FDA Enforcement
Class II
Terminated
DePuy Mitek LATARJET EXPERIENCE Top Hat Tap Product Code: 288203 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
Recall: Z-2618-2017
·
Reported June 28, 2017
Enforcement
- Recall Number
- Z-2618-2017
- Event ID
- 77252
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- DePuy Mitek, Inc., a Johnson & Johnson Co.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- June 28, 2017
- Initiation Date
- May 12, 2017
- Classification Date
- June 20, 2017
- Termination Date
- October 3, 2018
- Address
- 325 Paramount Dr, Raynham, MA, 02767-5199, United States
Description
DePuy Mitek LATARJET EXPERIENCE Top Hat Tap Product Code: 288203 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft
Reason
Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw
Code Info
GTN: (01)10886705026814 Lot codes: 16D01, 16D02, 16E01, 16J01, 16R01, 17A01, 17A02, 17C03
Distribution
US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK
Quantity
153 units