FDA Enforcement Class II Terminated

DxA 5000 (DxA Automation System, DxA 5000 fit), "For in Vitro Diagnostic Use", Reference Numbers: B87352 with B57634, UDI: 15099590707118

Recall: Z-2602-2021 · Reported October 6, 2021

Enforcement

Recall Number
Z-2602-2021
Event ID
88519
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 6, 2021
Initiation Date
July 21, 2021
Classification Date
September 30, 2021
Termination Date
October 10, 2024
Address
250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States

Description

DxA 5000 (DxA Automation System, DxA 5000 fit), "For in Vitro Diagnostic Use", Reference Numbers: B87352 with B57634, UDI: 15099590707118

Reason

There is a potential that sample carriers (w or w/o samples) within automated sampling system may transport samples to the outer lane which does not contain a physical barrier to restrain samples. Samples are dropped when they reach the end of the outer lane which could result in Delay in providing patient with appropriate medical treatment. New sample needs to be drawn. Substantial discharge of biohazards material - Infection with a human transmissible blood-borne or animal pathogen which could result in Infection with a human transmissible blood-borne pathogen (e.g. infectious HBV, HCV, HIV).

Code Info

Affected US Serial Numbers: T23000077, T23000075, T23000076, T23000074, T23000069, T23000070, T23000066, T23000099, T23000096, T23000098, T23000100, T23000095, T23000097, T23000101, T23000092, T23000091, T23000090, T23000142, T23000045, T23000040, T23000039, T23000038, T23000152, T23000154, T23000155, T23000153, T23000106, T23000108, T23000107, T23000109, T23000128, T23000124, T23000125, T23000126, T23000115, T23000116, T23000113, T23000112, T23000117, T23000114, T23000132, T23000133, T23000131, T23000134, T23000064, T23000065, T23000063, T23000062, T23000163, T23000164, T23000162, T23000173, T23000169, T23000161, T23000138, T23000136, T23000137, T23000094, T23000093, T23000135, T23000141, T23000139, T23000140, T23000166, T23000168, T23000167, T23000082, T23000081, T23000080, T23000085, T23000084, T23000083, T23000143, T23000146, T23000145, T23000144

Distribution

US Nationwide distribution in the states of AL, CA, IL, In, KY, MN, MI, OH, PA, TX, WV.

Quantity

123 systems