3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482V; designed to qualify and monitor dynamic-air removal steam sterilization processes
Enforcement
- Recall Number
- Z-2639-2023
- Event ID
- 92915
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- 3M Company - Health Care Business
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- N/A
- Report Date
- October 4, 2023
- Initiation Date
- August 22, 2023
- Classification Date
- September 26, 2023
- Address
- 3m Center 2510 Conway Ave, Bldg 275-5w-06, Saint Paul, MN, 55144-0001, United States
Description
3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482V; designed to qualify and monitor dynamic-air removal steam sterilization processes
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.
UDI/DI: 50707387787276( (shipper); Lot Codes: 33P3EF, exp. 12/27/2024; 33P4ND, exp. 01/01/2025; 33P4NR, exp. 01/01/2025; 33P5CA, exp. 01/03/2025; 33P5XT, exp. 01/06/2025
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.
172,464 devices