FDA Enforcement Class II Ongoing

Aurora 2 Series, Aurora 7 lighthead; Models: AUR7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.

Recall: Z-2619-2024 · Reported August 21, 2024

Enforcement

Recall Number
Z-2619-2024
Event ID
94979
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
DKK Dai-Ichi Shomei Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 21, 2024
Initiation Date
July 9, 2024
Classification Date
August 14, 2024
Address
1 Chome 32-26, Sakashita, Itabashi, N/A, Japan

Description

Aurora 2 Series, Aurora 7 lighthead; Models: AUR7 (including TV models). The Aurora Series is intended to illuminate the surgical field and ensure visibility of the surgical site.

Reason

The weld on the light head frame can crack which could lead to problems such as the light head frame sagging, the plastic front face housing cracking, or the light head detaching.

Code Info

Model Number: AUR7-LH/7TV-LH; Serial Numbers: 09010002, 9020003, 9020004, 9020007, 9020008, 9020009, 9020010, 9020011, 9020012, 9020013, 9020014, 9020015, 9020016, 9030017, 9030018, 9030019, 9030020, 9030021, 9030022, 9040023, 9040024, 9040025, 9040026, 9040027, 9040028, 9040029, 9040030, 9040031, 9040032, 9040033, 9040034, 9040035, 9040036, 9040037, 9040038, 9040039, 9040040, 9040041, 9040042, 9050043, 9050044, 9050045, 9050046, 9050047, 9050048, 9050049, 9050050, 9050051, 9050052, 9050053, 9050054, 9050055, 9050056, 9050057, 9050058, 9050059, 9050060, 9050061, 9050062, 9050063, 9050064, 9050065, 9050066, 9050067, 9050068, 9050069, 9050070, 9050071, 9050072, 9060073, 9060074, 9060075, 9060076, 9060077, 9060078, 9060079, 9060080, 9060081, 9060082, 9060083, 9060084, 9060085, 9060086, 9060087, 9060088, 9060089, 9060090, 9060091, 9060092, 9080093, 9080094, 9080095, 9080097, 9080098, 9080099, 9080100, 9080101, 9080102, 9080104, 9080105, 9080106, 9080107, 9090108, 9090109, 9090110, 9090111, 9090112, 9090113, 9090114, 9090115, 9090116, 9090117, 9090118, 9090119, 9110120, 9110121, 9110122, 9110123, 9110124, 9110125, 9110126, 9110127, 9110128, 9110129, 9110130, 9120131, 9120132, 9120133, 9120134, 9120135, 9120136, 9120137, 9120138, 9120139, 9120140, 9120141, 09120142, 9120143, 9120144, 10010145, 10010146, 10010147, 10010148, 10010149, 10010150, 10010151, 10010152, 10010153, 10010154, 10010155, 10010156, 10010157, 10010158, 10020159, 10020160, 10020161, 10020162, 10020163, 10020164, 10020165, 10020166, 10030167, 10030168, 10030169, 10030170, 10030171, 10030172, 10030173, 10030174, 10030175, 10040176, 10040177, 10040178, 10040179, 10040180, 10040181, 10040182, 10040183, 10040184, 10040185, 10040186, 10040187, 10040188, 10040189, 10040190, 10040191, 10040192, 10040193, 10040194, 10040195, 10040196, 10040197, 10040198, 10040199, 10040200, 10040201, 10040202, 10040203, 10040204, 10040205, 10050206, 10050207, 10050208, 10050209, 10050210, 10050211, 10050212, 10050213, 10050214, 10050215, 10050216, 10050217, 10050218, 10050219, 10050220, 10050221, 10050222, 10050223, 10050224, 10050225, 10050226, 10050227, 10050228, 10050229, 10050230, 10050231, 10050232, 10050233, 10050234, 10050235, 10050236, 10050237, 10050238, 10050239, 10050240, 10050241, 10050242, 10060243, 10060244, 10060245, 10060246, 10060247, 10060248, 10060249, 10060250, 10060251, 10060252, 10060253, 10060254, 10060255, 10070256, 10070257, 10070258, 10070259, 10070260, 10070261, 10070262, 10070263, 10070264, 10080265, 10080266, 10080267, 10080268, 10080269, 10080270, 10080271, 10080272, 10080273, 10080274, 10080275, 10080276, 10080277, 10080290, 10090278, 10090279, 10090280, 10090281, 10090282, 10090283, 10090284, 10090285, 10090286, 10090287, 10090288, 10090289, 10090291, 10100292, 10100293, 10100294, 10100295, 10110296, 10110297, 10110298, 10110299, 10110300, 10110301, 10110302, 11010303, 11010304, 11010305, 11010306, 11010307, 11010308, 11010309, 11010310, 11010311, 11010312, 11010313, 11010314, 11010315, 11010316, 11010317, 11010318, 11010319, 11010320, 11020321, 11020322, 11020323, 11020324, 11020325, 11020326, 11020327, 11020328, 11020329, 11020330, 11020331, 11020332, 11020333, 11020334, 11020335, 11030336, 11030337, 11030338, 11030339, 11030340, 11030341, 11030342, 11030343, 11030344, 11030345, 11030346, 11030347, 11030348, 11030349, 11030350, 11030351, 11030352, 11030353, 11040354, 11040355, 11040356, 11040357, 11040358, 11040359, 11040360, 11040361, 11040362, 11040363, 11040364, 11040365, 11040366, 11040367, 11040368, 11040369, 11040370, 11040371, 11040372, 11040373, 11050374, 11050375, 11050376, 11050377, 11050378, 11060379, 11060380, 11060381, 11060382, 11060383, 11060384, 11060385, 11060386, 11060387, 11060388, 