FDA Enforcement Class II Terminated

ROTH MOBELI Grab Handles used with EOS System: MOBELI Stabi Vario Grab Handle, ROTH Catalogue Number 14002 96 S, EOS Imaging Catalogue Number 9000111773; and MOBELI Dual Grip Grab Handle, ROTH Catalogue Number 14002 25 S, EOS Imaging Catalogue Number 9000111774.

Recall: Z-2626-2017 · Reported July 5, 2017

Enforcement

Recall Number
Z-2626-2017
Event ID
77424
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
EOS Imaging
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 5, 2017
Initiation Date
May 30, 2017
Classification Date
June 23, 2017
Termination Date
October 9, 2018
Address
4 Ieme Etage, 10 rue Mercoeur, Paris, N/A, N/A, France

Description

ROTH MOBELI Grab Handles used with EOS System: MOBELI Stabi Vario Grab Handle, ROTH Catalogue Number 14002 96 S, EOS Imaging Catalogue Number 9000111773; and MOBELI Dual Grip Grab Handle, ROTH Catalogue Number 14002 25 S, EOS Imaging Catalogue Number 9000111774.

Reason

ROTH MOBELI grab handles could detach and patient could fall if the EOS system operator does not follow instructions. Before each use, the operator must examine the vacuum safety indicator to check the vacuum is monitored and sufficient and the grab handle correctly attached. Do not use grab handles to help patient up. The handles are only for stabilizing the patients hand or arm in the EOS cabin.

Code Info

ROTH Catalogue Numbers 14002 96 S and 14002 25 S only,

Distribution

US Distribution to the states of : Pennsylvania, Massachusetts, Minnesota, Alabama, Michigan, Texas and Washington.

Quantity

8 units each