FDA Enforcement
Class I
Completed
GENOSYL DS (Delivery System) console, for use as a vasodilator.
Recall: Z-2587-2021
·
Reported October 6, 2021
Enforcement
- Recall Number
- Z-2587-2021
- Event ID
- 88558
- Classification
- Class I
- Status
- Completed
- Product Type
- Devices
- Firm
- Vero Biotech, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- October 6, 2021
- Initiation Date
- September 9, 2021
- Classification Date
- September 30, 2021
- Address
- 387 Technology Cir Nw Ste 125, N/A, Atlanta, GA, 30313-2411, United States
Description
GENOSYL DS (Delivery System) console, for use as a vasodilator.
Reason
Manufacturing defect involving reversal of the inlet and outlet gas lines, resulting in delivery of nitrogen dioxide (NO2) instead of nitric oxide (NO).
Code Info
Model Number: 601588-01, Serial Numbers G0294 (UDI (01)00850000962026(11)200728(21)G0294); G0378 (UDI (01)00850000962026(11)200917(21)G0378); and MA000065 (UDI (01)00850000962026(11)210511(21)MA000065)
Distribution
Distributed within US to Texas and Illinois.
Quantity
120 consoles (3 consoles affected)