ZAP-X Radiosurgery System, Model: 300150
Enforcement
- Recall Number
- Z-2639-2025
- Event ID
- 97413
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Zap Surgical Systems
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 8, 2025
- Initiation Date
- July 25, 2025
- Classification Date
- September 26, 2025
- Address
- 590 Taylor Way, N/A, San Carlos, CA, 94070-6307, United States
Description
ZAP-X Radiosurgery System, Model: 300150
If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.
UDI-DI: 00860183001504, Serial Numbers: ZUC17153, ZUC20156, ZUC20157, ZUC20159, ZUC21162, ZUC21164, ZUC22166, ZUC22169, ZUC22170, ZUC22171, ZUC22173, ZUC22174, ZUC23176, ZUC23177, ZUC24179, ZUC24180, ZUC24181. Software version: TDS V1.10.1
Worldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.