181220,191114,200912,200313,200929,200910,200908,200803,200904,200519,200304,190426,190912,190720,181124,200418,191021,190306,191220,190109,190725,190402,200103,191127,190121,190121,190628,190528,200615,190419,191025,200319,181219,200811,190119,200222,190813,190318,190320,200417,191018,191212,200513,200313,190103,200107,190119,190510,181221,190118,181113,190219,190319,190325,190426,190108,181109,181205,190318,190318,190116,231231,231212,231029,231015,220823,220823,231006,230920,230828,230823,230810,230606,230606,230519,230424,230209,230308,230309,230216,230307,221206,221024,221114,220905,220904,220823,220808,220824,220509,220423,220602,220422,220422,220317,220305,220324,220214,220214,220212,220212,220212,211203,211214,200413,180910,200108,190415,190116,200518,200511,190416,190218,200728,190826,200103,180917,190506,190422,190819,190729,190215,211122,211108,211026,211007,211007,210908,210826,210718,210718,210709,210802,210610,210609,210609,210520,210520,210317,210328,210328,210323,210111,201113,201028,201029,201028,200803,200803,200611,200904,200514,191012,190720,190218,200320,200329,190419,191012,190528,191023,190724,181219,190102,200911,191226,200420,190327,200317,191125,181113,210622,181218,191025,200421,221021,220912,220904,220507,220509,220419,211216,211215,200413,200304,190515,191111,200331,200313,190725,190823,200313,200305,211021,210926,210723,210703,210626,210611,210317,210324,210121,201022,200906,201023,190912,191112,200611,181219,190318,190227,190528,190118,190107,190724,181106,181130,190325,190306,190114,190108,230706,190320,181130,181218,181220,181221,181219,190102,181107,190510,181214,200415,200511,191218,190912,200704,200704,190510,190319,200403,200403,200304,191112,191217,191226,190118,190118,200203,200413,191213,191021,191021,190802,200305,200323,190409,190409,200305,200204,181113,190402,200331,190724,190506,181221,190826,200616,200616,190306,200418,190515,200103,200106,200222,190114,191016,190322,191206,191206,200319,200203,200326,200304,200313,211206,211204,211106,211106,211115,211102,211026,211026,211018,211015,211006,211126,211012,211007,210913,210914,210925,211115,210914,210902,210812,210809,210806,210810,210824,210716,210810,210622,210720,210811,210726,210728,210802,210720,210728,210802,210704,210622,210625,210622,210621,210622,210620,210709,210704,210703,210618,210618,210610,210613,210612,210618,210518,210519,210420,210520,210518,210518,210515,210511,210505,210508,210509,210318,210317,210422,210409,210405,210405,210403,210328,210326,210324,210322,210304,210304,210302,210310,210127,210117,210103,201208,201203,201202,201202,201127,201126,201106,201102,201102,201027,201028,201014,201023,201102,200929,200929,200929,200908,200803,200907,200907,200803,200803,200911,201022,200906,200907,200907,200907,200514,190720,181128,191127,200312,190119,190119,200615,200616,200616,200616,190620,190729,191019,200326,190425,181218,181218,190823,191004,191109,190219,200410,191023,191023,190215,190306,191217,191226,200313,200326,190620,200103,200221,181109,190108,190702,190321,200512,200513,190725,190107,200319,200320,191202,200911,200611,190422,200306,190813,190225,200409,200409,200518,200312,190103,190103,190809,190520,190913,200420,200420,191017,181228,181228,200221,200221,191223,190704,181106,181106,190416,200304,200108,200604,200911,200317,200317,181120,191023,190805,190805,190624,190702,190226,190219,191025,231231,231225,231221,231212,231207,231205,231124,231121,231123,231106,231102,231020,231008,231007,230915,230918,230914,230825,230825,230906,230706,230809,230622,230815,230815,230725,230721,230711,230711,230706,230613,230512,230512,230518,230523,230529,230421,230421,230424,230415,230415,230418,230321,230321,230209,230309,230309,230306,230302,230224,230216,221014,221015,221019,221206,220521,221023,221204,221022,221115,221025,221026,221026,221026,221026,221113,220510,220905,221009,220912,220912,220904,220825,220823,220825,220502,220503,220506,220522,220522,220808,220824,220824,220517,220517,220818,220818,220509,220818,220818,220818,220818,220422,0002621107,220429,220419,220419,220418,220305,220305,220305,220302,220302,220325,211221,220324,220207,220121,220111,220111,211213,230810,230724,230512,230410,230131,220907,221009,220517,220418,211215,191228,221021,211011,210622,201022,190426,200409,200304,190723,200409,231231,231221,231220,231125,231129,231124,231129,231029,231023,231012,231015,231012,231012,231008,230822,230828,230814,230811,230811,230811,230706,230706,230821,230523,230523,230529,230529,230421,230421,230424,230511,230419,230415,230309,230309,230402,230309,230309,230309,230302,230224,221203,221203,221202,221202,220912,221116,221204,220925,221007,220905,220905,220720,221204,220904,220912,220904,220902,220823,220522,220802,220818,220421,220325,220718,220421,220418,220305,220314,220314,220422,220418,0002621107,220418,220718,220429,220413,220419,220416,220415,220416,220302,220214,220324,211217,220326,211217,220326,211221,220324,220212,220111,220224,220223,220214,220212,211202,211214,211215,211215,211214,211214,200420,190415,200422,200403,190416,190915,191127,190827,200302,200728,190422,211115,211011,211011,211026,211026,210907,210913,210908,210820,210812,210826,210806,210729,210810,210823,210729,210802,210718,210613,210618,210610,210619,210609,210612,210520,200912,210517,210508,210502,210407,210409,210329,210404,210327,210324,210323,210305,210121,210111,201202,201208,201113,201102,201023,201007,201007,200928,200906,201023,191025,190805,190416,200514,190109,200803,181129,191220,181124,200320,191111,190327,181219,190725,190619,200611,181129,191125,200403,190813,200421,200415,190109,200615,181213,181120,200422,181221,200312,190409,191112,181220,190826,211107,211018,211014,210926,210820,210826,210729,210723,210620,210619,210612,210507,210318,210404,210326,210303,210111,210103,210111,201113,201014,200910,200928,200329,190111,190215,181128,200417,190121,190826,190215,190109,191109,190419,191023,191127,190628,200512,191226,181109,191212,190219,231216,231216,231216,231212,231205,231129,231015,231008,231006,230711,230828,230721,230720,230711,230511,230511,230510,230323,230303,220524,221202,221128,221024,221116,221114,220906,220904,220502,220808,220509,220414,211220,220104,211221,211216,220121,220212,220223,220223,211213,231223,231125,231029,230814,230809,230508,230424,230

More Code Info

321,230216,221208,221025,221114,220904,220506,220517,220602,220602,220303,220104,220121,211214,230519,220904,220902,220314,200221,190426,190114,210611,200928,190913,190805,181213,181214,190415,181218,190416,200107,200305,200103,190826,190515,190702,200313,190426,181128,190306,200203,200203,190109,190108,190723,191023,191023,190724,190724,190318,181205,200305,190520,200413,190109,200418,210420,210408,210505,210317,210317,210405,210422,210328,210323,210304,210121,210127,201023,201104,201109,200929,200908,200904,200513,200326,200616,200409,181113,181204,181204,181129,191016,190915,191016,191016,190809,190619,181219,190102,190227,181107,210327,210326,210405,210326,210324,200305,200227,210127,190826,200204,191004,200319,200323,190621,200421,200704,200403,181225,181113,181225,191021,190510,190624,191217,190507,190624,190729,200519,200519,181220,181221,181213,190402,211206,211106,211012,211102,211013,211006,210820,210809,210904,210718,210726,210720,210723,210622,210622,210517,210511,210408,210508,210405,210403,210305,210402,210127,210127,210103,210117,210127,201202,201102,201103,201102,201102,200610,190425,190426,200513,190823,200604,200410,200611,200302,200214,191025,200320,191202,190915,190327,200102,200403,190111,191223,191226,190415,190119,231123,220825,220912,220509,220818,220509,211219,220418,220418,220418,211217,220414,220302,220305,220223,211213,211213,211202,211213,220303,220303,211220,211221,220326,211221,220323,220323,220207,220121,220111,220304,220212,220212,220212,211216,211213,211214,211213,190520,190515,190226,190226,200615,191218,190510,200511,200514,190409,190802,181113,200603,200511,190729,190827,190426,181130,191016,190111,190827,211206,211204,211122,211021,211115,211108,211115,211012,211021,211126,211017,211015,211008,210826,210926,210926,210926,210925,210914,210908,210907,210903,210902,211008,210907,210820,210820,210812,210826,210715,210823,210729,210811,210802,210802,210802,210802,210802,210726,210620,210420,210408,210409,210409,210329,210328,210326,210402,210310,210310,210220,201104,201126,201102,201106,201028,201027,201029,201014,201026,201102,200929,200908,200928,200803,200907,200329,200513,190628,200326,200329,191127,191212,190327,200417,200911,190227,190306,190218,190218,190306,190103,181109,191127,190107,200611,200611,181107,181130,181219,190716,190422,200107,190119,190528,200416,191226,190218,180814,191018,191202,191125,191127,181228,200610,191112,200204,191223,190916,200604,200604,230511,230422,230220,221010,220904,220303,190515,191223,200422,200514,200518,181204,190520,181220,190119,191021,191021,190913,190802,190226,200305,200306,200331,190117,200418,190506,200511,190506,190218,200418,181218,181220,181220,211204,211105,211106,211105,211115,211027,211011,211011,211011,211108,211108,211026,211022,211015,211017,211015,211013,211115,210925,210925,210914,210826,210820,210820,210625,210806,210718,210718,210824,210806,210709,210726,210626,210802,210811,210802,210802,210802,210728,210719,210719,210718,210622,210704,210710,210611,210520,210518,210518,210517,210505,210408,210520,210317,210502,210318,210406,210405,210405,210403,210327,210326,210323,210402,210402,210304,210303,210111,210111,210103,210102,201102,201208,201207,201203,201127,200421,201028,201027,201022,201014,201014,201007,200910,200906,200910,200803,200929,200904,191112,191228,200329,191220,190114,181218,200911,190819,190823,190823,190215,190215,200513,190108,190108,190108,190227,190528,191023,191109,181109,200203,190107,190107,200312,190819,190813,200928,190813,191226,190318,190318,200421,191017,190716,190716,181107,200320,200420,200319,190119,190416,181120,190704,200415,200416,230511,230524,230419,230323,230323,230303,221014,221208,221206,221208,221204,220524,220521,220524,221128,221203,221022,221115,221026,221203,221114,220905,220908,220915,220517,220506,220825,220523,220522,220502,220508,220808,220818,220829,220824,211219,220509,220718,220423,220421,220421,220602,220602,220415,220429,220414,220421,220305,220305,220302,211217,220326,211221,220324,220324,220207,220111,220304,211216,211203,211213,230517,230424,230307,230216,221016,220524,220502,221203,221113,221009,220523,0002626702,220415,220302,220105,220304,220223,210324,200403,211206,211105,211102,210926,211013,211006,210809,210810,210726,210723,210802,210704,210420,210520,210519,210518,210518,210517,210502,210407,210404,210403,210304,210304,201013,210111,201026,201007,200929,200929,200413,200511,200511,240104,231205,231029,231015,231015,230511,230510,230508,230512,220510,221014,221013,221127,221127,220915,211218,0002621107,220418,220326,220223,211213,191218,200108,200102,191217,191019,190320,190515,181221,200329,190916,190108,190227,200320,191004,200221,191004,190809,191108,190327,211204,210812,210323,230616,230616,230616,230606,230323,220510,230216,221008,221007,220818,220104,220214,210305,190226,190515,190218,190422,190226,181113,200420,200904,190422,211008,211015,210913,210729,210621,210618,210518,210404,210326,210305,210303,201208,201202,201029,201028,191206,181105,200329,190720,190724,190114,190114,190826,191023,181213,200222,190510,190320,191025,200518,231108,231106,230912,230619,230512,221014,221015,220908,220522,220121,220214,211203,221016,210310,210402,210402,210303,210303,210302,210220,210121,210121,210111,210111,210117,210102,201205,201205,201203,201203,201013,201127,201026,210903,210926,210718,210407,210409,210408,210810,210812,210802,211122,211108,211026,211014,210925,210902,210820,210810,210719,210709,210716,210719,210729,210719,210703,210622,210620,210610,210609,210420,210420,210317,210406,210406,210305,210323,201203,201106,201208,201102,210520,201013,201208,210507,210505,210317,210420,210422,210406,210406,210324,210305,210121,210111,210102,201203,201127,201113,201102,210720,210117,210103,210103,190325,190823,190912,190425,200617,200214,201023,200513,191226,200325,190119,190813,191025,191111,190402,190805,190724,190226,200413,200413,190507,190827,200617,200107,181210,181221,190624,200107,190418,190307,191111,190624,191019,191019,191019,190802,200305,191217,191217,191217,190117,190117,190402,190402,190402,200204,190218,200313,200329,200329,190506,190415,190514,181220,191212,191021,181220,200319,190628,200103,190111,190306,190306,191111,191111,201023,201023,201023,200910,200908,200906,200928,200911,200911,200911,200103,200304,190628,190628,191220,191220,190111,191012,190720,190720,190322,190322,191004,190121,181204,200320,200326,190425,190425,181204,181213,200227,200611,200203,190819,191004,190215,190215,191109,191109,190724,191109,190619,190619,200102,200102,190102,190121,190102,200302,190321,191023,191023,191025,191018,191025,190723,190819,190321,200421,200421,190107,190107,190107,190723,190320,190320,190418,190416,190418,190418,190418,181129,200222,200222,180925,200222,190813,190813,190225,190225,191109,191108,200417,190515,190514,181228,190913,191223,190915,190915,190915,190704,190704,181106,200420,200304,200313,200420,191111,200304,200107,200912,200907,200513,200403,190621,191108,200518,181213,211115,210717,190724,191111,191114,200519,200728,200610,191111,191111,200421,211204,210903,210618,210618,210509,210303,210303,210111,201013,200928,191025,190111,190720,190916,200326,190327,190723,190103,231124,231020,230508,230308,221021,220506,220508,220121,220212,190506,190121,190121,190121,181219,190108,190102,231216,231103,230914,230306,221115,230906,230906,220419,181213,190809,200403,200403,200422,200108,200227,200617,200305,200305,191217,190118,191019,191206,191206,200204,200306,181113,200227,200320,200331,190506,190506,200511,190226,190108,191213,200418,191017,190307,190307,190307,200616,191111,191111,191111,190623,200313,190813,191223,200416,190325,190325,200514,190823,211109,211011,211102,211022,211013,211012,211013,211007,211007,210903,210824,210824,210806,210802,210802,210723,210621,210621,210621,210720,210610,210520,210520,210505,210407,210405,210323,210310,210304,210304,210117,210117,210303,210117,210102,201204,201204,201013,201202,201126,201126,201102,201102,201027,201028,201023,201026,201023,200906,200908,200906,200906,200911,191205,200203,200319,191012,200811,200410,181124,190108,190108,190109,200302,200512,190725,191202,191202,190107,190121,190121,200319,190419,190307,181205,181217,200222,200227,200611,190318,200409,190227,190520,190528,190916,190724,190724,200416,191223,191213,191017,191017,191212,191212,190813,190716,191112,200323,200320,200325,200325,200325,190416,181120,200604,200603,190702,190219,220414,220416,220302,220324,220223,220304,220223,211215,211215,211215,200313,210728,210612,210612,210509,210305,191017,200704,190507,230419,230323,211216,190319,190319,190912,190809,200107,181130,200418,191025,191228,211204,210318,210317,210327,210326,210304,210304,210111,201208,201126,201104,200910,200911,200911,200513,190720,200514,190425,200306,190823,190621,190215,191111,190327,200513,191220,200302,191112,200103,181109,200312,190723,191226,200329,190102,181120,200305,190813,190809,190107,190307,190322,200107,200420,191108,190215,190307,211206,210519,200803,200102,230622,211213,230517,190913,191202,200422,190528,200518,181210,190226,190409,181120,200413,200603,191218,191022,200617,200615,190319,190319,190913,191021,191017,200615,200107,190327,190729,191226,200326,190402,200519,200317,200203,181130,181220,190218,190218,200420,181220,191212,200904,200222,190306,190419,211204,211107,211027,211102,211107,211105,211109,211022,211109,211027,211011,211011,211021,211026,211017,211021,211126,211017,211013,211007,211012,210926,210904,210914,210907,210719,210719,210718,210718,210517,210626,210703,210622,210720,210625,210703,210610,210710,210618,200906,210420,200910,210618,210512,210515,210502,210509,210317,210111,210502,210420,210406,210404,210502,210327,210326,210327,210405,210305,201026,210220,210220,210303,210121,210111,210102,201205,210121,201203,201013,201207,201203,201028,201203,201014,201126,200928,201126,200912,200912,200928,200908,200908,200928,201022,200907,201023,190322,190510,190628,190628,190111,190915,191114,200417,200514,200513,191220,200313,190121,190409,190409,190916,190725,181124,190729,181219,181204,181204,181204,200319,190628,190704,190724,200615,200603,200409,190528,190102,190116,190915,190321,200204,200417,190409,200421,191018,191025,190327,190107,190716,181107,181107,200319,200319,190320,181205,181205,200603,200928,190813,190225,190225,200403,200221,181214,190913,190704,191223,190915,200420,190327,190109,190109,181106,191125,191226,231229,231229,231225,231221,231212,231207,231205,231205,231124,231125,231123,231123,231124,231106,231102,231025,231020,231020,231008,231007,230925,230920,230915,230906,230825,230622,230622,230815,230815,230725,230721,230711,230706,230613,230511,230512,230519,230523,230523,230421,230424,230415,230415,230209,230321,230304,230303,230303,230224,230224,230216,221208,221015,221019,221206,221206,220521,221023,221113,221010,221022,221024,221025,221026,221007,221114,220905,221128,221009,220915,220904,220825,220912,220503,220506,220522,220808,220829,220517,211218,220718,220718,220718,220602,220317,0002621107,0002621107,220429,220419,220418,220421,220303,220303,220303,211220,220325,211217,220104,211216,220121,220111,220304,220304,220223,211214,181214,211105,210826,210703,210420,210318,210328,210111,201013,201013,201013,200908,200320,190915,200518,190418,181102,230705,230510,220502,220314,220105,190109,190103,200518,190802,181120,191218,190827,200811,200418,190307,190402,191228,181221,211215,211026,211013,211007,210826,210728,210703,210723,210408,210520,210402,201203,201013,201014,200928,190111,191012,200418,200513,191127,190425,190621,190227,190716,190107,200513,190515,200420,231129,231008,230705,230705,230606,230616,230508,230415,230321,230306,230131,221016,221024,221115,220906,220825,220503,220503,220808,220824,220315,220418,220303,211217,220323,220224,211214,190226,181214,190117,200305,190520,190506,211014,210902,210719,210710,210613,210420,210507,210404,210405,210323,210323,201208,201208,201102,200910,200305,200604,190725,200611,191109,181107,190913,191111,240104,230825,230402,220510,221024,220904,220506,220423,220305,211203,220522,220523,220423,220315,220421,211217,211221,200319,200317,200317,190415,191104,200603,200518,190913,190510,191218,181113,190725,181210,190915,191017,181210,190827,191228,200306,190307,190623,200325,190117,190506,190114,190409,190729,181130,191228,191212,190325,211105,211011,211011,211126,211026,211026,211013,211007,211006,211006,210611,210812,210806,210728,210802,210626,210622,210625,210622,210420,210519,210508,210512,210511,210508,210502,210318,210317,210409,210409,210406,210404,210403,210325,210324,210323,210303,201026,200421,210111,210102,201102,201204,210121,201202,201102,201106,201029,201027,201026,201013,201013,200910,200928,200928,200803,200803,200906,200417,190119,190729,190823,190823,191127,200306,200911,190419,190826,200409,191220,190723,190108,191018,191202,200304,190318,190318,181219,190716,200409,200610,200320,181129,181228,190915,181214,200610,200403,190219,220506,220823,220509,220423,220314,220315,220421,220305,220305,220302,211220,211217,211213,220304,220304,220214,220212,220212,220214,211203,211214,211202,231231,231225,231223,231216,231121,231106,231108,231103,231029,231029,231006,230925,230920,230915,230914,230914,230828,230823,230706,230630,230821,230809,230724,230720,230720,230710,230510,230510,230522,230522,230523,230529,230508,230422,230422,230422,230419,230419,230131,230131,221204,221204,221202,221202,220524,220524,220524,221022,221022,221013,221024,221024,221007,221113,221114,221008,221023,221023,221205,221114,220907,220907,220908,221127,221127,221127,221009,220906,221009,220906,220904,220915,220904,220503,220915,220825,220823,220829,220502,220507,220522,220508,220508,220523,220823,220802,211218,220509,220315,220415,220315,220429,220418,220429,220415,220415,220302,220421,220303,220302,220303,220302,220302,220413,220413,220105,220304,220207,220214,220214,220214,211215,211215,211214,211213,211027,211027,211011,211011,211108,211108,211026,211126,211021,211017,211007,210925,210908,210904,210902,210820,210820,210810,210810,210719,210812,210717,210716,210716,210709,210802,210719,210907,210823,210728,210723,210719,210721,210709,210709,210703,210703,210625,210710,210710,210620,210620,210610,210609,210609,210612,210420,210408,210407,210517,210517,210512,210512,210507,210507,210505,210505,210505,210502,210508,210520,210318,210318,210509,210407,210422,210406,210404,210420,210404,210404,210422,210329,210329,210327,210327,210326,210324,210305,210310,210402,210402,210303,210303,210327,210302,210302,210121,210121,210111,210111,210111,210102,210102,201207,201207,201205,201203,201203,201203,201029,201029,201022,201022,201026,191111,190805,190805,190507,200518,200617,200704,181120,181120,200413,190507,181210,190827,200107,200302,181210,190827,200107,200107,190307,191111,190624,190624,191217,190117,190117,190327,191019,190729,190802,190913,190306,191217,190117,200305,190402,190402,190506,200204,200204,200305,200305,181130,181130,181130,200329,190823,190826,200904,190506,191111,200106,200106,191111,201007,200929,200908,200906,200906,200912,200928,200928,200911,200907,200904,190628,190628,200103,190111,200304,200410,200304,200304,181124,200513,200513,200514,200514,191220,191220,190111,191012,190725,190322,190322,191127,191004,191004,191004,190720,181124,200320,190425,191127,190425,181204,181204,190819,190819,191004,200611,200611,191109,190215,190619,190619,200102,200102,190108,191226,181109,200421,191018,190102,190723,190723,190321,200421,191125,190723,181109,181109,190320,181129,181129,200610,190418,190418,191125,200319,200222,200222,200222,200611,190416,181129,190813,190813,190813,190225,191108,191108,191108,200513,190515,190514,190514,200102,181214,190913,190915,190913,190704,190704,190704,190915,190915,200108,181106,181106,200306,200306,200420,200304,200108,200312,190415,190415,190415,200214,200220,200518,191025,200420,200904,211011,211014,210914,210914,210715,210811,210720,210610,210515,210508,210304,210405,201013,200616,200421,200611,231207,231220,231125,231029,231007,230925,230725,230518,230523,230523,230424,230308,230304,221206,221128,221116,220908,221127,220523,220507,220509,220416,220416,211217,211217,220323,220207,211210,211215,231231,231224,231124,231115,231121,231129,231108,231102,231015,231008,231008,231008,231006,230421,231006,230920,230925,230915,230915,230912,230912,230821,230724,230710,230616,230511,230512,230524,230524,230529,230508,230422,230422,230422,230420,230420,230415,230131,230131,221205,220524,220524,220524,221202,221202,221022,221010,221008,221008,221023,221024,221114,220907,220906,220906,220904,220503,220823,220825,220823,220829,220503,220506,220507,220522,220522,220720,220824,211219,220818,220509,220509,220315,220315,220603,220314,220422,220429,220416,220303,220324,220324,220207,220207,220111,220111,220111,220111,220105,220224,220224,220212,220212,211216,211215,211108,211115,211108,211115,211126,211026,211021,211021,211017,211017,211013,211014,211017,211008,211014,210926,210925,210925,210908,210908,210904,210904,210826,210820,210820,210810,210810,210718,210718,210717,210716,210717,210907,210809,210806,210709,210709,210719,210719,210729,210729,210728,210728,210723,210723,210703,210719,210709,210709,210703,210703,210709,210618,210625,210710,210710,210620,210620,210610,210610,210609,210609,210520,210420,210612,210404,210405,210420,210512,210512,210507,210507,210507,210317,210318,210508,210422,210422,210407,210406,210420,210404,210329,210329,210327,210327,210326,210325,210403,210403,210305,210402,210402,210327,210303,210302,210220,210121,210121,210111,210111,210117,210102,201205,201205,201203,201203,201013,201103,201127,201127,201113,201113,201013,201103,201029,201022,201026,190402,190805,190724,190507,181120,200413,200617,200617,200403,200107,181221,190827,200418,200107,190624,190624,190117,191019,191019,191019,190913,200519,200615,200519,191217,191228,190402,200313,200313,181130,200329,200329,190506,191205,200728,190823,200417,200610,200617,190623,190628,200103,190111,191111,201007,201007,200910,200929,200908,200908,200910,200906,201022,200928,200911,200911,200907,200904,200103,200304,200304,181124,200514,200513,191220,191012,190725,190720,190724,191004,200329,190425,181204,190425,190819,181204,191004,191004,200615,200611,191109,200518,200409,200302,200416,200302,191226,190319,190619,200102,200102,191226,181109,200421,191018,191018,190102,181124,190723,191127,190723,190723,190418,190107,181129,181129,190225,191108,200416,200417,200417,200416,190515,190514,200102,181228,190913,191223,191223,190702,190729,191212,190913,190913,190704,190915,200108,191111,200214,190109,190805,181102,200422,200312,190415,200518,200518,231121,231103,231012,230925,230925,230906,230706,230510,230402,230424,230529,230420,230321,230216,221016,221202,221023,221021,221009,220503,220508,220424,220421,220302,220324,220224,200421,190422,191019,181220,181220,200313,190111,211202,211027,211022,211022,211017,210926,210902,210715,210717,210709,210721,210721,210625,210612,210519,210517,210507,210318,210505,210502,210326,201026,201204,201106,201023,200803,200907,191111,200106,191114,190306,191109,190419,200611,190716,191025,190813,190813,190515,191223,191223,200317,200313,181120,200317,231229,231207,231124,231106,231029,231015,231008,231012,231008,230825,230811,230811,230706,230705,230822,190915,181113,190402,200221,181220,190506,181218,210720,210515,210404,201113,200911,200904,200513,181109,191125,190227,191108,190716,181217,210729,190121,231123,231006,230828,230524,221013,220509,220104,231229,231124,231121,231015,231008,231006,230925,230809,230710,230710,230508,230420,230411,230220,220509,231129,231115,231106,231102,231008,230925,230925,230920,230915,230912,230706,230821,230809,230512,230508,230524,230508,230508,230424,230131,221202,220521,221022,221024,221008,220905,221127,220503,220506,220507,220506,211218,211219,220509,220314,220418,220424,220302,211217,220104,211221,220324,220207,220121,220207,220214,211216,211202,211215,211126,211213,211204,211105,211021,211017,211013,211007,210820,210810,210802,210806,210709,210716,210709,210719,210703,210720,210625,210710,210620,210618,210609,210520,210610,210612,210420,210408,210618,210512,210507,210505,210502,210407,210422,210406,210404,210422,210327,210422,210326,210324,210402,210303,210220,210121,210111,210102,210102,201203,201013,201127,201126,201127,201103,201104,201102,201029,201026,191025,190805,190507,191021,200617,200418,200108,181221,200413,190624,191228,191019,190729,200519,190402,200313,181130,191205,191213,201007,200910,200908,201022,200911,200908,200906,200904,200904,190823,190628,200103,190111,200304,191220,190916,181124,190322,191004,190725,190425,191127,181204,200302,200617,200611,191109,190619,200102,190102,191226,200421,190723,181113,181129,191125,200611,190409,190813,190225,191019,190515,181228,191223,181106,190704,190915,200204,190619,200420,181120,200317,200306,200214,200518,200420,240104,231231,231225,231103,230823,230606,221204,220524,220524,221202,220907,220908,220912,211213,220214,220212,211214,200306,200615,211027,210729,210726,201106,201029,190108,191025,200910,190827,191025,190117,190117,200306,191021,191111,190325,200418,200418,181130,200418,190826,211107,211011,211115,211126,211022,211007,211007,210914,210908,210820,210907,210812,210809,210715,210826,210718,210723,210802,210802,210703,210519,210519,210519,210519,210511,210515,210408,210404,210404,210403,210403,210329,210304,210402,210402,210127,210121,210102,201127,201102,201102,200616,201014,201007,200803,200904,200904,191016,190729,190729,200616,190720,191220,190425,190621,191127,191205,200103,200203,200409,200513,190215,190108,190419,191202,190107,190116,190716,181228,200610,200317,200103,181214,231221,231212,231108,231129,231121,231008,231006,230925,230811,230809,230711,230606,230510,230511,230512,230529,230424,230424,230323,230321,230303,230216,221016,221204,221008,221114,221113,221113,221127,220720,220506,220823,220829,211218,211218,220315,220315,0002621107,220414,220305,220305,220324,220207,211203,211214,211213,211213,211115,181213,190809,200403,200403,181228,181213,200312,200312,190118,191206,200331,200331,190507,200413,200227,191213,191022,190307,191111,190114,190114,190114,190823,190729,190729,211210,211210,211109,211011,211011,211011,211115,211022,211018,211018,211014,211013,211006,211007,210914,210903,210806,210729,210716,210823,210719,210721,210721,210610,210626,210723,210625,210621,210621,210621,210618,210610,210610,210519,210515,210509,210409,210409,210409,210405,210310,210322,210304,210127,210117,201202,201202,201109,201204,201102,201102,201028,200929,201007,201007,200906,200911,190325,200514,191205,190119,181218,200203,190215,190826,190620,190620,181109,191012,191023,191202,191202,200319,190419,190419,181217,200611,191218,191109,191109,190103,181101,181228,190916,190916,191226,200221,191213,200221,191112,190716,190416,200603,200603,190702,190702,190219,190219,190219,200413,190510,190510,190515,231231,231231,231221,231207,231124,231115,231129,231108,231102,231020,231012,231012,230925,230918,230825,230825,230814,230814,230705,230706,230815,230726,230512,230518,230529,230420,230424,230415,230418,230209,230321,230308,230304,230302,230224,230224,230216,230131,221015,221015,221019,230131,221206,220521,221023,221115,221021,220904,221007,221113,221128,221203,220906,221203,220823,220823,220823,220507,220808,0002626702,220824,220818,220602,220315,220422,220315,220718,220718,220416,220323,220302,220323,220302,220415,211221,211221,220323,220105,220304,211216,211202,211202,211202,211202,220305,220508,0002626702,220508,220824,211218,220718,220422,220315,220429,220416,220421,220303,211220,211221,220323,220207,220105,220304,211214,210908,211027,231216,231124,231121,231123,231129,231106,231029,231025,231008,230920,230914,230912,230821,230209,230410,230220,230220,221016,221208,221009,220912,220808,220207,211214,211213,230308,221113,220523,220324,191016,211014,210612,210121,190321,200222,190913,200306,200422,190510,181221,191217,191019,181113,200511,181213,200403,200103,191022,190729,190506,190623,190826,200912,190325,200417,190111,211115,211017,211022,211115,211011,211018,211026,211105,211015,211008,210926,210908,210903,210902,210826,210820,210812,210826,210806,210716,210810,210811,210811,210723,210704,210613,210619,210610,210618,210420,210611,210520,210517,210520,210509,210317,210407,210405,210403,210422,210326,210323,210305,201026,210220,210121,210103,201204,201013,201127,201113,201027,201029,201014,201007,200910,200910,200928,201014,200906,190306,181218,191127,200304,181124,200326,190425,190121,191112,190621,190327,191012,191004,191111,191023,190725,200615,200320,200512,190321,190320,190102,200513,190515,190704,190418,191112,200415,190119,190915,190913,190422,190116,191112,191212,190528,190912,190510,190415,191108,191112,200811,200403,190702,180917,181218,200107,190802,200403,200511,190623,190506,191218,191213,190218,181130,190402,190307,181130,190419,211105,211105,211012,211026,211026,211008,211013,211013,210914,210913,210824,210904,210718,210823,210721,210802,210703,210620,210625,210618,210618,210520,210609,210517,210518,210517,210507,210502,210502,210407,210404,210403,210329,210327,210323,210402,210402,210127,210111,210102,210121,201203,201013,201127,201103,201027,201029,201028,201023,201007,201007,201007,200928,201023,201023,201007,190426,191213,200304,200513,191220,191220,200615,190823,190823,190621,181204,191012,200417,190819,191111,190227,200512,190528,181109,200319,190321,191202,190107,190107,190422,191018,181120,200221,190510,190102,190102,200317,190716,190102,200604,181120,181214,190915,190716,200108,190805,181102,200312,190415,200422,181213,200317,190409,190912,190510,200704,200108,200403,190827,191021,200312,200312,190319,191112,191212,190118,190307,200511,200227,190506,200617,200511,190218,190827,181113,181221,191212,200418,190623,190402,190325,200222,200912,190402,190218,190322,190416,191025,181128,200329,181218,190425,200513,200615,200615,190422,181124,190819,190620,191004,190121,190716,190725,190702,200611,200319,200319,190422,191202,190102,200512,181205,191018,190227,191019,191109,190620,190620,200222,200222,200417,191017,191226,190514,200204,190716,181106,190704,200420,191017,190619,181120,190219,200108,200302,190109,191025,231123,231123,231020,231129,231029,220823,230912,230912,230630,230815,230725,230522,230522,230523,230523,230415,230410,230410,230220,230216,230131,221208,221204,220524,211216,221007,221025,221007,221203,220908,220908,220523,220502,220720,220912,220829,220829,220912,220912,220818,220818,211218,220423,220314,220303,220303,211217,220326,211221,220323,220212,211213,211214,211213,190219,190724,190913,200306,191025,200704,200511,200108,190912,191017,200107,200413,200728,200519,190729,181220,190823,191016,211105,211011,211011,211021,211021,211013,210908,210908,210914,210826,210802,210619,210618,210517,210317,210317,210317,210406,210405,210324,210127,210121,210121,210121,201013,201126,201126,201102,201022,201014,200910,200928,191012,181218,190327,191111,190425,190419,191127,200611,190506,181109,190419,200512,200417,190318,190102,191223,200416,200421,190227,190321,190528,190102,191226,200222,190520,190915,190418,190102,181106,181106,200312,191021,200415,200302,190510,191019,211204,211006,210902,210902,210820,210811,210802,210703,210622,210613,210324,201208,201203,201127,201026,200803,191025,191111,190915,231231,231225,231221,231227,231205,231129,231124,231123,231121,231106,231102,231102,231015,231012,230915,230912,230925,230814,230705,230619,230721,230724,230606,230512,230523,230524,230508,230424,230415,230209,230321,230309,230224,230216,221208,220510,221203,221128,220906,220915,220902,220823,220502,220506,220808,220824,220818,220314,220602,220418,220429,220416,220421,220303,220303,220104,220325,211210,220121,211213,220212,211017,210512,210508,210508,190415,190415,200221,200107,190118,190915,200616,200617,181220,191205,191023,191111,190620,190620,190419,190306,211206,211215,211105,211102,211017,210913,210903,210812,210809,210824,210811,210726,210802,210703,210703,210723,210620,210609,210420,210518,210515,210505,210520,210502,210420,210329,210403,210403,210327,210405,210402,200611,210220,210127,210111,201027,201023,200908,190729,190720,200304,200514,200514,190119,181128,200329,191212,190425,190102,181129,190813,190318,200420,191109,200417,190219,190219,231121,231006,230915,230925,230705,230710,230725,230720,230710,230613,230508,230606,230511,230508,230512,230519,230523,230523,230529,230422,230422,230422,230424,230418,230411,230410,230323,230209,230208,230321,230302,220510,230131,230216,230216,221014,221019,221206,221203,220524,221128,221013,221013,221010,221007,221010,221116,220825,220902,220906,220915,220906,220523,220523,220720,220503,220507,220808,220829,220718,220429,220423,220602,0002621107,220416,220421,220424,220104,220324,211216,220104,220121,220223,211213,211214,191212,200204,211204,211122,211107,211014,211013,210925,210913,210520,210518,210612,210507,210329,210304,201208,200929,191125,231229,231216,231205,231220,231121,231121,231123,231129,231029,231023,231015,231012,231006,230920,230912,230810,230705,230822,230809,230724,230720,230711,230402,230410,230209,221023,220510,230220,230131,221014,221206,220524,221013,221114,221205,220908,221127,220502,220507,220829,0002626702,220808,211218,220818,220315,220718,220303,220305,220302,211221,211221,220324,220121,211210,220223,220214,211202,211107,211109,211027,211122,211102,211022,211015,211014,211012,211006,210926,211006,210907,210902,210824,210812,210718,210718,210824,210726,210719,210621,210703,210618,210619,210519,210420,210520,210518,210612,210509,210318,210409,210407,210405,210403,210328,210326,210324,210304,210304,210127,210103,201208,201203,201202,201202,201126,201027,201028,201014,201023,200929,200616,200912,200910,201014,200512,231231,231225,231221,231212,231207,231205,231124,231106,231102,231025,231015,231008,230925,230915,230914,230814,230622,230815,230725,230721,230725,230706,230613,230512,230519,230523,230529,230421,230424,230415,230415,230209,230321,230309,230304,230302,230224,230224,230216,221015,221019,221206,220521,220521,221023,221204,221024,221025,221026,221026,221205,220905,221128,220906,220904,220825,220502,220503,220506,220522,220824,220517,220802,220818,220418,220423,220602,220317,0002621107,220419,220415,220305,220303,220104,220324,220207,220105,220214,190402,181225,190119,211013,210904,210903,210823,210723,210710,210611,210611,210611,210518,210502,210317,210422,201013,201027,200410,200417,181124,200409,200513,190725,190121,190422,191112,191114,190109,211213,220104,220212,211006,210507,210326,210402,210127,201205,201013,210127,210220,210102,230510,230510,230519,230529,230529,230422,230419,230419,230323,230323,230208,230216,181214,181210,181113,190418,190117,190117,191212,200302,211013,210914,210826,210721,210626,210612,210511,210405,210329,210327,210303,210103,201204,201026,200911,200512,200514,191127,191205,200326,190724,200611,190422,181129,191004,200102,190520,211012,191004,230512,231029,231012,230914,230829,230823,230811,230706,230630,230815,230809,230724,230721,230710,230613,230508,230512,230529,230529,230422,230422,230411,221208,221205,221014,221025,221114,221128,221128,220823,220424,220314,220421,220414,220104,220105,220105,220224,211213,211213,211107,211115,210520,190218,210729,190515,190121,181109,190109,190515,220223,181218,181220,180925,190716,190912,190419,190419,190510,190225,191018,190116,231125,231123,230912,181128,181205,200220,190227,190102,181217,190226,190425,200420,200415,191125,190805,190724,190226,191112,191112,231015,230823,230619,190307,190307,200617,200416,211126,210716,210422,210327,191016,191016,181217,181217,231227,230308,221113,220509,211221,200403,200306,181221,190827,190510,200904,200413,200413,191025,191025,191017,200108,191218,191218,190325,181113,200413,200403,181221,200422,190409,190415,190624,190809,191019,190802,200306,200305,200326,190409,190507,190507,200511,200511,190623,190218,190219,191016,190321,190114,200617,191111,200331,200604,191213,190826,190327,190912,200617,190729,190729,181130,190426,200222,200107,190121,190218,190109,190219,190422,200418,200418,181218,200313,200313,190325,200519,211206,211109,211109,211115,211107,211105,211027,211115,211018,211021,211017,211013,211007,210926,210926,210914,210913,210904,210902,210902,210902,210826,210809,210826,210718,210717,210824,210806,210709,210810,210823,210811,210802,210726,210802,210723,210704,210704,210618,210612,210611,210310,210323,210326,210327,210327,210517,210518,210517,210515,210318,210502,210511,210507,210509,210404,210404,210502,210318,210317,210317,210407,210422,210403,210404,210409,210406,210405,210404,210404,210403,210329,210328 ,210327,210327,210327,210310,210327,210326,210324,210323,210322,210305,210305,210305,210402,210402,210304,210303,210302,210220,210127,210127,210121,210117,210111,210103,210103,210102,201208,201205,201203,201202,201013,201013,201127,201208,201127,201126,201113,201104,201106,201113,201028,201103,201106,201027,201027,201029,201028,201023,201022,201026,201023,201026,201023,200929,200910,200908,200910,200908,200908,200911,200803,200803,200907,201023,200906,200907,190227,190318,190322,200417,191206,191114,200317,190419,190306,190306,200329,200514,190111,191111,200615,200615,181113,181128,191114,191112,200326,200417,191220,191220,190823,200615,190725,190725,190621,190621,190623,200103,191212,190409,200320,181219,181219,200416,200302,200304,200513,190320,191109,190827,190916,190227,190321,200313,191127,181109,190419,200421,200728,190528,200615,200611,190506,190318,190102,190724,200615,190720,190716,200319,191218,200403,181205,191018,200102,200102,200319,190116,190225,190225,200403,190219,190225,190225,191226,190520,180814,190827,190915,191108,191017,191125,191125,190704,200610,181031,181107,181214,191223,200312,200306,200317,200416,200604,200102,200220,190116,190219,231108,231108,231029,231024,231023,231020,231006,231006,231006,230815,230815,230523,230418,230418,230418,230418,230411,230411,230410,230410,230323,230323,230323,230323,230308,230308,230309,230302,230220,230216,221014,221014,221015,221203,220524,211216,221114,221205,221202,221013,221013,221022,221022,221010,221010,221021,221204,221008,221114,220907,220908,220906,220906,220906,220904,220825,220912,220508,220502,220507,220522,220523,220824,220517,211218,220509,220509,220509,220423,220314,220422,220315,220414,220414,220416,220416,220303,220302,220305,220305,220303,211220,211217,220121,220111,220111,220304,220223,220214,220212,211216,211213,191217,191022,190528,191218,181113,200108,191017,191021,191021,191112,190624,200325,190402,190327,190426,181130,181221,181228,200907,190426,200313,190823,190823,190306,200409,200203,190507,190826,190729,191004,200304,190103,200320,181107,181205,190108,190225,200102,191018,200513,191112,191112,181106,191017,191022,191111,220904,221022,190416,190218,211108,210519,210318,210318,210318,190725,190723,200610,231020,221007,220825,220507,220429,190307,190307,181220,190506,190506,200106,191228,211202,211026,210904,210622,210626,210704,210618,210515,210405,210405,210323,210323,210102,201106,201013,201026,200912,190218,200227,190723,191109,190724,200420,190916,191017,181120,231123,231012,230823,230706,230418,221208,221010,220906,220508,220602,220413,211216,200418,200107,191205,191021,191111,231207,231115,231029,230823,230705,230821,230809,230606,230424,230308,221013,221010,221115,231220,231124,231029,231227,231229,231221,231212,231115,231029,230912,230912,230811,230720,230710,230522,230523,230524,230321,230308,230303,221206,221202,221206,221024,221021,220908,220915,220522,220502,220502,220508,220517,220422,220421,220314,220303,220323,220121,220111,220223,200617,200704,191217,190226,190507,190507,181130,181220,190307,191016,211108,211108,210926,210809,210806,210715,210826,210704,210610,210517,210318,210329,210329,210322,210127,201113,201126,201126,201023,200803,200519,181218,191109,200616,200214,191127,200221,191018,190716,191226,200420,190809,200416,191226,200317,231221,231205,231129,231125,231108,231108,231102,231007,230918,230918,230915,230912,230925,230925,230828,230828,230825,230823,230811,230706,230622,230817,230809,230724,230721,230720,230711,230616,230606,230511,230511,230523,230508,230402,230424,230410,230410,230302,230302,221014,230131,220524,220524,220502,221013,221013,221013,221024,221024,221115,221114,221008,221114,220905,220905,221128,221127,221009,220902,220915,220912,220902,220823,220522,220522,220523,220824,220305,220315,220315,220429,220419,220416,220111,211217,220104,220325,220325,220324,220212,220207,220121,220121,220105,220224,220224,220223,220212,220212,220214,220212,211215,211204,211215,211213,211204,211204,211106,211107,211011,211122,211018,211018,211126,211021,211017,211013,211013,211008,211008,210926,210926,210925,210914,210926,210904,210904,210820,210902,210820,210820,210810,210810,210809,210709,210709,210907,210717,210625,210625,210716,210716,210709,210729,210728,210728,210723,210723,210719,210802,210719,210703,210709,210703,210625,210710,210620,210620,210619,210610,210620,210609,210519,210420,210327,210405,210408,210408,210329,210329,210403,210329,210507,210505,210508,210508,210508,210408,210317,210502,210502,210422,210406,210404,210408,210329,210403,210329,210422,210327,210327,210327,210327,210327,210327,210325,210325,210405,210324,210324,210305,210305,210402,210402,210303,210327,210304,210303,210302,210220,210121,210121,210117,210102,210102,210102,201205,201103,201113,201106,201102,201029,201029,201022,201026,201026,190402,190402,191104,190912,200617,181120,190507,190507,200214,191218,190827,200108,200403,190624,191019,190802,190913,190402,190506,190218,190218,190506,191205,181124,181113,191212,190826,200616,181204,190111,190306,201023,201007,200908,200906,201022,200928,201022,201022,200907,200904,190628,191012,200513,200513,191220,200304,181204,190425,181204,200203,200227,200617,190215,190215,190619,200204,190724,190724,191226,181109,181102,191025,190103,190723,190819,191226,200421,200421,190628,190320,190418,181129,191125,200222,200928,200610,190418,190813,190813,190225,190514,190514,191018,190913,190704,191223,191223,200220,200220,200304,181106,190415,200304,190805,200214,200518,220212,191111,191111,190826,181130,200313,190327,191004,190227,210716,210403,210121,181129,200305,200418,211011,210823,210704,210613,210102,191205,191205,200513,191023,181214,190716,190704,211214,210908,191127,200409,190802,190219,230309,230306,230306,230303,220510,230131,221208,221015,221015,221015,221206,221206,220524,221013,221024,221010,221021,221026,221115,221114,221113,220905,221128,220902,220902,220912,220902,220523,220503,220503,220507,220507,220508,220523,220823,220517,220509,220423,220603,220314,220415,220415,0002621107,220414,220416,220416,220421,220305,220305,211220,220324,220121,211210,220223,220212,220223,220212,211202,211202,211214,211215,211215,211215,211214,190620,190515,190620,190515,181214,200421,191025,191025,191218,200704,190724,190226,190912,181113,190322,190321,190528,190520,191217,190307,200403,200511,200305,190724,200204,181204,200519,191111,191111,200418,190624,200325,190809,190813,190623,190114,200417,200519,200320,211206,211105,211109,211106,211102,211011,211011,211011,211115,211026,211021,211022,211017,211008,211006,211006,210913,210914,210907,211115,210902,210820,210809,210717,210824,210826,210709,210518,210626,210726,210811,210802,210723,210728,210728,210802,210626,210704,210619,210610,210610,210609,210420,210611,210611,210611,210518,210518,210515,210520,210318,210508,210505,210505,210317,200518,201113,210409,210409,210406,210405,210403,210304,210326,210323,210325,210305,210305,210402,210302,210303,210220,210220,210127,210117,201027,201202,201202,201202,201013,201104,201013,201113,201104,201102,201102,201028,201022,201014,201023,201007,200910,200910,200910,200803,200803,201023,200907,200929,200317,191112,181130,190322,190515,190620,191025,190119,200514,181218,181218,190507,200302,200203,200615,190826,190826,200103,190215,200410,200513,190725,190725,200204,200320,191018,191023,191213,190915,200911,200611,181024,181105,181205,181204,191202,200306,190813,200403,200409,200222,190813,200513,200511,200325,200325,200413,191021,190219,190813,190813,190809,190913,190913,190913,191223,200302,191111,190506,181102,220904,211215,190327,211008,210806,200331,211026,211026,210718,210622,210117,221014,220223,220212,200617,231207,230810,230720,200403,191019,190623,190826,190729,210719,210703,210508,210406,210405,190725,200320,200304,190724,190520,181106,230711,230424,230323,230304,221208,220524,221026,220522,220523,220912,220802,220419,211216,190619,190619,191112,191018,190102,190916,200421,191112,190813,190515,181228,181106,200312,191109,191228,190218,190422,211105,211108,210826,210325,210328,190327,190322,191018,191025,191018,200319,190215,190107,190218,181120,181107,191112,191125,190102,190107,190716,190704,200107,200102,191217,200511,200227,210806,210823,210613,210517,210509,210509,210325,201204,201027,200514,191125,191023,190702,200604,200603,190226,191206,181218,210709,201013,190119,190319,190108,181205,181217,190416,190702,190510,190510,190422,190510,181213,191218,191104,190912,200422,200422,200108,190319,181113,181214,191019,200617,200617,200306,200312,191217,191217,191217,190118,191213,190409,200214,190226,200227,200331,191206,191206,191206,190111,190121,200418,200418,200616,191016,191016,200514,200519,200313,211206,211206,211210,211107,210502,211122,211106,211106,211108,210111,211115,211012,211011,211115,211122,211108,211102,211015,211018,211015,211013,211007,211015,210404,210422,210926,210926,210925,210913,210907,210907,210903,210824,210908,210908,210809,210715,210806,210824,210802,210726,210729,211021,210811,210719,210726,210721,210721,210318,210704,210703,210625,210626,210626,210718,210622,210622,210621,210625,210618,210619,210613,210613,210613,210519,210420,210520,210518,210515,210517,210511,210511,210511,210505,210318,210404,210318,210409,210509,210318,210318,210407,210407,210408,210409,210405,210405,210405,191217,181113,190507,181221,181221,190109,190322,190415,190827,190802,190226,190307,190422,191016,191111,181113,190325,190729,190623,190218,190426,190114,181218,191228,190823,190306,190108,191012,190327,190621,190121,181124,181124,190419,181219,181219,191220,190819,190725,190227,190528,190318,190102,190102,191109,191018,190716,181107,181129,190320,190325,190225,190416,190416,190306,191021,181214,181225,190103,190103,190912,190805,190702,190515,191112,231102,230706,231129,231125,231103,231025,231015,230918,230912,230914,230809,230809,230724,230711,230616,230606,230510,230522,230524,230424,230424,230418,230302,221014,221022,221115,221008,220925,221026,221113,221113,220905,221127,220902,220825,220503,220325,220325,211204,211215,211213,211204,211106,211011,211011,211126,211017,211008,211007,210908,210904,210820,210810,210716,210716,210709,210719,210729,210703,210625,210710,210408,210618,210512,210408,210317,210502,210502,210406,210404,210329,210303,210302,210111,201127,201113,201103,201102,201026,191025,190805,190409,181120,200413,190827,190507,181221,191218,190307,190306,191228,190117,191019,190402,190218,200313,181130,200420,190826,190826,191212,200616,190111,201102,201022,200911,200904,191012,200513,190826,191220,200304,190322,190409,190425,191127,181204,200611,200227,191004,190306,200518,190415,190102,190720,191226,181102,200421,190107,181101,191112,200611,200222,190813,190225,190514,200102,181228,191223,200220,181106,200312,200214,190515,181218,181204,190321,231025,230828,230512,230518,230522,230524,230419,230410,220510,230216,221016,221016,221026,221026,221114,220908,221128,220902,220904,220523,220825,220825,220503,220503,220603,220423,220603,220315,220418,220111,220105,220224,211215,211204,211106,211011,211115,211126,211021,211008,210716,210914,210907,210904,210903,210820,210810,210718,210826,210716,210709,210729,210719,210703,210703,210622,210710,210610,210610,210618,210610,210408,210317,210422,210422,210406,210404,210329,210327,210326,210405,210310,210402,210220,210220,210220,201205,201103,201106,201113,201102,201026,190402,190215,190805,190507,181221,190827,190507,181210,181210,190307,190307,190624,191228,191019,190913,190913,190306,190306,190422,190624,190624,191217,190117,190402,190218,200313,181130,190402,190506,190506,181221,181220,181220,191212,191111,200929,200908,200908,201022,200928,201022,200911,190823,190628,190111,191012,200304,191220,190720,190720,190720,190322,190121,191004,190425,200227,190819,190819,181204,190215,191109,190619,191226,190619,190619,191018,190102,190723,190723,181109,190321,190107,190320,190320,190418,190416,190225,190418,191004,190813,191108,190514,200102,181228,181228,181228,191223,190704,190915,190219,181106,200312,190805,190415,190415,200518,200518,231123,231129,231231,231227,231224,231225,231221,231221,240108,231216,231221,231205,231129,231129,231125,231125,231129,231129,231115,231115,231115,231108,231108,231103,231020,231020,231020,231008,231007,231007,230925,230918,230918,230915,230915,230925,230925,230925,230814,230814,230809,230809,230725,230809,230725,230725,230725,230725,230721,230724,230809,230724,230711,230508,230529,230421,230421,230424,230402,230424,230418,230411,230411,230411,230323,230321,230302,230220,230224,230224,230131,230131,221208,221208,221015,221113,211219,220510,221019,220521,220524,221022,221022,221022,221024,221008,221010,221010,221115,221114,220908,221127,220915,220915,220915,220825,221009,220912,220508,220522,220522,220517,211220,220422,220314,220422,220602,220602,220718,220429,220718,220419,220419,220305,220415,220416,220111,220302,220302,220223,220326,220324,220207,220207,220121,220105,211210,220304,211203,220223,220223,220212,220212,211215,211215,211204,211204,211107,211122,211105,211105,211105,211115,211027,211011,211011,211018,211021,211017,211013,211007,211008,210925,210914,210820,210810,210718,210902,210810,210810,210826,210826,210806,210717,210716,210716,210726,210719,210709,210729,210802,210719,210719,210729,210728,210723,210719,210709,210720,210720,210703,210625,210710,210620,210620,210720,210622,210610,210609,210520,210519,210408,210408,200929,201023,201026,210517,210517,210507,210505,210403,210403,210328,210328,210405,210323,210324,210324,210305,210402,210102,210102,201207,201203,201106,201106,200912,190226,200704,200704,200413,190510,190510,200107,181210,200306,200306,190307,200418,190624,191217,191019,190802,190802,190402,190117,190117,200305,200204,200331,200331,200329,200519,190215,190218,190506,181220,201007,200106,190306,201023,200929,200929,200908,200906,200906,200929,200906,200912,200904,200904,190628,191220,191220,200304,200304,190322,190111,190720,190121,191004,190720,200326,200326,190425,190425,181204,181204,191004,191004,190215,190215,190619,200102,191023,190102,190321,190723,190704,190418,190107,200222,180925,181129,190813,190225,190225,190320,190320,190514,200102,200102,181228,200413,200319,200319,200331,200617,190913,190704,190915,200108,181106,190805,200306,200422,190415,190415,200220,191025,200420,200420,190507,190507,190913,191112,190409,190409,190624,190624,190623,190623,210420,210420,210304,210304,210127,201104,201104,201026,201026,200907,200907,191220,191220,191127,191127,200615,200615,200409,200304,200304,181109,181128,181101,191226,191226,191226,200305,200417,201014,200929,231227,231207,231125,231102,230912,230811,230811,230619,230606,230508,230302,230220,221208,221205,221204,220521,221023,221116,221023,221026,221025,221009,220523,220522,220718,220429,220718,220718,220718,220305,220223,211204,211017,211006,210914,210908,210826,210718,210626,210622,210710,210618,210518,210613,210408,210304,210102,220915,200928,200611,200704,181113,191218,200403,190827,190827,190118,190528,190402,190623,190218,190307,200106,191111,211108,211021,211017,211013,211008,211008,211006,210926,210913,210908,211115,210907,210903,210826,210826,210907,210902,210806,210806,210824,210810,210728,210802,210723,210719,210626,210720,210626,210610,210618,210420,210408,210611,210618,210518,210517,210512,210507,210520,210317,210422,210409,210404,210329,210303,210404,210304,210304,210304,210220,210117,210302,201208,201203,201202,201106,201029,200929,200928,200911,190510,191206,200304,181218,190425,190725,191127,190108,190121,181109,191109,200512,200511,200222,190227,190813,190809,200319,181214,191021,200604,200415,181102,201126,230418,230303,221022,220506,230906,230906,230906,230817,231216,231123,231123,231029,231102,230915,230418,230821,230616,230613,230616,230402,230307,230307,221023,220521,221019,221128,221202,220908,220904,220508,220419,211221,200704,191212,200511,190402,191228,191212,200107,190912,211108,211021,211105,210826,210826,210826,210823,210704,210622,210408,210511,210111,210406,210305,210121,201127,201027,200908,190218,190218,190912,190619,190108,190419,200403,190827,181106,190704,240104,240104,231229,231220,231220,231129,231125,231129,231124,231125,231106,231106,231029,231015,231008,230814,230705,230815,230710,230616,230606,230619,230619,230508,230518,230518,230523,230524,230529,230421,230421,230418,230309,230224,230220,230304,230131,221015,221019,221206,221205,220524,220521,221013,221013,221116,220510,221021,221010,221010,221010,221007,220908,221205,220907,221128,221009,220906,220915,220523,220720,220502,220503,0002626702,220522,220523,220802,220720,220315,0002621107,220429,220416,220424,220424,220302,211220,211220,211217,220304,220214,220214,211215,211215,211213,211204,211105,211105,211122,211011,211108,211026,211021,211017,211014,211017,210925,210926,211006,210914,210904,210902,210820,210810,210716,210709,210719,210728,210620,210609,210613,210520,210420,210408,210517,210512,210507,210505,210508,210520,210508,210318,210502,210422,210406,210404,210329,210327,210324,210305,210310,210310,210402,210402,210327,210303,210310,210302,210121,210121,210111,210102,201207,201203,201103,201127,201113,201106,201029,201022,201026,191025,190409,200617,181120,200413,200413,190510,181221,181225,191218,190827,200107,191111,190624,191019,190802,191217,190402,200204,181130,200329,200329,190506,200615,200904,190826,200103,191111,201023,200929,200908,200908,200910,200912,200911,200907,200904,190628,181124,190322,200513,191220,191012,190720,190720,191004,200326,190425,191127,191127,181204,190819,200611,191004,191109,190619,181109,191018,190723,191125,200912,190318,190318,190318,200610,200911,190418,181129,200222,191108,190813,190514,190913,191223,190915,190704,200108,200108,190805,190415,200220,200518,240104,231124,231205,231121,231123,231102,231029,231007,230912,230828,230821,230809,230721,230706,230512,230524,230420,230321,230131,230131,221014,221014,220524,221013,221114,220907,221127,221009,220906,220825,220522,220517,220315,220429,220423,190528,190325,190827,190409,190307,181130,190426,191022,211204,210718,210121,210111,201022,190729,190322,191025,190108,181113,190227,190528,190716,220104,231221,231121,231129,231108,231108,231103,230925,230920,230915,230914,230906,230810,230720,230415,230410,221016,221022,221008,220907,220902,0002626702,220418,220429,220421,220303,220104,220323,211216,211216,211213,211214,211214,211107,211106,211011,211115,211011,211018,211021,211017,211013,211007,210810,210718,210716,210709,210719,210823,210728,210723,210719,210703,210723,210710,210620,210620,210620,210612,210420,210408,210327,210618,210512,210507,210505,210508,210317,210509,210422,210407,210404,210329,210327,210326,210303,201205,201203,201013,201127,201113,201106,201029,201022,201026,191025,190507,190507,190226,200617,181120,200511,200107,181221,181210,190827,190307,190418,191111,190624,191019,190729,190402,191217,191205,200313,200305,181130,200329,190716,181220,181214,190823,200103,200617,200103,200103,190306,190306,201007,200929,200908,201022,200912,200912,200911,200904,190628,191202,200304,190322,200513,200728,191220,190725,191004,181124,200418,190425,191127,181204,200203,190819,200611,191004,190227,191109,190619,190322,190102,190320,190723,181109,191125,190415,190628,200222,181129,190813,190225,191019,190515,181228,190913,190704,190915,191111,190805,200312,190415,200214,190418,190307,210512,230519,221205,220720,220818,220315,220104,211203,200403,200420,190520,190704,191021,181130,190506,190426,211106,211013,210809,210728,210620,210509,210502,210406,210303,201207,201028,201028,200907,190307,200912,200319,191202,190102,190724,191017,191108,190322,190215,200302,200108,210327,210325,210325,210322,210323,210305,210403,210303,210302,210117,210117,210103,201208,201204,201204,201104,201104,201109,201106,201027,201028,201014,201026,201026,200908,200910,200928,200928,200907,190325,191205,191205,190426,200203,200203,191212,200811,181218,200410,181109,231205,191217,190118,190226,190306,190507,190218,190218,181113,200617,200616,211013,211008,211021,211007,210710,210703,210613,210622,210511,210507,210505,210508,210327,210326,210326,210324,201126,201202,201113,201113,201106,201102,201022,200904,200514,181218,200319,190318,190321,190321,200512,191017,200106,191127,200222,181228,231225,231220,231124,231106,231103,230912,230912,230914,230809,230725,230724,230606,230606,230517,230529,230419,230209,230323,230308,230308,221023,221013,221013,221010,220907,220825,0002626702,220522,220523,220212,211214,211213,231231,231225,231221,231216,231129,231108,231106,231103,231029,231020,231015,231007,230920,230918,230918,230925,230925,230823,230705,230619,230721,230711,230616,230510,230523,230524,230523,230402,230422,230419,230411,230410,230308,230308,230308,221014,221019,221204,220521,220521,220925,220925,221023,221007,221205,220907,220818,220818,220315,220603,220602,220314,220212,220111,220105,220224,211202,211214,211214,211204,211105,211021,211109,211115,211011,211018,211021,211017,211017,211013,211007,210720,210715,210907,210904,210904,210903,210820,210810,210718,210806,210709,210719,210728,210719,210703,210703,210512,210507,210505,210505,210502,210327,210326,210305,210402,210127,210111,201203,201127,201113,201102,201026,190226,181120,200413,200214,200617,200617,200418,200107,190307,190823,190729,181130,200519,200420,200420,201007,200908,200906,200911,190623,200304,200513,191205,190322,190121,181204,181204,200617,191112,190103,190720,191226,181129,200222,190813,190813,191111,190515,181217,200518,181113,200313,190426,190823,211206,211107,211012,211021,211013,210914,210904,210826,210718,210823,210622,210610,210618,210420,210507,210507,210407,210407,210302,190306,190723,190912,190321,191125,231020,231008,230920,230915,230828,230606,230519,230524,230419,230415,230209,220510,230131,221015,221016,221024,221116,221203,220907,221009,220906,220825,220825,220808,220422,220418,220414,220302,211221,211216,220121,220111,211213,210619,210619,210520,210512,200415,200518,200306,191223,181225,200603,191213,200413,190827,200418,191016,190623,191228,200616,200416,200103,200519,190729,181130,181107,200603,191226,181218,190620,200319,190823,201007,190623,211012,210704,211017,211008,211007,211115,210914,210908,210903,210826,210809,210718,210726,210720,210723,210704,210621,210625,210710,210710,210609,210517,210420,210519,210613,210420,210611,210519,210517,210512,210511,210509,210317,210407,210405,210404,210329,210327,210328,210325,210405,210305,210220,210303,210302,210103,210102,201202,201104,201113,201113,201027,201029,201027,201126,200929,201022,201026,201026,201106,200908,201023,200803,200907,200907,201014,200417,200103,191109,181228,190111,190227,190912,200415,191111,200513,191220,190729,180514,190725,181124,190819,200912,200611,200513,191223,200514,190528,200204,200102,190102,190103,190915,191109,200421,190716,181107,190813,200610,191023,191223,191104,200416,200416,190515,190515,190809,190809,191109,190702,200302,231231,231225,231221,231212,231121,231205,231125,231121,230512,230518,230321,230304,230302,230224,230216,221015,221019,221204,220521,221128,221022,221024,221025,221026,221026,221026,220522,220522,220808,220824,220517,211218,220509,220305,220415,220419,220415,220303,220303,220305,220121,211210,220304,231121,231121,230724,230420,221021,220720,220824,220324,220207,191218,190117,200422,200422,200220,200107,191022,200420,190623,200704,211105,210913,210718,210709,210729,210802,210704,210612,210404,210329,210303,210121,210111,201126,201102,201102,201023,200929,200904,200803,200907,190327,181218,190121,190322,190325,181124,190725,190619,191023,191127,181219,181213,190322,200604,191104,181124,191111,190528,181213,190226,190409,200413,200728,200617,190507,200413,190827,190307,190510,190624,190802,191019,190402,191226,200305,200511,200317,190409,181130,200519,191212,191021,200616,200417,190425,190426,200222,191012,211204,211022,211109,210811,211106,211027,211105,211105,211122,211027,211021,211026,211122,211021,211017,211014,210926,210904,210908,210904,210902,210820,210810,210806,210715,210709,210728,210723,210721,210703,210720,210710,210625,210621,210622,210618,210618,210612,210420,210609,210406,210408,210512,210507,210507,210408,210520,210508,210317,210302,210509,210422,210406,210404,210403,210326,210324,210305,210402,210303,210220,210310,210111,201205,201202,210117,201203,201203,201113,201202,210111,201013,201113,201113,201102,201022,200910,201109,200929,200908,200906,201022,201023,190623,190121,200410,190322,191220,191025,200313,181124,190425,190723,190819,200611,200409,190619,190724,191112,190723,181128,200421,190102,200319,190320,190418,200107,200911,190813,191004,181106,191212,190225,191018,200409,190520,190915,191223,200312,190704,190916,200312,200518,200220,231225,231207,231205,231207,231124,231129,231108,231102,231102,231015,231015,231012,231012,231006,230920,230912,230906,230906,230811,230705,230822,230809,230809,230711,230710,230606,230511,230512,230511,230522,230419,230419,230415,230209,230411,230308,230309,230306,230302,230224,230216,221014,221014,221208,221204,220524,221023,221008,221023,221116,221114,220510,221009,220908,220902,220823,220825,221009,220503,220503,220808,220522,220720,211219,220818,220818,220422,220317,220429,220421,220419,220325,220413,211219,211216,220104,220121,211210,211210,211210,220223,211204,220212,181113,190623,211115,200222,190416,191017,210704,210619,210515,231008,230711,230216,230304,220720,220303,211210,211210,200107,181129,190702,210726,210625,210305,201014,200320,190108,181109,190915,190619,190619,190624,230529,220422,191228,190826,210914,210515,201014,190108,190108,230925,190510,211126,210609,210328,201106,200304,190813,200420,190805,190528,201007,190510,191223,191218,181113,181210,181221,181214,190912,190319,190702,190623,190624,200107,200107,190116,200305,200312,191112,191212,190226,190226,190114,190109,190117,200418,200420,200413,191112,190307,181130,181130,200107,181220,190823,191016,190729,190623,190515,200313,200305,190319,211204,211115,211026,211202,211014,211006,210914,210907,210907,210826,210810,210826,210824,210726,210728,210719,210704,210622,210626,210619,210609,210420,210512,210515,210511,210509,210318,210318,210317,210302,210406,210329,210305,210324,210121,210305,210305,210304,210121,210121,210102,201202,201202,201203,201127,201104,201113,201113,201113,201102,201022,200929,201026,201014,201007,200908,200803,200911,190322,200417,200610,200106,191206,200203,190111,190119,191012,190114,200418,190322,200203,191112,181124,190425,181204,200203,190724,200416,200304,190109,200417,200513,200513,190415,191112,200103,191023,190121,200320,200320,190528,190107,190325,190716,200416,200403,190628,200319,191127,191112,181107,190620,181130,190227,181107,200420,190515,190515,190915,181214,200911,191108,200317,231227,231224,231216,231227,231124,231205,231124,231129,231106,231102,230918,230912,230914,230828,230823,230811,230706,230821,230821,230725,230724,230720,230720,230720,230616,230606,230511,230518,230523,230508,230424,230418,230411,230410,230402,230307,230306,230224,230216,221208,221014,221016,221203,220524,221021,221010,221010,221204,221008,221007,221114,221113,221128,220902,220904,220823,220915,220912,220507,220522,220823,220829,220818,211218,220423,220415,220429,220418,220415,220303,220303,220105,220224,220223,211203,211115,211218,220302,211217,220324,240108,231229,231223,231216,231216,240104,231220,231121,231125,231106,231029,231025,231020,231012,231025,231012,231008,231006,230925,230920,230918,230914,230823,230823,230810,230809,230809,230706,230809,230809,230726,230726,230710,230616,230424,230419,230410,230307,230304,230302,230220,230224,230131,221016,221208,221022,221024,221024,221008,221008,221023,220905,220907,220908,221127,220915,220915,220506,220823,220823,220502,220506,220912,220522,220522,220808,220808,211218,220517,220517,220517,220720,211219,220429,220718,220423,220423,220423,220305,220421,220317,220212,211106,220418,220416,220416,220415,220303,220302,211204,211217,220104,220326,220324,220121,220224,220224,220214,220223,220214,220214,211215,211215,211215,211214,211215,211204,211105,211027,211011,211108,211026,211014,211021,211021,211017,211014,211008,210925,211007,210925,211115,210908,210904,210902,210820,210810,210810,210719,210717,210823,210806,210709,210709,210719,210729,210729,210811,210728,210720,210625,210625,210710,210620,210620,210720,210610,210609,210520,210420,210408,210507,210511,210511,210511,210327,210517,210512,210507,210505,210508,210318,210407,210422,210406,210420,210404,210329,210327,210327,210326,210324,210305,201205,210402,210303,210302,210302,210121,210111,210102,201205,201207,201203,201103,190805,200518,190226,200617,200728,200413,181210,181225,191218,200107,190827,200418,190624,190307,191111,200615,190118,190402,191212,190218,200313,200329,190506,181220,190823,200417,200617,200103,190111,190306,191111,200906,201023,201023,200929,200929,200929,200908,200906,200803,200912,200911,200907,190628,200304,200513,200518,191012,190725,190720,190322,190121,191004,191112,200320,191127,190425,181204,200611,200417,200227,190819,190227,191109,190321,190506,190619,200102,191226,190102,190723,190723,200421,191125,190107,181109,190320,200610,200610,190416,190416,200222,190813,190813,190225,191019,191109,190515,200102,190913,191223,190704,191206,191223,191018,191018,200108,190109,190103,190619,190415,200220,200220,181102,200312,240104,240104,231231,231229,231225,231220,231216,231216,231212,231205,231205,231220,231123,231121,231123,231121,231123,231106,231103,231015,231015,230925,230915,230914,230828,230823,230823,230705,230705,230622,230822,230721,230720,230606,230606,230512,230512,230517,230522,230524,230402,230420,230420,230424,230422,230410,230323,230209,230323,230308,230304,230303,221014,230224,230220,230216,221208,221014,221015,221019,221019,221206,221203,221202,221204,221203,221008,221008,221025,221025,221008,220908,220906,221009,220824,220912,220502,220502,220507,220522,220824,220808,220517,211218,220818,220314,220415,220317,220418,220414,220415,220303,220104,211221,220323,220212,220214,211216,211214,211202,211214,211213,211214,181214,200413,200421,200704,201007,190802,181128,181120,181225,190319,181113,200403,200108,190102,191018,191021,200728,200222,200305,200420,190802,190307,181113,190507,200511,191212,200403,190402,190402,190506,181130,200331,181220,191205,191212,191212,190218,200417,200417,211204,211122,211126,211115,211126,211026,211105,211017,211014,211006,210913,210913,210903,210904,210826,210812,210809,210717,210824,210824,210810,210726,210704,210626,210622,210622,210420,210620,210613,210612,210519,210611,210408,210512,210507,210502,210318,210502,210422,210409,210405,210422,210403,210329,210404,210327,210326,210324,210323,210322,210305,210402,210302,210304,210220,210121,210121,210111,210302,210102,210102,201204,201203,201202,201202,201127,201113,201106,201102,201102,201102,201028,201028,201028,201022,201014,201026,200912,200929,200912,200928,200908,200803,200906,201022,200907,201023,200907,201023,190729,200222,200203,200803,181218,191212,190114,190215,190425,190823,200409,190320,200319,190819,200615,200513,200410,181109,191127,190723,190425,200227,191223,181101,200320,200421,190819,190109,190813,200409,190528,190318,181228,190716,191226,190520,191108,190913,200610,200610,200415,200420,181214,190813,200312,190416,181120,200317,191025,191025,200317,181221,181221,200704,200704,200403,200306,200420,190702,191025,191021,191217,181218,200204,191019,191022,190827,191212,211108,211102,211018,211022,211015,210902,210716,210723,210704,210720,210518,210318,210317,210407,210509,210422,210403,210329,210310,210103,201127,210111,201126,201022,201103,201013,201026,200928,200803,200911,190823,190623,190322,200329,190121,190912,200102,181128,181124,190318,190321,190528,200512,181219,181219,190528,190510,190809,190515,191112,181214,190219,190119,190114,200518,200518,231231,231207,231129,231115,231106,231008,230828,230705,230705,230725,230724,230726,230523,230420,230208,230306,221015,221206,221202,221114,221204,221009,220906,220503,220523,220523,220808,220602,220422,220414,220305,211221,220121,220223,211215,240104,231207,231108,231015,231007,230906,230809,230711,230424,230308,221206,221010,221009,220502,220603,220315,220305,220224,191112,200312,181220,191025,211204,211021,210908,210810,210704,210619,210507,210409,210323,210111,201127,200910,200904,190306,190628,190325,200421,200911,200320,200222,190913,190704,220302,211217,211220,211217,211217,211217,211217,211217,220212,211217,220104,211217,220224,211213,211213,211115,211126,211102,211022,211105,211018,211017,211014,210926,210913,210907,210904,210812,210806,210709,210809,210726,210811,210802,210728,210703,210626,210622,210718,210620,210609,210520,2 