FDA Enforcement Class II Ongoing

VITROS XT Chemistry Products ALKP Slides

Recall: Z-2624-2023 · Reported September 27, 2023

Enforcement

Recall Number
Z-2624-2023
Event ID
92928
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ortho-Clinical Diagnostics, INc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 27, 2023
Initiation Date
August 5, 2023
Classification Date
September 21, 2023
Address
513 Technology Blvd, Rochester, NY, 14626-3601, United States

Description

VITROS XT Chemistry Products ALKP Slides

Reason

Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in the System reporting an incorrect value (lower than the actual value), which may result in missed or delayed diagnosis of liver disease.

Code Info

Product Code: 1053180 UDI:(10758750008346) GEN #21; Expiry Date Range 01 Jul 2023 - 01 June 2024 GEN #23; Expiry Date Range 01 Oct 2023 - 01 Sep 2024 GEN #26; Expiry Date Range 01 Aug 2023 - 01 Dec 2024 GEN #28; Expiry Date Range 01 Aug 2024 GEN #29; Expiry Date Range 01 Sep 2024 - 01 Dec 2024 GEN #30; Expiry Date Range 02 Jan 2025

Distribution

Worldwide Distribution: US (nationwide) and OUS (foreign) distribution to countries of: Bermuda, Belgium, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, Italy, India, Japan, Mexico, Netherlands, Norway, Singapore, Spain, Sweden, Portugal, Russia, United Kingdom

Quantity

402,966 units