FDA Enforcement Class II Ongoing

Stryker ACM (Advanced Cement Mixing) System: Break-Away Femoral Nozzle-, Allows the user to mix bone cement with a constant high vacuum through the process. Catalog number: 0206-512-000.

Recall: Z-2596-2023 · Reported September 27, 2023

Enforcement

Recall Number
Z-2596-2023
Event ID
92841
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 27, 2023
Initiation Date
August 7, 2023
Classification Date
September 18, 2023
Address
1941 Stryker Way, Portage, MI, 49002-9711, United States

Description

Stryker ACM (Advanced Cement Mixing) System: Break-Away Femoral Nozzle-, Allows the user to mix bone cement with a constant high vacuum through the process. Catalog number: 0206-512-000.

Reason

Potential for the Break-Away Femoral Nozzle the applicator tip connecting the cement cartridge, to disassemble or break off during use, can lead to the potential for loss of function of the ACM system and bio-incompatibility due to components falling into surgical site

Code Info

GTIN: 04546540039415 Lot Numbers: 22314012 22336012

Distribution

Nationwide Foreign: Austria Belgium Canada France Germany Hong Kong Italy Malaysia Poland Singapore South Korea Spain Taiwan Thailand UK

Quantity

109 units US; 290 units OUS