10520,210519,210612,210518,210518,210511,210508,210329,210328,210302,210220,210121,210121,210103,201208,201113,201102,210903,190108,200302,190321,190723,240104,231231,231229,231225,231225,231225,231221,231216,231216,231227,231207,231129,231129,231220,231220,231125,231124,231124,231129,231129,231108,231106,231103,231102,231029,231015,231008,230915,230912,230925,230828,230825,230823,230814,230822,230814,230814,230705,230619,230619,230821,230725,230725,230721,230721,230720,230710,230710,230710,230710,230616,230613,230512,230512,230522,230523,230508,230421,230421,230422,230419,230419,230418,230411,230411,230415,230323,230410,230209,230323,230323,230321,230321,230309,230307,230224,230131,221208,221205,221019,221206,221206,220521,221013,221022,221022,221024,221024,221010,221008,221008,221010,221010,220925,221021,221021,221203,221115,221114,220907,220907,221127,220912,220915,220915,220825,220825,220506,220506,220507,220507,220508,220508,220508,220522,220523,220808,220824,220517,220802,220517,220509,220423,220421,220422,220422,220315,220315,220305,220718,220429,220419,220419,220416,220303,220302,220325,211217,211220,220104,211217,211217,220326,220326,211221,220324,220121,220111,220105,220224,220304,220304,220304,211202,220214,211216,220212,211216,211215,211215,211214,211204,211107,211106,211106,211105,211105,211027,211115,211011,211011,211108,211108,211026,211026,211021,211021,211017,211017,211014,211014,211008,211008,210926,210926,210925,210925,210908,210908,210904,210903,210902,210820,210820,210810,210810,210718,210402,210717,210717,210716,210716,210716,210709,210709,210719,210719,210729,210729,210728,210728,210723,210721,210329,210709,210703,210703,210710,210625,210625,210718,210619,210620,210609,210609,210520,210612,210609,210520,210420,210408,210405,210517,210517,210512,210508,210327,210317,210502,210327,210502,210406,210404,210408,210405,210404,210329,210327,210326,210111,210324,210305,210402,210303,210302,201203,210121,210111,210102,201205,201203,210121,201013,201127,201113,201106,201102,190507,190226,200413,181210,200107,190827,200418,190624,190826,190307,191111,191019,190802,191217,190117,200204,190218,200329,190506,190506,181220,190823,200417,190306,201023,200929,200908,200906,200912,200911,200907,200907,200904,200906,191012,200304,191220,190111,190725,190322,190322,190121,191004,191004,191004,200418,190425,190425,181204,200203,190819,190215,191109,191109,190425,190619,190102,190102,200302,200302,190321,190321,191018,190108,190320,190723,190107,190416,181129,200222,190813,190225,191108,190515,181228,191223,200102,200106,191205,200204,200214,200220,190915,191127,190913,190704,181106,200108,200108,190109,190109,190805,190805,200312,200220,191025,191226,190415,190319,190419,181109,190225,190515,190226,200704,190510,181210,190912,200420,191218,190319,191112,190624,191019,190402,200511,191228,200305,200305,200317,190218,190729,181113,200331,190506,200617,181220,191212,190826,191111,190306,190111,211204,211022,211105,211115,211011,211115,211102,211022,211013,210926,210903,210902,210806,210806,210802,210717,210717,210809,210709,210726,210726,210811,210728,210703,210621,210622,210618,210710,210610,210612,210408,210520,210518,210515,210515,210507,210520,210505,210508,210318,210407,210502,210404,210420,210403,210328,210326,210326,210220,210127,210111,210103,201208,201203,201208,201127,201126,201204,201103,201027,201022,201022,201014,200929,200910,200803,200803,200907,200904,191016,191213,190322,200514,191025,190720,190114,190425,181218,191127,200413,200615,200409,200319,200203,190215,190619,190619,200102,200204,190724,190325,200512,191023,190723,191125,190107,181109,200319,190422,181205,190528,200319,200222,190813,200611,190225,200417,190320,190520,181228,190915,200415,191223,200313,190915,200108,181106,190109,200317,191212,191226,220121,220304,211213,211213,200413,200415,191218,201007,200331,200413,200108,190912,200313,200519,211206,211115,211022,211018,211018,211007,210926,210908,210820,210809,210716,210729,210728,210420,210618,210519,210612,210408,210508,210512,210509,210318,210317,210403,210328,210328,210323,210402,210302,210220,210102,201013,201104,201022,201014,200929,200907,200910,200907,201023,191012,191220,200410,191127,200320,190716,191112,200611,200222,190915,200204,190813,190813,191226,191023,190805,200518,230711,230606,230523,230420,230420,230421,230402,230421,230402,230209,230224,230220,220524,220524,221010,221026,221026,220825,220915,220523,220912,220503,211218,211218,220423,220429,220415,220302,220326,220323,220105,211216,211214,211213,200413,200421,190510,200511,200511,181210,200221,200415,190912,200704,181120,181120,200420,190319,191217,200102,190624,200107,190117,191019,200305,200305,190409,191217,200204,190218,181130,200107,190729,200519,200331,190506,190506,190507,181220,181228,200728,200616,191212,191212,190826,200610,200417,200106,200222,190306,200103,190628,200326,191212,211202,211210,211107,211105,211027,211027,211027,211011,211108,211102,211022,211126,211017,211014,211008,211006,210926,210913,210903,210806,210902,210907,210812,210826,210826,210715,210824,210806,210729,210726,211021,210811,210728,210802,210723,210719,210823,210703,210625,210622,210622,210618,210619,210329,210612,210612,210511,210408,210519,210518,210512,210511,210505,210508,210508,210318,210408,210502,210422,210406,210409,210409,210404,210408,210403,210328,210326,210404,210324,210310,210405,210402,210220,210310,210127,210111,210103,201208,201203,201208,201127,201126,201126,201106,201102,201029,201022,201014,201023,201102,200906,200908,200908,200803,201022,200512,200519,200911,190322,191220,191012,190720,200514,191004,181128,181129,181218,191127,190119,190119,200418,190425,200320,200203,191021,200911,190215,200410,190619,191125,200103,190114,190320,200302,200416,190321,191023,191023,181109,200512,200512,190723,190723,191202,190107,200912,200611,200319,190320,190320,190422,190422,200306,181205,190416,200409,200604,190520,191108,190912,191223,181228,181228,190704,190912,200420,200420,190915,190915,200108,200317,190103,190619,190415,200317,190802,190913,191223,190402,231229,231225,231221,231212,231212,231207,231125,231121,231125,231124,231106,231108,231102,231025,231015,231008,231008,231007,230925,230925,230920,230918,230918,230906,230906,230825,230906,230706,230705,230622,230815,230725,230721,230711,230706,230706,230613,230613,230511,230512,230519,230518,230523,230524,230524,230421,230424,230424,230415,230415,230415,230209,230321,230309,230306,230304,230303,230224,230216,230216,221208,221015,221019,221206,220521,221023,221022,221024,221025,221116,221026,221026,221114,220905,221203,220908,221128,221009,220915,220904,220825,220912,220503,220506,220508,220808,220829,220829,220517,220818,211220,220718,220603,220317,220314,0002621107,220429,220416,220415,220416,220418,220303,220305,211220,211221,211219,220325,220111,220111,220111,220304,220223,211215,230721,230512,230518,230524,230411,230303,221206,220524,221010,221116,221113,220720,220908,221009,220915,220825,220912,220522,220824,220324,220207,211214,200413,191228,211012,210913,210824,210729,210611,210512,210409,210326,210310,210303,201202,201106,200803,190107,181124,200513,190108,190409,200615,191109,191217,191127,190318,200227,191004,190227,190716,231227,231216,231229,231231,231123,231216,230622,230619,230619,230619,230622,230821,230726,230508,230220,220904,200305,200305,190418,200312,200511,200728,200904,200331,211204,211014,211015,210925,210820,210812,210729,210721,210703,210610,210610,210408,210408,210508,210409,210325,210326,210121,210117,210302,201204,201104,201022,200929,200908,200302,200222,191025,190111,190724,190823,190725,200616,190912,200102,200203,190528,181109,190912,181024,181106,200325,200416,190725,220224,211213,190117,181113,210420,191127,200227,181130,211026,211026,210719,210720,201205,231015,200518,190102,190724,191212,191212,190117,190827,200108,190318,190402,190117,190826,190409,181220,200319,190422,200222,200908,201023,200904,200304,190121,190425,200513,200513,181204,191127,190819,200611,190227,200227,191109,200102,190724,200204,191018,191018,181128,191212,190107,181129,190409,190225,190318,200222,190520,181214,190913,191112,190704,191223,181106,190415,190109,210515,200422,190226,200312,200326,200331,200421,191025,181128,190118,181210,181213,181213,181213,200305,200214,190912,190912,200603,190510,190319,190528,190528,191213,191218,190109,190802,200615,190624,190823,191112,191112,190117,191114,191019,200305,200326,190507,200511,190623,190623,190624,191205,191205,191016,191019,181129,181204,181130,191228,191228,200616,190729,190729,190506,191212,190218,190218,190826,181130,200610,200331,200417,190426,190306,200518,200519,190111,190111,200320,191212,190823,190823,190819,190819,191022,200304,211109,211106,211027,211115,211026,211022,211017,211018,211014,211008,211008,211006,210926,210926,210903,210913,210907,210902,210824,210728,210820,210826,210812,210715,210729,210716,210729,210726,210820,210802,210823,210811,210811,210729,210723,210723,210723,210719,210728,210518,210704,210703,210626,210625,210625,210622,210618,210609,210619,210618,210710,210520,210520,210612,210408,210612,210408,210518,210517,210512,210511,210511,210511,210507,210507,210509,210508,210502,210502,210317,210420,210502,210422,210409,210405,210404,210403,210404,210403,210329,210328,210327,210326,210324,210310,210305,210402,210402,210303,210303,200421,210127,210302,210111,210102,210102,201207,201203,201202,201203,201013,201127,201106,201102,201127,201113,201103,201106,201103,201102,201029,201022,201022,201014,201014,201023,201102,201007,200929,200929,200910,200908,201028,200908,190813,191223,191205,191218,200928,200928,200803,200803,200417,200417,191114,191012,190915,200329,190510,200514,190826,200222,200222,191220,190720,190720,190720,181218,191112,191127,190114,190109,190119,190109,190915,191004,190520,181128,190425,190723,200513,191220,190306,200326,191125,200319,190121,190819,200329,190619,200409,200214,200417,190108,191226,200417,190227,200512,191023,190724,190723,191125,190109,190327,200304,181109,190318,190318,190318,190307,190422,190422,190422,200421,191018,200611,200616,200610,200611,190716,200227,200312,200319,200102,181205,191125,191125,200221,200222,200222,190813,190813,200409,200409,190119,200610,190416,190809,190307,190520,190119,190915,190219,190916,190913,190916,190510,190320,181228,181228,191223,190418,200417,200418,200417,200409,200420,200421,190915,181106,181106,200409,190915,190109,190321,190109,191108,190702,191108,190619,190415,191114,191021,200317,190805,200518,200518,190515,200107,200103,200108,191112,200317,240116,240104,231231,231227,231225,231221,231216,231216,231207,231205,231129,231220,231124,231121,231123,231124,231106,231029,231029,231025,231015,231008,231008,220823,230918,230925,230912,230828,230823,230622,230821,230821,230809,230725,230725,230724,230721,230720,230724,230720,230720,230710,230710,230616,230616,230710,230616,230606,230511,230512,230512,230523,230529,230421,230424,230424,230415,230415,230415,230209,230321,230309,230304,230306,230303,220825,230307,230224,230224,230216,221208,221015,221014,221019,221208,221206,220524,221128,221128,221013,221024,221025,221026,221026,221114,221113,221113,221128,221127,221203,220912,220912,220904,220523,220915,220503,220507,220522,220808,220808,220818,220829,220517,211218,220802,220423,220421,220602,220317,220317,220429,220419,220302,220302,220303,220302,211220,211217,220325,220324,220121,220111,211210,220105,220224,220214,220214,211216,211202,211203,211202,211115,231216,231212,231205,231207,231220,231124,231124,231123,231121,231121,231123,231129,231129,231115,231106,231103,231029,231029,231020,231020,231020,231015,230524,220315,190415,190528,190802,200418,190623,190218,191016,211206,211115,211106,211027,211115,211018,211015,211007,210908,210902,210809,210824,210726,210726,210704,210704,210626,210622,210703,210704,210613,210420,210610,210517,210507,210502,210508,210406,210403,210323,201204,210117,201027,200928,200928,200616,191004,190419,190725,190103,200421,190716,200413,220522,211218,220818,220315,220305,220602,0002621107,220418,220415,220303,220326,211221,220207,220304,220214,240116,240118,231205,231207,231123,231025,231007,230914,230914,230710,230616,230522,230402,230410,220825,221206,221204,221013,221026,221026,221127,220825,220825,220825,0002626702,220808,220802,220802,220315,220424,220303,220325,220104,220121,220223,211214,190805,200420,200728,200617,200413,200511,191112,191112,200323,190409,200325,190219,200416,181221,200403,190426,190419,190419,211115,211021,211015,210908,210826,210809,210826,210810,210811,210719,210704,210618,210610,210520,210420,210517,201202,210111,210406,210403,210403,210328,210328,210402,210220,210302,210121,210302,201203,201013,201104,201102,201027,201023,200929,200904,200928,200911,190823,190628,191016,181130,190322,190515,191127,190823,191127,190111,200513,181204,190916,190725,191025,190121,190121,200326,191023,200409,200221,200611,200611,200221,190318,191112,200102,191112,191108,181228,200416,200409,190716,191021,191021,200518,190725,190219,221008,220212,211214,211202,211213,190515,191218,201007,190320,191112,181204,190514,190415,190912,190319,181221,190520,190624,191218,190915,190916,200222,190117,200331,200511,191228,191228,200204,191017,190802,200331,200107,200617,191111,200418,181210,190826,190823,190422,190402,211206,211202,211105,211115,211027,211011,211126,211026,211013,211013,211008,211006,210926,210907,210820,210812,210812,210902,210806,210717,210709,210810,210729,210811,210723,210719,210728,210703,210802,210703,210704,210618,210610,210610,210519,210612,210612,210408,210318,210317,210408,200511,210407,210409,210405,210328,210327,210310,210310,210402,210304,210302,210220,210220,210127,210117,210117,201207,201207,201013,201106,201126,201106,201102,201102,201013,201028,201022,201013,201023,200617,201007,200803,201023,200907,200417,190823,200317,191016,190306,190322,181128,181218,190121,190720,190322,200616,200513,191127,200326,190621,200302,190121,190823,190327,200417,190108,181120,190724,200203,200320,191213,190724,190422,181219,191023,190723,181109,190628,190528,181129,180928,200222,200102,190215,200417,200421,191017,190225,200421,200204,200610,181106,190620,190915,190620,200108,200108,190809,200604,200604,190119,190320,190510,190912,191112,200107,200331,181130,181130,190623,190114,191228,211206,211215,211026,211021,211017,211008,210926,210908,210908,210904,210820,210820,210726,210718,210716,210802,210810,210728,210703,210625,210622,210710,210619,210613,210420,210517,210509,210406,210404,210402,210111,201208,210117,201202,201029,201026,200910,200912,191114,191016,190119,181218,190121,200410,191205,200704,190327,190425,190716,190219,200420,181214,190416,190913,191223,191223,181120,231231,231205,231123,231108,231103,231102,231020,231029,231006,230920,230918,230823,230906,230705,230821,230809,230720,230616,230510,230508,230523,230402,230422,230209,230209,230323,230307,230306,220510,230220,221208,221019,221203,221128,221010,221024,221024,221021,221114,220908,220906,220825,220507,220522,220808,220829,220517,220517,220718,220429,220414,220305,220303,211217,220323,220323,211203,211213,190515,200403,200422,200422,200617,200107,200304,190226,181113,181221,191021,190409,191217,190724,200323,200312,200306,190802,190422,200511,200331,190507,190327,190111,190111,190915,191228,191213,200904,190916,191111,190827,200421,190623,190409,181130,190426,190716,200106,200107,181218,191016,190402,190823,200313,211206,211109,211122,211106,211106,211027,211126,211012,211115,211102,211026,211108,211014,211015,211012,211007,210926,210913,210914,210925,210913,210925,210914,210826,210812,210820,210809,210810,210824,210716,210810,210823,210811,210726,210720,210802,210704,210626,210622,210704,210619,210619,210618,210612,210420,210611,210518,210517,210512,210507,210509,210502,210317,210407,210409,210405,210403,210329,210327,210327,210325,210323,210305,200906,210302,210220,210121,210103,210102,201204,201202,201104,201127,201113,201102,201027,201028,201014,201026,200929,201007,200910,200906,200928,200912,200907,200906,200410,200519,200317,190225,190422,190306,200326,200329,190109,190215,190109,200616,190425,191025,191212,190725,190725,191111,181120,191212,190409,181124,190621,200214,200103,181219,200302,200416,200513,190416,190225,190419,191202,181225,200320,190724,200615,200611,190318,191018,190813,190716,181107,181107,200409,190528,200102,191112,181129,190320,191226,191226,200421,190322,190415,200319,200107,200610,190416,181225,191108,190624,190624,190704,191218,200415,200220,190913,200415,190116,231231,231225,231231,231227,231121,231205,231124,231115,231121,231106,231029,231102,231029,231015,231008,231007,230920,230915,230925,230823,230814,230624,230815,230725,230721,230720,230706,230606,230511,230512,230518,230523,230529,230421,230424,230418,230410,230209,230321,230309,230306,230306,230303,230224,230216,221014,221015,221019,221204,220521,221023,221013,221203,221115,221025,221026,221116,221007,221113,220905,221128,221009,220915,220904,220902,220823,220915,220503,220506,220508,220522,220808,220824,220517,220818,220317,220818,220423,220602,0002621107,220429,220429,220416,220305,220302,220302,220326,211221,220324,220207,220121,220111,220304,220223,220223,211203,210515,210422,201022,190515,220908,220718,190226,181221,181128,181219,190108,190107,190102,190107,181228,190116,181120,190507,201023,191025,190805,191226,200420,190623,191228,200313,211107,211115,211018,211126,211006,210903,210810,210720,210720,210621,210420,210408,210403,210102,201109,201007,190425,190119,190321,190107,200222,190318,231227,231123,231216,231216,231220,231216,231212,231207,231205,231220,231207,231129,231106,231106,231103,231029,231024,231008,231007,230920,230918,230914,230828,230825,230823,230711,230821,230809,230724,230721,230720,230710,230512,230519,230522,230522,230524,230421,230421,230424,230419,210728,230323,230323,230308,230307,221023,230303,220521,221206,221128,221013,221013,221013,221022,221022,221025,221013,221010,221021,221021,221021,221026,221114,221115,221114,221009,221009,220906,221009,220904,220825,220720,211218,220603,220718,220423,220421,220603,220302,220207,220224,220212,211213,190416,200518,181221,190319,211122,211107,211027,211108,211202,211006,210913,210806,210717,210709,210811,210704,210618,210612,210611,210511,210407,210329,210324,210402,210121,210102,201202,201029,200928,200911,200907,200417,181213,200304,200203,190621,190915,200611,190107,191018,190716,200420,190619,200221,231224,231205,231205,231106,231029,231024,220823,230914,230829,230825,230809,230518,230522,230523,230524,230209,230131,221205,220521,221204,221010,221021,221128,221127,221127,220825,220720,211218,211218,220509,220422,220315,220421,220303,211216,220214,220223,201007,240104,240104,231212,231106,231103,231024,231008,230925,230918,230823,230622,230724,230606,230606,230519,230420,230418,230410,230308,220510,230216,221019,221023,221008,221008,221026,220908,220906,220906,220906,221009,220915,220825,220825,220720,220503,220502,220508,220522,220602,220315,220718,220423,220603,220416,220421,220421,220303,220207,220121,220223,220212,211213,200422,190226,200906,200906,200413,190215,200413,200108,181113,200403,191218,191021,200422,200422,191112,191112,190418,190624,190118,191019,200305,181113,190624,200326,190507,200227,190218,190729,191111,191022,200317,190307,181130,181130,200417,200417,200102,211206,211107,211106,211115,211115,211012,211021,211126,211017,211015,211014,211013,211007,210925,210914,210908,210908,210908,210902,210826,210809,210717,210716,210824,210709,210810,210729,210728,210621,210621,210621,210620,210610,210612,210612,210612,210611,210518,210408,210408,210511,210508,210502,210317,210407,210405,210405,210403,210328,210327,210327,210325,210405,210305,210304,210303,210303,210220,210121,210111,201208,201208,201202,201127,201113,201106,201029,201027,201014,201102,200910,200908,200906,201023,200417,191228,190215,190218,190729,190322,190322,200514,191012,190720,190425,190215,200320,200513,200514,191025,190108,190121,190227,200102,190108,190327,190723,200421,191004,190418,190418,190528,190620,200912,200610,190318,190102,190915,190225,190809,191125,190219,190418,190102,190620,191018,181214,190913,190416,190218,190219,200422,190520,200317,200518,191226,231124,231125,231106,230912,230706,230822,230822,230710,230510,230524,230422,230323,230306,220521,221203,221208,220323,220223,220212,211213,231212,230422,230308,220907,220908,220908,220303,220303,220302,200413,211012,210907,210420,200908,200326,190619,190227,190227,181219,210407,191206,231212,231020,230721,230420,230309,221016,221026,220825,231129,231123,231015,231006,230912,230823,230616,230522,230524,221015,221008,220906,220915,220503,220315,220315,220424,220302,220421,211202,211213,211115,211122,211026,211102,211021,211017,211008,211013,211014,210914,210820,210810,210719,210902,210717,210716,210709,210719,210729,210723,210719,210703,210618,210710,210609,210520,210420,210408,210618,210512,210507,210505,210508,210317,210407,210422,210326,210324,210305,210402,210302,210121,210111,210111,201205,201203,201013,201127,201113,201103,201029,201014,191223,190507,200518,200617,181120,200413,200511,190416,190827,200514,200107,200108,190624,191019,190729,190402,200519,191228,191217,190117,200204,200305,181130,200329,190506,200616,200904,190823,200222,191111,200911,200907,200904,190628,190111,200304,200513,191019,191220,191012,191012,190322,191127,200313,191004,190725,181124,200326,190419,190425,190425,181204,190819,200611,200108,191004,191109,190619,200102,190102,191226,190321,190704,190723,190723,200421,191125,181109,190416,181129,200222,190813,191108,191109,190515,181228,191223,191021,191021,191023,190913,190915,190702,190704,190915,200108,200312,190805,181109,190415,200214,211218,210329,200906,210925,210925,210811,210111,200403,200409,210512,190624,190802,211204,211122,211102,211011,210812,210811,210811,210723,210802,210612,210408,210405,201026,210127,201208,201126,201106,201027,191025,191023,191004,190723,231231,231225,231221,231121,231025,231015,231008,231007,231006,230912,230925,230823,230810,230624,230606,230512,230518,230523,230508,230424,230415,230410,230308,230303,230216,221014,221015,221204,220524,221013,221025,221007,221205,221128,0002626702,220506,220507,220522,220808,220824,220818,220818,220718,0002621107,220419,220415,220305,211220,211221,220105,211210,220223,220214,211216,211202,211206,211210,211109,211109,211107,211109,211027,211011,211011,211122,211122,211122,211122,211108,211102,211102,211022,211022,211015,211018,211014,211013,211013,211007,210926,210925,210914,210907,210903,210903,210903,210824,210806,210806,210717,210806,210806,210729,210823,210809,210810,210823,210904,210823,210721,210719,210802,210721,210721,210723,210723,210621,210625,210619,210613,210519,210611,210609,210520,210518,210515,210511,210511,210505,210520,210509,210509,210317,210317,210409,210409,210405,210405,210403,210422,210403,210328,210327,210325,210305,210117,210103,210103,231227,231227,231227,231227,231227,231225,231216,231231,231231,231216,231216,231221,231205,231205,231205,231124,231124,231124,231124,231115,231121,231123,231123,231129,231129,231108,231103,231103,231015,231015,231015,231012,231012,231012,231012,231008,231007,230920,230918,230918,230912,230912,230906,230906,230906,230828,230825,230825,230811,230810,230811,230705,230705,230705,230705,230630,230809,230809,230724,230724,230721,230711,230725,230710,230710,230613,230606,230606,230606,230606,230606,230606,230510,230512,230512,230518,230524,230508,230508,230508,230422,230419,230419,230415,230415,230411,230415,230402,230411,230209,230209,230321,230321,230321,230309,230309,230306,230306,230302,230302,230224,230224,230131,230131,221014,221014,230131,221208,220510,220510,221204,220524,220524,220524,221202,221203,221203,221010,221116,221024,220720,221025,220720,221021,221021,221115,221115,221113,221114,221114,221115,220907,221127,221127,220906,220906,220906,220906,220906,220906,220825,220825,220823,220825,220912,220912,220502,220502,220506,220506,220507,220523,220523,220424,220424,220808,220824,220824,220824,220824,211218,220818,220818,220423,220314,220314,220314,220315,220315,220315,0002621107,220418,220418,220418,220418,220418,220416,220424,220303,220303,211220,220325,220325,211220,220325,220325,211220,220325,220325,220325,211219,211219,211217,220111,211213,211213,220214,211216,211216,211216,211214,211203,231220,231212,231124,231029,231024,230914,230810,230711,230606,230511,230511,230508,230216,230308,230216,220524,221008,220908,221009,220823,220825,220720,220506,220517,220523,220718,220603,220414,220302,220302,220207,220223,211204,211214,191104,181213,181218,191022,200305,211107,211109,211018,210904,210904,210902,210826,210810,210718,210520,210515,191114,190327,181218,190621,191226,190322,200421,191218,190912,200728,200403,181228,191112,200302,190118,190409,191226,200413,190823,190826,190802,200323,181113,181113,190416,200204,190219,200331,190506,190826,200103,200418,200518,181220,190805,190422,190515,200106,191016,190322,191206,200203,200326,211210,211215,211109,211109,211027,211011,211115,211108,211022,211015,211014,211013,211006,210926,210913,210907,210903,210806,210824,210812,210806,210717,210824,210709,210726,210811,210907,210723,210719,210720,210704,210626,210621,210618,210610,210612,210519,210408,210520,210518,210518,210612,210511,210317,210505,210508,210318,210317,210422,210409,210405,210403,210328,210326,210328,210324,210322,210304,210304,210304,210117,210302,210127,210117,210103,201208,201203,201204,201013,210111,201127,201126,201204,201106,201027,201028,201014,200929,201013,200929,200910,200803,201014,200803,201022,200906,200514,181205,190215,190720,181128,200103,191205,190111,200329,190425,190823,191109,200811,191212,190422,200410,200410,181219,181109,190619,200302,190321,190321,200512,191023,190102,190415,190723,200512,190422,191202,200912,200222,190813,200518,190321,190416,190520,191109,191017,181214,190102,191212,191212,190520,190704,191223,200329,200108,181106,200604,190416,181220,190916,190805,190624,190219,190514,190510,231227,231225,231221,231212,231207,231129,231125,231121,231115,231106,231102,231025,231015,231008,230814,230925,230920,230906,230823,230814,230622,230622,230815,230725,230721,230724,230711,230706,230613,230511,230511,230508,230523,230421,230424,230415,230410,230209,230321,230304,230302,230224,230216,221014,221015,221019,221206,220521,221128,221013,221115,221025,221026,221026,220905,221128,221009,220915,220904,220825,220912,220503,220506,220508,220522,220808,220829,220517,220802,220818,220423,220423,220317,220602,220315,220429,220419,220415,220305,220303,211220,220104,220104,220324,220207,211210,220105,220304,220223,220212,211122,211115,211126,211108,211026,211026,211015,211017,211013,211007,210925,210908,210810,210826,210824,210812,210812,210809,210820,210718,210824,210806,210810,210823,210728,210802,210520,210318,210404,210403,210328,210302,210302,210103,201203,201127,201127,201113,220111,190528,211012,220105,211203,211214,190510,190912,181214,190306,190823,190819,181213,191218,191218,190809,190319,200403,200403,191021,191021,191114,191217,190118,190118,190802,200214,190226,190226,190226,181113,190506,200511,200227,200331,181221,181221,191206,191213,191213,190826,200616,200617,190307,190307,190623,190111,200313,190813,200514,200514,200511,200203,200203,200313,211206,211102,211021,211115,211115,211102,211102,211022,211022,211015,211018,211018,211013,211006,211006,210926,210914,210907,210903,210824,210717,210824,210826,210716,210716,210823,210626,210904,210721,210802,210802,210621,210621,210626,210626,210619,210613,210519,210511,210511,210512,210509,210318,210318,210317,210409,210403,210329,210328,210325,210325,210305,210304,210304,210127,200910,210117,210103,210103,201208,201204,201204,201202,201104,201126,201106,201027,201027,201028,201014,201014,200906,201026,201007,200908,200908,200910,200928,200928,190325,190325,191127,191205,181128,181128,200326,200326,200513,190426,190426,190119,200616,181218,181218,190823,191004,191012,190416,200410,200410,190620,190620,190108,200302,200302,200302,191023,191023,191023,190321,200512,200518,191202,200319,200319,190107,190121,200227,200227,200611,190318,190318,200409,200403,181217,181217,190225,190225,190103,190119,191218,191109,181228,200221,190916,190916,200420,191017,190813,190813,190704,200103,200313,190716,190716,200317,200317,200325,190416,190416,190416,190416,181120,181120,200604,200603,200603,190702,190702,200331,190510,190510,190515,190515,231231,231225,231224,231225,231212,231205,231205,231205,231220,231220,231124,231115,231124,231106,231106,231102,231025,231025,230918,230914,230914,230920,230906,230825,230720,230720,230724,230711,230711,230726,230710,230710,230711,230613,230613,230613,230510,230619,230512,230510,230519,230524,230524,230508,230422,230422,230517,230424,230415,230415,230209,230209,230321,230309,230304,230302,230131,230131,230131,230131,230131,221014,221015,220510,220510,220510,220521,221205,221128,221013,221116,221010,221010,221010,221010,221113,220720,220510,220510,220906,221203,220915,220915,220902,220902,220823,220825,220506,220506,220508,220808,220808,220829,220517,211218,220509,220802,220509,220305,220314,220317,220415,220415,220429,220429,220419,220424,220415,220302,220302,220325,220325,220104,220111,211220,220413,220323,220323,211216,211216,211202,190716,230421,200617,200617,200320,200320,200420,200511,191217,200107,200107,200108,190409,181120,190507,181221,190322,200422,190325,190528,191213,190827,200615,190422,200403,200214,190111,200615,200929,200616,190802,190515,190325,200519,190729,190426,190623,190623,181124,181130,181113,181128,200319,200603,200106,181213,200305,200313,191228,200320,190823,211206,211109,211108,211106,211107,211106,211027,211027,211115,211102,211126,211026,211105,211126,211015,211012,211017,211007,210913,210926,210925,210914,210826,210809,210806,210812,210820,210726,210729,210619,210626,210908,210728,210720,210811,210728,210704,210626,210622,210704,210620,210619,210618,210618,210612,210420,210611,210518,210517,210512,210512,210507,210509,210502,210502,210520,210317,210407,210409,210405,210403,210329,210329,210327,210325,210323,210305,201013,201104,210327,210111,210121,210302,210220,210121,210103,210102,201204,210121,201104,201127,201113,201102,201027,201014,201026,201102,201007,200910,200906,201028,200912,200929,200616,200616,200910,200417,191205,191205,190218,190107,190215,190419,190306,200616,190218,191004,200417,190327,190506,191220,190322,190621,200221,200103,191127,200306,200615,181120,200410,181220,181219,200513,200320,190725,200409,190121,190415,190227,190419,190528,181225,190102,190318,191109,191226,191018,190716,191112,181105,190418,190416,200610,200610,200421,200421,200222,190320,190418,190219,200325,200319,181214,190102,181113,181129,181214,191108,200413,200415,191021,190913,190624,190515,190725,231231,231225,231221,231216,231207,231205,231125,231124,231124,231106,231103,231025,231020,231008,230825,230925,230915,230925,230622,230815,230725,230721,230720,230720,230706,230613,230511,230512,230519,230523,230523,230524,230131,230421,230508,230424,230415,230415,230209,230306,230321,230323,220905,230309,230304,230303,230224,230216,221014,221015,221019,221019,221206,220521,220521,221208,221022,221115,221007,221007,221008,221008,221026,221114,221128,221113,221113,221009,220912,220912,220904,220502,220825,220503,220503,220506,220508,220522,220808,220824,220517,211220,220818,220423,220317,220602,0002621107,220429,220429,220416,220421,220303,220302,220302,211220,211220,220104,220207,220104,220111,211214,230510,230512,211217,220326,220104,181220,211018,210625,210329,210304,200803,191114,190425,190416,190215,191019,190826,211102,211022,211018,210726,210802,210723,210326,210326,210111,210103,201208,201203,201013,190819,191021,191023,191017,190915,240104,231229,231221,231212,231205,231121,231129,231106,231029,231025,231020,231012,231006,230914,230912,230828,230705,230220,230216,221015,221206,221025,221021,221116,220908,221127,220522,220915,220517,220825,220503,220507,220508,220808,220802,220509,220423,220718,220419,220303,220305,211221,211221,220326,220324,220121,211210,220105,220223,220214,211213,231015,230724,230724,221016,200331,200512,200512,191202,191202,200319,190121,190121,181205,200320,200611,190225,181228,181228,190916,200416,200422,190415,181213,191017,190827,200418,190623,200519,200616,200417,191212,211206,211122,211022,211027,211122,211122,211008,210926,210914,210903,210826,210718,211021,210625,210619,210612,210611,210517,210515,210502,210407,210405,210327,210405,210303,210121,210111,201205,201113,200908,190621,190327,190725,180629,190419,190416,190318,190724,191226,200421,190116,200319,19081 3,200513,191223,231225,231121,231205,231123,231102,231008,231007,230906,230622,230725,230721,230706,230613,230511,230517,230418,230402,230309,230304,230303,230224,230216,221208,221015,221023,221009,220904,220506,220823,220522,220829,220517,220818,220317,0002621107,220421,211221,220121,200420,231008,230918,230622,191213,191017,230828,230821,230706,221021,221021,221116,220912,220422,231025,231023,230914,230914,230726,230415,230220,220414,220325,220223,200422,190915,190226,200519,190724,190724,181113,191212,210826,210811,210618,210613,210325,210326,210325,210310,201202,201126,201104,201026,200908,190322,190322,181218,200304,191025,190415,200421,190724,200421,200313,200320,181120,200422,200617,200422,200511,191017,191223,181113,181221,181221,181228,200305,200603,190507,190109,190415,190415,200413,191213,191213,191021,190802,190418,190418,200108,200403,200107,191114,190623,200326,191228,190321,190729,190325,200511,181113,200317,200603,190426,181218,191021,200912,200616,190620,211206,211122,211107,211027,211018,211026,211017,211008,211007,210820,201023,200410,190218,190422,190422,190320,190912,181102,191213,191112,200304,200102,190327,200513,191220,200306,190725,181124,200103,181219,181210,181219,190117,190409,200410,200317,190419,200203,190819,190819,200611,200102,200410,181220,181221,190102,190416,190528,190819,191213,191109,191226,190318,200421,190716,200326,200319,200107,191125,190321,181107,181107,190111,191212,200604,200610,180814,190813,190418,200513,190515,190809,181214,181225,190702,191114,191108,200413,200102,191111,231229,231221,231212,231207,231125,231121,231115,231125,231108,231102,231025,231020,231020,231007,230925,230918,230825,230622,230815,230725,230721,230720,230613,230613,230511,230512,230309,230304,230302,230224,230216,221015,221019,221010,221024,221115,221025,221007,221114,220905,221009,220915,220912,220503,220506,220508,220522,220808,220517,220422,0002621107,220418,220416,220419,220421,220305,220326,220325,220207,211210,220121,220223,211214,190912,190704,190102,190219,230420,190716,190716,231007,200317,200325,231123,231029,231008,230822,230706,230710,230421,230209,221206,221010,220908,220808,220517,220415,220303,220323,211216,211215,200106,211115,211108,211006,210820,210723,210723,210613,210612,210405,210405,210305,210103,201013,201014,200911,200911,190322,200514,190725,190724,190915,191109,190102,200222,200416,200416,200610,200518,200603,191205,210326,230814,230810,230529,221127,220906,220825,221205,220829,220829,220423,220718,220305,220303,220302,211216,220212,221113,190306,190729,211204,211115,211011,211011,211022,211017,211014,211007,210904,210902,210824,210811,210703,210720,210619,210610,210610,210610,210519,210508,210505,210405,210403,210405,210111,200928,181219,200305,211021,210318,200304,200518,190415,190415,191025,191022,191218,191218,190226,190912,181120,191114,200704,200928,181113,200420,181221,190702,200107,200107,190827,191019,200331,190507,190507,190507,200326,200403,200103,200227,191226,200617,190506,181220,181220,191205,190218,190425,200222,191212,190318,190325,190515,190111,190823,190823,190419,211122,211106,211106,211011,211115,211108,211018,211018,211105,211014,211008,210926,210908,210908,210913,210903,210902,210810,210809,210809,210809,210626,210823,210823,210729,210729,210620,210811,210723,210719,210719,210728,210703,210613,210610,210613,210519,210420,210611,210517,210515,210613,210511,210511,210511,210507,210507,210318,210407,210509,210310,210406,210405,210403,210327,210323,210402,210402,210302,210127,210121,210121,201205,201203,201202,201113,201109,201106,201103,201027,201022,200929,200910,200906,200928,201023,200929,191016,200304,190322,190426,190720,181124,200417,191127,190121,190725,190306,190306,181124,190318,200409,200513,190819,190321,190619,190819,190102,190724,190915,190723,190723,200320,190915,190716,200416,200326,190119,190510,200320,200420,191223,181106,190915,181214,200610,200416,200102,191112,190103,190219,190219,220823,220824,220824,211219,220603,220315,220418,220418,220414,220305,220421,220305,220303,220302,211221,211213,220304,220214,220214,220214,220212,220212,211216,211202,211202,211214,211213,210519,191218,181106,190515,190805,190219,190219,190510,190114,190507,190528,190729,181130,181204,190623,190218,190426,181210,190114,190114,190307,190327,190322,181124,190227,190528,190102,190107,190723,190419,181219,190318,181214,190702,190702,190116,190102,181106,190620,190415,200617,190702,200107,191016,181219,181228,200420,190416,231231,231220,231212,231125,231129,231115,231103,231020,231012,230925,230828,230906,230822,230710,230523,230419,230410,230411,230411,230306,220510,230307,221015,220524,221025,221026,221114,221113,220912,220523,220502,220508,220424,220808,220422,220315,220418,220415,220303,220104,220104,220324,220207,220224,220223,220223,211203,190510,190402,190912,200108,190702,200615,200331,190402,200418,211122,211107,211027,211021,210824,210812,210718,210806,210729,210823,210802,210625,210610,210420,210612,210520,210508,210508,210326,210304,210103,201205,201127,201014,200617,200910,200906,200811,200615,191205,181219,191202,190724,191019,191012,181129,191125,190318,200320,200102,200204,190102,190704,210913,231216,231221,231123,231205,231106,231008,231024,231103,231015,230918,230828,230622,230823,230726,230616,230706,230511,230511,230523,230418,230411,230309,230307,230304,230302,221115,221021,221116,220524,220502,221022,221010,221203,221127,220906,220523,220503,220423,220423,220509,220818,220415,220603,220303,220303,220324,211210,211203,200728,201007,181214,190109,200811,191021,190307,190624,190107,190402,190318,181217,190422,211206,211012,211108,211026,211021,211008,210926,210810,210726,210809,210806,210709,210823,210728,210723,210520,210517,210317,210403,210305,210302,210111,201126,201127,201023,201014,190321,190322,190515,190620,181204,191021,181210,190121,200513,190121,190528,190102,200416,191112,200319,191025,190322,200409,190624,190723,191108,181106,190402,181129,190402,200422,191218,181221,190702,190118,190409,200511,200331,200331,190729,181217,190307,190307,181221,200106,200928,201023,200907,190218,190426,190623,200329,190823,190823,200305,181128,191127,191023,190321,190915,200611,181219,181219,200107,200421,200421,190915,191108,181106,200317,191025,231221,231015,231216,231221,231207,231129,231124,231125,231108,231103,231025,231015,231008,231008,230915,230925,230619,230809,230724,230721,230711,230616,230613,230606,230606,230512,230522,230524,230529,230508,230422,230523,230209,230321,230306,230306,221016,221208,221204,221022,221024,221115,221008,221007,221114,220510,220907,220908,221127,220523,220912,220503,220506,220502,220522,220508,220720,220824,211219,220418,220414,220418,220413,220305,220416,220302,211220,211217,220326,211219,220121,220111,211213,211213,220224,211215,211215,211107,211210,211204,211107,211122,211022,211107,211107,211105,211027,211022,211011,211126,211026,211021,211017,211014,211008,210925,210914,210907,210902,210902,210820,210719,210729,210729,210728,210723,210721,210709,210703,210620,210620,210609,210519,210420,210408,210511,210512,210512,210517,210517,210512,210507,210508,210317,210407,210422,210406,210404,210329,210327,210326,210324,210305,210402,210303,210220,210121,210111,210102,210127,201205,201203,201013,201127,201102,201113,201127,201106,201102,201102,201029,201203,201022,201026,191025,190805,190510,200518,200214,181120,200617,200107,200108,181213,181213,191218,190307,191111,190624,191019,190729,191217,190117,200204,200204,200313,181130,181220,200904,190628,200103,191111,201013,200910,200908,200908,200928,201022,201022,200906,200907,200904,190111,191202,200513,191220,200304,191004,181124,181204,181210,191127,191127,200611,191004,191111,190619,200102,190103,190321,191018,190723,191226,200421,191125,191127,181109,190320,181130,200222,191111,190515,181228,190915,190704,191223,190913,191111,190620,231129,231129,231020,230822,230823,230711,230810,230810,230308,190510,191019,190325,190624,191016,191025,211204,210904,210806,210709,210902,210318,210502,210422,210422,210406,210404,210324,210127,210117,210102,210103,201208,201127,201126,201106,201103,201029,201027,201022,200929,190623,190322,190826,190827,190915,190409,240104,231225,231220,231227,231221,231212,231205,231124,231121,231108,231103,231029,231023,231015,231006,230914,230912,230823,230810,230828,230821,230821,230726,230721,230720,230606,230511,230510,230517,230523,230306,230309,230220,221205,221019,221202,221203,220720,221127,220823,220502,220506,220507,0002626702,220508,220829,211218,220423,220509,220602,220415,220315,220429,220419,220302,220303,220305,220121,211204,220323,211202,181214,200415,191218,200704,190510,200403,191019,200221,191217,190118,190409,200326,181113,200325,190826,191022,200331,181221,200617,200616,190307,191212,200420,200106,191016,200304,211115,211022,211014,211013,211006,211102,210926,210908,210904,210820,210809,210824,210709,210811,210907,210719,210703,210620,210621,210620,210718,210609,210408,210408,210310,210518,210515,210408,210508,210407,210317,210422,210407,210405,210405,210420,210420,210329,210310,210326,210326,210326,210304,210304,210302,210127,210117,210103,201208,201203,201127,201126,201106,201109,201028,201023,200929,200908,200803,200904,200911,190724,181128,190322,200514,190415,181213,191127,200220,200326,200418,200911,190226,200103,190425,190103,190723,190415,200302,200417,181113,200512,200512,190422,190628,181205,191202,190103,200319,200512,200319,200306,200611,190416,200409,180724,190520,190915,191112,191017,200220,200221,190520,200420,200108,191223,181107,200317,181129,200604,190624,190215,191025,211217,210902,210729,210716,210709,210823,210823,210823,190409,200403,211126,211011,211026,211013,210914,210902,210826,210726,210704,210610,210408,210515,210507,210317,210328,210329,210305,210326,210402,210402,210310,210127,200611,210102,201013,201103,201022,200910,200912,200904,200514,200514,200928,200409,200415,200403,200420,231231,231231,231207,231106,231020,220823,220823,230809,230705,230721,230606,230512,230529,230418,230308,230220,220521,221116,221023,221007,221024,220908,220906,220503,220508,220829,220317,220418,220413,220324,211210,220223,181225,190319,190520,190415,181130,190422,181220,190729,190426,181210,190720,181124,190621,190425,190215,190108,190528,190528,190103,181220,200312,181213,200728,181221,190322,190415,190915,190103,190507,191228,190507,190827,181113,191228,190114,201007,190307,190307,190111,200313,190419,211204,211115,211021,211102,211026,211017,210826,210904,210903,210709,210721,210719,210703,210613,210420,210408,210403,210507,210505,210509,210317,210403,210117,210121,210103,210111,201102,201102,200616,200908,200906,201014,190306,191025,200616,200803,200304,190426,190426,181124,