FDA Enforcement Class II Ongoing

MOSAIQ Oncology Information System

Recall: Z-2591-2021 · Reported October 6, 2021

Enforcement

Recall Number
Z-2591-2021
Event ID
88599
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Elekta, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
October 6, 2021
Initiation Date
September 2, 2021
Classification Date
September 29, 2021
Address
100 Mathilda Pl, 5th Floor, Sunnyvale, CA, 94086-6076, United States

Description

MOSAIQ Oncology Information System

Reason

A drug strength in MOSAIQ can be changed during the ordering process and approved, but the saved drug strength reverts to the original value.

Code Info

MOSAIQ 2.83 when run with First Data Bank (FDB), UDI/GTIN MOSAIQ 2.83 (0858164002367)

Distribution

Distribution was made to CA, CO, FL, ID, LA, MO, MS, NC, NH, NM, NV, NY, OH, OK, PA, SD, and WA. There was no military/government distribution. Foreign distribution was made to Australia, Canada, and the United Kingdom.

Quantity

31