FDA Enforcement
Class II
Ongoing
MEDLINE STERILE IRIS SCISSORS CVD/STD, REF DYNJ04049
Recall: Z-2595-2023
·
Reported September 27, 2023
Enforcement
- Recall Number
- Z-2595-2023
- Event ID
- 92881
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- MEDLINE INDUSTRIES, LP - Northfield
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- September 27, 2023
- Initiation Date
- August 3, 2023
- Classification Date
- September 18, 2023
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
MEDLINE STERILE IRIS SCISSORS CVD/STD, REF DYNJ04049
Reason
4.5" Iris Scissors, Sterile, Curved (DYNJ04049) are being recalled due to the potential of the tip protector to fall off. The sharp tip of the scissors could compromise the sterile barrier system by puncturing large and/or small holes that may not be visible to the user.
Code Info
UDI/DI 10080196847273, all lots
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Canada, Hong Kong, Taiwan, Singapore
Quantity
324,451 devices