Flow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI, GTIN 0080682102339; (2) Aestiva 7900; (3) Aestiva 7100; (4) Aespire System Carestation, S/5 Protiva Config., Aespire 100 Compact Wall Mount (Aespire 7100); (5) S/5 Aespire 7900; (6) Aespire View; (7) Avance; (8) Avance CS2, GTIN 00840682102292; (9) Aisys; (10) Aisys CS2, GTIN 00840682102322; (11) Amingo; (12) 9100 NXT Config, Aelite NXT; (13) Protiva 7100;
Enforcement
- Recall Number
- Z-2596-2021
- Event ID
- 88400
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- GE Healthcare, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 6, 2021
- Initiation Date
- August 2, 2021
- Classification Date
- September 30, 2021
- Address
- 3000 N Grandview Blvd, N/A, Waukesha, WI, 53188-1615, United States
Description
Flow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI, GTIN 0080682102339; (2) Aestiva 7900; (3) Aestiva 7100; (4) Aespire System Carestation, S/5 Protiva Config., Aespire 100 Compact Wall Mount (Aespire 7100); (5) S/5 Aespire 7900; (6) Aespire View; (7) Avance; (8) Avance CS2, GTIN 00840682102292; (9) Aisys; (10) Aisys CS2, GTIN 00840682102322; (11) Amingo; (12) 9100 NXT Config, Aelite NXT; (13) Protiva 7100;
The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cause leaks resulting in incorrect anesthesia machine tidal volumes potentially leading to over-delivery of volume to the patient.
The flow sensors built prior to June 2021 that are installed in the devices/anesthesia machines.
Distribution was nationwide, including PR. Foreign distribution was made to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote D'lvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Fiji, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Laos, Latvia, Lebanon, Libya, Lithuania, Macedonia, Madagascar, Malaysia, Maldives, Mali, Malta, Mauritius, Mexico, Moldova, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Niger, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom Uruguay, Venezuela, Vietnam, and Yemen.
Estimated 310,072 flow sensors