FDA Enforcement Class II Terminated

FluoroSet Radiographic Tubal Assessment Set Cannula, Manipulator/ Injector, Uterine Used for instillation of contrast media into the uterine cavity for radiographic evaluation of the uterine cavity and for injection of appropriate contrast media into the fallopian tubes for evaluation of tubal patency.

Recall: Z-2615-2016 · Reported August 31, 2016

Enforcement

Recall Number
Z-2615-2016
Event ID
74010
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cook Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 31, 2016
Initiation Date
April 15, 2016
Classification Date
August 22, 2016
Termination Date
August 4, 2017
Address
750 N Daniels Way, Bloomington, IN, 47404-9120, United States

Description

FluoroSet Radiographic Tubal Assessment Set Cannula, Manipulator/ Injector, Uterine Used for instillation of contrast media into the uterine cavity for radiographic evaluation of the uterine cavity and for injection of appropriate contrast media into the fallopian tubes for evaluation of tubal patency.

Reason

Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.

Code Info

All Lots. Catalog/product number; Global/order number J-RTAS-100 G51608

Distribution

Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, "Palestinian Territory, Occupied", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.

Quantity

1,881