11060389, 11060390, 11060391, 11060392, 11060393, 11060394, 11060395, 11060396, 11060397, 11060398, 11060399, 11060400, 11060401, 11060402, 11060403, 11060404, 11060405, 11060406, 11060407, 11060408, 11060409, 11060410, 11060411, 11060412, 11070413, 11070414, 11070415, 11070416, 11070417, 11070418, 11070419, 11070420, 11070421, 11070422, 11080423, 11080424, 11080425, 11080426, 11080427, 11080428, 11080429, 11080430, 11080431, 11080432, 11080433, 11080434, 11090435, 11090436, 11090437, 11090438, 11090439, 11090440, 11090441, 11090442, 11090443, 11090444, 11090445, 11090446, 11090447, 11090448, 11090449, 11090450, 11090451, 11090452, 11090453, 11090454, 11090455, 11100456, 11100457, 11100458, 11100459, 11100460, 11100461, 11110463, 11110464, 11110465, 11110466, 11110467, 11110468, 11110469, 11110470, 11110471, 11110472, 11110473, 11110474, 11110475, 11110476, 11110477, 11110478, 11110479, 11110480, 11110481, 11110482, 11110483, 11110484, 11110485, 11110486, 11110487, 11110488, 11110489, 11120490, 11120491, 11120492, 11120493, 11120494, 11120495, 11120496, 11120497, 11120498, 11120499, 11120500, 11120501, 11120502, 11120503, 11120504, 12010505, 12010506, 12010507, 12010508, 12010509, 12010510, 12010511, 12010512, 12010513, 12010514, 12020515, 12020516, 12020517, 12020518, 12020519, 12020520, 12020521, 12020522, 12020523, 12020524, 12020525, 12020526, 12020527, 12020528, 12020529, 12040530, 12040531, 12040532, 12040533, 12040534, 12040535, 12040536, 12040537, 12040538, 12040539, 12040540, 12040541, 12040542, 12040543, 12040544, 12040545, 12040546, 12040547, 12040548, 12040549, 12040550, 12040551, 12040552, 12040553, 12040554, 12040556, 12040557, 12040558, 12040559, 12050560, 12050561, 12050562, 12050563, 12050564, 12050565, 12050566, 12050567, 12050568, 12050569, 12050570, 12050571, 12050572, 12060573, 12060574, 12060575, 12060576, 12060577, 12060578, 12060579, 12060580, 12060581, 12060582, 12060583, 12060584, 12060585, 12060586, 12060587, 12070588, 12070589, 12070590, 12070591, 12070592, 12070593, 12070594, 12070595, 12070596, 12070597, 12070598, 12080599, 12080600, 12080601, 12080602, 12080603, 12080604, 12080605, 12080606, 12080607, 12080608, 12080609, 12080610, 12080611, 12080612, 12080613, 12080614, 12080615, 12080616, 12080617, 12080618, 12080619, 12080620, 12080621, 12080622, 12080623, 12080624, 12080625, 12080626, 12080627, 12080628, 12080629, 12080630, 12080631, 12090632, 12090633, 12090634, 12090635, 12090636, 12090637, 12090638, 12090639, 12090640, 12090641, 12090642, 12090643, 12100644, 12100645, 12100646, 12100647, 12100648, 12100649, 12100650, 12100651, 12100652, 12100653, 12100654, 12100655, 12100832, 12110656, 12110657, 12110658, 12110659, 12110660, 12110661, 12110662, 12110663, 12110664, 12110665, 12110666, 12110667, 12110668, 12110669, 12110670, 12110671, 12110672, 12110673, 12110674, 12110675, 12110676, 12110677, 12110678, 12110679, 12110680, 12110681, 12110682, 12110683, 12110684, 12110685, 12120686, 12120687, 12120688, 12120689, 12120690, 12120691, 12120692, 12120693, 12120694, 12120695, 12120696, 12120697, 13010698, 13010699, 13010700, 13010701, 13010702, 13010703, 13010704, 13010705, 13010706, 13010707, 13010708, 13010709, 13010710, 13010711, 13010712, 13020713, 13020714, 13020715, 13020716, 13020717, 13020718, 13020719, 13020720, 13020721, 13020722, 13020723, 13020724, 13020725, 13020726, 13020727, 13020728, 13020729, 13020730, 13020731, 13020732, 13030735, 13030736, 13030737, 13030738, 13030739, 13030740, 13030741, 13030742, 13030743, 13030744, 13030745, 13030746, 13030747, 13030748, 13030749, 13030750, 13030751, 13030752, 13030753, 13060754, 13060755, 13060756, 13060757, 13060758, 13060759, 13060760, 13060761, 13060762, 13060763, 13060764, 13060765, 13060766, 13060767, 13060768, 13060769, 13060770, 13070771, 13070772, 13070773, 13070774, 13070775, 13070776, 13070777, 13070778, 13070779, 13070780, 13070781, 13070782, 13070783, 13070784, 13070785, 13070786, 13070787, 13080788, 13080789, 13080790, 13080791, 13080792, 13080793, 13080794, 13080795, 13080796, 13080797, 13100798, 13100799, 13100800, 13100801, 13100802, 13100803, 13100804, 13100805, 13100806, 13100807, 13100808, 13100809, 13100810, 13100811, 13100812, 13100813, 13100814, 13100815, 13100816, 13100817, 13100818, 13100819, 13100820, 13100821, 13100822, 13100823, 13100824, 13100825, 13100826, 13100827, 13100828, 20040831, 08120001 (PROTO),

Distribution

Domestic: Nationwide Distribution International: Australia, Canada, Egypt, Hong Kong, Indonesia, Japan, Jordan, Netherlands, New Zealand, Philippines, Singapore, South America, South Korea, Thailand, UAE.

Quantity

822 units