FDA Enforcement Class II Terminated

Dimension Vista Estradiol (E2) Flex reagent cartridge

Recall: Z-2608-2017 · Reported June 28, 2017

Enforcement

Recall Number
Z-2608-2017
Event ID
76915
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 28, 2017
Initiation Date
March 21, 2017
Classification Date
June 16, 2017
Termination Date
August 3, 2018
Address
500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States

Description

Dimension Vista Estradiol (E2) Flex reagent cartridge

Reason

The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.

Code Info

Device Listing Number # D121541 Lot # 16148AA, 16190BA, 16201BA, 16228BC, 16256BA, 16277BB 16298AB, 16343BE,17005BC

Distribution

Product was distributed throughout the United States and Canada, United Arab Emirates, Argentina, Austria,Australia,Bangladesh,Belgium,Brazil,Canada Switzerland,China,Columbia,Czech Republic,Germany,Denmark,Algeria Egypt,Spain,Finland,France,Great Britain,Greece,Croatia,Indonesia,Ireland India,Italy,Japan,Canary Islands,Republic of Korea,Kuwait,Lebanon,Latvia Macedonia,Netherlands,Norway,New Zealand,Philippines,Poland,Portugal Qatar,Saudi Arabia,Slovenia,Slovakia,Turkey,Taiwan,South Africa

Quantity

1,697 units