FDA Enforcement Class II Ongoing

Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

Recall: Z-2635-2023 · Reported October 4, 2023

Enforcement

Recall Number
Z-2635-2023
Event ID
92973
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Fresenius Kabi USA, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
October 4, 2023
Initiation Date
April 22, 2022
Classification Date
September 25, 2023
Address
50 High St Ste 50, N/A, North Andover, MA, 01845-2620, United States

Description

Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

Reason

Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set

Code Info

(UDI): 00811505030122 Software Version: 5.0.0 and 5.0.1

Distribution

US Nationwide distribution in the states of NJ, WI.

Quantity

1344 pumps