FDA Enforcement Class II Terminated

Corin Trinity Screw, diameter 6.5mm 35mm Length, Cancellous Bone Screw, REF 321.035, Sterile; and Corin Trinity Screw, diameter 6.5mm 50mm Length, Cancellous Bone Screw, REF 321.050, Sterile, both are components of the Trinity Acetabular System.

Recall: Z-2604-2021 · Reported October 6, 2021

Enforcement

Recall Number
Z-2604-2021
Event ID
88589
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Corin Ltd
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
October 6, 2021
Initiation Date
July 28, 2021
Classification Date
September 30, 2021
Termination Date
June 12, 2024
Address
Unit 1-57-8, Corinium Centre, Cirencester, N/A, N/A, United Kingdom

Description

Corin Trinity Screw, diameter 6.5mm 35mm Length, Cancellous Bone Screw, REF 321.035, Sterile; and Corin Trinity Screw, diameter 6.5mm 50mm Length, Cancellous Bone Screw, REF 321.050, Sterile, both are components of the Trinity Acetabular System.

Reason

The packaging system on the devices have potential physical and water damage resulting in damage to the devices, loss of sterility, or contamination of the devices.

Code Info

The affected products are shipment specific and were delivered to the consignee between 7/22-26/2021. Model 321.035 - Lot number 481070 was delivered 7/23/2021 and 7/26/2021; and Model 321.050 - Lot number 478892 was delivered 7/23/2021.

Distribution

US Nationwide distribution in the states of AZ, CA, FL, IL, OK, PA, TX, UT, and WI.

Quantity

6 units