181210,200417,200221,200203,200320,200214,190819,190628,190322,190121,181124,191127,190723,190327,200611,181205,191125,190716,200409,190620,191226,190318,200421,191223,200319,190915,190813,181106,220503,231231,231225,231216,231227,231125,231121,231123,231106,231103,231029,231008,231012,230920,230914,230914,230823,230810,230619,230821,230724,230724,230720,230710,230724,230616,230510,230522,230522,230508,230424,230529,230419,230419,230415,221010,220905,220905,220915,220915,220904,220823,220825,220522,220522,220522,220824,220414,220413,211106,220303,211217,220325,220214,220214,220214,181113,211122,221204,230624,230529,221013,220121,220111,220104,190507,191016,210913,210220,191111,190227,200611,181107,210304,210304,210302,200320,190225,190226,230914,230914,230814,230711,220421,231216,231106,200413,191025,191226,181221,200312,190109,191017,200107,200603,191019,200416,200331,190507,181130,190623,191213,190307,200413,200418,191212,190729,200313,191111,210719,210520,210611,210520,210507,210406,210402,201026,210111,201203,201208,201028,200616,201014,201026,200928,200911,201023,200417,190218,200304,181128,181128,200514,200514,181218,191220,190823,190623,190623,190103,191012,200320,190321,200320,190318,181219,190422,191226,190416,200102,191018,190813,200416,191125,190716,191025,191025,181106,220904,220915,220825,220825,220506,220507,220508,220508,220824,220509,220421,220422,220418,220413,220104,211219,220325,211221,220324,220111,220224,220214,211215,211204,211107,211107,211027,211115,211018,211126,211013,211007,210720,210907,210908,210904,210820,210820,210719,210709,210709,210729,210728,210723,210719,210709,210703,210720,210710,210625,210625,210710,210620,210610,210609,210520,210420,210408,210406,210407,210517,210512,210507,210505,210508,210317,210407,210422,210404,210329,210327,210326,210324,210324,210305,210402,210302,210121,210111,210102,201205,201203,201203,201013,201127,201113,201106,201106,201102,201029,201026,191111,190805,190507,200214,190226,200704,200603,181221,191218,190827,200107,190307,190307,191111,191019,190802,200519,191217,190117,190218,200313,190402,191205,200329,181220,190823,190506,191212,200617,190628,190306,201013,200910,200908,200928,200803,200911,200904,200103,190111,190111,191012,191220,200304,190322,200513,190725,190121,200320,190425,191127,190819,190320,200203,191004,191004,200617,200611,191109,190227,200102,190321,190619,191018,190102,190723,200421,190107,190320,191125,200610,190416,190813,190225,190225,191108,200222,200102,190915,190915,190515,190729,200108,190109,181105,181102,200312,190415,200518,231231,230912,221128,220503,220522,220802,211218,220818,220421,220315,220419,211210,211014,211022,231220,231212,231124,231220,231220,231123,231108,231106,231106,231029,231008,231008,231007,220823,230918,230914,230906,230814,230809,230821,230809,230724,230720,230711,230710,230616,230613,230606,230512,230422,230419,230424,230321,230308,230304,230303,230220,230131,220521,221019,221206,220521,220521,221202,221013,220502,221007,221025,221021,220522,221205,220908,220908,221127,220824,220825,220523,220413,220304,220214,211213,200413,200403,200305,190802,200227,231216,231115,231029,230920,230918,230711,230510,230529,230424,230418,230209,230209,230308,230410,220524,221204,221025,211027,211109,211105,211011,211021,211015,211014,211013,210926,210913,210904,210914,210820,210718,210802,210806,210823,210904,210729,210723,210626,210723,210703,210718,210612,210609,210420,210518,210517,210507,210502,210409,210329,210403,210403,210329,210323,210322,210303,210220,201203,201203,201013,201104,201027,181220,181214,190402,200704,191218,190507,190319,190624,190109,191022,190819,191019,190415,200326,200418,200418,200604,190415,200615,190729,190827,190623,200222,200329,201026,201023,200908,200912,200519,190621,191127,190111,190416,190425,200319,200513,190823,200616,181124,190826,190621,190103,200409,181219,190227,190506,191023,191004,190813,191018,190107,200319,190318,191127,190528,200420,181219,181219,190515,190915,200610,181129,181120,200221,200107,181102,220104,220104,220104,220104,220207,220207,220207,220223,220212,211214,211213,190415,200519,200518,190108,190724,181214,190218,181113,181120,181221,191017,200728,200422,200102,200102,200615,200305,190802,190306,190307,200511,190623,191016,200306,200511,200511,200511,190327,191213,190409,190507,191226,191228,200313,190813,191022,200420,200302,181220,181128,181210,191212,200106,190306,190306,211206,211107,211106,211011,211012,211102,211022,211013,211014,211014,211014,211014,210914,210908,210810,210903,210812,210826,210625,210610,210806,210729,210811,210728,210719,210723,210802,210718,210703,210622,210704,210622,210619,210609,210519,210420,210517,210518,210518,210517,210518,210512,210507,210508,210502,210317,210422,210409,210405,210403,210310,210326,210326,210405,210323,210402,210303,210220,210111,210102,210103,210102,201203,201013,201127,201113,201102,201027,201028,201022,201022,201022,201007,200929,200908,200928,201022,200512,190628,190823,190823,190111,190720,191226,191226,200305,200331,190322,190426,200616,200616,190322,190325,191212,190121,191004,181210,190415,190725,190916,190306,200326,191127,190823,200615,200511,181128,190419,200417,200103,191220,190107,181204,190915,190619,190723,181109,200611,200302,200102,190724,191202,191025,190422,190422,191018,190225,190724,191111,200102,180814,190813,200603,200421,200420,191017,190510,191104,190103,190322,190418,190102,190102,200415,200420,200304,181106,200317,200317,200415,190515,200108,200221,190109,191025,191025,200421,181113,200107,191218,200214,191016,191111,211204,211011,211013,210913,210809,210729,210704,210618,210610,210408,210507,210404,210324,210127,201013,201022,200928,191025,200304,200304,190108,190121,191004,190619,200319,190422,190813,190716,190320,190109,200604,220207,220223,211203,191206,200418,200418,190114,200514,210310,210117,201104,201027,201023,200928,191127,191017,191012,190729,190916,211027,210719,210621,210512,210422,201102,190823,200617,200616,200417,200320,230920,230811,230606,230422,230411,230411,230304,230216,221016,221208,221202,221007,221113,221204,220523,220523,220419,220415,211216,220223,231227,231212,231121,231106,181225,181210,181120,191226,191228,181130,190114,190327,200519,190506,190826,211204,211122,211126,211022,211017,211014,211008,211006,210925,210907,210826,210716,210811,210622,210329,210422,210402,210327,201102,201007,190415,181128,181109,190107,200615,200306,181205,190416,181106,181113,231008,220823,230814,230817,231006,230915,230915,230925,230823,230825,230622,230815,230725,230721,230720,230706,230613,230511,230512,230518,230523,230529,230421,230424,230410,230209,230209,230321,230309,230306,230302,230224,230216,221015,221019,221208,221206,220521,221128,221204,221010,221115,221025,221026,221205,220905,220905,220908,221009,221009,220906,220904,220904,220823,220502,220503,220506,220506,220508,220808,220808,0002626702,220522,220317,220829,220517,220602,220818,220818,220423,0002621107,220429,220419,220421,220303,220303,211221,211220,211221,211210,220324,220111,211203,211214,200514,210508,210806,210517,210612,210517,210305,200929,200221,200908,211221,181205,231229,231121,231121,231012,220823,230711,220510,220908,211213,190510,191018,190813,190510,200403,200422,200305,200305,190409,191217,191213,190802,200204,190515,200511,181220,191206,190827,200418,200320,190623,190114,190215,190114,190114,200313,211210,211204,211107,211107,211106,211011,211021,211021,211102,211126,211202,211022,211018,211006,211008,211105,211015,211015,210925,210913,210908,210903,210902,210902,210812,210806,210902,210806,210824,210810,210726,210823,210904,210723,210721,210626,210703,210704,210621,210621,210704,210619,210613,210613,210519,210611,210611,210517,210511,210507,210317,210317,210407,210322,210322,210305,210303,210127,210127,210117,201208,201204,201202,201103,201028,201007,200910,200906,200906,200803,201014,200911,200912,200514,200326,200513,200513,191012,200302,200302,191023,190725,191202,200320,200320,200320,190419,190419,200320,200227,190227,190520,190819,190520,190916,190724,190724,200420,200420,191017,191017,191018,191018,191018,190716,190716,190704,200313,200604,200604,200604,231231,231225,231223,231220,231115,231115,231103,231029,231025,231020,231012,231012,231006,230920,230914,230829,230828,230823,230810,230810,230822,230725,230710,230616,230606,230510,230510,230518,230522,230524,230402,230402,230422,230411,230410,230208,230308,230309,230306,230304,230302,230220,230220,230131,221014,221016,221016,221203,221113,221128,221021,221024,221024,221021,221008,221026,221024,220720,220907,220902,220904,220904,220825,220824,211218,220818,220818,220718,220418,220418,220418,220413,220421,220421,220424,220302,220305,211217,211219,220324,220323,211213,220224,220214,220214,211216,211203,211203,211213,190119,231220,231108,230828,230918,230918,200422,200908,200728,200312,190415,181113,181120,190507,200214,181210,191213,191218,191218,190827,190402,190117,191228,181130,191114,200603,200329,190114,181218,181220,181205,190823,211206,211204,211106,211107,211022,211027,211012,211026,211202,211013,210914,210926,210914,210826,210820,210809,210810,210723,210626,210619,210618,210420,210517,210509,210502,210404,210326,210324,210402,210121,210121,210102,201202,200908,201208,201127,201013,201022,201026,200908,200803,190628,191012,191012,190107,190121,200227,200409,191226,191018,200421,200421,200421,181109,190119,200610,190320,191213,200222,190418,190219,181214,190702,231225,231106,231025,231008,231008,230918,230912,230705,230705,230510,230517,230522,230519,230524,230422,230415,230410,230410,230309,230302,230306,230220,230131,221208,221023,221021,221114,220906,220906,220825,220823,220506,220507,220823,220823,211218,220422,220603,220315,220315,220429,220421,220326,220326,220325,220111,220304,220212,220214,200422,200305,190226,190215,181120,181120,181210,181210,191218,191022,190827,200108,191111,190624,191218,191019,190402,200305,200519,200204,200204,200317,200204,190218,190409,200329,200329,190426,191212,200616,200904,200417,200103,200222,190306,211204,211105,211011,211026,211021,211008,211006,210913,210902,210820,210826,210716,210823,210802,210703,210723,210620,210612,210613,210517,210518,210507,210318,210422,210406,210329,210403,210326,210326,210310,210305,210304,210220,210111,210111,201207,201013,201013,201113,201102,201102,201102,201022,201102,201007,201007,200908,200803,200907,200907,190628,190215,190111,200304,181124,181128,190322,200513,191220,190121,190725,190916,181204,200319,200912,200611,190819,190227,191111,191112,190724,190724,191109,200417,200221,191226,190108,190723,190723,190320,190320,190418,190322,190107,200319,200319,200222,200410,200416,200102,190225,190515,190515,181214,181214,200415,200420,181106,181106,200108,190109,190805,200317,191226,191025,231225,231221,231125,231123,231108,231108,231025,231008,231008,230920,230705,230724,230920,230912,230828,230828,230828,230624,230619,230624,230724,230724,230711,230613,230511,230522,230518,230508,230508,230508,230420,230410,230321,230306,230303,230216,221015,221115,221208,220524,220521,220925,221114,221114,221128,220915,220912,220825,220503,220522,220824,220824,211218,220802,220423,220423,220418,220418,211106,220326,220323,220418,220419,220302,211221,220121,220121,220121,211202,190215,190325,211108,210925,210913,210802,210703,210318,210302,210323,210127,210102,201208,201203,201104,201013,201126,201106,201103,191025,191025,190912,191023,181120,231221,231207,231205,231124,231115,231121,231106,231102,231102,231015,231008,231007,231006,230920,230912,230823,230814,230815,230725,230720,230706,230606,230511,230511,230512,230518,230529,230421,230419,230418,230209,230208,230303,230224,221019,221204,220521,221202,221007,221113,220907,221128,220915,0002626702,220808,220522,220829,220517,220423,220317,0002621107,0002621107,220419,220305,220302,211220,220104,220323,211210,211202,211202,181213,191218,190809,200403,200108,200617,191019,200305,190409,191206,191212,191213,190802,190226,200604,200511,190306,200326,190623,190111,190802,191022,211210,211021,211122,211108,211021,211027,211108,211108,211115,211102,211102,211102,211115,211014,211022,211014,211007,210903,210903,210812,210704,210823,210721,210728,210704,210621,210618,210613,210519,210518,210505,210407,210310,210304,210103,201027,201027,201007,200911,190325,191127,190426,181218,191012,190620,190826,200103,191112,200302,200512,190725,191202,190107,200320,190819,190225,191109,191109,200320,200416,191218,200420,191017,190704,200317,200603,190624,190702,200413,200416,231229,231216,231123,231102,231102,231020,230914,230825,230810,230810,230705,230624,230630,230720,230711,230510,230508,230529,230524,230420,230419,230410,230308,230304,230216,221014,221014,221014,221023,221024,221025,221114,220523,220502,220506,211218,211218,220718,220418,220423,220603,220603,220421,220414,220424,220104,220324,211107,211210,211203,200422,200312,190528,191223,200317,181220,181113,190916,181113,181225,190225,191012,200331,190109,200413,200107,190827,200418,190422,190507,190507,190623,191228,190114,191016,200617,200313,190325,200728,200519,190218,190415,190729,181130,191114,190426,190813,191021,200616,181218,190620,210726,210720,210723,210626,210625,210720,210610,210519,210512,210509,210317,210409,210405,210327,210310,210305,210303,210121,210220,210102,201205,210121,201202,201104,201103,201103,201027,201029,201026,201007,200928,200803,201023,200410,200417,190306,190621,190227,191108,200304,190327,200513,191220,200306,190121,190725,180611,180614,180622,190419,190419,200221,181219,181219,200320,200317,200319,200203,190819,200912,200514,190416,190915,190318,191018,191213,190716,190720,191226,190506,200421,190116,200326,190528,190528,200107,190620,191125,181031,181107,190108,190111,200221,190813,200604,180906,180913,180925,180912,180912,190418,200415,190915,181214,191109,190702,200413,190515,200102,191111,200302,230724,221202,231216,231125,231125,231129,230823,230810,230705,230821,230726,230711,230510,230420,230422,230422,230424,230424,230303,220510,220510,220510,221023,221022,221113,221009,220523,220927,220418,220305,220326,220323,220214,190215,190510,200306,190307,190307,181113,200331,191213,181204,200221,200728,190506,181220,191212,191228,211106,211015,210913,210908,210826,210809,210718,210716,210811,210811,210704,210626,210618,210517,210317,210329,210117,210117,201203,201013,201027,201026,200803,200908,200421,190823,200106,190416,190119,200418,191212,190425,190108,200203,190819,190520,191017,200416,190109,190514,191112,181102,190409,190912,190409,200331,200305,200106,191212,211204,211122,211021,211006,210820,210718,210823,210718,210618,210408,210317,210103,210403,210405,210127,201104,201022,201023,201022,200514,191127,200616,200417,190725,191112,181219,190102,190915,181217,190725,240104,240104,240104,231231,231212,231121,231020,231008,230825,230821,230720,230720,230720,230711,230512,230420,230323,230306,230216,221204,221013,221021,220912,220904,220503,220522,220509,220415,220419,220223,220207,220212,231227,231224,231224,231216,231221,231129,231125,231124,231125,231108,231108,230821,231029,231025,231008,230925,230918,230914,230828,230906,230814,230705,230822,230809,230724,230711,230711,230606,230517,230522,230508,230508,230422,230508,230415,230321,230302,230131,230224,221014,221014,220521,221208,221204,220524,221204,221204,221023,221114,220907,220902,220906,220825,220825,220506,220808,220824,211218,211218,220603,211221,220325,220415,211220,211217,220303,220326,211219,211006,210914,210908,210904,210717,210709,210719,210729,210723,210723,210709,210704,210420,210502,210502,210404,210404,210329,210327,210326,210405,210402,210220,210121,210111,210127,201208,201203,201013,201127,191223,190805,190226,200413,200413,200403,200107,190307,191019,190729,190117,190402,190218,200519,200329,191021,200610,191212,190419,201007,200929,200912,200911,190823,200304,200304,191220,190916,191004,200326,181204,200227,190819,190227,200518,200409,190619,191226,191018,191018,190320,181129,200222,190813,190225,200409,190515,191018,181228,190426,190702,191111,200312,190620,200214,200203,190620,181109,191012,190327,190425,190725,200302,200319,200227,200227,190227,181228,190716,190416,200317,181120,200603,190118,190506,190729,190215,190321,191012,190628,190318,190506,190819,190520,190704,190416,191226,190702,190402,210403,210327,200409,181128,181218,181217,181228,200331,200704,200617,200305,190226,200511,200227,181221,200616,191022,211210,211215,211108,211122,211022,211013,210914,210907,210903,210824,210726,210902,210823,210726,210802,210626,210723,210621,210622,210619,210613,210613,210519,200616,200616,210520,210511,210317,210310,210310,210220,210220,210103,201204,201204,201126,190913,190802,200421,190510,200511,181210,181210,191019,201007,181120,191223,200108,200403,190319,190319,190510,190624,190624,200221,200107,190322,200306,191019,190409,190409,190402,190415,191217,200326,181130,200317,190114,190218,190729,200413,200331,200331,190506,190215,181220,190823,191022,190823,190826,200616,200610,200325,200302,200106,190306,190306,191111,190628,200326,190215,190215,190318,211204,211202,211107,210926,210926,211105,211027,211115,211027,211011,211108,211102,211126,211122,211022,210508,211017,211014,211008,211006,210926,210826,211006,210903,210903,210902,210820,210810,210812,210806,210824,210716,210729,210726,210802,210729,210811,210811,210728,210723,210802,210728,210703,210625,210622,210622,210618,210710,210329,210310,210612,210408,210519,210520,210609,210518,210515,210515,210511,210507,210408,210505,210505,210508,210318,210502,210407,210406,210404,210420,210404,210404,210403,210328,210326,210324,210310,210402,210327,210220,210127,210111,210103,201208,201203,210117,201126,201013,201127,201126,201106,201103,201203,201029,201022,201014,201007,201023,201106,200910,200610,200514,200603,200907,200420,200803,200803,200519,200811,201022,200907,201023,190322,191220,191012,191004,191004,191017,190720,200513,200514,190119,190119,181128,181128,181218,181218,191127,200313,190425,190425,200203,191212,200203,191021,200911,200410,200306,190619,200102,200103,200214,190103,190108,190108,200416,191226,190321,200512,191023,191025,191112,191125,200312,191025,191125,190723,181109,200323,200320,190107,190624,200611,190422,190422,200416,200518,181205,200302,190225,200409,190320,190318,190520,190520,191108,181228,190103,190103,190704,190704,191226,191223,200420,191017,190915,191125,200325,181107,181106,181106,200317,190415,200317,191205,200518,191223,190226,190402,231231,231225,231221,231212,231227,231212,231207,231207,231205,231124,231121,231121,231106,231103,231102,231025,231020,231008,231007,230925,230823,230823,230828,230814,230906,230622,230815,230725,230725,230721,230720,230720,230706,230613,230511,230512,230517,230523,230420,230421,230424,230415,230415,230323,230209,230321,221127,221127,230304,230304,230302,230224,230216,221208,221019,221206,220521,220521,221023,221022,221024,221025,221026,221007,221114,220510,220908,221009,221203,220904,221204,220825,220502,220503,220506,220508,220522,220808,220824,220517,211218,220818,220423,220423,220602,220317,220315,220429,220419,220421,220305,220303,211217,211220,211221,220323,220111,211210,220304,220223,211106,211126,211106,211126,190225,190520,190912,200617,191213,200222,200323,190623,190623,190826,190402,190402,190623,190623,200222,181220,211105,211006,210826,210806,210823,210728,210625,210718,210613,210517,210518,210505,210407,210329,210305,210127,201203,201126,201013,201029,200910,200512,200518,200514,191127,190111,181128,200410,181219,191023,181124,181128,190416,190320,191108,191108,190514,200420,190103,190109,221205,221127,221127,220506,220720,220720,220423,220602,220416,220121,190510,181213,181213,200421,200511,200415,191016,200103,190912,190912,191019,200611,200704,200403,190510,191112,200302,190322,191217,191019,191226,200304,190409,200203,200305,181113,190402,200331,190506,181221,191212,190826,200617,200103,200418,190416,200106,181205,190321,190321,191016,190322,200320,200302,200304,211210,211215,211011,211115,211108,211022,211015,211014,211013,211006,210926,210913,210907,210902,210902,210903,210824,210812,210806,210717,210824,210902,210709,210726,210811,210723,210723,210719,210720,210704,210625,210621,210621,210618,210610,210612,210519,210408,210520,210408,210515,210511,210505,210508,210317,210407,210422,210409,210405,210322,210403,210328,210326,210324,210322,210304,210304,210302,210127,210117,210117,201208,201203,201126,201013,201127,201126,201104,201113,201102,201027,201028,201014,201007,201007,201007,200929,200929,200904,200803,201022,201023,200514,200518,191206,181128,200312,200513,200326,200320,190425,181218,200811,190823,181109,191226,200410,190215,191016,191019,200103,190103,200203,190619,200103,190916,190318,190321,191023,200512,200512,191202,191127,190107,190422,181205,190819,190813,200912,200306,200518,190321,191109,190520,200204,191109,191017,190102,200220,200413,200329,200422,200420,181106,200317,190416,181120,181120,181129,200204,200604,190219,191025,230814,230809,230815,230810,230624,230613,230421,230209,230321,230224,230224,230216,221014,221014,221015,221019,221204,220524,221023,221013,221115,221007,221026,221007,221113,220905,221128,220808,220829,220818,220718,220302,220413,211217,220207,220304,190320,201027,190720,231103,231029,231015,231006,230920,230821,221113,220829,190729,181213,190319,190510,190321,181113,190402,181205,190418,190111,200304,211115,211115,210904,210726,210719,210515,210404,181128,190119,181218,190215,190108,190225,190318,190415,180727,181109,181109,200512,181205,190318,181228,190322,181107,190402,231231,231225,231221,231227,231207,231205,231121,231102,231025,231008,230920,230915,230906,230906,230823,230814,230725,230725,200422,191218,190409,200422,181210,200108,190624,190624,200107,191217,190109,190307,181113,200511,191016,200519,190729,190916,190729,190506,200617,200107,191212,190823,190823,190515,200313,211122,211012,211026,211105,211008,210903,210907,210902,210820,210718,210823,210802,210720,210622,210609,21 0408,210518,210511,210318,210407,210404,210328,210323,210305,200421,210127,210103,201126,201208,201202,201104,201113,201104,201022,200929,201007,200928,200911,200929,190720,190111,181128,191127,200331,200410,181204,200513,190227,190121,191004,200417,200214,191023,200615,181024,181109,190321,191111,181205,190528,200222,190320,181228,181106,181219,181219,191114,191017,190415,200317,230510,230510,230510,230508,230512,230518,230524,230422,230415,230411,230410,230323,230209,230307,230304,230224,230309,221015,221019,220524,221202,221202,221013,221025,221010,221026,221114,221114,220908,221009,220523,220720,220507,220522,220824,211218,220423,220315,220317,220416,220303,220104,220207,220207,220121,220121,220304,211214,190510,200306,191213,180509,181130,190426,190723,190729,200603,211107,211027,211015,211014,210913,210709,210719,210512,210515,210318,210317,210403,210328,210327,210305,200910,200908,201022,200811,200514,191012,190724,190322,190422,200409,190227,191023,191226,200518,190409,190321,200518,190215,231106,231015,230823,230309,210728,210728,230306,230224,221007,220925,221127,220824,220904,220507,220508,220418,211217,220413,211202,220314,220413,201027,201014,230619,200415,231020,200929,200928,200906,200803,200803,200911,200907,200907,200907,200514,181128,200513,200329,190426,200203,200811,190108,200319,200305,200416,200221,200513,190725,190318,190225,190520,200221,200416,190716,200323,200319,190702,200604,190515,190515,181213,191228,191228,191206,191218,191218,200415,200422,200511,190912,190319,200108,200108,200108,190107,190109,200305,200306,190320,190409,191217,191021,191021,190802,200214,200214,200204,181113,200331,200420,190507,200511,191021,190226,181221,181221,191213,191213,190827,190827,200604,190307,190307,190307,190307,190623,190623,190623,190426,200616,200616,191111,190114,190114,191012,200514,200514,200313,191021,190325,190325,200519,200319,200203,190823,190729,211206,211206,211026,211210,211204,211021,211202,211215,211106,211109,211106,211108,211126,211108,211122,211012,211021,211012,211021,211102,211126,211018,211008,211015,211012,211006,211006,210925,210925,210913,210914,210925,210907,210903,210824,210824,210812,210611,210726,210820,210809,210717,210715,210717,210824,210824,210824,210729,210726,210726,210823,210904,210823,210904,210726,210723,210719,210802,210721,210721,210704,210704,210626,210621,210621,210622,210625,210621,210520,210622,210622,210622,210610,210519,210502,210611,210611,210520,210515,210511,210511,210507,210509,210318,210502,210502,210317,210407,210409,210403,210403,210328,210329,210327,210325,210325,210323,210310,210304,210304,210304,210304,210303,210326,201204,210127,210117,210103,210103,201208,201208,201204,201204,201204,201202,201104,201126,201126,201126,201109,201109,201102,201103,201102,201027,201028,201028,201028,201022,201022,201026,201007,200910,200910,200910,200906,200421,200803,200911,200907,200907,191205,190119,190119,181128,181128,200326,200326,190426,200615,181101,181218,181218,191012,191109,191012,200102,200410,200410,190620,190306,181124,200302,190327,200512,200512,181128,191023,200319,191202,191202,191202,200320,200320,200320,190419,190419,190419,231227,231227,231220,231220,231220,231216,231216,231221,231221,231205,231205,231220,231220,231125,231124,231124,231123,231123,231205,231205,231029,231029,231025,231025,231015,231015,231015,231015,231006,231006,230920,230915,230915,230829,230829,230829,230828,230825,230825,230823,230822,230823,230828,230810,230810,230906,230706,230706,230821,230821,230725,230725,230726,230724,230724,230720,230711,230711,230616,230616,230606,230606,230510,230508,230510,230518,230518,230522,230522,230529,230524,230524,230402,230402,230424,230424,230418,230418,230418,230418,230411,230411,230411,230411,230410,230410,230308,230321,230308,230402,230307,230306,230306,230306,220510,220510,220510,230220,230220,230220,221205,230131,221205,221208,230131,221016,221016,221204,221204,221206,221206,220524,221023,221023,221023,220502,221115,221013,221013,221013,221007,221007,221010,221010,221021,221204,221008,221007,221007,220908,220907,221113,220908,221127,220906,220906,220902,220904,220904,220904,220825,220912,220902,220503,220503,220506,220507,220502,220522,220720,220720,220823,220802,220802,220818,220818,220802,220802,220314,220314,220602,220315,220305,220418,220418,220418,220718,220419,220419,220418,220424,220421,220424,211220,211220,211220,211220,211220,220324,220104,220325,220325,220104,220326,220324,220207,220207,220207,211213,211210,220223,220214,220223,211203,220104,220214,220214,211216,211216,211216,211203,211203,211203,211203,181214,181220,190402,190510,200422,200413,191021,190319,190823,200403,191019,191019,190729,200416,181220,191228,191212,200617,190119,210903,210812,210812,210717,210709,210811,210721,210721,210728,210703,210718,210625,210619,210610,210613,210420,210520,210408,210408,210520,210318,210317,210407,210329,210405,210403,210327,210323,210402,210220,210121,210121,210121,201208,201203,201113,201027,201014,201026,200908,200911,200519,191220,181124,190327,190425,190109,191021,200928,200611,181109,200403,190716,191125,181217,200420,200511,190109,211106,211013,210703,210515,210507,210103,210111,210325,210103,210103,200912,190215,191226,200512,190327,190116,200403,190813,191004,200518,200519,231106,231007,230726,230616,221204,220829,211218,211218,211217,211014,190724,230828,190107,200611,231216,231123,230619,230822,230815,230706,230518,230508,230418,200227,200227,200511,201127,200910,201007,200417,200203,200320,200421,200512,190510,190510,190415,190107,190415,190226,190219,191111,190419,190425,190426,190402,191019,191019,190507,191228,200325,190118,200325,200103,190729,190422,190515,181218,190306,190327,181221,181220,210926,210810,210810,210728,210728,210704,210618,210408,210611,210613,210511,210405,210404,210328,200421,200611,210127,210220,210117,210121,210117,210121,210302,201208,201203,201013,210117,201202,201204,201113,181129,190306,190327,190506,190623,190121,191111,181124,190723,190321,190321,191018,190520,191218,200413,191223,190623,190418,190418,190915,190915,200108,181214,181225,190215,191025,200317,190520,191223,231216,231020,230914,230914,230616,230518,230510,230402,230306,220521,221022,221010,221127,220906,220906,220912,220912,220523,220720,220507,220523,220808,220423,220423,220602,220429,220302,220323,220223,220104,211219,220111,220105,181120,181120,190119,181113,190409,190409,211008,190121,190121,190623,190623,191217,190402,191021,200728,211206,211021,210926,210810,210811,210618,210402,210502,210405,210304,210127,201013,200616,201007,200904,190215,191212,190425,190215,181124,191012,190915,181109,181107,190624,200325,200416,231212,230524,230912,230809,230303,230220,220824,220720,220507,220718,220414,211213,211213,181214,190422,190226,181109,181128,190319,190702,190307,190118,190325,190325,190117,190215,190307,181220,190506,190716,190426,190623,181220,181213,181225,190121,190325,190515,190218,190416,181128,181128,190121,190215,181128,181124,181109,190327,181210,190108,190215,190227,190108,190227,190419,190121,190321,190321,190419,181109,190318,190318,190422,181205,190322,181129,190225,190320,190520,190514,190116,190515,190416,190619,190219,181102,230914,230822,221015,221019,221208,221127,220902,220507,220912,220305,211221,211220,211217,220104,211221,211221,211221,220207,220207,220325,220207,220324,220324,220324,211214,220207,220105,220223,220212,211203,211213,211202,211202,211214,211214,211213,211214,211213,211213,211214,211214,200317,200422,191218,200108,191112,200403,190827,190913,190802,190802,191112,200420,190802,200331,200511,200511,200617,200331,200519,200331,200519,200728,190729,190729,191205,200617,200107,190826,200617,200417,200222,191212,191212,190729,211206,211206,211206,211202,211204,211107,211106,211106,211106,211106,211105,211109,211122,211105,211105,211022,211022,211022,211027,211012,211011,211012,211115,211115,211012,211012,211108,211108,211018,211108,211026,211026,211021,211021,211014,211014,211013,211013,211012,211008,211007,211006,211006,211015,211006,211006,210913,210913,210908,210925,210903,210907,210820,210812,210812,210809,210709,210810,210810,210726,210726,210824,210723,210811,210723,210703,210723,210728,210728,210703,210709,210704,210704,210703,210703,210626,210625,210621,210710,210610,210619,210610,210610,210610,210618,210618,210520,210408,210408,210611,210611,210520,210517,210518,210518,210517,210508,210517,210515,210612,210408,210507,210507,210507,210520,210318,210318,210318,210317,210317,210422,210407,210409,210405,210404,210404,210328,210327,210327,210327,210327,210323,210310,210220,210220,210220,210127,201203,210111,210302,210103,210102,210102,201207,201207,201203,210121,201202,201013,201013,201104,201127,201106,201113,201109,201104,201027,201027,201029,201029,201028,201022,200616,201014,201026,200929,201007,200910,200910,200928,200906,200910,200928,200803,200907,201014,200610,201023,201023,200907,200106,191114,191205,200222,191025,191127,200513,191220,191127,191220,191112,200319,200203,200615,200604,191220,200513,200512,191023,191004,191004,191202,200416,200416,200417,190418,191218,200409,190915,200403,190915,190913,191226,200204,190716,190704,191226,200420,190915,200415,200312,191023,190809,190704,190704,200320,200317,200604,200518,240219,231231,240104,231231,231124,231225,231223,231231,231220,231216,231124,231207,231205,231205,231125,231220,231123,231121,231121,231123,231121,231129,231115,231106,231106,231115,231103,231029,231020,231015,231024,230823,231008,231012,231008,230925,230920,230829,230914,230906,230825,230823,230814,230810,230706,230619,230821,230821,230724,230724,230724,230724,230710,230721,230721,230720,230616,230606,230510,230510,230508,230523,230524,230524,230308,230303,230224,230224,230216,230131,221019,221206,221206,221206,220524,221203,220521,220524,211216,211216,221116,221013,221013,221013,221013,221013,221013,221013,221022,221022,221010,221010,221010,221024,221116,221024,221010,221008,220905,220905,220905,221128,220908,220908,221127,221009,220902,220906,220523,220523,220823,220818,220824,220315,220602,220315,211218,211218,211218,211220,211218,220422,220603,220314,220418,220718,220429,220302,220302,220302,220326,220324,211213,211213,211012,211210,211213,211012,211215,211210,190318,190318,200409,181205,191017,190227,191109,191109,190116,190116,190116,190103,181228,181228,200221,190724,200416,200421,200421,200420,200616,190419,190716,190704,190103,191112,191112,200325,200325,200325,190416,190416,200610,190318,181120,181129,181113,181120,200604,191108,191025,191226,190219,180724,191218,191104,190912,190912,190319,200305,200305,200312,200312,230420,221114,190624,200418,210826,210515,210327,201203,200928,190306,191017,191017,190915,181120,200317,220522,211213,231227,231029,230912,230710,230421,230421,230321,230410,221208,221014,221015,220523,220802,220305,211221,200421,200108,190422,211106,210812,210826,210618,210420,210403,210302,201127,200928,200312,191021,200417,181218,190121,200403,190724,200417,190102,181107,200928,190704,190915,190510,201007,190916,190916,181210,200615,191217,190118,181204,200305,190507,190402,190307,191112,200331,190114,211115,211126,211115,211022,211013,210913,210907,210826,210826,210810,210729,210823,210729,210726,210728,210703,210710,210619,210618,210609,210408,210518,210408,210520,210317,210422,210405,210403,210323,210305,210303,210127,201205,201203,201013,201127,201127,201026,201007,200910,200803,201023,200519,190520,190426,181204,191220,190121,190729,200615,190621,191205,191112,190419,181204,200513,190619,181109,200320,190102,190724,200911,190318,190227,191004,190409,181109,200222,190716,191223,191021,200304,200221,200415,231216,231205,231125,231124,231123,231121,231103,231025,231008,230914,230906,230906,230810,230821,230720,230616,230511,230517,230519,230421,230419,230410,230209,230308,230306,230302,230220,221208,221016,221206,221013,221008,221202,220907,220906,220904,220825,220720,220507,220523,220517,220509,220421,220315,220418,220416,220421,220303,211220,220325,220207,220105,211213,191217,190118,190118,191021,191021,200204,200331,181221,181221,181221,191206,190114,191213,200418,200418,191125,200616,200616,190623,200514,200313,200313,200313,190813,200514,200514,200514,210403,210323,210324,210305,210302,210127,210117,210117,201208,201208,201204,201013,201104,201104,201104,201126,201109,201106,201102,201027,201014,201014,201007,201109,201007,200908,200910,200910,200928,200803,200911,200912,200912,200907,190325,191127,191205,181128,181128,190426,190119,190119,200912,200326,200326,191004,191012,200410,190422,190215,181109,200221,190108,200416,200416,200617,191023,191023,200313,200319,190419,190419,200611,190318,200227,181205,181217,181217,181205,181213,190520,190520,181228,200221,190916,200416,200416,200420,200420,191017,190813,190704,191112,200317,200604,200610,200610,190416,190416,181120,200928,200518,190702,190702,190702,190219,190219,191025,190510,181210,190409,190118,191206,190620,220414,220304,211204,210902,210806,210612,181221,181213,200108,200617,181225,191019,190118,181113,200511,200511,200227,200227,191206,191206,191213,200604,200604,191022,200616,200302,191111,190114,190114,191012,190729,200313,200313,211215,211107,211012,211027,211017,210913,210715,210715,210719,210704,210621,210621,210420,210420,210515,210318,210409,210409,210403,210404,210328,210323,210323,210127,210117,201204,201126,201106,201028,201027,200911,200911,200514,181128,200326,181218,181218,181221,200203,200811,200811,200410,181109,190108,190108,190108,200302,200512,200512,191202,200512,200611,200403,200409,190520,200302,200302,190916,191112,191112,200317,181120,181120,200603,231225,240104,231124,231115,231106,231106,231029,231012,231012,231012,230920,230918,230925,230906,230825,230814,230810,230630,230726,230721,230710,230511,230508,230508,230511,230518,230523,230529,230402,230424,230411,230323,230308,230308,230309,230303,230216,220521,221019,220521,221115,221115,221023,220907,221203,221203,220915,220912,220502,220502,220506,0002626702,220720,220508,211218,211218,220509,220314,220315,220424,220413,220413,220323,220323,211216,220105,191202,191114,200317,230705,230705,230624,230624,230624,230625,230625,230809,230821,230821,230815,230817,230725,230725,190912,190912,190109,200616,200616,200603,200329,200329,211021,211109,211011,211106,211107,211027,211017,211115,211006,210718,210728,210728,210802,210802,210721,210620,210720,210512,210512,210409,210409,210325,210325,201204,201205,201026,201026,190116,190119,190119,190409,181204,190114,190107,190119,191114,191109,200103,200103,190528,190528,190619,191108,191108,191226,181107,181107,231225,231123,231225,231008,230828,230821,230822,230508,230508,230419,230419,230220,221014,221014,221015,221015,221015,221025,220825,220825,220523,220824,220824,220315,220104,220104,220224,230828,211213,211213,230510,230410,220326,190117,210619,210619,200514,181102,190716,200728,200728,190725,181210,211012,211115,211021,210322,210619,210908,210907,210716,210723,210703,210704,210515,210318,210422,210327,210323,210402,210220,210121,201013,201104,201022,200908,200514,191205,190121,181219,190318,200409,191202,200415,181217,200107,200518,200107,200420,191019,190319,190826,200728,200325,201026,201026,200410,190306,191212,190409,190318,181107,191021,191025,191223,200420,231231,231216,231216,231224,231216,231220,231029,230914,230825,230821,230809,230710,230512,230518,230308,230323,220510,230216,220521,221022,221021,220502,220424,211218,220314,220314,220323,211214,211219,220105,211202,190111,190827,190827,200222,191025,211008,210812,210824,210704,210704,210620,210505,210422,210327,210220,201202,201029,200911,200911,190720,200417,200304,190319,191112,190819,190320,190809,200317,190409,191019,191213,191218,200222,200214,190103,190226,191019,190915,181113,190624,200331,191213,190826,200904,181221,190422,191111,200518,191016,211204,211022,211108,211026,211026,211013,211013,210926,210903,210907,210810,210809,210716,210720,210718,210622,210609,210520,210408,210507,210511,210317,210508,210422,210409,210409,210405,210405,210326,210326,210220,210127,210121,201205,201013,201126,201027,201023,201026,200910,200910,201023,200904,190325,200417,200106,200302,190729,190729,190729,190426,200222,200329,191127,200410,190425,190621,181204,190103,191112,200409,190619,191023,200326,200512,200306,190327,191218,191018,190107,190307,181106,181107,200417,190619,190318,180814,191213,190225,190520,200420,190218