FDA Enforcement Class II Ongoing

SureStep Foley Tray System Bardex I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303316A

Recall: Z-2589-2025 · Reported September 24, 2025

Enforcement

Recall Number
Z-2589-2025
Event ID
97273
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
C.R. Bard Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
September 24, 2025
Initiation Date
July 17, 2025
Classification Date
September 18, 2025
Address
8195 Industrial Blvd Ne, N/A, Covington, GA, 30014-1497, United States

Description

SureStep Foley Tray System Bardex I.C. Complete Care Infection Control Foley Catheter Tray, REF: A303316A

Reason

Foley catheter trays have incorrect inserts, user may think:1)Silicone, when it's natural rubber latex, or 2)Natural rubber latex with infection control coating, when it's silicone hydrogel with antimicrobial coating, or 3)Natural latex with infection control, when it's silicone hydrogel coated without antimicrobial coating, which may cause infection and/or local and/or systemic allergic reaction

Code Info

Lot#NGJX6399

Distribution

US Nationwide distribution including in the states of NC, FL, NY, MA, MN, TX, CA, OH, MI, WA, NJ, IL, ID, SC, IN, OK, OR, AL, VA, HI, MO, MS, MD, AZ, KY, GA, IA, CO, NM, PA, ME, UT, NE.

Quantity

4,300