FDA Enforcement
Class II
Terminated
Discovery MI, labeled as Discovery DMI 15 cm 3 ring, and Discovery DMI 20cm 4 ring
Recall: Z-2580-2018
·
Reported August 8, 2018
Enforcement
- Recall Number
- Z-2580-2018
- Event ID
- 80484
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- GE Healthcare, LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- August 8, 2018
- Initiation Date
- May 24, 2018
- Classification Date
- July 31, 2018
- Termination Date
- February 8, 2019
- Address
- 3000 N Grandview Blvd, N/A, Waukesha, WI, 53188-1615, United States
Description
Discovery MI, labeled as Discovery DMI 15 cm 3 ring, and Discovery DMI 20cm 4 ring
Reason
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Code Info
Model Number 5454001-170. Serial Number (System ID Number): CJRPX1800016CN ( 4550966PT11), CJRPX1800017CN (561422MI), CJRPX1800018CN (617726PET1EB), CJRPX1800021CN (404616DMI)
Distribution
Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Quantity
42 units total