,220207,220111,220224,211215,211213,211105,231121,230616,230402,230209,221016,221019,220906,220802,220509,220414,220303,211217,211213,210903,210718,210729,210408,210508,191213,220503,220429,200305,200704,190507,200108,181225,181221,190702,200511,200313,190826,191212,200418,200103,200103,200417,211115,211126,211011,211102,211202,211015,210914,210908,210913,210820,210820,210826,210811,210907,210704,210704,210610,210613,210618,210518,210408,210508,210422,210310,210405,210322,210304,210303,210117,200704,201023,201127,201013,201028,201029,201007,200908,200906,200519,200106,190729,190628,190119,190720,181128,190322,200417,200326,200418,191220,191012,190422,190121,190724,181219,181219,200409,190813,190813,190318,200102,190619,231223,231216,231121,231125,231115,231106,231008,231006,230912,230809,230823,230814,230814,230619,230821,230821,230720,230606,230511,230510,230508,230402,230424,230424,230323,230303,220524,221208,221128,221013,221010,221025,221026,221116,220907,220908,220523,220915,220523,220823,220829,220509,220509,220315,220423,220418,220303,220302,220302,220324,220304,220224,211215,211214,200416,191012,211021,210618,210609,210327,210403,210327,181218,190826,211220,231227,231224,231216,231227,231125,231125,231124,231123,231129,231102,231102,231015,231006,230912,230906,230828,230811,230705,230619,230809,230725,230721,230720,230726,230606,230512,230517,230523,230517,230529,230424,230422,230209,230419,230411,230323,230308,230308,230307,230304,230216,220510,230220,230131,221014,221019,221208,220524,220925,220902,220818,220818,220818,220818,220818,220509,230323,190813,200331,210409,210117,201007,200417,190218,190716,191004,191018,200221,190623,191012,190725,190107,190415,190227,190416,181120,190704,211107,211027,211102,211015,211018,211013,210926,210904,210703,201013,180917,181101,181130,181130,181109,181205,190507,181130,200519,190823,211206,210304,190419,190425,220421,220302,211213,200203,200306,190528,181120,190226,190813,200305,200306,200306,190510,181214,190724,200615,191218,200403,190307,190402,191228,190121,191012,191019,190729,200313,200519,181220,181130,190426,191213,200603,181220,191212,190823,191021,201007,200417,200102,211206,211107,211106,211109,211105,211105,211027,211021,211026,211017,211014,211013,210926,210914,210908,210907,210903,210902,210820,210809,210718,210626,210716,210716,210811,210728,210719,210718,210622,210625,210620,210610,200906,210420,210609,210220,210304,210304,210310,210512,210507,210408,210508,210318,210422,210406,210404,210404,210117,210323,210402,201023,210220,210121,210111,201205,201013,201204,201013,201106,200421,201029,201022,201026,201007,200906,201028,200928,201022,200907,200907,201023,200904,200904,190419,190628,200222,190107,191111,191220,191114,200513,190725,181124,190418,190425,181210,191127,190409,191223,200317,200417,200319,190724,200912,191111,190227,190102,191109,191226,190318,191018,190720,181107,200108,200227,181130,200928,200222,190813,190219,190510,200513,190515,181214,190913,191223,190915,200420,190702,200108,190116,181204,231231,231115,231115,231115,231115,231108,231102,231015,231008,230309,231007,230912,230912,230925,230906,230823,230823,230823,230619,230725,230720,230613,230512,230523,230422,230418,230323,230323,230302,230304,230216,230216,221206,220521,221023,221115,221013,221024,220925,221114,221009,220908,220502,220912,220912,220823,220823,220517,220517,220429,220423,220422,220315,220423,0002621107,220429,220416,220421,211115,220323,220302,211217,220324,220324,220207,220121,220121,220212,220212,220223,220212,211213,211213,200107,191228,211021,210726,210609,210328,210328,201204,201204,201022,190325,191114,191205,181218,190422,191004,190819,231125,230720,230419,230216,221114,220314,220429,220105,220522,220223,181213,190912,200728,200403,200108,190319,200312,190409,190409,190118,200214,190226,181113,190506,200331,191206,191213,191213,181221,190325,211206,211122,211115,211022,211018,211012,210925,210907,210903,210812,210806,210726,210704,210626,210619,210720,210613,210519,210420,210517,210505,210318,210317,210407,210407,210409,210409,210405,210328,210328,210323,210323,210303,210303,210117,210102,201204,201126,201109,201027,201027,201014,201014,201007,200603,200603,200514,191205,190119,190119,181128,181218,191012,200615,200410,190620,190306,181109,181124,200302,200512,190108,190108,191202,200319,190422,190121,190121,200227,191109,181228,191226,191226,191017,200221,190716,190716,200610,190702,190702,190219,190515,190515,240108,231229,231225,231216,231207,231205,231205,231220,231121,231129,231106,231102,231020,231007,230925,230915,230828,230912,230912,230823,230825,230705,230822,230817,230724,230710,230710,230711,230512,230518,230529,230424,230424,230411,230308,230304,230304,230224,230216,221205,221016,221206,221208,221202,221021,221021,221007,221021,221115,220908,220904,220603,220317,220602,220419,220302,220302,211221,211210,220214,230822,230705,230508,230306,221206,221114,220523,220423,220302,220214,181214,191112,190624,181113,190826,190506,211026,210826,210721,210719,210618,210509,210327,201027,200928,191114,190218,200513,190321,190425,200410,200409,200409,200416,191025,191025,181107,181107,200422,200422,200518,200728,200904,200312,190415,190226,200603,200413,190507,190307,190623,190117,190327,190402,190415,191228,200302,200107,190108,181130,190426,181220,190823,211107,211026,211007,210826,210718,210908,210811,210626,210619,210611,210420,210512,210512,210612,210317,210329,210329,210305,210403,210121,201208,201022,201205,201013,201113,201022,200906,200907,191114,190826,190912,190725,190121,181219,200227,190416,190528,200204,190318,191018,191125,190118,181107,190119,200610,200610,190219,190225,180814,191021,190915,181120,181214,190813,191108,191226,231229,231216,231205,231125,231125,231123,231108,231108,231025,231025,231008,230920,230912,230828,230828,230906,230823,230906,230705,230811,230821,230821,230809,230809,230720,230711,230606,230510,230512,230524,230519,230422,230422,230419,230415,230410,230410,230209,230308,230309,230306,230304,220510,230220,220908,221009,220915,221009,221022,220912,221022,221023,220906,221116,221019,221016,230306,221022,220912,221009,221022,220915,221009,221023,220908,221016,221019,221116,220906,230306,221014,221208,221208,221203,221202,220524,220524,221203,221115,221021,221024,220905,220906,220503,220522,220508,220818,220824,211218,220421,220315,220422,220414,211213,211213,211202,230810,230821,181213,191218,200422,191217,181113,181113,190725,200511,200511,181220,190826,200421,181128,191016,190805,211210,211109,211108,211122,211018,211018,211014,211013,210914,210913,210907,210903,210903,210824,210806,210809,210726,210904,210721,210721,210723,210621,210610,210519,210515,210507,210405,210405,210322,210302,210326,210127,201204,201106,201023,200929,200910,200906,190325,200514,200514,190426,190119,200203,200728,190620,190321,190215,191023,191202,191202,181205,181228,191226,191125,190416,190416,190416,231229,231015,231015,231007,230915,230915,230912,230912,230823,230811,230705,230619,230619,230809,230809,230725,230721,230711,230710,230511,230511,230518,230518,230508,230508,230402,230420,230402,230402,230424,230419,230411,230411,230321,230304,230309,230220,230216,230216,230216,221208,221015,221019,221019,220524,221208,221022,221128,221202,221116,221116,221203,221205,221114,221114,220908,221127,221009,221009,221009,221009,221009,220904,220902,220912,220507,220522,220424,220808,211218,220314,220418,220418,220429,220429,220305,220223,211217,211214,220212,220104,220105,220214,211214,211203,190422,181213,181129,190507,190913,190306,190826,181218,190114,190322,190425,181120,181219,190102,181113,181205,190318,190225,190912,190119,190103,190327,190102,190327,200611,190702,190510,190510,190319,210517,220603,200403,200403,191217,191019,191021,190226,190418,211006,211007,210518,210328,210328,210329,201013,201104,201028,200908,200912,200912,190119,200203,190215,190422,200410,190108,191012,190725,191202,200222,191109,190813,190724,181120,191104,231227,231220,231125,231020,231108,231029,231012,231006,230914,230828,230906,230706,230630,230721,230508,230424,221014,220524,221022,221008,221021,221115,221113,220718,220429,220207,200413,191112,190226,181221,190109,191021,200313,200422,190307,190624,190402,200420,191226,191212,181102,181130,200616,211204,211012,211014,211008,211007,210907,200906,200906,210810,210811,210610,210618,210507,210507,210329,210329,210404,210322,210305,210127,210127,210121,201104,201208,201106,201022,201022,201007,201022,201013,200803,200912,200907,200907,200417,191016,200513,200514,191220,191205,181218,190729,190322,200203,191021,190915,200204,190419,191202,191125,191109,190318,200421,181205,190515,190826,181106,200610,191226,191226,190109,190215,200108,190118,191019,200305,200305,231225,231121,231123,231029,231015,231007,230825,230821,230711,230510,230517,230529,230209,230216,221014,221024,220907,220915,211219,220603,220415,211221,220121,220121,190121,190507,210903,201202,190813,191111,200704,190117,190319,211122,211106,211108,211027,211011,211026,211021,211018,211017,211015,211008,211013,211006,211007,210926,210913,210908,210810,210902,210907,210812,210809,210718,210717,210717,210718,210824,210716,210809,210709,210726,210823,210811,210907,210728,210704,210704,210703,210710,210718,210619,210612,210612,210508,210403,210121,210310,201022,201102,201028,190729,190426,200203,200610,190725,190108,191213,190716,191025,190318,200420,200512,231227,231124,231221,231231,231231,231216,231227,231205,231205,231207,231220,231124,231124,231123,231121,231129,231029,231015,231015,231008,231008,231007,231007,231006,230914,230914,230912,230828,230906,230823,230814,230705,230619,230821,230710,230725,230721,230720,230710,230613,230710,230606,230512,230510,230512,230517,230523,230524,230421,230411,230209,230209,230209,230321,230308,230309,221023,230306,220510,230220,221208,221016,221206,221022,221204,220524,221128,220524,220521,220524,221010,221008,221116,221116,221021,221021,221116,220720,220908,221009,220906,220912,220904,220904,220825,220915,220506,220912,220507,220508,220808,220517,220509,220818,220315,220315,220818,220602,220603,220423,220429,220416,220421,220305,220302,211217,211221,211221,220324,220207,220207,220121,220121,220304,220223,220214,211203,211203,211213,211214,211213,211215,211213,191218,200305,200704,200704,190319,190507,190415,191019,190802,190528,190402,190402,190117,190117,190327,191111,200331,190912,200418,200421,190623,200107,200107,191212,181130,181130,181130,191228,211206,211106,211115,211122,211018,211022,211014,210913,210903,210820,210809,210718,210729,210726,210728,210811,210718,210626,210720,210621,210619,210610,210420,210408,210520,210515,210512,210507,210318,210407,210409,210404,210329,210422,210326,210323,210305,210303,210121,210103,201205,201127,201113,201022,201102,201027,201014,201026,200910,200910,200928,190520,190725,191016,190426,190826,191205,181218,191220,190119,181214,181204,200413,191112,200409,190819,200416,190725,190725,190628,190619,190619,190724,200221,200421,191109,190416,180814,190102,190716,191125,200102,200102,191127,181225,181214,200420,191109,181106,200422,200421,200518,211214,190218,200616,201013,200910,200928,200811,200305,190912,190624,190912,190916,191019,190117,181113,190409,200305,190321,200417,211022,211108,211027,211012,211018,211021,210508,210914,210913,210908,210902,210715,210717,210726,210719,210704,210703,210620,210618,210518,210508,210323,210310,210304,210322,210310,201203,201127,201127,201127,201203,201102,201027,201023,200906,200906,200803,201023,200904,191016,190720,200514,200329,200320,191220,190306,200409,190620,200214,200417,190321,181107,190422,190422,181205,200107,200221,190320,191223,200416,190912,200304,191226,190809,190809,200518,200317,231106,211220,210220,181204,190215,190108,190121,181109,231231,231225,231216,231216,231205,231121,231115,231115,231123,231123,231123,231108,231103,231029,231024,231015,231008,230920,230914,230914,230524,230828,230828,230705,230624,230619,230822,230724,230710,230721,230720,230720,230711,230710,230512,230517,230518,230524,230529,230511,230422,230422,230419,230411,230410,230323,230308,230307,230306,230306,220510,230224,230220,221014,221014,221016,221019,220524,220524,221023,221010,221007,221026,220522,221202,220908,220906,220906,220904,220523,220503,220503,220502,220508,220508,220808,220517,220509,220317,220429,220422,220429,220303,220305,220303,220302,220104,220104,220324,220323,220304,220304,220304,220212,211213,211213,211213,210914,200313,191206,190802,200204,190226,190226,190226,200331,200511,200331,200511,190725,181220,191213,191213,191022,190826,200418,190307,190306,190623,200106,191111,190114,200514,200514,190823,200313,210806,210824,210729,210904,210904,210721,210721,210704,210626,210626,210723,210621,210619,210610,210613,210613,210420,210420,210515,210511,210509,210509,210317,210508,210405,210405,210328,210327,210325,210304,210304,210302,210302,210117,201202,201013,201126,201109,201102,201028,200929,201007,201014,200912,200912,190325,190325,190119,200615,200305,191004,191012,191012,191012,190215,200302,200512,190321,190321,200319,200319,190819,190819,200611,190318,191017,191017,200102,181228,190724,200420,190704,190416,181120,200603,190510,191218,190912,181210,190319,190319,191217,191217,190118,191021,200214,181120,190506,190506,191206,210703,190515,181210,200912,190322,191212,190823,200326,190425,190725,190108,200305,190520,190103,191125,200420,190107,190702,191223,200312,200511,181210,181225,190409,190802,200413,190827,200418,190624,190528,190327,200617,190729,200313,200519,191025,181130,190426,200222,200329,181214,190823,211206,211106,211027,211012,211018,211021,211013,211007,210914,210904,210826,210809,210718,210718,210826,210810,210823,210720,210723,210709,210626,210622,210625,210619,210610,210618,210420,200906,200910,210420,210509,210502,210407,210420,210406,210404,210324,210303,210220,210127,210111,210127,201202,201202,201127,201113,201102,201029,201007,200908,200928,200803,200907,200907,191114,190419,190306,190111,190912,190725,200304,190725,181219,191127,200410,200320,190419,200227,200227,200417,190528,190724,190716,191226,181217,190320,191108,190419,190320,181124,191125,191104,200313,190114,190109,231225,231124,231220,231123,231129,231029,231015,231008,230920,230912,230705,230822,230809,230720,230616,230606,230510,230519,230524,230508,230422,230415,230410,230209,230402,230306,230220,230131,221016,221128,221022,221024,221025,221021,220908,221116,221114,221009,220906,220503,220507,220522,211218,220509,220423,220423,220422,220418,220416,220421,220326,220326,220323,220325,220121,220304,230822,230619,190109,181130,181218,190111,181210,181218,181124,181217,210824,210709,210729,210726,231123,230814,230724,230307,221021,220908,220904,220825,220324,220304,220223,211214,230705,220905,231227,231231,231207,231125,231123,231108,231015,231007,230918,230906,230811,230811,230811,230811,230725,230817,230809,230809,230809,230809,230720,230616,230616,230511,230511,230522,230522,230523,230524,230508,230420,230410,230410,230321,230306,230302,230302,230302,230220,221014,221014,221014,221208,221203,220521,221023,221023,221013,221024,221008,221026,221025,221113,221116,221113,220905,220718,220522,220906,220906,220825,221009,220506,220823,220720,220823,220829,220829,220829,220517,220517,220718,220718,220718,220718,220718,220602,220602,220315,220423,220418,220418,220414,220414,220414,220415,220415,220418,220305,211221,211221,220325,220111,220121,220105,220105,220224,220223,220212,220212,200312,200418,200418,200418,200313,200313,200519,200420,200329,200410,200417,211204,211105,211106,211109,211022,211027,211027,211021,211021,211018,211021,211022,211021,211017,211007,211018,210914,210709,210904,210826,210809,210809,210717,210709,210709,210729,210903,210728,210823,210723,210723,210723,210718,210709,210622,210618,210720,210709,210610,210710,210720,210710,210420,210611,210613,210512,210507,210408,210509,210509,210505,210502,210317,210502,210502,210409,210422,210422,210404,210404,210405,210404,210329,210403,210323,210305,210402,210220,210220,210121,210102,201205,201205,201202,210121,201203,201127,201127,201102,201203,201029,201027,201029,201104,201203,201022,201007,200910,200928,201022,200912,200907,200904,200616,191114,200417,200513,200403,200320,200911,200610,200611,200514,200409,191212,191218,20 0326,200418,200421,200420,191223,191223,191223,200416,200415,200518,200616,200511,200518,191218,190226,190226,190507,200108,181210,191218,200107,200221,190418,190307,190117,190117,190402,191228,191228,190729,190729,200214,190103,190415,181130,181130,181220,181220,181220,191212,191212,191212,190826,190826,190823,190628,190628,190823,190322,191220,181124,181124,181124,190419,191109,190102,190102,190724,190724,190318,200320,181107,200222,191226,190515,190507,190809,190515,190219,190915,191109,190913,191109,190915,190915,190720,190702,190702,190620,190109,200422,200704,200312,190415,190528,181210,190226,191025,190409,181113,190507,200305,200413,191213,191019,190624,200519,191228,190117,190327,200329,200603,190426,191212,190823,201007,200417,211106,211107,211022,211027,211018,211202,211017,211007,210926,210914,210914,210903,210904,210826,210809,210809,210810,210718,210718,210810,210716,210806,210810,210810,210824,210823,210811,210823,210720,210719,210723,210709,210622,210625,210703,210610,210618,210612,210611,200910,210612,210507,210505,210317,210509,210409,210405,210406,210329,210327,210326,210324,210305,210402,210303,210220,201208,210121,210111,201208,210102,201202,201202,201127,201022,201113,201102,201029,201022,201126,201026,201007,200910,201023,190621,190823,190628,191114,190306,190912,200304,191112,190327,200513,181124,191127,181219,200319,190819,190108,190108,200409,190528,190724,191205,190716,191019,190320,200421,200615,190116,191025,190119,190218,190219,200107,191125,200319,191004,190813,200610,190418,200513,190809,190702,191021,190913,191223,191108,191104,200107,231223,231106,231008,230920,230920,230912,230828,230705,230625,230822,230809,230724,230710,230606,230510,230512,230517,230523,230508,230422,230415,230209,230307,230220,221016,221015,221208,220524,221023,221022,221024,221025,221021,221116,221114,220907,221009,221203,220912,220825,220825,220506,220503,220507,0002626702,220808,220829,211218,220818,220418,220509,220421,220315,220315,220421,220303,211220,211220,211220,220121,211221,220324,211210,220212,220214,220224,220223,211204,211214,210811,210111,201029,190716,191108,191022,190827,190402,191104,190510,181213,181213,190912,191019,190702,190623,190325,190111,190729,191212,190425,191127,190826,191220,190320,190107,181107,190620,190510,190826,190716,181120,200221,220509,231129,231220,231123,231123,231205,231115,231102,231102,231025,231020,230918,230915,230925,230822,230809,230822,230821,230725,230724,230724,230809,230724,230721,230721,230724,230726,230710,230619,230613,230606,230518,230519,230524,230524,230524,230508,230421,230420,230422,230410,230302,230220,221206,221116,221022,221022,221204,221024,221204,221025,221010,221010,221007,221021,221203,221203,221116,220720,220907,220510,221204,220506,220507,220507,220508,0002626702,220912,220508,220424,220808,211218,220818,220802,220423,220718,220421,220317,220429,211220,220424,211018,220224,211220,220104,220324,211214,220207,220111,220207,220212,211202,211215,211215,211215,211215,211204,211204,211107,211106,211105,211006,211115,211018,211026,211021,211017,211017,211014,211014,211007,211008,211006,210925,210904,210903,210810,210718,210717,210717,210716,210716,210709,210709,210719,210719,210729,210703,210625,210710,210710,210620,210620,210718,210610,210520,210612,210519,210420,210408,210408,210422,210515,210512,210408,210408,210512,210512,210507,210508,210318,210502,210502,210422,210422,210406,210406,210404,210404,210329,210327,210327,210326,210326,210324,210324,210305,210305,210402,210303,210220,210220,210121,210121,210111,210111,210102,210102,201207,201103,201127,201113,201113,201102,201103,201102,201029,201029,201026,201026,191022,191025,190409,200617,200617,181120,181120,190507,190507,181221,181213,200413,191218,191218,200403,200403,190307,191111,191019,191019,190802,200511,200519,191217,181130,190402,200317,200317,200204,200313,200313,181130,190506,190218,191212,190826,191021,200616,181124,181124,200610,200417,191111,201007,201007,200929,200910,200908,200906,200928,200928,200911,200911,200907,200904,200904,190111,190111,200304,200304,190322,191220,190111,191012,190913,191012,190720,190720,191004,190425,191127,190425,200227,200227,200611,181204,181204,190819,191004,191109,190619,200102,190102,191226,191226,190320,191018,190102,181109,191125,191127,190418,181129,181129,181129,200222,200222,200928,200928,190813,190225,190514,190915,181228,181228,190913,191223,191223,200302,200313,200312,190704,200220,190915,200108,181106,181106,200317,200214,190805,190805,190913,190415,200214,200518,200518,211206,211202,211215,211109,211106,211126,211108,211012,211012,211102,211022,211014,211018,211014,210926,210926,210925,210914,210913,210907,210903,210908,210902,210812,210812,210717,210716,210716,210823,210723,210721,210802,210823,210721,210621,210718,210610,210609,210518,210511,210511,231227,231227,231225,231221,231221,231221,231205,231121,231121,231124,231121,231106,231102,231106,231020,231020,231012,231012,230920,230918,230918,230914,230914,230914,230825,230906,230810,230817,230809,230726,230720,230711,230711,230619,230726,230512,230518,230522,230522,230522,230402,230402,230422,230420,230415,230415,230410,230209,230323,230308,220905,230306,230306,230306,230302,230302,230220,230220,230220,221014,230131,221014,221014,221208,221204,221204,221204,220524,220524,221128,220502,220907,220908,220912,221008,221008,221205,220907,220907,220907,221127,221127,220902,220902,220906,221009,220823,221009,220503,220503,220506,220506,220507,220507,220508,220808,220824,220802,220824,220802,211218,220429,220423,220423,220421,220421,220421,220602,0002621107,220415,220419,220419,220419,220424,220416,220424,220421,220413,220413,211221,220325,220323,220323,211210,211210,211210,220224,220224,220214,220214,220223,211216,211216,211215,211215,211214,211203,211203,211203,231216,231207,230705,230906,230821,230511,230415,190307,190119,190409,190623,190623,190623,190114,191016,190325,190325,190325,191205,191212,181128,190426,190426,181218,181218,190823,181220,190215,190620,190620,190108,200220,200416,190321,191023,190419,190422,190419,181205,190520,190520,190916,190426,190813,190716,190704,211011,181128,181128,190422,190319,200403,181113,191019,190802,200204,200204,181220,181221,211210,211122,211109,211027,211108,211012,211022,211013,211022,211013,211006,211007,210726,210823,210310,210721,210721,210802,210720,210704,210621,210613,210613,210519,210611,210611,210518,210515,210505,210505,210407,210310,210310,210302,201208,201013,201126,201028,201023,200514,191127,190426,200410,181109,190108,200512,190121,190819,190520,190916,191212,190716,191019,200603,231229,231212,231205,231115,231115,231102,231023,231012,230920,230914,230828,230906,230823,230810,230624,230822,230724,230711,230616,230606,230511,230518,230518,230522,230508,230402,230422,230419,230411,230410,230209,230308,230307,220510,221019,221202,221022,221010,221008,220908,220825,220502,220522,220802,211218,220603,220418,220315,220315,220416,220421,220416,211217,211221,220324,220323,220323,211213,211213,220214,211203,191023,191018,191104,191218,200511,190218,211122,210729,210507,201202,181205,190520,190620,191022,231106,230914,230616,230220,221008,220302,230524,190813,190916,190510,191019,200331,191226,211206,211106,211105,210908,210802,210728,210703,210618,210407,210304,200611,200615,200616,200417,200403,191127,191127,190528,190102,181106,211214,210518,190119,181120,181213,181214,190702,190702,190219,200615,200511,200313,190912,200417,200410,191109,220808,211115,211106,211108,211115,211026,211102,211022,210926,210908,210908,210903,210826,210826,200906,210622,210622,210507,210505,210509,210502,210502,210405,210324,210310,210305,210117,201104,201208,201027,201027,201014,201026,200704,220524,221128,221127,220503,220503,220503,220508,220517,220423,220718,220414,220416,211217,220212,211203,211215,211213,191218,191218,200415,200415,200312,200312,200214,200331,191228,191228,191213,200313,200313,191205,200329,200302,191023,200227,200221,200416,191017,191112,191112,200320,231229,240104,240104,231229,240104,240104,231229,231229,231229,231229,231225,231223,231224,231224,231231,231220,231223,231221,231231,231231,231216,231212,231212,231212,231212,231205,231205,231205,231207,231207,231220,231212,231123,231115,231115,231115,231115,231124,231129,231129,231129,231108,231108,231108,231108,231108,231106,231103,231102,231029,231025,231025,231023,231023,231023,231020,231020,231020,231015,231012,231015,231012,231012,231007,231006,231006,231006,230920,230920,230918,230915,230915,230914,230914,230912,230912,230912,230912,230829,230829,230829,230825,230906,230825,230828,230825,230825,230825,230825,230906,230823,230823,230822,230823,230823,230810,230810,230811,230828,230811,230809,230705,230624,230624,230706,230828,230706,230828,230624,230822,230624,230822,230822,230821,230821,230809,230809,230809,230726,230726,230724,230724,230724,230720,230725,230710,230711,230711,230616,230616,230616,230606,230706,230606,230606,230606,230606,230606,230511,230511,230510,230510,230510,230510,230510,230518,230518,230518,230518,230522,230522,230522,230522,230522,230523,230523,230523,230529,230529,230524,230529,230524,230508,230524,230508,230402,230402,230422,230424,230420,230420,230420,230419,230402,230402,230415,230411,230411,230402,230411,230410,230410,230402,230410,230410,230410,230323,230402,230209,230209,230209,230321,230402,230308,230308,230309,230309,230309,230309,230306,230304,230304,230306,230302,230302,230302,230224,230304,230304,230131,221205,230131,221205,230131,221014,221014,230131,230131,230131,221015,221015,221019,221016,221206,220510,221206,221203,221203,221202,221202,221203,221113,221206,221023,221202,221116,221204,221023,221013,221013,221013,221013,221013,221203,221204,221204,221007,221203,221025,221203,221010,221010,221025,221204,221010,221022,221025,221010,220925,221008,221025,221026,221007,221115,221021,221026,221007,221008,221007,221115,221115,221203,221115,221113,221113,220720,221113,220907,220510,220908,220908,220908,220908,220510,220908,221127,221204,221203,220906,220522,220906,220906,220904,220902,220825,220502,220523,221203,220915,220912,220720,220718,220720,220503,220503,220502,220507,220507,220508,220508,220508,220720,220823,220829,220829,220829,220824,220824,220808,220802,220517,220517,220802,220802,211218,211218,211220,220818,220509,220509,220718,220603,220418,220421,220603,220314,220314,220302,220421,220315,220603,220305,220603,220314,220314,220314,220314,220422,220422,220314,220415,220422,220415,220418,220418,220429,220429,220429,220414,220416,220104,220326,220415,220416,220416,220416,220416,220424,220424,220416,220416,220424,220424,220421,220302,220302,220302,211220,220223,211220,211220,220413,220326,220325,211221,220104,211221,220324,220325,220325,220325,220325,220324,220324,220323,220323,220323,220323,220323,220323,220207,211204,220105,220105,220105,220105,220105,220304,220224,220224,220224,220214,220214,220214,220214,220214,211216,220212,211216,220212,211203,211203,211203,211203,211202,211215,211214,200317,190628,200319,190528,181204,200305,190415,190109,200603,200617,200728,200617,191217,190118,190319,200511,190624,191012,190620,191114,191012,190916,200305,190826,190426,200222,200227,191212,200227,200418,200418,211115,211027,211012,211108,211017,211008,210913,210907,210926,210820,210806,210729,210810,210729,210729,210811,210719,210720,210720,210610,210612,210612,210611,210507,210511,210515,210507,210318,210422,210403,210305,201205,210304,210121,201208,201013,201204,201127,201104,201028,201023,200910,201023,200911,201023,181124,181113,181210,190422,200409,190528,191023,191213,190916,190720,190723,190121,190107,190318,200611,200416,200421,190307,190307,190320,191127,190813,190724,200222,190915,200319,191112,181225,181214,190102,220522,220802,220818,220818,220509,220818,220422,220418,220416,220421,220302,220326,211217,220325,220207,220223,231231,231212,231220,231108,231024,231029,231015,231015,230920,230925,230925,230725,230711,230710,230710,230522,230524,230420,230410,230308,230309,230306,220510,221208,221208,220522,221113,220907,221114,221127,220904,220523,211218,211220,211218,220718,231220,231115,231012,230829,230829,230822,230815,230725,230724,230511,230511,230529,230415,230415,230309,230216,221014,221023,221008,220907,220823,220506,211218,220314,220315,220424,211221,220324,220323,211210,220207,220212,211216,211202,211213,181130,181124,181124,181204,181205,191025,181120,181210,190827,200413,190426,211206,211106,211022,211012,211026,211013,210926,210914,210826,210809,210902,210718,210809,210811,210719,210709,210622,210720,210619,210610,210612,200910,210408,210318,210502,210409,210405,210406,210404,210329,210326,210305,210402,210402,210220,210111,201205,201202,201007,201106,201113,201029,201026,200910,200928,200907,200908,191114,190111,190111,190912,191016,200906,200616,181124,200227,200410,190318,190318,191018,200512,200511,190116,190116,190528,190225,190702,190826,191223,231229,231224,231216,231124,231123,231129,231106,231025,231015,231008,230920,230619,230822,230809,230724,230720,230616,230510,230512,230524,230422,230419,230209,230321,230306,230220,230131,221014,221014,221208,221208,221203,221021,221025,221025,221025,221021,221116,221114,221009,220906,220508,220503,220503,220507,220720,220808,211218,220509,220422,220422,220414,220303,220326,220325,220121,220111,220223,211204,190805,200220,190319,190319,181221,191112,190219,190725,190402,200331,191022,180917,191212,200616,200617,190515,190623,190419,211107,211026,211008,210903,210729,210729,210613,210613,210703,210611,210611,210502,210404,210310,210303,210111,201205,201104,201014,201007,200906,190322,190720,200305,181219,191220,190320,181120,191109,200320,181109,181107,190225,190704,190514,200604,181106,200512,190416,200304,190109,231229,231229,231227,231108,231006,230920,230706,230809,230518,230421,230304,221204,221203,220912,220522,220415,220429,220324,220324,220224,220212,211202,211204,211126,211027,211108,211108,211018,211013,210907,210902,210903,210914,210809,210709,210902,210824,210729,210811,210811,210802,210719,210703,210703,210621,210622,210618,210609,210613,210612,210420,210611,210508,210505,210508,210317,210317,210406,210403,210117,210324,210322,210402,210302,210121,210103,201208,201013,201207,201106,201106,201027,201028,200616,231231,231020,230524,230724,230529,230224,221019,221013,221128,220522,220818,220802,211220,220421,220602,220429,220305,220305,220302,211220,220323,220121,220223,211213,190415,190506,190510,190409,200704,190916,181124,181213,190102,200221,200221,190827,190319,200403,190118,200331,181204,191228,190415,190506,190623,190218,181220,190306,191228,190111,200929,200912,200911,201022,200912,201023,200811,190823,191016,190325,200513,200513,191205,191212,191213,181129,200320,190725,190306,181128,181124,190425,190823,191112,200410,181204,190915,190415,190321,190724,190724,190321,200512,200305,200203,200615,191018,200107,181120,190813,200928,200417,191017,191223,190416,190514,191223,181106,190729,190913,191114,200108,200420,200317,191112,200221,200518,191213,211017,190218,190218,181220,190915,191220,190725,190422,190318,230517,230619,221013,221013,220720,220326,220326,210809,210518,210324,210121,210117,181113,181219,220507,230410,200928,200928,220808,220121,210703,181225,191218,190729,190802,181221,191223,181113,191022,190916,190418,190623,190114,190510,211108,211115,211014,210926,210717,210716,210811,210720,210611,210511,210409,210304,210305,210302,201208,201205,201202,201022,201102,200908,201014,200907,190426,181218,200421,200511,190725,181124,200409,190912,190121,200320,200320,190325,181219,191023,190819,191109,191025,181205,191112,191018,190723,190219,181107,200320,181228,190915,190915,191112,190321,190704,190219,190514,191212,231212,231025,231015,230925,230915,230822,230706,230822,230821,230720,230422,230306,230302,230131,221014,221202,221202,221013,220506,221114,220904,220523,220424,220523,220424,220829,220317,220315,220303,211216,220207,220214,230424,231115,230823,230724,230529,230410,221015,221016,220908,221128,220825,220507,220507,220912,211219,220302,220302,220104,220304,211214,200728,190507,190226,191016,200331,190402,190826,190826,190729,190506,190306,190111,211108,211013,211007,210902,210824,210611,210719,210420,210507,210507,210508,210422,210404,210327,210121,201113,201109,200928,190628,200513,200410,190103,190823,200512,200222,191125,200222,181219,190318,191125,231231,231123,231205,231205,231124,231124,231123,231123,231106,231106,231106,231115,231103,231029,231025,231106,231008,231015,231008,231008,230724,230720,230619,230914,230809,230823,230705,230705,230906,230622,231220,230606,230131,190307,220303,220302,220303,211217,220324,220304,220223,211106,211027,211011,211011,211011,210903,210903,210826,210726,210823,210811,210611,210518,210515,210505,210422,210422,210422,210111,201203,201013,201104,230411,211105,210811,210619,181124,190116,221021,220824,210327,200403,181210,181210,200317,200317,200518,200422,200617,200103,200222,200222,211126,210908,210824,210902,210715,210811,210704,210618,210519,210520,210508,210407,210327,210310,210111,201203,201109,201023,200928,200227,190724,190915,211219,220121,211202,211214,211213,211213,211214,211213,190805,190724,190114,191016,200420,181210,190322,191017,190826,190624,200325,190215,200331,190418,190510,210323,200908,200514,191220,190119,191012,190426,181214,200616,190322,190218,191111,200403,200409,190620,200912,200304,190318,200416,191025,181107,190225,190108,180814,190102,190520,191017,190915,200103,200107,190219,200421,200518,191025,211204,211122,211105,211027,211026,211006,211007,210913,210806,210806,210729,210619,210408,210518,210618,210512,210511,210505,210318,210502,210422,210324,210323,201208,201203,201013,201127,201126,231231,240104,231216,231207,231205,231125,231121,231125,231108,231103,231029,231020,231012,231008,231007,230920,230914,230912,230829,230825,230814,230810,230725,230721,230720,230616,230613,230511,230523,230529,230424,230411,230323,230323,230306,230303,220510,230216,221208,221019,220510,221202,221202,221013,221116,221025,220925,221205,221127,220915,220912,220825,220825,220503,220502,220507,220508,220802,220808,220517,220829,220523,220818,220509,220422,220315,220718,220419,220415,220305,220303,220302,220111,220326,220105,220121,211210,220214,211108,191218,200513,200304,200403,211107,210926,210517,210613,210408,210518,210518,210517,210407,200511,210328,210327,231020,230906,230821,230510,230512,230422,230420,230208,220415,200929,200910,200615,200615,191111,190619,240116,231224,231216,231205,231124,231129,231106,231008,230925,230915,230828,230619,230809,230809,230724,230711,230616,230606,230512,230524,230508,230422,230420,230418,230306,230302,230220,230216,221023,221115,220503,220823,220829,220506,220507,220808,220720,220829,211220,220509,220603,220224,220418,220416,220104,211221,211215,210720,210810,210709,210622,210620,210609,210408,210420,210507,210505,210405,210422,210326,210325,210121,210111,201127,201026,190805,190510,190226,190912,200306,181221,190729,190103,190218,191205,181220,191021,201007,200908,200928,200904,190111,190322,200304,190916,181124,190419,190121,200413,200319,200615,200417,190321,191226,200421,181113,190225,190704,181217,200312,200518,200306,200518,181204,200422,181213,190827,200603,190510,190624,190624,190118,191019,200511,190422,190425,181220,190111,211017,211014,210913,211006,210908,210907,210812,210715,210718,210729,210729,210802,210728,210613,210720,210618,210609,210519,210515,210518,210509,210508,210318,210407,210404,210324,210220,210121,210117,210102,210103,201013,201106,200811,201022,201007,200906,200904,191016,191202,200203,200409,200227,200203,190227,200513,190528,191109,181107,200302,190813,190320,200204,190809,191223,190619,190116,181102,190510,190805,190510,190702,191114,200313,200305,181113,181120,190103,190402,190506,191212,200616,200418,211204,211122,211012,211102,211102,211015,201023,201007,200928,200514,190425,181218,200410,191004,191023,181109,190107,190422,191202,200611,190813,191109,190318,200102,190102,200513,190116,190704,200415,200317,200317,200204,200108,200317,240104,231225,231216,231124,231205,231121,231125,231020,231008,231006,230920,230825,230825,230821,230809,230523,230424,230529,230420,230418,230410,230308,230306,221208,220524,221202,221202,221010,221127,220906,220904,220502,211218,220423,220429,220414,220305,220303,220413,220413,211221,211221,211210,220214,211213,190510,200617,200617,190319,191021,200312,191213,200214,191206,191206,190826,190826,200418,200418,191016,200514,211210,211109,211012,211115,211022,211015,211014,211013,210913,210809,210904,210721,210721,210625,210626,210619,210515,210515,210408,210509,210310,210127,210127,210220,201202,201126,201028,201028,201007,200906,200906,191205,200912,200615,191206,190321,190425,200319,200227,200221,200221,200420,190813,190813,191112,200603,200811,200403,231231,231229,231223,231221,231216,231212,231205,231205,231121,231115,231115,231103,231029,231023,231029,231015,231012,230920,230914,230829,230825,230822,230810,230810,230624,230821,230725,230710,230710,230616,230606,230508,230518,230522,230529,230420,230420,230424,230418,230411,230323,230208,230308,230306,230304,230224,230302,221205,221016,221113,221023,221021,221013,221007,221008,221007,221021,221010,221026,221115,221113,220908,221009,220718,221009,220502,220912,220507,220802,220315,220415,220424,220326,220324,211220,220224,211216,211203,211203,231020,230914,181214,190913,190402,201007,200929,200911,200513,200103,200103,190215,200513,191023,200304,190327,191004,181109,190620,190528,200317,200317,240104,240104,231205,230705,220915,190520,190515,190805,191218,191112,200108,190307,181113,190325,191228,200325,200511,190117,181220,200418,200617,200616,191202,211206,211115,211018,211017,211006,210908,210907,210806,210718,210811,210622,210620,210519,210612,210515,210502,210407,210407,210406,210325,210310,210121,210102,201127,201106,201029,201113,200908,200910,200912,201023,200907,190119,200329,191127,191021,190725,200203,190425,200611,200302,181219,190619,191213,181128,190318,200421,190116,190724,200204,190418,181214,190702,200312,200415