FDA Enforcement Class II Terminated

Shuttle Select Slip-Cath Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques

Recall: Z-2613-2016 · Reported August 31, 2016

Enforcement

Recall Number
Z-2613-2016
Event ID
74010
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Cook Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 31, 2016
Initiation Date
April 15, 2016
Classification Date
August 22, 2016
Termination Date
August 4, 2017
Address
750 N Daniels Way, Bloomington, IN, 47404-9120, United States

Description

Shuttle Select Slip-Cath Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques

Reason

Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.

Code Info

All Lots. Catalog prefixes/suffixes SCBR5.5-/-SHTL SCBR6.5-/-SHTL Catalog/product number; Global/order number SCBR5.5-35-125-P-NS-H1-SHTL G31129 SCBR5.5-35-125-P-NS-JB1-SHTL G31130 SCBR5.5-35-125-P-NS-JB2-SHTL G36361 SCBR5.5-35-125-P-NS-MPA-SHTL G23697 SCBR5.5-35-125-P-NS-SIM1-SHTL G23699 SCBR5.5-35-125-P-NS-SIM2-SHTL G36360 SCBR5.5-35-125-P-NS-VERT-SHTL G23698 SCBR5.5-35-125-P-NS-VTK-SHTL G36359 SCBR5.5-35-135-P-NS-ANG-SHTL G48154 SCBR5.5-35-150-P-NS-0-SHTL G23705 SCBR5.5-35-150-P-NS-ANG-SHTL G23704 SCBR5.5-35-75-P-NS-ANG-SHTL G48152 SCBR6.5-35-125-P-NS-H1-SHTL G31134 SCBR6.5-35-125-P-NS-JB1-SHTL G31135 SCBR6.5-35-125-P-NS-JB2-SHTL G36364 SCBR6.5-35-125-P-NS-SIM1-SHTL G23701 SCBR6.5-35-125-P-NS-SIM2-SHTL G36363 SCBR6.5-35-125-P-NS-VERT-SHTL G23700 SCBR6.5-35-125-P-NS-VTK-SHTL G36362

Distribution

Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Djibouti, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Monaco, Mongolia, Morocco, Myanmar, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, "Palestinian Territory, Occupied", Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, U.S. Virgin Islands, Yugoslavia.

Quantity

60,302