Philips CT systems labeled as the following: 1. IQon Spectral CT, Model Number: 728332.

Recall: Z-2601-2025 · Reported October 1, 2025

Enforcement

Recall Number
Z-2601-2025
Event ID
97468
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips North America Llc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 1, 2025
Initiation Date
August 13, 2025
Classification Date
September 19, 2025
Address
222 Jacobs St, N/A, Cambridge, MA, 02141-2289, United States

Description

Philips CT systems labeled as the following: 1. IQon Spectral CT, Model Number: 728332.

Reason

The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

Code Info

1. Model Number: 728332; UDI-DI: 00884838059542; Serial Numbers: 860088, 860023, 860077, 860086, 60019, 860152, 860215, 60031, 60123, 60129, 60152, 860011, 860049, 860019, 860008, 860030, 860025, 860009, 860022, 860013, 860024, 860020, 860036, 860041, 860046, 860055, 860058, 860054, 860073, 860065, 860070, 860072, 860066, 860103, 860076, 860085, 860095, 860102, 860101, 860104, 860175, 860136, 860117, 860120, 860121, 860124, 860125, 860191, 860131, 860150, 860164, 60041, 860207, 60003, 860157, 860167, 860158, 860196, 60012, 860176, 860185, 860168, 860179, 860194, 60010, 60013, 60007, 60001, 395003, 860209, 860208, 60036, 60030, 60027, 60054, 60055, 60053, 60058, 60079, 395005, 60145.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauritius, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Vietnam.

Quantity

81 units