,230914,230914,230823,230822,230720,230616,230519,230529,230421,230410,230306,220510,221019,221128,221128,221010,221115,221127,220823,220506,220522,211218,220317,220418,220302,220325,220224,220224,211214,231231,240116,231216,231221,231205,231129,231124,231123,231129,231108,231025,231015,231006,230915,230925,230823,230619,230809,230809,230725,230721,230720,230706,230606,230511,230512,230518,230518,230522,230523,230524,230421,230424,230419,230418,230410,230410,230306,221016,221023,221010,221115,221008,220925,221007,220905,221128,220506,220506,220522,220522,220824,220509,220509,220421,220603,220314,220315,220718,220416,220303,220303,220302,220302,211217,211221,220325,220304,211216,220323,220111,220224,220214,220212,211214,211213,211214,211204,211105,211021,211115,211108,211021,211017,211014,211013,210926,210914,210908,210904,210826,210820,210810,210902,210716,210709,210719,210729,210728,210723,210721,210709,210703,210625,210710,210620,210609,210520,210420,210408,210405,210507,210502,210317,210407,210422,210408,210406,210404,210329,210327,210326,210324,210310,210310,210402,210327,210302,210121,210121,201205,210111,210102,201205,201203,201013,201013,201208,201113,201106,201102,191025,190805,190507,200518,200415,191218,200617,200107,200107,181210,181210,200418,191111,190624,191019,190802,200511,191217,191217,200305,200305,200204,181220,190506,191212,200617,190628,190111,190215,191111,200911,200907,200103,200304,200513,191220,191012,191004,190725,190425,181204,191127,200611,191109,200203,200302,190619,200102,200421,191018,190102,191125,190723,190416,200610,200222,191108,190514,200102,181228,191223,191226,190704,200420,200420,190915,200108,190409,200220,200617,190319,191019,231231,231227,231225,231216,231212,231205,231121,231125,231106,231029,231008,231008,220823,231006,231006,230925,230809,230630,230822,230815,230724,230721,230720,230711,230512,230510,230512,230518,230524,230529,230422,230424,230419,230419,230410,230216,230131,221208,221016,220524,220521,221013,221022,221021,221007,220908,220908,221116,221114,220908,220908,220904,220825,220523,220523,220506,220507,220912,220808,220802,220517,211218,220422,220315,0002621107,220718,220224,220224,220223,220223,220223,220223,220214,220214,190319,191217,190114,211115,211012,211017,210914,210913,210908,210907,210810,210726,210823,210710,210625,210704,210621,210420,210420,210420,210509,210409,210404,210325,210103,210103,210103,201127,200910,200928,200304,191018,191104,200329,200329,191127,200203,200615,190723,190723,190723,190628,181205,190813,190520,190716,200420,190805,190809,231221,240108,231227,231207,231124,231123,231129,231129,231025,231015,231008,231008,230920,230920,230915,230914,230925,230823,230823,230823,230705,230809,230724,230721,230616,230613,230606,230512,230522,230524,230508,230422,230420,230418,230411,230410,230321,230302,230131,220524,221202,220524,221022,221010,221008,221203,221114,221114,220907,220907,220908,220522,220915,220825,220825,220502,220503,220823,220829,211214,220802,211218,211218,220419,220509,220415,220415,220315,220315,220718,220305,220305,220419,220302,211217,220325,211221,220325,220105,220105,220105,220304,220304,220214,220224,220214,220207,211215,211115,211204,211106,211107,211109,211105,211107,211011,211027,211027,211011,211108,211108,211026,211021,211017,211014,211008,211007,210925,210908,210904,210719,210716,210709,210719,210729,210728,210723,210721,210709,210802,210720,210625,210710,210620,210620,210612,210610,210609,210612,210610,210408,210509,210507,210517,210512,210507,210505,210508,210318,210407,210502,210422,210502,210406,210404,210329,210327,210326,210324,210305,210402,210327,210302,210121,210111,201106,201102,201029,201022,201026,191025,190805,200214,191218,200617,200107,181210,191111,191019,190802,190802,190802,200615,190624,191217,190402,200305,200204,191205,181220,191212,200617,190628,190628,191111,201023,200910,200908,200906,201023,200911,200907,200908,200904,200103,190111,200304,200513,191220,190725,191127,200611,200203,191111,190619,200102,191018,190102,190723,190720,191226,200421,191125,200222,200611,191108,200102,181228,181228,190704,191223,191023,191025,200102,190915,200108,181102,200518,220824,220523,190724,181214,191218,191218,190624,190118,191021,190319,190409,200511,190218,200313,181129,181113,200331,191226,190729,201007,191016,191111,211204,211126,211012,211012,211021,211015,211017,211013,211006,210926,210903,210902,210810,210806,210824,210729,210810,210811,210728,210723,210704,210703,210613,210621,210618,210519,210502,210318,210404,210327,210402,210117,201203,201127,201027,201007,200929,200928,200904,200514,181218,191127,190215,190819,190108,191213,190107,200911,191202,190318,190116,200417,200409,190325,190225,191108,190704,191226,191226,200329,200604,231205,231103,231008,230619,230825,230825,230821,230809,230710,230517,230419,230419,230410,230308,230220,221016,221202,221013,221021,221114,221114,221127,221009,220904,220904,220507,220502,220508,220508,220829,220802,211218,211218,220818,220509,220314,220315,220603,220718,220414,220418,220302,220305,220305,220303,220303,220325,211213,211213,220105,211204,211213,211214,231231,231225,231221,231216,231207,231129,231125,231220,231103,231103,231106,231029,231029,231029,231020,231020,231012,231008,231012,231012,231008,231015,230925,231006,230920,230920,230918,230915,230914,230914,230925,230925,230821,230821,230725,230725,230721,230721,230725,230720,230523,230524,230424,230418,230308,230309,230302,221208,220521,221016,221015,221204,221204,220524,220521,220521,221022,221022,221024,221024,221115,221010,221008,221203,221008,221021,221021,221203,221205,220907,220906,220906,220522,220522,220915,220902,220823,220503,220502,220506,220503,220502,220506,220522,220508,220424,220808,220824,220829,220802,211218,220818,220818,220818,220315,220421,0002621107,220419,220419,220419,220418,220421,220424,220305,220325,220324,220324,220207,220121,211210,220224,220223,220214,220212,220214,220214,220214,211216,211215,211215,211215,211115,211202,211107,211215,211204,211204,211105,211027,211018,211108,211026,211026,211026,211021,211021,211017,211017,211013,211008,211008,211008,210926,210925,210925,210913,210908,210904,210904,210902,210826,210907,210820,210810,210718,210719,210717,210717,210709,210716,210709,210709,210723,210802,210719,210729,210729,210823,210721,210719,210723,210709,210709,210720,210720,210704,210710,210620,210610,210609,210609,210520,210520,210420,210420,210408,210327,210512,210512,210613,210507,210508,210317,210406,210404,210328,210328,210326,210324,210310,210402,210303,210302,210121,210111,210102,201207,201203,201013,201127,201113,201204,200617,200413,200511,190507,181210,191218,190827,200107,191111,191111,200418,190624,191019,190802,190802,200305,190402,191217,190117,200204,200329,190218,190506,190506,181220,190121,190121,190826,200417,191111,200103,201023,200929,200908,200906,200912,200911,200907,200904,190628,200410,200513,191220,190111,191012,191127,191004,191004,190720,200326,190327,190425,181204,190819,200615,191109,190619,200102,200102,191226,181109,191018,191023,190723,191125,190107,200610,190418,190418,191108,190813,190225,190515,200102,190913,191223,190915,190704,181106,200108,190109,200306,200312,200409,200306,200306,190805,190415,200220,200518,200616,191025,200422,200518,200511,191217,200617,200617,191112,200422,191213,190827,190827,200615,190802,200306,191114,200929,191111,200519,200317,200313,191228,200320,190823,211206,211109,211106,211106,211027,211012,211102,211026,211021,211015,211017,211007,210913,210913,210926,211115,210914,210826,210812,210809,210812,210810,210824,210716,210810,210622,210811,210726,210720,210728,210802,210626,210704,210704,210619,210619,210618,210611,210420,210611,210518,210517,210512,210507,210509,210502,210317,210508,210407,210422,210404,210405,210403,210329,210329,210404,210325,210323,210404,210327,210305,200906,210302,210303,210121,210103,210102,201202,201204,210117,201013,201202,201104,201127,201113,201102,201027,201027,201014,201026,200929,200910,201028,200912,201023,200417,200912,200616,191012,200302,200513,191220,200103,200306,200320,200320,200912,200511,200320,190813,190819,200409,191226,191018,200319,200222,200610,200610,200421,191112,191021,191108,200413,200302,231231,231221,231216,231207,231205,231124,231121,231123,231106,231103,231025,231015,231008,230925,230920,230906,230906,230825,230814,230814,230815,230622,230725,230720,230720,230706,230613,230511,230512,230518,230523,230421,230424,230415,230411,230306,230304,230224,230216,221208,221015,221019,221206,220521,221128,221010,221115,221025,221025,221026,221026,221007,221116,221204,221114,221113,221128,221009,221009,220904,220825,220912,220503,220506,220507,220522,220808,220829,220517,220818,220818,220818,220602,220602,220315,220317,220315,220429,220419,220415,220303,220303,220305,220302,220325,220207,220325,220104,211210,211214,200305,190409,190118,190118,191206,191206,191206,200214,200214,200214,200214,190226,181113,200227,190507,190506,200511,200511,200511,190117,191213,200418,181221,181221,190307,190623,190623,200302,200616,181129,191016,191016,181213,181213,211210,211215,211109,211108,211012,211102,211122,211022,211022,211018,211014,211013,211006,210926,210926,210914,230828,230823,230512,230517,230522,221023,221008,210826,210826,210619,210304,211213,190802,200511,211204,211204,211011,211014,210806,210811,210811,210723,210723,210622,210622,210618,210619,210520,210520,210505,210502,210404,210127,201203,201106,200313,200403,190422,190422,200302,231216,231220,231220,231123,231115,231029,231020,231015,230810,230710,230710,230510,230510,230529,230419,230418,230411,230410,230410,230308,230209,230308,230306,220510,221014,221014,221016,221019,221019,221204,221208,220524,220521,220524,221202,221013,221013,221022,221024,221007,221205,221113,221205,221113,221113,221009,220912,220904,220825,220523,220503,220502,220507,220507,0002626702,220522,220829,220829,220824,220802,220517,220517,220517,211218,220423,220423,220422,220603,220602,220602,220602,220602,220317,0002621107,0002621107,220718,220419,220421,220305,220303,220305,220305,220302,220413,220413,220413,220326,220324,220323,220121,211210,220304,220223,220223,220212,220214,211202,211213,200728,200511,191218,200403,190624,190827,190827,191212,200106,200106,191228,211027,211115,211026,211115,211014,211021,211022,211014,211022,211021,211108,211017,211105,211126,211013,210926,211006,210913,210902,210812,210824,210907,210809,210709,210729,210809,210823,210908,210720,210625,210621,210620,210619,210619,210519,210611,210611,210515,210512,210509,210318,210407,210405,210422,210324,210322,210303,210127,210103,201202,201203 ,201127,210111,201126,201202,201106,201106,201027,201026,200906,200906,200912,200928,200907,200907,200811,200615,200410,200604,200410,191111,190621,190623,200103,200214,190819,190725,200512,200227,200227,191023,191018,200409,191112,200102,191111,190915,190724,190716,191112,191112,200304,191021,191021,191226,191104,190913,190805,190416,190515,231227,231221,231212,231125,231129,231115,231020,231020,231015,231020,231020,230920,230914,230914,230810,230906,230821,230721,230710,230710,230511,230512,230529,230420,230418,230418,230411,230323,230306,230303,230216,230216,221019,221128,221206,221128,221204,221025,221024,221007,221113,220907,220908,220825,220904,220508,220912,220502,220508,220829,220829,211218,220603,220429,220429,220414,220305,220413,220324,220105,220212,211202,230823,230508,220121,190321,200403,200418,191217,200227,200604,200418,200420,191111,200329,200329,211204,211105,211102,211017,211014,211013,210904,210820,210826,210709,210823,210811,210728,210802,210625,210618,210610,200906,210518,210512,210612,210507,210505,210407,210329,210327,210127,210127,210121,201205,210117,201013,201102,201106,200417,200417,200409,191220,191023,191023,200403,191112,200221,211203,211202,211213,211214,190805,200107,200326,190827,211122,210508,211015,210718,210802,210802,210420,210512,210409,210323,210305,201208,210102,201007,200928,190510,190306,190111,200302,200302,191228,190318,200911,190716,190320,190915,181120,190514,231220,231207,231029,230912,230828,230809,230706,230710,230523,230402,230422,230216,221016,221206,221007,221115,221127,220503,220503,220915,220315,220305,220104,211202,211213,200413,190805,190507,200617,200313,190319,190109,190827,190915,200420,200305,200611,191228,200103,190729,181210,190402,190402,200331,181214,190218,181130,190426,190823,200403,190623,210407,210409,210327,210405,210403,210422,210328,210404,210305,210305,210304,210220,210127,210111,210103,201126,201203,201202,201013,201127,201113,201106,201027,201029,201029,201026,201026,200910,200928,200803,200803,200610,200907,200904,190325,190218,191016,200313,190826,191112,190215,190415,181128,200513,181218,190111,200417,191220,190823,200403,200320,181218,200203,191205,190306,200906,191111,200313,191004,191217,190227,190121,181109,200326,191018,191202,190724,191226,190318,200409,181205,200306,200222,190528,190813,200320,190520,200204,200420,190624,191114,190102,191218,191218,190619,200317,190218,190702,190219,200403,200302,200518,191226,190409,190111,211021,211022,210902,210716,210719,210519,210404,210327,210220,210220,201208,201113,201027,200908,190322,190623,191016,191021,190720,191127,171011,191220,200416,200302,190321,190325,191018,200319,200611,200603,180814,180830,200319,190704,190219,230421,230418,230323,230303,221115,221021,220825,220507,211220,220207,220207,211216,231229,231231,231212,231129,231015,230920,230925,230823,230828,230710,230720,230711,230606,230529,230508,230323,220510,220510,230309,230220,221013,221013,221008,221008,220905,220825,220829,220503,220507,220824,211218,220423,220414,220415,220303,211221,211221,181120,190415,190402,200418,200519,190419,210326,210326,210220,210220,210303,210117,201204,201013,201113,201014,201014,200908,201028,200906,191127,200329,190515,190515,200103,190819,190819,190916,190725,200203,190619,190619,190528,190724,200417,181214,190915,190109,190109,190116,181213,190912,200421,200511,200518,200415,191218,190416,200403,200617,200304,200904,200108,191217,190118,190118,191019,190802,200329,190409,200204,190219,190218,200519,200331,190506,190506,181210,181220,181228,181210,190116,200302,191212,191213,190826,190826,200616,200418,200421,200103,190515,200106,191016,200417,200305,200305,211210,211215,211109,211105,211027,211108,211011,211115,211108,211022,211006,211108,211014,211006,211013,211006,210926,210913,210907,210903,210824,210812,210806,210717,210718,210824,210806,210726,210811,210721,210907,210723,210719,210720,210704,210622,210621,210621,210618,210610,210612,210509,210519,210408,210520,210520,210408,210520,210518,210515,210612,210511,210505,210508,210519,210317,210322,210422,210422,210502,210409,210405,210403,210422,210328,210326,210324,210324,210404,210322,210403,210304,210304,210327,210302,210127,210117,210103,210117,210121,210117,201203,201027,201013,201202,201202,201106,201126,201126,201104,201106,201027,201028,201127,201014,201102,201022,200929,200929,200803,201014,200906,200911,201023,200514,200302,191206,190720,200103,191127,191127,190121,190119,190415,200513,200329,190425,200203,190823,200811,190215,200410,190912,190108,200416,200203,190619,190916,190723,190321,190318,200512,191023,181220,200512,190422,191202,190107,200911,190813,200518,200222,190103,190416,190225,190215,190215,190520,191109,190913,191017,191017,181228,190107,181214,181228,200221,200221,190704,190704,200413,200329,200108,200317,200604,200604,190805,190624,190402,190226,190226,190510,231231,231225,231221,231212,231216,231207,231129,231125,231123,231124,231015,231008,231008,230925,230915,230906,230823,230809,230809,230828,230622,230815,230724,230721,230720,230706,230613,230511,230508,230518,230523,230529,230421,230424,230418,230415,230415,230209,230321,230308,230308,230309,230302,230224,230216,230216,221208,221015,221019,221019,221206,220521,221128,221022,221021,221025,221026,221026,221026,221205,221113,221128,221204,220904,220915,220823,220502,220503,220506,220508,220522,0002626702,220829,220517,220517,211218,220818,220818,220422,220317,220317,220423,220419,220415,220303,220305,220302,211217,211220,220104,211220,211210,211220,220207,220105,220105,220223,220212,220214,220212,211204,190226,210703,201014,210509,181107,191017,181221,200107,191217,190624,181113,200331,191213,190826,190425,190720,200203,190307,191111,191206,190321,190723,190628,200409,200222,190915,190109,200415,200317,181214,190805,190215,210404,190102,200403,200514,200403,200617,190118,190409,190802,190802,200204,190507,190507,190827,190827,200418,190623,191016,200519,190823,190823,211206,211107,211021,211126,211122,211126,211109,211215,211122,211122,211102,211022,210903,210914,211018,211015,211007,210926,210925,210913,210903,210926,210914,210726,210812,210717,210806,210824,210824,210716,210823,210823,210823,210726,210728,210613,210626,210621,210626,210619,210619,210610,210613,210610,210519,210611,210520,200616,210611,210611,210517,210613,210508,210507,210507,210520,210509,210322,210318,210317,210407,210407,210409,210409,210322,210322,210323,210305,210304,210304,210303,210127,210117,210303,210103,210103,210103,201204,201202,201104,201126,201109,201109,201028,201014,201023,201026,201007,201007,200910,200906,200803,190325,191016,200615,181218,200203,190327,200221,191112,190725,190419,200409,190227,190227,190528,190528,181228,190102,190724,190724,190716,200610,200604,231231,231227,231220,231212,231205,231220,231220,231123,231115,231115,231103,231025,231025,231025,231025,231012,231007,230920,230918,230915,230912,230912,230828,230823,230906,230828,230828,230705,230822,230822,230815,230725,230809,230720,230706,230613,230511,230511,230510,230522,230523,230524,230524,230421,230421,230424,230415,230415,230418,230323,230323,230308,230308,230306,230307,230304,230220,230220,230131,221208,221014,221208,221206,221206,221202,221203,221202,221013,221202,221021,221022,221025,221025,221022,221114,221115,220510,221127,221009,220915,220915,220503,220823,220823,220808,220808,220720,220824,220818,220509,220423,220314,220314,220421,220314,220422,220419,220419,220419,220424,220421,211220,211220,220325,220325,220325,220324,220323,220413,220324,211213,220323,220111,211210,220104,220304,220223,220212,220212,211214,211214,211203,211203,190916,211022,211108,211008,210908,210826,210723,210704,210613,210509,210509,201022,220414,220325,211202,230705,220302,231227,231207,231220,231029,230912,230809,230726,230724,230706,230606,230523,230411,230306,221019,221114,220904,220522,211218,220315,220413,220104,211221,220214,210127,200319,191218,200107,200420,190827,200313,200313,211106,211015,210824,210717,210802,210612,210511,210507,210407,210327,210324,210103,201113,201103,201103,201029,200803,200912,190426,190426,190912,200811,200320,200320,200320,190725,191023,190819,190819,190716,190716,190318,190325,200304,191112,200413,190912,181221,200325,190528,191228,190409,200904,190306,190306,200417,211206,211115,211107,211012,211026,211026,211014,211013,210913,211115,210904,210812,210809,210806,210810,210721,210704,210622,210621,210519,210420,200910,210511,210317,210409,210403,210324,210325,210304,210326,210121,210102,201208,210102,210111,201113,201102,200929,201013,200929,200928,201023,190915,191016,200513,190723,181128,190111,190109,181221,200302,200304,190327,190729,190725,191023,191004,200222,200928,200421,190514,200108,200413,191223,200317,191223,231225,240104,231212,231125,231123,231106,231025,231012,230914,230809,230815,230724,230821,230809,230710,230512,230522,230529,230424,230411,230323,230306,220510,230304,220521,221206,221202,221115,221026,221113,221127,221127,220523,220508,0002626702,220508,220808,220602,220423,220602,220429,220303,220207,220304,220212,211213,190623,190117,190119,220906,220523,220523,220503,220523,211218,220818,220423,220418,220302,220303,220326,220324,220323,220121,220111,220304,220223,220214,211214,211213,190802,191021,200305,190117,191212,191212,200418,200417,200610,211202,211122,211107,211106,211012,211026,211105,211015,211013,211015,210926,210913,210925,210820,210809,210718,210824,210811,210723,210703,210723,210613,210622,210704,210619,210610,210612,210420,210420,210505,210317,210406,210403,210328,210324,210305,210302,210117,201022,190628,190809,190621,181218,190425,190327,190227,191202,200911,200611,200611,190327,190107,200911,180814,191223,191108,191104,181107,181214,181106,191108,190109,190107,190307,190218,181218,181124,190515,190402,200704,190805,200306,200421,190912,191112,200422,191021,190702,200214,200305,200214,190121,200511,191019,200331,190506,190507,200511,181221,200421,200222,181221,181221,190103,210519,210519,210407,210405,210327,210323,210402,210220,201027,201113,201022,201027,201014,200910,201026,201014,201026,200929,200928,200912,190325,190325,181120,190823,190704,191127,190725,181128,190826,190507,200103,190619,181109,200320,181219,181219,190322,190416,200319,190620,200204,191112,200611,200403,191112,190418,191108,181106,181228,200415,190915,190912,191004,190219,181106,190912,190827,181101,180827,190913,190418,190226,190402,190507,200511,190118,190218,190218,200519,200728,200418,190623,181220,181218,200617,190320,191016,190729,210407,210304,210324,210405,210220,210121,210121,210102,201202,201202,201113,201127,201127,201102,201029,201007,200929,200928,200317,200103,191025,190720,190912,191025,191127,181128,200102,200320,200320,181109,200421,190528,200417,200409,200416,191226,190318,190520,191112,191112,200304,200415,191223,200319,190416,200107,190321,191022,190506,200728,211021,210907,210718,210811,210612,210515,210326,210323,201113,201102,201027,200312,190819,191111,200417,190107,181205,200911,200511,200415,200317,190119,181109,200331,210502,200929,191016,210502,220207,231205,230510,231205,230914,191019,190826,211021,210902,210826,210622,210610,210505,210409,210409,210409,210220,210220,210117,201013,201027,200910,200910,190425,190425,191220,191220,190724,190227,190227,190227,190422,200305,200320,190620,240104,240104,231231,231231,231216,231205,231207,231220,231115,231106,231025,231015,231020,231020,231020,231006,230925,230829,230828,230825,230906,230821,230815,230721,230616,230517,230522,230422,230422,230418,230411,230321,230302,230302,220510,221208,221015,221204,221013,221008,221010,220908,220915,220507,220507,220507,220522,220508,220523,220720,220824,220509,220818,220509,220423,220415,220418,220415,220207,220111,220212,200306,200422,191025,190319,200603,200617,190624,190624,190827,190819,191022,190624,190507,200511,190623,190623,190506,190320,191022,200103,200103,191212,200615,190729,191212,190218,190623,190826,191112,211106,211115,211018,211018,211021,211021,211018,211015,211006,210908,210903,210902,210902,210826,210806,210810,210823,210823,210703,210703,210626,210626,210625,210625,210620,210619,210619,210612,210517,210515,210515,210512,210302,210317,210317,210404,210329,210403,210326,210305,210305,210304,210127,210127,210117,201208,201202,201127,201109,201029,201022,201026,200908,200908,190325,190325,191114,200513,190515,181124,190327,181124,190121,190215,200320,200320,181204,191228,200102,200512,190724,181128,200320,200611,190419,181225,181225,190723,190318,190318,190318,200403,200403,200403,191109,191109,190227,191226,191017,190416,190913,181217,181106,200317,231223,231220,231216,231205,231123,231123,231025,231024,220823,231006,231006,230823,230711,230705,230630,230726,230711,230616,230512,230522,230522,230522,230410,230323,230323,230131,221208,221208,221204,220502,221010,221010,221007,221114,221114,220904,220904,220507,220912,220508,211219,220415,220315,220315,220315,220414,211217,220326,220104,220207,220223,220223,220212,211203,181210,181102,231231,231216,231216,231115,231102,231008,230912,230912,230705,230721,230510,230421,230209,191019,190415,190226,190802,190325,190111,191022,191012,190419,190306,190628,190628,181124,190415,190422,181109,181205,191018,190320,190520,190520,190520,190716,181228,210812,210806,210715,210824,210824,210726,210403,210904,210811,210721,210802,210721,210626,210723,210709,210625,210609,210619,210610,210519,210519,210518,210517,210511,210509,210509,210509,210317,210409,210403,210325,210322,210322,210305,210302,210127,210103,210103,210302,201204,201204,201202,201013,201104,201126,201106,201027,201027,201028,190219,200403,191021,200306,191112,200221,200418,200615,190624,200325,200214,200331,190826,190506,190716,211206,211206,211206,211206,211206,211206,211202,211204,211204,211115,211107,211106,211107,210718,210720,210703,210626,210626,210613,210613,210613,210406,210422,210328,210304,210303,210326,210324,210304,210310,210220,201202,210121,201106,201023,200911,200904,190119,190720,190720,191212,190725,190916,190725,191112,200203,191012,191004,190619,191109,190723,200512,191202,200203,190724,190227,190528,190219,190116,200420,200604,190219,220413,201102,201023,231015,191212,200319,201027,201027,201023,200422,200221,200107,190409,190117,200519,190826,190729,191212,190422,200418,211105,211021,211122,211011,211202,211014,210926,210914,210907,210718,210717,210806,210823,210726,210729,210811,210703,210710,210704,210618,210518,210408,210408,210502,210302,210405,210327,210304,210323,210303,201027,210111,201203,201013,201106,201013,201026,200910,200911,191016,181210,181124,200616,190327,190108,190107,191023,190107,200319,200512,190422,191213,191226,200911,200611,190227,200518,200222,191218,190520,191017,190704,181129,200420,190520,230411,230208,230307,230304,220510,221204,221128,221024,221025,221113,221113,221127,220904,220829,220720,220507,220808,211218,220423,0002621107,220421,220305,211217,220104,220207,220105,220214,220326,211204,211115,211027,211011,211102,211018,210913,210904,210902,210806,210709,210810,210612,210612,210408,210518,210507,210505,210318,210407,210405,210404,210304,210303,210303,201204,201203,201203,201203,201203,201202,201102,201106,201102,201027,230828,211220,220418,220418,220418,220424,220305,211217,220104,220105,220224,220304,220223,210729,200513,190823,200908,200910,200928,200912,190325,200514,191127,200313,190426,190119,200203,191012,200615,200410,190215,190107,190108,200302,190321,191023,200512,200512,191202,191202,191202,191202,200319,200319,200319,190419,190422,200320,200227,200409,190225,190227,190227,191114,191109,200102,190520,190520,190528,190528,190528,191109,200221,191223,191226,191226,181120,191017,200313,190716,200317,200325,200325,190416,200604,181225,181213,200603,200604,200604,190826,190704,191104,191104,190219,200331,200413,200421,181213,181214,181214,190118,190118,181113,181113,190111,190114,181109,181124,181124,190108,190109,190107,181219,181219,181219,190111,190321,211109,211122,211014,210907,210806,210823,211021,210721,210610,210409,190510,181213,200403,190118,200617,200305,190802,191206,191213,191022,200418,200616,200616,211109,211102,211108,211006,210903,210806,210820,210823,210721,210621,210618,210519,210518,210420,210403,210304,210305,210117,201208,201204,201204,201013,201126,201106,201027,201023,200929,200908,201028,200911,200728,200326,200410,200203,191023,190725,200512,191202,190419,200222,200320,190318,200320,200221,200221,200420,191017,191018,190716,200317,200604,191025,200413,190912,200413,200421,200415,200422,200422,190319,200617,191021,190409,191217,191228,191228,191213,200214,200214,190226,190506,200331,200418,200418,190623,191016,191016,200313,190325,190402,200519,200519,190823,210117,210117,210121,201203,201204,201202,201104,201109,201109,201028,201106,201007,201007,200803,201014,191205,191205,190119,200331,200329,200326,200418,190823,211108,210728,181102,221015,190506,190506,211014,210327,201102,200912,231207,230209,220517,190912,200403,200107,200107,191212,190325,191019,200518,200904,190729,200420,190826,191212,190422,200417,190515,200106,181214,190823,210703,210621,210610,210613,210520,210612,210612,210420,210611,210408,201106,210517,210518,210517,210512,210511,210505,210408,210318,210317,210317,210317,210407,210422,210409,210405,210404,210403,210328,210327,210326,210323,210402,210402,210304,210220,210127,210121,201207,201203,201103,201127,201113,201109,201106,201027,201027,201022,200616,201026,201023,200910,200906,201022,200907,191206,191205,200326,200326,191127,190114,190425,190103,190419,190819,190912,200416,190121,190916,200512,200615,190723,190321,191202,191025,200222,200102,181107,200102,181107,181109,191202,200204,191108,200319,200416,190419,190426,200604,190515,190219,181130,181221,190506,190826,190416,190702,190322,211210,211215,211108,211108,211126,211108,211011,211115,211108,211022,211006,211014,211013,211006,210926,211007,210925,210904,210820,210812,210811,210907,210704,210723,210720,210518,210518,210704,210622,210621,210618,210520,210515,181128,181220,190422,190103,181113,181205,190103,190318,200420,181107,181214,231231,231225,231223,231227,231207,231129,231205,231124,231115,231121,231102,231029,231025,231015,231008,231007,231007,231007,231007,230925,230915,230925,230825,230809,230622,230815,230725,230721,230711,230706,230613,230511,230508,230519,230510,230523,230529,230421,230424,230415,230415,230209,230321,230308,230304,230304,230302,230224,230216,221014,221015,221019,221206,220521,221023,221010,221115,221025,221026,221026,221114,220905,220908,221009,220915,220912,220906,220904,220523,220502,220503,220503,220506,220508,220522,220808,220824,220517,211220,220818,220423,220314,220602,220317,0002621107,0002621107,220429,220416,220421,220302,220305,220302,220305,220302,220104,211221,211221,220324,220207,220207,220121,220105,220224,220212,211216,200320,240104,231224,231216,231220,231129,231029,231015,220823,230912,230823,230721,230721,230321,230216,221015,221204,221202,221007,220925,221115,221114,220908,220904,220503,220508,220912,220829,220509,220603,220314,220418,220416,220303,211214,211106,220325,220111,220223,211215,190515,200422,190912,191019,190319,191112,190802,181113,190528,190402,200325,181221,200331,200604,200418,200418,191228,191212,190218,191111,191111,190114,200313,211107,211027,211011,211026,211017,211007,210914,210907,210806,210811,210802,210625,210703,210610,210612,210518,210511,210511,210519,210502,210409,210409,210327,210322,210304,210121,210102,201013,201202,201127,201106,201022,201027,201014,201007,200906,200912,200906,190426,200912,200616,200514,181218,200203,190823,190725,190306,190215,181124,200320,181218,190321,190226,190321,190619,190620,190819,191109,190916,191018,190102,190819,190116,190107,200409,190325,190409,200611,191125,191218,191226,190809,200222,200421,190704,190219,200422,200305,191218,190415,181228,190724,191019,190628,190702,200107,190409,190623,200617,200519,190506,190306,181130,200417,211204,211027,211027,210902,210809,210715,210810,210810,210729,210728,210621,210610,210609,210420,210518,210515,210318,210409,210404,210404,210327,210323,210303,210111,201208,210102,201013,201127,201127,201027,200929,201126,201026,200928,200907,200304,190720,200326,190215,190322,191212,190823,181218,191220,200313,190121,190724,181109,181109,190528,190321,190102,190422,190227,200102,200518,200912,181129,190813,200319,190510,190514,200415,190109,190109,191025,231231,231212,231020,231029,231012,230822,230823,230822,230724,230411,230304,221206,221010,220508,220720,211218,220418,220413,220207,211210,220214,210726,210520,181109,231007,230711,220421,220603,211011,211011,211011,211014,230529,230402,221022,221021,220912,220912,220912,220902,220509,220418,220423,220429,211220,220304,200422,190805,190805,181225,181213,190409,200107,191019,190624,191217,190729,191111,190419,211204,211108,211006,210826,210715,210728,210723,210619,210618,210517,210520,210508,210405,210327,201205,201202,201208,201102,201014,200928,200616,190121,190425,200320,190108,191127,191205,190318,181213,190716,200326,191017,191017,191226,190913,191218,190219,230711,230616,221113,220902,220902,220720,220718,191226,181113,210811,210721,210618,210310,210611,210520,210509,210402,201106,200803,191012,190916,191127,190320,190515,190515,190510,190912,190912,200415,200415,200403,190419,190319,190319,200617,200617,200617,190118,190118,191021,191021,200305,200305,190802,191206,191228,191206,200204,200214,200214,190226,190226,181204,181113,200331,200331,190507,200511,200511,181221,181221,191213,191213,191218,190827,190912,190623,200106,190114,190114,190114,190114,200313,200313,190325,190325,190325,211210,211210,211126,211126,211105,211109,211126,211126,211027,211022,211122,211122,211102,211026,211202,211202,211018,211018,211014,210903,211007,211015,210913,210913,210907,210903,210907,210908,210809,210824,210729,210823,211021,210823,210904,210726,210802,210802,210728,210728,210626,210703,210621,210710,210703,210621,210621,210621,210619,210619,210613,210613,210420,210611,210507,210507,210520,210518,210515,210420,210511,210507,210509,210509,210317,210317,210409,210409,210403,210328,210327,210327,210325,210404,210323,210328,210322,210322,210304,210302,210303,210303,210303,201204,210117,210117,210117,201208,201208,201204,201202,201202,201104,201126,201013,201013,201109,201109,201102,201102,201028,201028,201023,201007,200910,191205,191205,191205,190119,190119,181128,181128,200329,200326,190426,200615,191012,191012,191012,191111,191111,200410,200410,200410,181218,190306,190307,190215,190307,190620,190620,190620,181124,200302,200302,200512,200512,191202,191202,200319,190419,190121,200227,200227,190227,190520,190520,181228,190102,181228,200221,200221,190724,191017,190716,191112,200323,200325,200325,200325,190416,190416,200610,181120,200603,190702,191025,191104,191104,190219,231229,231229,231229,231229,231223,231216,231216,231216,231216,231212,231212,231212,231205,231220,231121,231115,231115,231115,231108,231108,231103,231103,231029,231029,231023,231023,231023,231015,231020,231020,231012,231012,231006,231006,230920,230915,230915,230915,230829,230829,230828,230828,230828,230822,230823,230823,230810,230810,230624,230624,230821,230821,230725,230725,230724,230724,230724,230720,230720,230711,230711,230616,230616,230616,230606,230606,230508,230508,230518,230518,230518,230522,230522,230522,230524,230524,230524,230402,230402,230424,230424,230418,230418,230402,230411,230411,230323,230323,230208,230208,230308,230308,230309,230309,230309,230309,230304,230304,230224,230224,230216,230216,221205,221205,221205,221014,221015,221015,221016,221016,221016,220510,221206,221206,220524,220524,220524,221128,221128,221023,221023,221116,221022,221022,220510,220510,221007,221007,221115,221025,221008,221008,221021,221021,221116,221203,220720,220720,220908,220908,220908,221127,221127,220906,220906,220904,220904,220904,220523,220523,220915,220912,220912,220502,220502,220507,220507,220507,220507,220508,220508,220603,220603,220508,220720,220829,220802,220802,211220,211220,220509,220509,220423,220423,220423,220602,220603,220603,220429,220429,220429,220414,220414,220415,220424,220424,220415,220424,220303,220302,220223,220104,220104,220104,220104,220324,220324,220323,220323,220207,220207,220111,220214,220214,211216,211216,211202,211202,211202,211122,211108,211108,211011,211108,211115,211108,211022,211014,211006,211014,211006,210926,210913,210907,210908,210820,210726,210802,210703,210622,210621,210621,210505,210318,210317,210317,210422,210407,210405,210322,210304,190720,190704,230706,231020,231108,231029,231008,231007,231006,230925,230918,230906,230823,230810,230624,221021,230216,210728,230309,210728,230720,230518,230510,230420,230424,230418,230402,230323,220510,221205,220925,230306,230303,230224,230216,221208,221015,221208,221208,220524,221013,221115,221026,220905,220510,220905,221127,220823,220912,220508,220829,220818,220818,220603,0002621107,220422,220418,220418,220415,220415,220305,220105,211213,220104,220105,211107,211012,210903,210820,210709,210728,210802,210703,210622,210621,210609,210609,210518,210317,210329,210327,210220,210127,210121,201204,201204,201104,201113,201027,220414,220121,220105,200326,200313,210903,210812,210723,210703,210511,210318,210121,201203,200928,190227,191127,191018,231020,231008,231006,230809,230622,230402,230422,230309,221014,221016,221021,221113,220523,220829,220829,220429,211216,220224,200418,211105,201022,201127,191004,200422,200312,200312,190528,200214,181113,200603,190725,190507,191212,200413,190827,200107,190307,191019,190528,190117,200617,200313,200416,191226,200519,190729,191212,190218,181130,200603,190426,200616,200616,200204,191212,190823,211206,211115,211109,211012,211021,211105,211027,211012,211018,211026,211021,211202,211017,211008,211013,210914,210926,210914,210904,210826,210508,210718,210718,211013,210823,210726,210811,210802,210720,210719,210823,210723,210709,210626,210621,210610,210625,210621,210619,210610,210618,210612,210612,210404,210420,210611,210611,210517,210509,210512,210509,210408,210317,210317,210407,210407,200616,210405,210329,210327,210404,210326,210404,210305,210220,210303,210117,210111,210111,210102,201202,201202,201104,201127,201113,201113,201026,201023,201007,200910,200908,200803,201014,191016,190826,190628,200103,190409,190912,190327,200306,191127,190419,200615,200513,200222,200320,190819,200514,200409,190227,191109,190724,200416,190318,200320,190704,190912,200107,181105,190119,200610,200221,191125,190418,191108,190826,190702,200108,190109,231229,231229,231225,231216,231129,231106,231029,231025,231008,230920,230920,230822,230724,230616,230522,230522,230522,230523,230508,230422,230415,230415,230323,230321,230321,230410,230209,230309,230304,230220,230131,220507,221024,221113,220508,221013,221113,221015,220521,221203,221024,221021,221114,220907,220906,220906,220825,220503,220808,220404,220315,220315,220421,220303,220303,220121,220304,220304,220223,220212,220212,211204,211214,211214,190913,200420,210404,201027,200420,181221,181221,201022,200302,240104,231221,231205,231121,231108,231108,190111,231216,231125,231220,231121,231123,231121,231108,190724,190319,190827,190802,190218,211202,211204,210812,210904,210715,210806,210518,210502,210422,210322,210304,210111,210102,201208,201127,201106,201103,201029,201022,200929,200615,200611,200611,181218,190819,190219,181205,190813,190225,190416,190103,190226,231225,231231,231227,231207,231205,231124,231115,231115,231121,231106,231102,231102,231015,231012,231007,230920,230918,230925,230906,230810,230624,230815,230725,230721,230706,230613,230511,230512,230518,230508,230421,230415,230321,230309,230224,230131,230216,230216,221014,221019,221204,220524,221023,221023,221205,221128,220915,220503,220823,220502,220506,220808,0002626702,220912,220824,220718,220315,220416,220424,220302,220323,220325,220325,220104,211203,220303,231216,230710,220315,220421,220302,190515,190510,190109,200305,190729,190402,200728,200519,211021,211026,211022,210926,210913,210904,210810,210823,210718,210407,210111,210102,201102,200906,200417,200417,191220,200417,181124,191004,190321,181109,191018,191202,181109,181107,190219,181218,220925,221205,220915,220915,220823,220522,220522,220523,220808,220720,220315,181213,191218,200403,200403,200617,200617,190319,190319,200305,200305,191019,181113,190507,190507,190506,190507,190226,191206,191213,191022,190826,200418,200616,190307,190114,190114,191016,190823,190729,200313,211210,211202,211210,211109,211012,211108,211018,211013,210925,210914,210914,210914,210826,210907,210824,210806,210806,210726,210726,210704,210626,210621,210618,210613,210511,210509,210509,210317,210405,210406,210403,210327,210121,210405,210127,210103,210103,210102,201204,201204,201013,201104,201104,201126,201106,201106,201102,201027,201027,201028,201014,201014,201102,201109,201109,200908,200912,201028,200928,200911,200907,190325,200514,200326,200514,181218,200203,200803,200102,200410,200410,190620,200302,200416,200512,191202,200319,190422,190107,200611,200409,191017,190225,190520,190520,200204,181228,181228,200320,200221,200204,200320,190813,190813,200416,200313,191112,190704,200317,200306,190416,200604,200413,200409,240108,231229,231216,231216,231216,231216,231207,231121,231125,231115,231115,231103,231103,231103,231025,231025,231025,231015,231012,231012,231012,231007,230920,230920,230915,230915,230912,230912,230912,230828,230828,230825,230906,230906,230705,230706,230822,230822,230809,230809,230724,230720,230710,230710,230711,230616,230616,230606,230616,230606,230510,230517,230517,230524,230524,230402,230402,230422,230422,230422,230418,230411,230411,230410,230410,230209,230209,230209,230308,230308,230303,230304,230304,220510,230304,230304,230131,230131,221014,221016,221016,221206,221206,221202,221202,221013,221013,221204,220415,220415,221204,220925,221008,221008,221114,221114,220510,220510,221127,220906,220904,220915,220915,220506,220912,220829,220802,211219,220818,220509,220509,220509,220422,220422,220423,220603,220423,220602,220315,220718,220718,220416,220415,220416,220303,220303,220303,220303,220104,211221,220323,220323,220323,220104,220111,220111,220224,220224,220207,220212,211203,211203,211216,211203,211203,231227,231221,231029,231029,231008,231008,190515,210507,191220,181120,190109,231207,231029,231025,220823,230823,230724,230720,230508,221023,221114,220915,231015,230308,181213,181214,181213,191218,190912,200617,200929,200312,191217,190118,191019,191021,191206,191228,200204,200214,190802,190507,190520,181221,191213,191213,190826,200418,200418,200418,190307,200107,190623,190114,190114,190116,191016,200313,181213,190823,211210,211204,211126,211106,211011,211027,211022,211122,211122,211102,211022,211018,211015,211007,210925,210913,210902,210907,210903,210903,210902,210812,210824,210826,210729,210904,210904,210726,210728,210626,210619,210718,210710,210610,210613,210613,210519,210610,210520,210517,210517,210511,210505,210509,210407,210305,210405,210323,210324,210402,210302,210117,210103,201208,201204,201205,201202,201102,201102,201103,201103,201028,201028,201023,201007,200910,200910,200910,200906,200906,200803,200803,190325,200514,200519,200519,191127,191205,190119,190119,190218,190218,200326,190426,190723,190426,200616,200615,200203,200203,200203,191004,191012,190620,190620,190306,181128,181124,190321,200512,191023,191023,191025,190725,190725,191202,200319,200320,190419,200227,190819,200409,190227,191112,191109,190321,190528,190724,190724,191226,191226,200416,200420,190716,191112,190704,200317,200326,200604,181120,200604,200604,190702,191104,190219,190219,190219,190515,231227,231231,231225,231229,231223,231220,231123,231115,231103,231029,231023,231020,231006,230914,230829,230828,230822,230810,230705,230822,230809,230809,230724,230710,230616,230606,230518,230522,230524,230402,230418,230411,230323,230208,230308,230308,230308,230309,230304,230304,220510,230220,221205,221014,221016,221206,220524,221202,221023,221203,221025,221116,220415,221021,220908,220925,221008,221113,220510,221127,221009,220802,211218,211218,220415,220509,220603,220603,220603,220603,220603,220422,220718,220718,220315,220414,220414,220418,220415,220415,220424,220305,211221,211217,220325,220325,220324,220207,220104,220111,220304,220224,220214,211216,211203,211203,211202,211203,190402,190226,190226,181228,181225,181220,190318,190117,181130,190415,181220,181220,190419,211021,211126,210914,210903,210811,210612,201113,201202,201205,201203,190327,190322,190322,190402,190111,190114,190121,181124,181219,190227,190102,190107,190425,190320,190116,190327,181107,181107,190320,181130,181113,190219,181214,190702,190702,190327,181102,181106,231207,220522,231006,230915,230823,230823,230823,230814,230630,230619,230619,230822,230823,230720,220503,220302,220323,220121,211216,211213,211214, 