FDA Enforcement Class II Ongoing

Allura Xper FD10/10 Product Codes: (1)722011, (2) 722027, (3) 722005; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Recall: Z-2624-2025 · Reported October 1, 2025

Enforcement

Recall Number
Z-2624-2025
Event ID
97501
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 1, 2025
Initiation Date
September 3, 2025
Classification Date
September 25, 2025
Address
Veenpluis 4-6, N/A, Best, N/A, N/A, Netherlands

Description

Allura Xper FD10/10 Product Codes: (1)722011, (2) 722027, (3) 722005; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Reason

Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

Code Info

Product Codes: (1)722011, (2) 722027, (3) 722005; UDI-DI: (1) 00884838059047, (2) 00884838054196, (3) N/A; Serial numbers: (1) 157, 268, 95, 107, 59, 319, 49, 170, 173, 101, 84, 155, 133, 234, 345, 332, 102, 270, 177, 132, 325, 15, 87, 255, 171, 362, 168, 275, 344, 324, 163, 103, 215, 328, 286, 81, 293, 294, 224, 117, 334, 338, 146, 145, 266, 21, 202, 165, 367, 372, 340, 318, 98, 211, 158, 271, 204, 246, 206, 321, 295, 104, 217, 265, 261, 125, 282, 216, 333, 329, 230, 198, 244, 292, 180, 183, 314, 337, 341, 376, 377, 378, 259, 347, 287, 99, 233, 156, 71, 77, 188, 223, 238, 153, 129, 278, 118, 186, 108, 89, 258, 310, 274, 139, 141, 86, 300, 88, 309, 227, 220, 242, 166, 299, 273, 190, 189, 260, 280, 74, 75, 251, 228, 100, 199, 194, 91, 116, 112, 114, 43, 313, (2) 123, 133, 134, 183, 106, 169, 107, 42, 101, 209, 153, 40, 156, 112, 1, 104, 71, 175, 116, 174, 29, 50, 56, 152, 43, 49, 202, 178, 127, 39, 27, 64, 78, 99, 6, 86, 7, 145, 102, 44, 142, 30, 79, 38, 85, 181, 144, 177, 165, 201, 3, 68, 51, 21, 23, 24, 90, 171, 45, 122, 33, 137, 163, 150, 96, 155, 20, 166, 147, 139, 91, 136, 193, 69, 65, 168, 206, 158, 66, 75, 62, 138, 121, 18, 31, 131, 195, 192, 125, 148, 149, 46, 47, 135, 159, 100, 103, 110, 57, 274, 61, 36, 151, 300, 167, 60, 81, 126, 58, 184, 5, 188, 94, 154, 172, 35, 76, 95, 187, 170, 146, 162, 82, 117, 48, 205, 182, 132, (3) 212, 435, 128, 323, 269, 342, 531, 403, 287, 457, 373, 332, 374, 428, 75, 499, 511, 473, TPE, 333, 237, 338, SN545, 159, 100071298SN408, 96;

Distribution

Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.

Quantity

286 units (67 US, 219 OUS)