SOMATOM Definition Edge with software version VA48A-SP2; Model # 8098027 computed tomography x-ray system

Recall: Z-2626-2016 · Reported August 31, 2016

Enforcement

Recall Number
Z-2626-2016
Event ID
74840
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 31, 2016
Initiation Date
July 13, 2016
Classification Date
August 23, 2016
Termination Date
January 24, 2017
Address
40 Liberty Blvd, Malvern, PA, 19355-1418, United States

Description

SOMATOM Definition Edge with software version VA48A-SP2; Model # 8098027 computed tomography x-ray system

Reason

Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and the desired examination result is not achieved. This error only affects the automatic mode or coupled mode and does not affect the manual control of the injector.

Code Info

Model # 8098027 Serial # 96020, 96047, 96011, and 96029

Distribution

Distributed to: MI, NY, CA, KY, ND, NE

Quantity

4 systems