200403,191218,191223,200305,190319,181213,191218,190322,190117,190307,211206,211012,211105,211017,211015,211008,210826,210718,210717,210704,210728,210704,210618,210408,210502,210407,210422,210409,210404,210304,210402,210304,210303,200611,210111,201202,201127,201022,201014,201026,200929,201007,200908,201023,200911,201014,201023,191025,200514,190912,190912,200103,191220,190819,191023,200313,200204,200214,200512,191127,200312,190915,190819,200417,190119,190528,190819,190320,200415,190520,191202,191108,190510,190107,191112,191223,191223,191223,200512,190725,191019,191022,211006,210913,210907,210811,210723,210703,210325,210322,210304,201026,210103,210103,201014,190915,191023,181228,231231,231229,231227,231212,231220,231115,231115,231106,231103,231102,231008,231006,230920,230918,230912,230914,230823,230810,230706,230815,230809,230721,230706,230606,230510,230518,230529,230508,230209,230402,230424,230424,230422,230418,230321,230308,221023,230224,230216,221205,221015,221019,221206,220524,221202,221013,221024,221203,220905,220908,220904,220503,220523,220506,220507,220508,220720,220517,220517,220423,0002621107,220429,220429,220415,220305,220302,220302,211220,211221,220207,220121,220105,220223,211202,211013,210913,230624,230624,230624,190805,200302,200221,190118,200519,190507,190422,211202,211014,210810,210420,210406,210117,201028,200911,181205,200611,191109,231123,230914,230517,221015,220906,220302,211217,190805,200422,200704,200603,200306,191021,181113,200511,190402,190307,181220,190114,191228,211206,211204,211021,211115,211115,211026,211105,211017,210913,210904,210902,210902,210809,210709,210823,210729,210728,210625,210622,210619,210612,210520,210511,210511,210507,210508,210422,210405,210405,210329,210323,210324,210305,201027,210121,210102,201204,201127,201113,201102,201022,201007,200908,200928,200907,190416,190723,190826,200103,190119,191205,190729,190823,190912,200617,200304,200513,190913,190725,191111,200320,200203,190215,190422,200417,191213,191025,190321,181205,181107,190119,190716,191108,190514,191112,181214,200413,191223,190109,191111,191112,200221,190219,230131,231225,231227,231125,231123,231129,231029,231015,230925,230925,230914,230711,230622,230725,230725,230720,230606,230508,230523,230419,230411,230411,230209,230308,220510,220524,221024,221022,221022,221116,221025,221114,221127,220904,220825,220912,220508,220829,211218,220421,220421,220315,220414,220305,211220,220324,220111,220214,211220,210811,210404,231221,231123,190510,190319,200325,190415,200617,190111,190109,210913,210809,210802,210609,210403,210404,210404,210402,201203,200910,200907,200907,200305,191213,191004,191023,191112,190409,190813,200421,180814,190716,190716,231231,231129,231129,230814,230720,230508,230511,230308,230308,221015,221116,220510,221127,221009,220508,220802,220419,220212,200413,200413,190510,190510,181210,190912,200704,181120,191218,191218,190510,190319,190319,200108,200102,200403,200403,190624,200107,191019,191019,190409,190409,191217,200305,200326,200331,200519,200511,200325,200204,181113,181113,181130,190506,190114,190117,190117,190218,190218,190729,190729,181220,200616,191212,190826,200103,200417,200106,200222,190306,190422,190623,200326,211206,211204,211008,211106,211115,211115,211215,211109,211109,211126,211115,211115,211102,211122,211108,211022,211015,211014,210826,211013,211015,211015,210926,210913,210908,210925,210907,210902,210820,210812,210809,210717,210826,210904,210709,210726,210811,210823,210625,210802,210811,210726,210719,210721,210728,210703,210625,210622,210622,210622,210618,210610,210618,210609,210518,210407,210420,210520,210612,210408,210519,210518,210518,210515,210512,210408,210511,210505,210505,210508,210318,210318,210502,210422,210409,210404,210310,210403,210111,210127,210328,210326,210324,210310,210310,210402,210305,210304,210304,210220,210127,210111,210103,210103,201208,210121,201203,201013,201208,201127,201126,201106,201102,201103,201029,201022,201014,201014,201023,201023,200929,201029,200908,200616,200611,200518,200514,200513,200610,200514,200910,200511,200420,200421,200906,200803,200911,200907,201022,200911,200907,201023,200907,190510,190322,191025,191220,191025,191012,191016,190720,190720,200514,200514,191127,190119,191004,181128,181219,200320,200329,181218,200203,200203,190425,190510,190819,200615,200410,190215,200214,200103,200103,191111,191111,191112,191111,191228,190619,190108,200302,200302,181109,181109,190322,190321,200512,200417,191023,190107,190325,190723,200319,200512,191125,191202,200611,200611,190506,190320,190422,181205,181205,200306,190225,190225,190416,200409,200204,190520,190520,190915,190915,191108,200413,191223,200420,200617,181228,190704,200108,181106,200317,200317,190103,190103,190415,200220,200317,190802,200518,191025,191108,191025,190226,231231,231225,231221,231216,231207,231207,231205,231125,231123,231124,231108,231102,231102,231025,231020,231008,231007,231007,230925,230918,230906,230906,230825,230814,230706,230622,230622,230815,230815,230725,230724,230711,230710,230613,230511,230511,230619,230508,230523,230508,230529,230529,230529,230529,230420,230424,230402,230402,230321,230321,230308,230306,230306,230306,230306,230303,230302,230224,230216,221208,221015,221015,221019,221206,221204,220521,221128,221128,221022,221022,221013,221013,221024,221024,221025,221025,221025,221026,221114,221113,220905,220908,220908,221009,220915,220904,220825,221009,220720,220912,220503,220506,220506,220508,220522,220808,220829,220824,220829,220517,220818,220818,220718,220314,220422,220602,220423,220423,0002621107,0002621107,220317,220429,220419,220418,220415,220305,220305,220302,211220,220325,220325,211217,211221,220207,211213,220121,220111,211210,220105,220223,211204,211215,211109,211012,211026,211018,211006,210907,210820,210721,210721,210622,210704,210703,210621,210618,210612,210318,210317,231108,231102,231020,231006,230825,230825,230619,230619,230710,230616,230619,230523,230323,230307,221208,221015,221023,221008,220904,220314,211221,220413,220415,220325,220224,220212,231231,231125,231102,230915,230809,230711,230606,230510,230421,230323,230306,220510,221015,221008,220925,220907,220912,220808,220421,220326,220104,220304,220214,190409,200403,200107,190119,190823,200313,181113,190624,191228,200204,181204,190422,190325,190111,211204,211011,211021,211013,210729,210726,210622,210519,210511,210317,210404,210323,210305,210220,201202,201127,201113,201022,200908,200907,190218,200728,200514,190823,200615,190724,200320,181218,190621,190419,190422,191125,190915,200416,200319,190318,191018,200306,190520,190520,190813,200604,181106,190515,181213,181214,190118,181113,181221,181220,181210,190321,210621,181128,190119,181218,200203,190108,181109,190325,190107,181205,190322,190322,190102,181106,181129,231020,190425,190426,191012,240104,231221,231121,231115,231015,230711,230510,230517,230524,230420,230418,230303,230216,221015,221019,221023,221114,220824,220507,220507,220517,220509,220423,220418,220303,211220,220207,220214,190805,191019,200422,200928,191217,190109,190218,200325,190623,200417,191228,211105,211027,211108,211026,211006,210719,210903,210810,210726,210802,210622,210408,210520,210515,210511,210509,210409,210328,210304,210127,210111,201208,201202,201103,201026,200910,200617,200803,200907,200929,190729,200304,200418,190416,191212,190724,191111,190425,200409,200302,190327,181129,190108,190724,190720,191127,200512,181219,200611,181109,190318,191018,190819,190227,200102,190102,191226,190520,181106,201014,191127,190121,200410,190515,190118,200107,200227,181220,181130,190724,211106,211011,210913,210820,210716,210810,210703,210420,210505,210317,210404,210326,210322,210304,210127,210102,201013,201113,201102,200906,200906,200906,200418,191205,200302,190321,190321,191004,190227,200911,190107,191111,181205,190724,190827,181228,190716,200421,190809,211204,211027,211011,211026,211126,211006,221015,221204,221113,221009,220503,220523,220507,220522,220824,211220,220423,220317,220315,220415,220305,220302,211221,220324,211210,220105,211202,220212,181213,191218,191218,190809,200421,200403,200108,190322,190118,200617,200204,191213,200421,190620,200511,211108,211022,211021,210812,210719,210610,210405,201204,201028,200911,181218,200616,200416,190321,190628,190819,191018,190702,221014,221014,221023,221116,220824,221009,220502,220802,220429,220424,220324,220105,231229,231212,231115,231103,231020,231023,231012,231006,230829,230829,230828,230822,230810,230810,230624,230821,230809,230721,230616,230510,230402,230424,230411,230410,230131,191205,191217,190507,181214,190322,200312,190218,190325,191213,200413,200214,190226,190307,200331,190327,200929,200604,191111,200329,190426,181214,200616,200313,190402,210925,210914,210826,210809,210824,210729,210716,210729,210729,211021,210719,210619,210609,210619,210710,210613,210611,210518,210517,210512,210612,210507,210509,210520,210502,210317,210407,210509,210422,210406,210420,210329,210327,210325,210403,210305,210305,200728,210302,210220,210121,210121,210103,210102,201204,201207,201104,201126,201113,201113,200910,200910,200906,200928,200410,200317,190218,190306,200302,190327,191220,191212,200306,181124,181124,200320,200421,190121,190227,190419,190107,200611,191226,190318,200319,200610,191212,190117,191125,191112,190109,190322,190418,190416,181214,200413,200415,200304,230512,230518,230523,230421,230424,230415,230410,230209,230321,221115,221025,221113,220905,221128,220503,220506,220508,220508,220808,220824,220517,220818,220818,220423,220602,220317,220416,220421,220302,220305,211220,211220,220324,211221,220325,220323,220121,231231,231207,231121,231106,231020,230814,230811,230725,230524,230410,230302,220524,220524,221008,220720,220503,220508,220422,220305,220324,220304,211203,220824,220207,220111,220105,220223,211202,211213,211213,211107,211109,211012,211027,211012,211102,211107,211013,211126,211105,211126,211126,211105,211015,211126,211011,211011,211015,211011,211011,211015,211102,211013,211007,211014,211008,210820,210908,211013,210802,210718,210716,210802,210806,210811,210626,210625,210625,210618,210720,210620,210710,210408,210609,210609,210613,210508,210509,210507,210406,210406,210407,210406,210420,210404,210325,210327,210325,210220,210220,210327,201207,201205,201207,201205,210102,201205,210127,201109,201109,201109,201202,200929,201026,201126,201026,201026,200928,201022,201023,200803,200811,200728,200906,201007,200811,200803,200803,200728,200728,200811,200728,200811,200811,201007,200728,200811,200803,200728,200906,200803,200803,200811,200811,200811,200728,200906,200728,200811,200803,200728,200811,200728,201007,200728,200610,200415,200415,200403,181128,190327,190327,190321,190418,180509,200108,200108,190418,190422,190418,190418,190716,190425,181205,181205,181219,181129,181129,181129,181219,181217,190118,181225,181225,190103,181225,181225,181214,181214,181225,190107,181228,181225,181225,181225,190109,200220,200221,200221,200220,200220,200103,200302,190809,190805,190809,190826,190805,190809,190809,190809,191022,190809,190819,190225,190416,190219,190416,191109,191108,191108,200102,191111,191104,191104,191104,191022,191104,191104,191022,191021,191017,191022,191022,191022,191018,191213,191022,191022,191017,181107,181107,181109,181107,190520,190620,190514,190514,190819,190620,190520,190514,190514,190813,190325,190116,190116,190116,190325,190116,190111,190119,190913,190916,191109,190913,190916,190916,190916,200319,200306,200611,200615,200912,200518,200518,191223,191223,200102,191111,200106,200106,200204,200107,200106,200422,200421,200616,231220,231223,231223,231223,231223,231223,231212,231207,231212,231207,231106,231103,231103,231103,231103,231007,231007,231007,230814,230814,230726,230822,230726,230809,230726,230726,230726,230726,230726,230613,230613,230519,230519,230519,230519,230519,230519,230421,230422,230421,230308,230308,230307,230307,230307,230307,230220,230220,221208,221113,221202,221116,221116,221116,221127,221116,221202,221010,221026,221026,221025,221203,220907,220907,220902,220902,220902,220902,211219,211219,220315,220317,220317,220317,220418,220414,220414,220414,220404,220404,220404,220404,220404,220404,220414,220413,211204,220214,220304,220304,220224,220224,220304,211021,211011,211013,210904,210902,210718,210622,210823,210719,210618,221013,200302,200512,200512,181129,190422,181128,190121,190119,231216,230920,230915,230912,230906,230720,230518,230323,221208,221208,221206,220825,220915,220517,220517,211218,220418,220224,220212,181220,181113,181219,190729,191025,200304,191125,191025,190102,190827,190226,190422,190219,190827,190218,190422,181130,181220,190510,191104,190306,211122,211105,211012,211102,211102,211026,211007,210926,210926,210913,210904,210826,210809,210812,210719,210802,210726,210729,210823,210802,210625,210622,210611,210517,210512,210518,210507,210507,210317,210317,210405,210322,210323,210402,210402,210303,210220,210127,210121,210121,210121,210121,210117,210127,210121,201126,201113,201104,201103,201028,201022,201014,200908,200906,200910,200907,200329,190916,190109,190720,190506,200513,191220,190318,200615,200615,190215,191127,191004,181124,190915,190723,200320,200512,200421,191018,191125,190227,191226,200416,200416,191226,190510,180814,180814,191213,200222,191213,190724,200422,191226,190219,200512,190402,190620,191112,231224,231216,231108,230823,230706,230726,230616,230606,230517,230518,230523,230424,230321,230323,221023,220510,221208,221008,221024,221114,220904,220720,220502,220506,220508,220508,220517,220509,220315,220414,220418,211220,211221,220207,220207,220223,220214,190912,181213,210402,191125,191023,191023,190321,210611,231225,231224,231102,231029,230814,230705,230524,230415,230411,221204,220521,221010,221114,220905,220522,220422,220418,211216,211210,211202,191025,200108,181113,190409,190219,181220,190623,190422,211012,211102,211021,211006,210903,210810,210611,210704,210622,210408,210508,210407,210405,210402,210111,201103,201023,200519,190720,190419,190827,191112,181219,190724,200512,190723,190321,191213,191226,190805,190702,200421,191218,190805,190409,191019,190724,181113,181220,191021,191228,200329,190425,191220,191212,200326,200214,191109,191127,190111,190227,190528,190109,200304,190321,190318,200306,180814,190102,181219,181204,191218,220902,221208,221206,220524,221021,221023,221026,221114,210407,210409,210406,210420,210326,210329,210325,210327,210303,210323,210324,210323,210324,210402,210402,210303,210326,210303,210326,210303,210302,201014,181113,181214,190215,181109,190109,231225,231205,230829,230814,230724,230508,230508,230522,230303,230303,220521,221128,221203,221116,220510,220906,220523,220603,220303,220111,200312,190507,190402,200313,190307,200420,190623,211115,211014,210908,210824,210703,210613,210512,210329,210329,210304,201104,200912,190915,191016,190729,200514,190111,200615,190409,190419,190723,190225,190620,191104,200107,191217,190510,190510,191213,190826,210710,210317,210317,210220,201203,201103,200929,190720,190121,190121,181109,200420,191125,200604,200611,240104,240104,231231,231231,231212,231125,231121,231129,231029,231020,230918,231006,230914,230828,230906,230706,230821,230721,230711,230606,230510,230519,230508,230424,230411,230209,230308,230306,220510,230131,221015,221204,221202,221010,221022,221010,221026,220907,220908,220503,220912,220522,220824,211220,220802,220315,220429,220416,220305,211217,220104,220324,220121,220304,220214,220212,211214,181214,200514,190510,200511,191228,200221,181130,191112,200107,190422,211202,211107,211011,211018,211105,211013,210914,210907,210820,210820,210810,210709,210811,210802,210720,210704,210621,210703,210610,210519,210611,210403,210507,210502,210407,210405,210329,210325,210305,210310,210117,201023,201113,201104,201029,201014,200908,200803,200616,190325,190510,190515,190218,200305,200313,190426,191127,190103,190111,181218,200803,190720,190823,191021,191004,200320,190319,200511,200409,190422,191202,200221,200302,190107,180731,191018,191109,191226,200102,181106,190813,190913,200604,190704,200317,190416,190219,190515,190515,181225,181221,181220,201203,200512,200103,190621,191218,191021,211102,210811,210703,201202,191114,200409,191212,181228,220823,230915,230619,230410,220524,220429,220718,200319,180928,211022,211202,201202,190327,230920,230711,230510,230410,230308,220524,220524,221022,220906,220507,211218,220418,211026,230925,190628,190422,200227,190819,190819,191109,190528,200221,191223,190813,190716,200325,200325,200326,231231,231216,231231,231227,231121,231205,231124,231123,231121,231106,231102,231102,231015,231008,230726,230726,230725,230915,230825,230828,230914,230705,230815,230725,230721,230720,230706,230606,230511,230512,230512,230518,230421,230424,230418,230410,230209,230321,230309,221023,230303,230224,230216,221014,221015,221019,221204,220521,221023,221013,221115,231220,230616,230522,221015,221203,220325,211213,200518,191213,191217,190515,190325,211105,211013,210729,210612,210519,210408,210305,210305,201102,190318,200415,190704,190913,190510,200728,200305,190826,200331,190506,200519,211107,211011,211026,211014,210926,210904,210809,210824,210824,210811,210720,210703,210626,210704,210626,210610,210420,210517,210318,210329,210403,210323,200421,210102,201127,201102,201023,200929,201022,191025,190725,200103,200409,181219,181219,190415,200417,181109,200302,200610,191202,190102,190723,190716,190219,190913,190913,200304,190619,200420,231227,231123,231025,231024,230915,230918,230918,230814,230711,230825,230815,230721,230616,230616,220906,220317,220317,220429,220415,220303,211221,220121,220214,211214,211214,191223,200511,200413,181213,181221,190507,200108,190322,200302,190827,190912,190915,191019,190416,200418,181113,200403,190218,191012,200313,190729,191111,200418,191212,190422,190623,190320,200615,200615,190823,211107,211027,211115,211026,211017,211007,210914,210904,210812,210902,210806,210716,210811,210802,210618,210420,210612,210511,210304,210121,210121,201204,201202,201106,201113,201104,201022,201029,201014,201023,200908,200803,201022,200907,200616,200519,200511,200103,191228,191025,191205,190119,190426,190327,190725,190725,190826,191127,200615,190215,190623,190620,200302,200409,181109,181124,190107,200320,190422,200421,190724,200416,181219,200222,200227,200221,190528,200610,190320,181129,191018,191018,190813,190225,190716,200317,191112,200304,191112,181214,190416,181106,231229,231212,231212,231207,231123,231129,231103,231029,231015,230925,230915,230726,230828,230810,230821,230724,230711,230606,230512,230522,230529,230424,230411,230209,230307,230303,220510,230131,221019,221206,220524,221205,221010,221021,221114,221114,220906,220904,220825,220507,220808,220720,220517,220818,220718,220314,220414,220415,211221,220121,220223,211213,211213,181113,190623,190729,200519,211206,211022,211017,211013,210908,210810,210812,210823,210611,210408,210404,210520,210317,210406,210305,210220,210127,201113,200908,200928,201023,191220,200513,200302,190108,190318,190528,190913,190813,191223,231029,231025,231008,230920,230821,230705,230724,230606,230616,230419,230307,220524,221025,221021,221114,220906,220825,220503,220507,220509,220423,220422,220416,220223,190510,200617,190117,211105,211007,210826,210704,210420,210422,210111,201022,190325,190325,190215,200320,190827,190724,190724,181120,200317,231129,230912,230710,230710,230307,221010,220823,220418,211221,231220,231216,231212,231129,231025,231015,231012,231008,231007,230920,230918,230918,230906,230906,230705,230822,230510,230511,230517,230522,230529,230524,230422,230421,230419,230419,230411,230131,230131,220521,230131,221023,221114,220907,220907,220907,220908,221127,220902,220906,220503,220503,220506,220522,220523,220808,211219,220305,220317,220418,211221,220416,220415,211106,220302,220303,211221,220111,220224,220214,211216,211215,211215,211214,211215,211215,211204,211106,211105,211027,211011,211108,211126,211014,211008,210720,210925,210908,210904,210902,210820,210810,210719,210904,210716,210709,210719,210729,210728,210723,210719,210625,210710,210505,210620,210610,210609,210612,210519,210408,210408,210517,210512,210507,210505,210508,210318,210407,210422,210406,210404,210329,210327,210326,210324,210324,210305,210402,210303,201106,201106,201102,201029,201029,201022,201026,191223,190805,190805,190805,190507,190724,190507,190226,200617,181120,200413,200413,191218,200107,181210,181221,190307,190307,190418,191111,191111,190624,190327,190729,190802,190117,200313,190402,190402,200204,190218,200313,181130,181130,200329,190506,190506,190426,181220,191212,200904,181220,190422,200103,200106,190306,191111,201023,200929,200928,200908,200928,200911,200907,200904,190628,190623,190111,190215,190322,190409,190725,190720,190720,181124,181124,200326,200320,190419,190419,181204,191127,190121,200911,190819,190227,191111,190619,190619,200102,190102,191226,190723,190321,200421,191125,191127,190716,190723,181107,181107,181109,190320,181129,200610,200610,190415,190415,200222,200222,200222,190409,181129,190813,190813,190225,190515,190514,190515,200102,191223,190704,190704,200304,190109,181102,190415,181120,200312,200312,190620,200220,200220,200214,200518,181214,191025,181120,181213,190510,190218,200416,201102,200907,200417,190912,190725,190916,190306,190318,190318,200305,181105,200610,190702,191218,200413,191226,201104,190528,200615,211018,201109,200603,200603,200603,200615,190702,190702,191104,190219,190510,211202,211107,211122,211027,211012,211122,211206,211109,211210,211204,211021,210715,210809,210810,211015,211011,211109,211202,211108,211109,211109,211115,211011,211115,211108,211022,210824,211105,211014,211013,211006,210926,211007,210907,210902,210820,210824,210812,210806,210824,210709,210820,210811,210802,210723,210719,210720,210704,210622,210621,210621,210618,210609,210610,210520,210612,210519,210508,210318,210317,210317,210422,210403,210304,191112,200518,200317,231006,230918,230814,230810,230624,230420,230424,230424,230411,230418,230308,230303,230224,230220,230216,221208,221208,221015,221019,221206,221206,220521,221206,221128,220521,221013,221115,221026,221116,221025,221008,221202,221026,221026,221026,221204,220905,220905,221127,220912,221009,220912,220829,220829,211218,220818,220818,220818,220305,220305,220305,211220,220223,220326,211220,211220,220326,210914,200513,200319,220223,210402,201127,190109,211122,211021,211014,211013,211006,210926,210914,210907,210903,210824,210611,210715,210806,231123,231015,231008,231006,231006,231006,231006,181130,181220,181219,181120,211105,211105,211106,211115,211102,211011,211021,211026,211017,211014,211006,210908,210902,210718,210716,210823,210907,210728,210721,210709,210625,210622,210610,210611,210512,210507,210515,210508,210422,210404,210404,210326,210325,210402,210220,210111,201207,201203,201127,201113,201028,201029,201029,201026,201014,201014,190418,190528,190805,190319,200306,190507,200220,181221,181221,191218,190827,200420,191112,190620,190416,190624,190823,200313,190528,190402,190402,200519,190218,190729,181130,190716,200107,200616,200904,200616,200329,191212,191212,190823,190306,200911,201022,200904,200904,190111,191220,190725,191127,190121,190121,190425,181204,200227,200319,200912,190227,200514,190418,190619,190619,190415,190528,191109,200421,191125,181107,181129,181129,181129,191109,190225,190225,190704,191223,191218,191223,200317,181130,190915,190109,190109,190109,181120,230925,230925,230619,230529,230420,230415,230209,230306,230307,230306,230302,230131,221014,221208,221204,220521,221023,221115,221008,221021,220907,220905,220905,220906,220902,220502,220915,220508,220823,220829,220818,220818,220423,220603,220302,220418,220415,211220,211220,211220,211210,220105,211216,211204,211210,211213,181106,200910,200906,201028,231121,231015,231220,231125,231124,231108,231115,231102,230616,230613,190515,181220,200413,190510,190912,190510,190319,181204,181113,190725,181210,191017,191022,200221,190827,200222,190415,191212,190118,190226,190307,190507,190520,200317,190729,191228,190415,190729,190307,181130,200418,190506,190218,200417,200106,190111,211206,211202,211107,211122,211106,211108,211022,211022,211022,211021,211018,211006,210907,210908,210908,210908,210907,210902,210809,210806,210826,210718,210718,210709,210723,210718,210704,210626,210703,210622,210620,210610,210710,210610,210609,210609,210520,200906,210519,210420,210511,210317,210502,210302,210407,210407,210422,210404,210422,210404,210327,210327,210305,210305,210327,210323,210324,210324,210323,210305,210402,210220,210220,210121,210117,210111,210111,210111,210111,210111,210102,201127,201127,201127,201027,201027,201029,201028,201014,201026,201026,201007,200616,200928,200928,200803,200911,200907,191206,191114,200513,200512,190515,191025,200312,200617,190720,190720,181130,191127,190119,181124,190327,190322,200513,191220,190725,190306,190215,181124,200320,200410,190425,191220,200227,190227,200410,190723,190227,200320,190321,200320,190506,200512,200203,200203,200403,191018,200319,200409,200222,200222,200221,180830,180906,191213,181228,200420,181214,191108,200221,190402,190515,191112,190109,240104,231231,231216,231221,231220,231115,231115,231106,231024,231020,231020,231015,231015,230912,230906,230823,230705,230705,230815,230809,230721,230710,230511,230510,230512,230512,230517,230410,230410,230323,230323,230208,230308,220510,230224,221016,221019,221019,221202,221128,221013,221010,220908,221008,221113,220823,221114,221114,221114,220907,221114,220908,221127,220912,220904,220523,220904,220506,220517,220904,220825,220502,220503,220927,220508,220522,220522,220522,220523,220829,220802,220517,220602,220415,220317,220423,220314,220415,220315,220415,220421,220305,220303,211221,220324,220324,220324,220323,220323,220323,211213,220212,220214,220212,211214,211202,211214,211215,211213,191218,200305,190114,200227,181221,191212,190111,211022,210824,210904,210610,210519,210408,210408,210409,210305,201203,201026,200906,191212,181128,200615,191023,200611,200611,181107,191018,190116,200409,191218,190913,181217,190418,191212,211202,200910,210404,201208,191012,200615,231007,230309,220915,231205,231125,231129,230925,230811,230422,230216,230308,230307,230306,221204,221008,221007,221116,220912,220522,220829,220829,220418,220303,211214,200305,200418,190506,211021,210904,210904,210718,210111,201127,201113,190823,190419,200911,190102,190704,200317,190915,181120,191218,190802,190118,190218,190306,210622,210622,210518,210519,200417,190425,181124,190321,200611,231029,231029,220823,230815,230821,210729,181130,211115,210716,210726,210622,210612,210517,210329,210325,201027,200610,231121,230825,230814,230622,230606,230424,230308,230306,220506,220818,220314,211202,181214,200704,200306,200214,200615,200511,190426,191111,200331,190402,190307,190319,190320,190114,190826,191016,211107,211011,211014,211008,210914,210903,210806,210729,210716,210811,210719,210728,210613,210520,210408,210518,210517,210618,210408,210509,210507,210508,210317,210405,210304,210324,210402,210305,210111,201202,201127,201106,201029,201014,201022,200910,200912,190809,181124,191127,200615,200615,200418,200418,200221,200103,200227,191220,190121,190422,190724,190723,190528,181130,181113,181130,181129,191125,200421,190724,190813,191108,191018,200108,190704,190219,200403,231225,231121,231125,231029,231006,230815,230815,230724,230606,230511,230517,230529,230419,230208,230131,221016,221206,220510,221010,221026,220907,221127,220912,220523,220523,220503,220508,220829,220422,220602,220422,220315,220315,211220,220104,220105,220111,220214,211213,231125,231029,230920,230809,230821,230711,230711,230518,230418,230308,221015,221202,221114,220523,220503,220503,220507,211218,220429,220429,220302,220207,220214,211214,191104,191025,190226,191112,200422,181221,200616,200704,200728,190415,190422,190624,200305,190916,211105,211011,211017,210914,210812,210810,210723,210625,210420,210512,210502,210403,210403,210323,210220,201207,201127,201029,201026,200929,200908,200302,191012,190425,200410,200320,190108,200911,181109,181109,190318,191018,190119,190107,181107,200518,191125,190520,190813,200610,191223,200317,191213,200514,200511,191217,191112,200305,190827,181130,181113,181204,190507,181210,190322,200422,190528,191213,190802,200615,200511,191004,191114,190103,200929,200615,190325,200519,190729,190827,190623,190802,200317,181130,200603,190426,191104,191022,200313,211206,211109,211109,211126,211106,211027,211115,211108,211126,211026,211014,211018,211015,211017,211007,210913,210913,211115,210925,210826,210820,210809,210810,210824,210809,210810,210622,210811,210726,210720,210802,210811,210704,210626,210622,210704,210618,210619,210610,210618,210611,210420,210611,210518,210517,210512,210507,210502,210317,210407,210407,210409,210405,210403,210329,210404,210325,210323,210305,200906,210327,210302,210220,210121,210103,210102,201204,201202,201104,201127,201113,201113,201027,201027,201027,201014,201026,200929,200910,200906,201014,200912,200907,200907,201023,200417,200616,190114,191012,200417,190402,200513,191025,191025,191212,190621,200306,181124,200320,200331,200214,181219,200320,200320,190819,190725,200415,200409,200514,190419,190528,200615,190724,190318,191018,191018,190716,181107,200319,190416,190102,200610,190510,181225,190109,191108,200413,200415,191021,200317,190624,190515,200302,231231,231020,231008,231007,230925,230915,230925,230823,230814,230622,230815,230725,230721,230720,230706,230613,230613,230511,230512,230518,230523,230529,230421,230424,230418,230415,230410,230209,230321,230308,230304,230302,230224,230216,221208,221015,221019,221204,220521,221023,221010,221115,221025,221026,221205,220905,221128,221009,220915,220823,220517,220823,220502,220503,220506,220912,220522,220808,220829,220829,220517,220818,220818,220423,220602,0002621107,220418,220416,220305,220302,220302,220325,220325,220324,220224,191019,190826,211014,211013,210925,210812,210726,210811,210408,210102,201208,201106,201014,201029,201023,190620,190416,190723,191018,190827,190827,231231,231225,231221,231227,231205,231129,231124,231121,231106,231102,231102,231020,231012,231007,230920,230920,230912,230925,230823,230814,230705,230815,230725,230721,230710,230724,230616,230606,230512,230508,230424,230410,230321,221023,230306,230306,221014,221015,220521,221023,221208,220925,221116,221128,220902,220915,220503,220502,220506,220522,220808,211219,220314,220314,220415,220421,220303,220302,211217,220323,220104,220105,220214,240104,231229,231229,231227,231207,231129,231129,231125,231121,231106,231102,231102,231029,231020,231015,231007,231006,230918,230925,230925,230925,230823,230817,230721,230710,230720,230616,230512,230529,230424,230424,221206,230418,230323,230308,230216,220925,221113,220907,220908,220825,220825,220523,220502,220502,220506,220502,220506,0002626702,220508,220508,220802,220602,220517,211218,211220,220422,220315,220305,220315,220315,220418,220429,220416,220223,220214,211202,211215,211214,211204,211106,211011,211115,211011,211011,211108,211026,211021,211017,211014,211008,210925,210925,210908,210926,210904,210902,210820,210810,210719,210717,210716,210709,210719,210729,210728,210723,210721,210709,210703,210618,210710,210621,210620,210609,210520,210420,210408,210407,210509,210420,210612,210507,210505,210508,210502,210317,210407,210422,210406,210404,210329,210327,210326,210324,210305,210402,210303,210302,210121,210111,210102,201205,201203,201013,201127,201113,201113,201103,191025,190507,200617,200511,190827,191021,200403,200107,181221,191217,190307,190319,191111,191019,190729,190402,190117,190117,191205,200317,190218,200313,200329,190506,190506,190716,190506,191021,200610,200617,190218,201007,200910,200908,200929,200912,200911,200907,200904,200103,191012,200304,190322,200513,200513,191220,190725,190121,200326,190425,181204,200227,190819,200611,191004,191004,190227,190215,191109,190619,190619,191217,190102,191226,190321,200611,191018,190723,190320,190107,190107,200928,190416,190416,181129,200222,191004,190813,190226,191019,190225,190514,200102,181228,181228,190913,191017,190913,190915,191111,190109,181217,190805,200518,200312,190415,200214,190624,191205,190623,190114,210329,201023,200417,190422,190225,200511,230402,230421,221026,220418,211215,181221,181214,190227,190624,190624,190416,190416,191202,191202,220915,220523,220304,211214,211214,200107,200420,200305,191022,191016,211109,211126,211017,211014,211007,210925,210904,210718,210823,210726,210625,210618,210609,210519,210612,210612,210612,210515,210403,210507,210318,210407,210407,210404,210326,210303,210102,210302,201207,201127,201113,201027,201023,201023,191228,191025,200329,200611,190813,200611,191226,191112,200108,200221,231227,231216,231121,231103,231012,230613,230522,230524,230411,230308,221015,221010,221113,220317,220414,220324,211203,211213,240219,230920,230920,230724,230711,230402,230402,230411,230323,221208,220524,221128,220720,220517,220314,211217,211210,211203,211213,231229,231227,23 