FDA Enforcement
Class II
Terminated
Optima CT 540
Recall: Z-2586-2018
·
Reported August 8, 2018
Enforcement
- Recall Number
- Z-2586-2018
- Event ID
- 80484
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- GE Healthcare, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 8, 2018
- Initiation Date
- May 24, 2018
- Classification Date
- July 31, 2018
- Termination Date
- February 8, 2019
- Address
- 3000 N Grandview Blvd, N/A, Waukesha, WI, 53188-1615, United States
Description
Optima CT 540
Reason
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Code Info
Serial number 500925HM3, System ID Number 615320OPT
Distribution
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Quantity
42 units total