FDA Enforcement Class II Ongoing

VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2

Recall: Z-2580-2024 · Reported August 21, 2024

Enforcement

Recall Number
Z-2580-2024
Event ID
95020
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Ortho-Clinical Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
August 21, 2024
Initiation Date
July 3, 2024
Classification Date
August 12, 2024
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101, United States

Description

VITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2

Reason

It was confirmed that fibrinogen in patient plasma samples precipitates out of solution upon the addition of folate stabilizer reagent as part of the pre-treatment process causing the increase in "TM5-4MB" condition codes which lead to a delay of results.

Code Info

Model No 1513266; UDI/GTIN# 110758750009237; Lots within expiry: 3380, 3390, 3400, 3410, 3422, 3430, 3440, 3450, 3461, 3471, 3480.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Sweden, United Kingdom.

Quantity

18,467 units