1125,231115,231020,231103,231023,231008,231008,230920,230828,230811,230821,230721,230616,230510,230523,230419,230209,230309,221023,230304,230307,230224,221208,220521,221116,221022,221010,221010,221008,221113,221128,220915,220507,220508,220829,211218,211218,220423,0002621107,220416,220416,220303,211221,220207,211216,181214,181213,190320,190507,190507,190827,190827,200403,200420,200107,191217,191217,200511,200325,200204,200204,190402,190226,190226,190226,200331,200331,200617,181128,181204,200417,200304,191017,191016,191016,211107,211026,211107,211107,211107,211106,211027,211102,211105,211008,210914,210903,210902,210903,210809,210907,210709,210810,210810,210726,210718,210720,210720,210710,210704,210723,210622,210620,210620,210620,210613,210420,210519,210612,210612,210611,201208,210102,210103,210121,210127,210220,210517,210517,210508,210508,210507,210508,210508,210502,210420,210406,210404,210404,210329,210329,210324,210326,210325,210310,210310,210304,210302,210102,210102,210102,201207,201208,201104,210111,201127,201113,201103,201106,201027,201028,201014,201026,200910,200910,200906,200912,200907,200904,190322,190426,190325,191111,200302,200103,200103,190729,190729,190111,200513,191220,200610,200410,191127,200912,190620,190620,191112,191104,191108,191109,190227,191226,191226,200319,191018,181219,191125,190107,190813,191004,190418,190628,200222,200319,200222,191212,200416,200417,200420,200420,200102,200102,191112,190702,200604,181102,181106,181106,181106,200422,200312,231029,231020,230906,231012,230925,230721,230706,230606,230512,230522,230523,230224,230220,230304,230131,220521,221016,220524,221202,221013,221204,220925,221114,220907,221203,220904,220502,220825,220825,220506,220503,220424,220315,220718,220415,220305,220303,211217,220121,220111,211215,211215,211214,211204,211105,211027,211115,211011,211018,210926,210925,210908,210904,210902,210820,210810,210718,210717,210716,210709,210719,210729,210625,210710,210620,210620,210710,210520,210612,210610,210512,210408,210515,210512,210515,210520,210517,210512,210507,210505,210406,210326,210310,210323,210303,210303,210302,210121,210220,210111,210102,201205,201203,201013,201127,201113,201106,201029,201022,201026,190507,200617,200413,200413,181210,190827,200107,191111,190624,190802,191217,190117,190402,200204,200305,200329,190506,181220,190826,191021,200904,200103,191016,191111,201023,200911,200908,200906,200928,200911,200911,200907,200904,190628,190322,190322,190322,200513,191220,190111,191004,191004,190720,200326,200326,191127,190425,200103,200103,181204,190819,191109,190215,190619,200102,190102,191018,190723,191125,190320,190418,190418,190628,181129,190813,191108,190514,200102,181228,190913,190913,190915,190915,190704,200108,190805,200306,200306,190415,191111,200220,200518,191025,190515,190306,190425,190111,190306,171011,190425,181204,181219,190318,181107,181120,190702,200214,191012,191021,201007,200320,191112,200420,211021,210903,210824,210703,210612,210403,201126,231212,231029,230914,230825,230721,230512,230306,230306,221208,221022,220908,220904,211219,220415,211217,220224,181113,190704,230711,230220,200320,230613,231106,231103,231008,230725,230920,231006,230918,230829,230705,230822,190515,181213,181213,181213,181214,191218,190809,190813,190826,190912,200403,200403,200403,200108,200305,200617,200617,190319,190319,191112,190109,190109,200312,191021,191021,191021,190409,200214,190226,181113,200511,191206,191206,200331,200331,181221,191213,191213,190826,190826,200418,190307,191111,200106,200616,200616,200616,190623,190623,190114,191016,200313,211202,211210,211215,211126,211109,211011,211115,211108,211012,211012,211122,211122,211122,211102,211102,211022,211018,211126,210926,211013,211015,211012,211013,211006,211006,210926,210926,210913,210914,210903,210903,210824,210812,210812,210809,210717,210715,210806,210802,210802,210907,210823,210823,210904,210726,210726,210802,210721,210721,210704,210620,210626,210626,210621,210723,210621,210621,210718,210619,210620,210610,210613,210613,210519,210611,210420,210613,210520,210518,210518,210408,210515,210511,210613,210505,210505,210502,210409,210318,210519,210317,210407,210407,210407,210409,210405,210405,210403,210403,210328,210327,210325,210323,210305,210310,210304,210302,210302,210127,210117,201208,201208,201204,201202,201104,201104,201126,201109,201102,201102,201027,201027,201028,201014,201023,201007,201007,201007,200908,200912,200928,200928,200615,200911,200907,200907,200906,200904,200519,200519,191127,190426,190426,190426,200615,181128,200326,181218,200203,191012,191012,200102,200410,181109,200302,200302,200512,191023,190108,190108,190321,190321,191202,191202,200319,200319,190419,190107,200928,200227,200320,200222,200409,200409,190619,181205,181217,181219,181219,181219,190227,190227,190227,191109,191109,191109,200603,190528,190528,190528,190916,191223,191223,191226,191226,200420,200420,191017,190704,191112,191112,200323,200325,200325,200610,200610,190702,190702,191226,190215,190219,240108,240108,231229,231229,231221,231221,231227,231227,231205,231205,231207,231207,231207,231125,231125,231121,231121,231125,231125,231121,231103,231103,231029,231029,231012,230920,230920,230920,230914,230914,230914,230912,230914,230914,230914,230828,230825,230814,230814,230809,230809,230619,230619,230821,230821,230725,230725,230724,230726,230726,230710,230725,230616,230511,230512,230131,230508,230508,230419,230419,230410,230410,230209,230321,230323,230309,230306,230306,230306,220510,220510,230216,230131,220521,220521,221016,221206,221206,221206,220524,220524,221010,221013,221013,221010,221115,221025,221021,221115,221114,221114,221127,221127,221127,220906,220906,220523,220523,220825,220502,220829,220829,220824,220824,211218,220315,220509,220315,220509,220422,220602,220418,220418,220415,220302,211220,211220,211220,211221,220323,211210,211210,220304,220207,220207,220212,220212,211214,211216,211216,190826,190624,191019,190913,181129,190409,190426,181220,200103,190729,190322,190619,190725,190103,190116,181204,190704,190915,191025,231229,190624,190802,211027,210925,210902,210826,210729,210726,210625,210622,210622,210618,210612,210103,210103,201203,201126,201027,201022,190720,191012,191023,190915,231231,231229,231225,231216,231212,231205,231220,231121,231123,231115,231115,231015,231020,231012,230821,230925,230918,230914,230828,230815,230725,230721,230720,230720,230710,230616,230511,230508,230518,230523,230508,230421,230209,230308,221023,230306,230306,221208,220521,221019,221116,220908,220829,220506,220912,220508,220720,220829,220423,220423,220315,220317,220429,220415,220415,220302,220413,211219,211221,220324,211202,211202,220223,231102,230906,230814,230811,230821,210719,210719,200331,200422,200312,190415,200511,181210,191025,190415,190802,181113,181120,190507,200413,200418,200221,191213,190827,191019,190307,190624,200519,191228,190117,200617,190729,200313,190327,181112,200313,181130,181107,190218,190823,200616,200417,181220,211206,211106,211107,211022,211027,211018,211026,211202,211017,211007,210926,210914,210914,210908,210904,210826,210820,210809,210718,210716,210810,210823,210726,210802,210720,210823,210723,210704,210626,210622,210619,210612,210618,210612,210611,200912,210517,210512,210509,210502,210317,210519,210509,210327,210404,210406,210404,210329,210327,210326,210405,210323,210305,210402,210303,210303,210121,210111,210102,201207,201126,201202,210121,201013,200908,201113,201113,201102,201029,201022,201026,200910,200908,200908,200928,200803,202616,201023,190111,190306,190227,200514,190628,190912,200304,191220,190327,190725,181124,190419,191127,190121,181219,190819,200615,200222,200320,190227,200514,200409,200409,190306,190318,191018,190724,190716,200319,200326,200325,200421,190322,190119,200102,191125,200319,200221,181107,181107,200610,191104,191226,190320,181124,190813,190418,200513,190515,190702,191021,190913,190913,191223,191212,200107,200108,190109,231124,231121,231106,231024,220823,230920,230912,230705,230822,230809,230720,230711,230606,230524,230422,230411,230321,230323,230308,210728,230308,230306,230309,220510,230131,221014,221203,221022,221024,221025,221021,220908,220908,221009,220503,220507,220823,220829,211218,220509,220422,220415,220419,220414,220421,211221,220325,220323,220111,220214,190422,210420,201007,190322,190723,191212,191202,181205,230705,220502,210802,231223,231121,210726,210719,210723,210719,210704,210625,210621,210621,210618,210610,210613,210519,210519,210420,210609,210515,210511,210505,210509,210318,210317,210408,210409,210322,210304,210303,210302,210127,210111,201208,201204,201104,201113,201113,191218,191218,191019,200617,200617,200305,191217,191213,200511,200227,191206,191206,191022,200106,200514,200514,210318,210511,210511,210318,210317,210317,210409,210325,210310,210403,210127,210117,201013,201013,201109,201127,201027,201028,200929,200422,200511,190322,200312,200305,190307,200331,200511,181221,190111,200519,191213,191016,200313,211206,210907,211122,211109,211106,211027,211115,211102,211026,210720,211018,211015,211008,211007,210913,210913,211115,210914,210902,210812,210809,210824,210716,210810,210810,210810,210622,210811,210726,210720,210802,210802,210704,210625,210704,210619,210619,210720,210618,210611,210612,210612,210611,210518,210517,210507,210507,210509,210502,210317,210409,210329,210325,210323,210305,210302,210220,210102,201204,201202,201202,201027,200929,190218,200302,190319,190327,200320,190415,190225,190227,200319,200610,200421,181225,231231,231225,231124,231106,231102,231102,231025,231015,231008,231007,230925,230925,230915,230915,230906,230823,230814,230622,230624,230622,230815,230725,230711,230706,230613,230511,230512,230519,230523,230529,230421,230424,230415,230411,230209,230321,230321,230309,230304,230302,230302,230224,230216,221019,221208,221015,221019,221206,220521,221023,221128,221022,221115,221025,221007,220908,221114,221205,220907,221009,221009,220915,220904,220823,220502,220503,220506,220508,0002626702,220522,220808,220829,220818,220818,220423,220602,220317,0002621107,220429,220419,220415,220305,220302,220302,211221,211221,220325,220324,220207,220121,211210,211213,190515,190510,181210,190702,190415,190415,191212,190827,190218,191212,200106,191206,190111,190111,190912,190724,190327,191127,181124,190827,190227,190419,181219,191109,200222,190116,190321,191112,191017,191112,190809,190809,200910,231207,230613,230303,220915,210824,210408,210302,190416,190510,191218,191218,200422,190319,190319,200313,200108,190118,191021,200305,200305,190409,190409,191213,191213,191213,191206,200214,200331,200331,200227,190507,190507,190507,190226,190226,190226,200418,200616,190307,190307,190623,190623,200106,200313,200313,200514,200519,200519,200514,191022,200313,211106,211106,211109,211109,211108,211012,211122,210908,211126,211216,211126,211122,211108,211102,211022,211022,211018,211014,211012,211013,211006,210926,210926,210903,210907,210903,210824,210824,210812,210806,210717,210806,210806,210726,210823,210726,210802,210721,210721,210704,210703,210626,210723,210619,210610,210611,210611,210518,210518,210515,210515,210511,210505,210507,210509,210318,210317,210407,210409,210405,210403,210502,210329,210329,210117,210117,210323,210323,210323,210305,210303,210304,210302,210127,210127,210117,201208,201204,201204,201202,201202,201126,201126,201109,201102,201027,201027,201028,201028,201014,201014,201023,200803,200803,200912,200912,200907,200907,191205,191205,190119,190620,181128,181128,200326,200326,190426,200615,200616,200615,200616,181218,181218,200203,200203,200102,191111,200410,200410,190306,190422,190826,190620,190621,190620,190620,190621,190621,190621,190621,190621,181109,190108,190108,200302,200302,190321,200512,200512,200512,191023,191202,200319,200320,200319,190419,200227,200320,190819,190318,200409,190103,190103,190103,190227,191109,191109,181129,181120,191223,200413,191017,190724,190813,181106,181107,200323,190416,190416,200610,181213,181225,200413,200331,231229,231227,231229,231229,231229,231223,231223,231216,231216,231212,231212,231205,231220,231220,231121,231115,231115,231108,231115,231106,231106,231029,231029,231023,231020,231015,231006,230920,230914,230914,230914,230829,230906,230906,230906,230906,230825,230825,230822,230822,230810,230810,230624,230624,230624,230821,230821,230726,230726,230724,230724,230724,230720,230710,230616,230616,230606,230606,230510,230518,230518,230522,230524,230524,230424,230424,230418,230418,230411,230411,230411,230323,230323,230323,230208,230208,230208,230308,230308,230306,230304,230224,230224,230216,230216,221205,221014,221014,221016,221016,221016,221206,221206,221206,220524,221113,221113,221023,221204,221010,221010,221021,221007,221008,221007,221021,221021,221116,221115,221113,220510,220510,221128,220906,220915,220915,221009,220720,220502,220522,220522,220829,220802,220802,220802,220509,220423,220603,220603,220415,220315,220315,220315,220429,220429,220429,220419,220419,220419,220424,220305,220415,220415,220424,220323,220305,220326,220323,220325,220305,220302,220302,220326,220326,211221,211217,220324,211210,211210,220224,220224,220214,211216,211214,211214,211203,211203,211203,190515,200511,201007,191112,200304,181129,190322,191217,190827,200323,190226,190307,190826,200325,200403,190507,181221,190116,191016,190327,191213,190218,190218,190415,190916,190827,181130,200603,190802,181218,181213,190823,211206,211115,211026,211015,211008,210913,210913,210908,210914,210826,210809,210716,210810,210622,210726,210723,210625,210625,210609,210618,210519,210611,210517,210512,210507,210408,210502,210407,210409,210403,210329,210327,210325,210323,210305,200906,210220,210121,210102,201204,201202,201104,201126,201113,201027,201028,201013,200929,200929,200906,200928,200912,200906,200410,200317,190306,190108,190108,190121,200616,190327,191018,191212,190621,181124,190419,200214,181219,200511,190227,190227,200312,200320,190318,191109,200912,191018,200306,190528,191112,190318,181129,190102,190716,191226,190109,190109,190418,200326,191127,200304,200610,181221,181228,190913,200413,191108,200317,200415,231231,231102,231008,231006,230920,230814,230624,230606,230410,230209,230302,230216,220521,221013,221026,220508,220802,220302,220325,211214,211214,211213,190319,190319,181129,190426,190624,190121,181220,181220,190325,190720,190425,190422,181109,190322,181210,181213,181205,190116,190219,190809,190506,190416,181113,190623,190813,190226,190528,200331,191213,190729,191111,200106,190506,190318,200304,190826,181109,181107,190819,200222,190915,191108,181106,181228,181214,191111,190109,211007,190426,231221,231123,231008,230828,230710,230418,230209,230216,221010,220904,220303,220111,231229,231224,231220,231227,231121,231207,231220,231121,231121,231106,231029,231029,231020,231008,231006,230914,230914,230508,230825,230814,230810,230724,230706,230510,230529,230420,230307,230306,230220,221016,221128,221008,221204,220905,220907,220904,220502,220507,220824,220314,220603,220315,220302,220323,220121,220304,220212,211215,211202,211202,211022,211011,211011,211122,211102,211102,211022,211015,210907,210914,210926,210908,210903,210826,210907,210902,210718,210719,210806,210726,210729,210723,210622,210621,210626,210622,210620,210622,210610,210519,210520,210518,210517,210515,210511,210507,210507,210502,210407,210409,210405,210403,210327,210405,210305,210304,210303,210303,210127,210127,210117,210103,201208,201208,201202,201104,201202,201113,201109,201103,201029,190422,200222,191111,200422,190913,190510,190510,181214,181214,181214,181221,181221,190322,191111,200102,200615,200107,200107,190219,190226,191019,190802,190624,190226,190226,200305,200519,200313,190729,181130,181130,200418,181220,200615,191228,191212,200320,190823,190826,190306,190306,190306,201026,200929,200906,200928,200928,200704,201007,200904,190915,200513,190426,190327,181218,181128,190327,190823,190327,200410,191112,191112,190108,191004,200512,190619,191127,190725,200417,200320,190628,200320,190327,200611,181107,200403,181219,191004,190813,190813,200928,190225,190227,190215,181130,190102,190318,191018,191018,191213,191108,191112,200415,191223,190704,190218,181106,200415,200421,201103,201014,190108,200422,190319,181210,190418,190827,200617,200422,190624,200107,200108,191218,191218,191111,190802,190422,190307,190507,191228,200325,200204,190402,191019,200331,190117,200313,181113,190402,200519,200418,190426,181218,191212,190823,211126,211107,211109,211027,211115,211108,211026,211018,211013,211007,210926,210914,210907,210904,210902,210820,210809,210904,210718,210716,210709,210823,210811,210723,210723,210802,210704,210710,210619,210610,200906,210519,210408,210518,210515,210507,210509,210318,210317,210422,210409,210404,210403,210327,210404,210117,210323,210402,210303,210220,210121,210111,210103,201208,201203,201013,201208,200803,201127,201204,201106,200616,200811,201027,201022,201026,201007,200910,200910,200928,200803,200908,200906,190628,190306,200103,191025,190119,190827,200803,200305,200329,200329,190322,190325,200514,191127,190109,190111,190912,190720,181105,181128,190826,200102,200514,191127,190723,200304,200512,191112,190107,191226,191226,181109,190102,181225,200227,190823,190819,190318,190422,190422,181205,190724,191018,200611,190227,190915,200222,190724,200513,181228,190520,200420,190913,190915,200204,190913,191223,190619,190109,200604,190515,190215,191226,181102,240104,231224,231216,231212,231220,231115,231108,231108,231029,231015,220823,230912,230825,230810,230810,230821,230724,230711,230606,230512,230523,230402,230418,230209,230309,230302,230216,221016,220521,221013,221024,221204,221021,220907,220907,220907,221127,220904,220503,220508,220808,220818,220314,220315,220414,220305,220212,220413,211214,211221,220207,220304,220304,220214,220121,220111,220223,220212,211215,211213,231205,210902,210612,210409,201208,191226,231221,230809,200511,190102,201007,181113,181204,181204,200511,190327,190117,191022,190402,181130,181130,181221,200418,211027,211102,211022,211008,210925,210810,210824,211007,210902,210820,210810,210806,210726,210908,210703,210703,210618,210610,210609,210518,210505,210508,210318,210508,210422,210406,210422,210405,210402,210327,210127,210127,210102,201207,201203,201202,201113,201022,201022,201023,201022,200514,190119,181128,181218,191127,190215,190422,200409,190103,181204,191112,181109,191023,190327,190107,200611,181205,190225,181228,190102,191025,190103,190321,200420,181107,190215,190226,231231,231123,231007,230814,230624,230529,230512,230519,230619,230822,230309,230511,230523,230421,230224,230216,221208,221015,221019,221206,220524,221128,221013,221010,221010,221026,220905,220905,221127,220829,220802,220418,220415,220416,220305,220305,211213,211213,211204,231115,230920,230606,230508,230306,221022,220720,220824,0002621107,220418,211203,211213,191021,190307,190307,181130,211021,210907,210612,200728,200912,210502,210403,201204,201026,200222,200319,200615,190619,200403,211011,211021,211021,181107,231125,210719,210327,200519,181124,200410,190619,200615,190103,191223,190809,191217,191213,191114,211021,211126,211109,211105,211115,211115,211022,211122,211021,211022,211022,211018,211014,211007,210914,210903,210914,210903,210726,210812,210806,210717,210823,210729,210721,210811,210802,210626,210621,210610,210611,210611,210515,210318,210407,210328,210310,210303,210303,210302,210127,201208,201104,201102,201022,201007,201029,200912,201022,200514,200616,200615,200410,200319,200227,200227,200221,200320,191226,200610,231229,231225,231225,231225,231216,231205,231125,231125,231123,231115,231115,231115,231023,231012,231007,231007,220823,220823,230914,230914,230912,230912,230828,230823,230810,230810,230906,230810,230705,230810,230822,230822,230809,230720,230710,230710,230606,230606,230510,230522,230522,230523,230402,230419,230419,230419,230323,230308,230308,230308,230307,230307,230303,220510,230216,221205,221205,221016,220524,220524,220524,221023,221013,221013,220925,220925,220720,221127,221127,221127,220904,220825,220824,220509,220509,220419,220424,211220,220424,220302,220424,211220,211217,211214,211221,220325,220323,220323,220105,220304,220223,220223,211203,190802,200326,190325,200102,211109,200513,220508,220508,220413,211203,211203,211213,211007,200418,211122,210621,210408,210611,210611,210611,210408,210520,210502,210407,210405,201023,200803,201022,200410,191218,191218,210806,210619,210611,210518,210317,210317,210220,190912,200616,200616,200403,211027,211026,210907,210408,210304,210304,210103,210103,200912,200912,200305,191112,190619,181109,191025,191025,200409,200222,190402,220904,220523,220315,211217,220214,220214,200704,190912,181120,181120,200305,200413,200603,190624,190117,181130,200313,200302,181220,191212,190823,191016,191016,211105,211106,211011,211011,211026,211013,211006,210914,210908,210826,210806,210824,210719,210723,210718,210625,210618,210620,210704,210619,210620,210620,200906,210520,210420,201026,201027,201027,201102,201103,210512,210507,210422,210404,210422,210326,210117,210326,210405,210305,201106,210303,210111,210102,210127,201013,201207,200615,201029,201026,200906,200929,200912,200912,200904,200904,190111,190111,200513,200304,181124,181128,190725,200203,200409,190227,190422,191025,191111,190419,190528,181109,191018,200319,190116,190318,190318,190321,190327,190327,190322,190322,190418,190117,190109,190107,190119,181129,200222,190813,200610,190320,191111,181214,190913,191017,191223,191223,200420,181102,181102,200317,200317,200317,231223,231212,231121,231020,231015,231006,230914,230616,230616,230616,230616,230606,230606,230522,230522,230420,230418,230418,230323,230309,230302,221016,220524,220524,220524,221023,221024,221116,221021,221114,221007,221114,221024,221024,220907,220907,221127,221009,220906,220825,220823,220502,220507,220507,220522,211218,220509,220509,220818,220423,220422,220315,220315,220418,220418,220413,220303,220413,211220,211217,220325,220323,220323,220323,220304,211213,220111,220223,220214,211216,211203,211213,211214,211214,181214,200904,190409,200615,200326,190218,200603,191212,211011,211018,211006,210904,210826,210810,210704,210618,210611,210509,210409,210329,210405,210304,210303,201102,200803,191114,190322,200514,181219,190215,191111,190826,191004,190723,190819,191112,191018,181228,200911,200518,190827,210907,190121,230711,240104,231212,231008,230823,230711,230706,230510,230512,230619,220524,221202,221202,221021,221025,221113,220523,220502,220808,220818,0002621107,220305,220326,220323,220224,211213,240104,231227,231205,231121,231020,231020,231103,231025,231012,230914,230829,230823,230821,230710,230508,230518,230420,230411,230208,230303,221014,220524,221021,221010,221114,221127,220825,220502,220508,220421,220302,200422,190510,190510,200420,200312,181204,211013,210508,210728,210408,210409,210111,201102,200616,200929,201023,181218,190215,181101,181205,230220,190623,190402,181128,190121,200313,200416,200331,181130,200603,190628,210925,210826,210703,210626,210515,210326,201113,201027,200929,201023,191114,191104,191212,200513,191220,190826,181218,190409,190327,190425,190826,200610,190915,231207,231129,230724,230424,230419,221014,221025,221026,220904,220523,220718,211221,220104,211204,211105,211105,211107,211012,211011,211108,211014,211008,211008,210925,210914,210907,210904,210826,210820,210810,210719,210723,210721,210703,210703,210625,210609,210520,210612,210420,210408,210517,210512,210512,210507,210505,210505,210508,210508,210318,210318,210406,210406,210404,210329,210329,210329,210328,210328,210326,210326,210310,210402,210303,210302,210302,210121,210102,201207,201207,201203,201013,201127,201127,201113,231225,231229,231229,231223,231216,231115,231115,231103,231029,230918,230918,230914,230925,230825,230828,230814,230809,230706,230821,230721,230720,230720,230710,230710,230616,230606,230309,230309,230303,230224,230131,221014,221114,220907,220908,220908,220915,220523,220823,220824,220808,220509,220509,220818,220314,220315,220418,220414,220416,220424,220303,211220,220325,220325,220323,220224,220224,211216,220212,211202,211215,211214,191025,191025,190219,200617,200617,181120,200413,200413,190510,190507,181221,191218,190827,190827,200107,200107,181221,190307,190307,200403,200107,191111,190624,190624,190624,191228,190117,190117,190802,190802,200519,200511,200511,191217,190117,190117,200305,190402,190506,191205,200204,200204,200313,181130,200329,200329,190218,190506,190506,190716,181220,191212,190826,191021,200904,200616,200417,190628,190306,190306,191111,191111,201007,200929,200929,200908,200908,200908,200906,200906,200928,200928,200912,200911,200911,200907,200907,200904,200519,200904,190628,200103,200304,200304,200513,200513,190111,190111,190111,190725,190121,190720,190720,200326,200326,190327,190327,191127,191127,190415,190425,190425,200227,200203,190819,190819,191021,200615,200615,200611,200611,181204,191127,191127,191004,191004,191004,190215,191109,191111,191109,190215,190215,190619,200204,190102,190108,200302,191226,190321,200421,200421,190102,190723,190723,190723,191125,190107,190320,181129,191125,191125,200603,190418,181129,200222,200610,191108,191218,191108,191218,190418,190813,190813,190225,190225,190225,190514,190514,181228,181228,190913,190913,190913,191223,191223,191223,200312,200317,200317,200317,200323,200325,190704,190704,190704,200304,191112,191111,190116,181106,181102,200306,200306,200409,190415,200220,190109,190109,190805,190805,200312,200312,200518,200518,191025,200803,191127,191127,200329,190620,200203,200203,200410,190620,190620,200302,200416,191012,191127,190819,200611,191017,200220,200220,200319,190916,200416,190704,190716,200420,200420,191112,200317,200317,190416,190416,200604,200603,200413,200331,190510,191019,190916,211006,210811,210515,210323,210322,210304,210220,210117,210302,210117,201013,201127,201126,201014,200320,200203,200409,190619,190620,191109,190119,190915,190805,190913,200518,210626,210326,210511,190226,190319,190422,190409,190118,190802,181113,190402,190327,190506,181220,181221,190422,190515,190823,190111,181113,190725,191228,181221,190426,181220,190802,190720,210926,210613,210422,201013,201126,201027,190415,181124,181124,190725,190102,181219,191125,190116,190116,181109,190520,190915,231216,231207,231129,231115,231008,230918,230918,230829,230815,230706,230323,220720,211216,211216,231108,231129,231115,200422,200305,200331,200617,200615,200728,200313,200320,191212,200203,211106,211027,211202,211115,211015,211014,211013,210903,210907,210812,210824,210723,210718,210622,210609,210612,210518,210408,210505,210317,210406,210328,210329,210324,210305,210220,210102,201202,201113,200421,201102,201026,201026,200910,200911,200410,200513,200415,200610,200317,220524,221013,221021,221008,221010,221114,220720,221127,220902,220825,220503,220522,220802,220421,220315,220419,211220,220323,220304,220105,220212,211214,190118,190515,200306,190219,211105,210324,230508,220906,191114,200108,200403,191217,190226,200326,181130,200418,181214,191016,211011,211013,211006,210904,210809,210620,210420,210317,210405,210323,210402,210302,210103,201203,201026,201202,201029,201029,201007,200928,200907,201023,190621,190121,191021,200320,191127,190318,190819,200611,190724,181204,200604,190620,200413,190327,231216,231231,231125,231121,231025,231015,230508,230209,230208,220521,221202,221009,220904,220904,220507,220824,211218,220423,220602,220303,220121,220304,220304,220223,211215,190702,190426,190322,190510,190720,181128,190119,190425,181218,190215,181124,190103,190619,190321,190723,190107,190422,190422,181205,190318,190416,190520,190520,190117,190107,190102,190704,181106,190416,181129,190619,190624,190402,231225,231224,231216,231123,231129,231102,231025,231015,231012,231006,230920,230912,230906,230823,230811,230705,230822,230809,230724,230721,230721,230711,230616,230606,230510,230512,230519,230524,230422,230419,230415,230209,230321,230307,230304,230302,230220,230131,230131,221015,221015,221016,221208,221206,220524,221202,220502,220523,220808,220423,200403,200108,191112,200511,200617,211210,211109,211011,211006,210824,210824,210812,210715,210810,210704,210511,210318,210317,210304,210117,201208,201104,201126,201027,201028,200929,201023,200803,200410,200512,200611,200102,200221,191223,200604,231227,231221,231207,231125,231125,231129,231129,231115,231115,231115,231115,231106,231106,231102,231102,231020,231012,231012,231006,231006,231006,230918,230918,230918,230828,230925,230920,230920,230810,230925,230825,230811,230705,230705,230822,230822,230809,230809,230809,230720,230720,230809,230809,230711,230508,230419,230410,230209,230209,230304,230131,221014,221206,221203,220524,221023,220502,221113,221204,221024,221024,221021,221021,221008,221021,221115,220907,220907,220912,220904,220904,220904,221009,220823,220915,220507,220522,220808,220808,220509,220718,220423,220418,220315,220418,220418,220414,220414,220421,220305,220305,211220,211221,211221,220323,220323,211210,211216,211202,190409,190528,220111,200518,211027,211026,210903,210903,210718,210726,210303,210303,200417,200320,191218,210328,210303,231123,231015,230705,230616,230529,230308,221116,211213,220304,190723,190402,181120,181228,181120,181120,190321,181204,181130,181220,190111,181124,190121,190415,181204,190102,181107,190116,190116,181113,181120,190218,190118,181106,210121,181225,190118,190409,181113,190724,181221,190422,190111,191104,181128,190415,190723,181218,191023,181120,190108,190107,190628,190318,181205,190322,191017,181228,181221,181106,190416,190416,181205,190119,190108,190102,190102,181205,180810,180814,231223,231220,231115,231103,231008,231008,230920,230706,230726,230616,230508,230508,230418,230418,230321,220510,221014,220524,221204,221116,220908,221127,220507,220517,220317,211218,220818,220415,220419,211217,220104,220121,220304,220223,181120,200421,200422,181214,190306,211122,211122,211108,211126,210913,210903,210709,210719,210622,210612,210612,210518,210507,210318,210407,210328,210324,210402,210220,210102,201127,201022,201022,201023,201023,190723,190119,190322,190913,191111,200320,190320,200302,200203,200416,191023,190813,190619,190724,191017,191125,200604,231227,231225,231225,231125,231124,231125,231108,231108,231102,231025,231020,231012,231008,230811,230619,230619,230809,230619,230619,230616,230606,230510,230518,230523,230524,230529,230402,230422,230419,230402,230323,230323,230308,230306,221202,221023,221013,221202,221010,221116,221021,221007,221010,221115,221115,221114,220907,220908,221127,220906,220825,220503,0002621107,0002621107,220429,220429,220424,220302,211220,220104,220214,211216,211216,211203,211215,211018,211026,211021,211017,211013,211007,210926,210925,210908,210904,210902,210820,210810,210718,210717,210716,210709,210729,210728,210723,210625,210710,210620,210620,210609,210520,210519,210406,210515,210329,210329,210327,210328,210328,210326,210325,210324,210305,210402,210402,210303,210302,210111,210102,210121,201207,201203,201013,201208,191202,190226,190724,200704,200617,200617,181120,200511,200107,181221,190827,190827,200418,190624,190418,191019,190729,191217,190117,191205,190218,190218,181130,200329,200103,191114,181220,190823,200904,200616,200616,200417,190426,190306,201007,200929,200908,201022,200904,190823,190111,191012,200304,200513,191220,191004,190916,191127,191004,181124,200320,200326,190419,190121,181204,181204,200227,190819,200611,200518,190215,190422,190320,190619,190102,200421,191025,181120,190723,191125,191125,181109,181109,200928,181129,200222,190415,190813,190225,200416,190515,190915,191202,190913,190704,190109,190327,200304,190805,181106,200312,200214,200518,200413,200421,200422,190912,190912,200107,200107,190117,181113,190307,181220,190729,211017,211204,211122,211026,211014,210908,210715,210728,210729,210802,210703,210619,210518,210519,210408,210507,210323,210323,210127,201127,201127,201029,201029,201007,200904,200904,191114,191114,190426,190322,190111,190111,191220,190121,200615,200611,190419,200302,190704,200610,200610,190111,231227,231212,231212,231129,231029,231012,231012,230914,230912,230829,230914,230810,230724,230606,230529,230411,230216,230321,230224,230224,221208,221208,221021,221010,220908,220904,220507,220802,220808,220422,220416,220324,220324,220212,231125,231020,231008,230828,230811,230810,230721,230606,221014,221016,221203,221021,221025,220908,221128,221128,221127,220508,220416,220324,220212,190912,190912,190623,211202,211105,211014,210508,210718,210519,210420,210324,210302,201208,201127,201027,201014,200906,200907,200519,190121,190121,200403,191226,191111,231008,191025,181120,190912,190218,181221,190724,190306,190306,211122,211011,211108,211026,211022,210508,211014,211013,210926,210903,210820,210806,210806,210729,210726,210723,210721,210728,210703,210704,210618,210520,210520,210612,210420,210511,210511,210505,210408,210318,210422,210409,210405,210403,210305,210324,210324,210402,210220,210103,201013,201013,201106,200803,200904,190215,190 725,191016,190426,181210,181128,190425,181109,190510,190416,231231,231221,231212,231125,231125,231006,230912,230823,230814,230726,230622,230725,230809,230720,230606,230512,230518,230523,230529,230421,230424,230418,230410,230209,230309,230303,220510,221014,220521,221019,220524,221128,221013,221010,221025,221114,220907,221009,220912,220904,220823,220506,220912,220912,220517,211218,220423,220415,0002621107,220418,220419,220415,220303,211217,220104,220105,220212,220224,220212,211213,211213,190805,200413,190510,191218,200511,201022,200929,200803,200417,191212,190720,200418,200615,200410,200410,191004,190619,191112,191023,190823,200512,191125,191111,190725,200420,190723,200317,231216,231223,230914,230725,230720,230424,230415,221007,220906,221009,220429,220415,210726,210824,210823,210721,210704,210618,210520,210518,210502,210318,210409,210326,210326,210304,210327,210302,210127,210111,210102,201208,201027,201022,200904,190805,190624,190319,200511,200511,190218,191212,200728,200616,200331,190506,190623,191212,200107,190422,190325,181220,181213,201007,200929,200906,200911,191016,190111,190426,200417,191127,191212,181124,200103,191220,190912,200326,200320,190422,190819,190819,190227,200102,191112,200417,190915,190813,191109,200107,200415,200511,200319,200403,181213,191213,200227,190506,190507,191206,191206,181128,190823,211108,211015,210719,210625,210812,210721,210621,210420,210420,210408,210318,210103,201028,190325,200514,191127,191205,200326,190426,200203,200410,191111,200302,190108,200319,190419,200320,200203,191017,200317,200323,200323,200518,200610,190704,200320,190219,190219,231223,231212,231123,231115,231029,231025,231020,230920,230914,230829,230823,230906,230825,230821,230725,230724,230720,230616,230606,230512,230508,230518,230522,230522,230523,230529,230420,230420,230418,230323,230308,230304,230131,221014,221206,221206,221128,221013,221007,220906,220904,220502,220802,220802,220603,220416,211220,220207,211215,190621,191114,200312,200325,191114,191114,191114,191114,200325,191125,191125,181129,200422,200617,200511,191108,200304,181129,190507,181214,200422,190528,190322,191213,190827,190802,200615,200305,190409,190226,200305,200326,190422,190319,200331,190121,191016,200904,200313,190325,200519,190729,181025,190623,200106,200317,181130,200603,200329,190426,181210,191228,200320,211206,211109,211106,211106,211106,211027,211115,211102,211026,211018,211015,211012,211007,210914,210826,210809,210718,210806,210729,210810,210622,210704,210703,210619,210609,210611,210612,210420,210618,210611,210518,210517,210512,210507,210509,210303,210502,210317,210407,210409,210406,210403,210329,210404,210325,210305,210305,210305,210302,201104,210220,210303,210121,210103,210102,201204,210121,201104,210117,201126,201127,201113,201102,201027,201103,201027,201014,201026,200929,200910,200906,200906,201028,200912,200907,200811,200811,200417,190218,190306,190306,200912,200616,190227,190823,191004,190402,200513,191220,190621,200103,200306,190121,181124,200320,181219,200320,190819,190725,200409,190227,190419,200912,190318,200416,200421,191018,190716,181107,190528,191112,190416,190102,200610,181129,200220,190510,190418,200319,190704,191021,181214,190109,191108,200415,190913,190515,221128,221009,220506,220424,220802,220802,220802,220802,220802,220517,220422,220602,211218,220603,0002621107,220429,220416,220421,220303,220303,220302,220302,220325,220325,211221,211203,191021,191212,190802,200313,181130,190913,211204,211027,211018,210903,210806,210520,210403,210102,201127,201208,201106,200906,190913,190419,190419,200513,200409,190913,190225,190520,200610,231227,231006,221014,220907,220902,220720,220720,220423,220414,220121,201202,201027,190422,200416,191111,200204,200907,200417,200611,190227,190227,200512,200319,181221,190402,190307,200317,190114,190111,190306,190419,181205,190704,210704,190515,181213,190319,190318,190226,190422,190416,191016,191017,190729,190506,190506,181204,200103,191111,190327,190327,190425,190320,190108,190725,190227,190227,190916,190723,190416,190422,181129,191018,191018,190116,190813,191125,191017,190510,181128,181213,190520,191218,190409,190827,200331,200103,200221,200331,181220,200617,200616,190515,190805,190111,211022,211021,211018,210913,210913,210908,210925,210907,210810,210622,210823,210618,210610,210610,210610,210611,210519,210409,210508,210326,210323,210302,210103,210103,201027,201027,201029,201029,200929,190111,190119,200912,190111,190720,190823,181124,190425,191023,181109,190422,200102,190416,191112,200417,200204,191108,240104,231220,231121,231121,231106,231020,231008,230825,230823,230828,230725,230613,230606,230606,230518,230420,230419,230323,230307,230304,221014,221015,221021,220502,221008,221026,220906,220904,220912,220818,220818,220315,220718,220602,220421,220207,220121,220111,220224,181225,191213,200222,200317,190826,190623,190716,210304,201109,201029,201023,200417,200106,190528,190720,190823,190826,190306,190813,190320,191125,190915,190809,191202,210515,230302,210127,200331,200416,200214,200214,181120,181221,200108,190402,200617,200519,200519,190729,181130,181220,211204,211106,211022,211109,211018,211017,211017,211014,211008,211006,210914,211115,210902,210902,210820,210812,210810,210826,210709,210823,210729,210728,210721,210626,210704,210710,210704,210609,210612,210420,210613,210310,210317,210502,210409,210406,210404,210404,210329,210329,210326,210310,210220,210121,210111,201207,201208,201203,210117,201208,201113,201203,201102,201029,201014,201007,200929,200908,200906,201022,200302,200304,200304,181124,181124,190121,181204,181204,190619,190619,190102,190107,191112,190724,190916,190724,191004,190724,181107,181129,190219,200417,200409,200409,200409,200416,190107,191112,190109,181102,231229,231229,231207,231124,231124,231108,231102,231025,220823,231006,230920,230920,230619,230825,230825,230706,230706,230809,230711,230710,230710,230711,230619,230606,230606,230524,230524,230529,230402,230424,230415,230410,230321,230321,230307,230304,230304,230302,230131,221014,221015,221019,221204,220524,221023,221116,221013,220912,221008,221205,221021,221115,220907,220915,220902,220825,220503,220503,220507,220507,220508,220424,220720,220823,220823,211219,220818,220509,220718,220422,220421,220421,220302,220413,220121,220121,220304,220304,211216,211214

Distribution

Distributed nationwide and internationally to Canada, Mexico, New Zealand, Belize, and Hong Kong.

Quantity

135,999 cases (6,799,950 units)